(74 days)
No
The summary explicitly states "Mentions AI, DNN, or ML: Not Found" and the description focuses on image processing and registration techniques without mentioning AI/ML algorithms.
No
This device is a software application for image visualization, registration, and fusion, used for mapping planning information. It does not directly provide therapy or therapeutic interventions.
Yes
The device is described as "Multi-Modality Image Fusion is a software application to be used by physicians in the clinic or hospital for 2-D and 3-D visualization, image registration, and fusion of MRI, CT and Ultrasound imaging modalities for mapping planning information across modalities." It aims to provide mapping planning information to physicians, which is a key function of diagnostic devices. Although the provided information does not explicitly state it as "diagnostic," the nature of processing and presenting medical images for physicians to use in mapping planning strongly implies a diagnostic purpose.
Yes
The device description explicitly states that the device is "software, which comprises of two software components". While performance studies mention "MMIF Hardware verification Test Procedure" and "complete system consisting of both the online and off line software installed on specified hardware", the core device being submitted for 510(k) is described as software. The hardware is likely the platform on which the software runs, not part of the regulated device itself.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is for "2-D and 3-D visualization, image registration, and fusion of MRI, CT and Ultrasound imaging modalities for mapping planning information across modalities." This describes a tool for image processing and planning, not for examining specimens derived from the human body to provide information for diagnosis, monitoring, or treatment.
- Device Description: The description focuses on software for image preparation, fusion, and visualization. It doesn't mention any components or processes related to analyzing biological samples.
- Lack of IVD Characteristics: There is no mention of analyzing biological specimens (blood, tissue, etc.), detecting analytes, or providing diagnostic information based on laboratory tests. The focus is entirely on medical imaging data.
Therefore, this device falls under the category of medical imaging software, not an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
Multi-Modality Image Fusion is a software application to be used by physicians in the clinic or hospital for 2-D and 3-D visualization, image registration, and fusion of MRI, CT and Ultrasound imaging modalities for mapping planning information across modalities. Additional software features include database management, data communication, surface rendering, segmentation, regions of interest (ROI) delineation, volumetric measurements, and data reporting.
Product codes
LLZ
Device Description
Multi-Modality Image Fusion (MMIF) is software, which comprises of two software components, which is referred to as offline and online. The offline component perfains to the preparation of gland and suspected lesion boundaries on a DICOM image file days or hours prior to the biopsy procedure. The online component fuses the DICOM image files, which were, prepared on the offline component, with a snap shot incoming TRUS image. Each of the two software components can work together or independently.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
MRI, CT, Ultrasound
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
physicians in the clinic or hospital
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
The conducted verification tests have confirmed static and dynamic performance of the complete system consisting of both the online and off line software installed on specified hardware. Furthermore these tests establish compliance to the specification under conditions simulation real environment of use.
Summary of Performance Studies
Bench performance tests, functionality tests, and system test were successfully completed.
Testing activities were conducted to establish the performance, safety, effectiveness, functionality usability, and reliability characteristics of the new device.
The purpose of the performance testing (bench testing) was to ensure the performance of the new MMIF software by verifying the accuracy of the specifications and simulating real customer data collected from the hospital setting.
Tests performed:
- MMIF Hardware verification Test Procedure
- MMIF Online Verification Test Procedure
- MMIF Offline Software Application Verification Test Procedure
- MMIF Risk Mitigation Verification Test Procedure
- Registration Accuracy for Clinical Data (bench test)
- Phantom Volume Measurements (bench test)
- Phantom Registration Accuracy (bench test)
Key Metrics
Not Found
Predicate Device(s)
K041182, K081987, K093672, K053610
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
0
510(k) Summary
APR - 6 2012
510(k) Summary
Submitter's Name: Submitter's Address: Submitter's Telephone: Contact Name:
Date Summary was Prepared: Trade or Proprietary Name: Common or Usual Name: Classification Name:
Predicate Devices:
IGT, LLC dba Eigen 13366 Grass Valley Avenue, Grass Valley, CA 95945 530-274-1240 . William Mandel 21 Oct 2011
Multi-Modality Image Fusion Multi-Modality Image Fusion
System, Image Processing, Radiological, LLZ
Picture Archiving and Communications, 21CFR 892.2050
Device Name | 510(k) Number |
---|---|
Syntegra | K041182 |
UniSyn | K081987 |
Aegis Navigation | K093672 |
Abaris | K053610 |
Intended Use
Multi-Modality Image Fusion is a software application to be used by physicians in the clinic or hospital for 2-D and 3-D visualization, image registration, and fusion of MRI, CT and Ultrasound imaging modalities for mapping planning information across modalities. Additional software features include database management, data communication, surface rendering, segmentation, regions of interest (ROI) delineation, volumetric measurements, and data reporting.
Description of the Device and Summary of the Technological Characteristics:
Multi-Modality Image Fusion (MMIF) is software, which comprises of two software components, which is referred to as offline and online. The offline component perfains to the preparation of gland and suspected lesion boundaries on a DICOM image file days or hours prior to the biopsy procedure. The online component fuses the DICOM image files, which were, prepared on the offline component, with a snap shot incoming TRUS image. Each of the two software components can work together or independently.
Offline Software
The offline software can delineate suspected areas of interest using visualization and save this data in a standard format, e.g. as a surface mesh in VTK or STL format. Furthermore, the
EIGEN
PREMARKET NOTIFICATION - MULTI-MODALITY IMAGE FUSION
1
offline software can ontionally delineate the gland boundary from a DICOM image, or other volumes where the boundary can be easily observed. This addition data is also saved in a standard format. e.g. surface mesh in VTK or STL or as a binary volume. This software would be typically executed by Radiologists.
Online Software
The online unit software is designed to fuse two images by entering a fusion mode and load the DICOM image and the boundary data provided by the offline unit. The online unit performs rigid alignment, either selecting globally align that orients the two DICOM volumes based on the known relative orientation difference between the two or specify landmarks to register images rigidly. The online unit then elastically warps the image to register on the incoming image from the offline unit.
The registered image volume is displayed on the incoming offline image frame of reference. Also overlaid on the incoming offline and registered image volumes annotating the suspected lesion boundaries. The two volumes are displayed side-by-side in a locked view. Rotation, slicing, panning, or zooming one of the volumes has the same effect on the other enabling visualization of similar features in both volumes simultaneously.
Plan targets on the volumes by selecting suitable locations within the suspected areas of interest boundary. When sufficient targets have been added, exiting fusion mode imports all targets planned in the offline frame of reference back to the original reconstructed ultrasound volume. This is done based on the deformation field that was estimated from registration (rigid followed by elastic registration).
Substantial Equivalence:
| Manufactu
rer | Eigen | ADAC
Laboratories | Convergent
Imaging
Solutions | Sentinelle Medical
Inc. | Traxtal
Technologies lhc. |
|------------------|------------------------------------------------------------------------|----------------------------------------------------------------|-----------------------------------------------------|-----------------------------------------------------------------------|---------------------------------------------------------|
| Product
Name | Multi-Modality
Image Fusion | Syntegra | UniSyn | Aegis Navigation | Abaris |
| 510(k)
number | pending | K041182 | K081987 | K093672 | K053610 |
| Intended
Use | Multi-Modality
Image Fusion is a
software
application used by | Syntegra is a
software
application for
multi-modality | UniSyn is a
software
application for
image | This device
provides two and
three-dimensional
image review. | ABARIS is a
stereotaxic
accessory for
Computed |
EIGEN
PREMARKET NOTIFICATION - MULTI-MODALITY IMAGE FUSION
2
physicians in the | image registration | registration and | manipulation, and | Tomography | ||||
---|---|---|---|---|---|---|---|---|
clinic or hospital | and diagnostic | fusion display | analysis tools to | (CT), Magnetic | ||||
for 2-D and 3-D | fusion. Images are | of scanned | assist users in | Resonance, (MR), | ||||
visualization, | registered and | image data | screening, | Ultrasound (US), | ||||
multi-modality | displayed in a | from CT, PET, | diagnosis, planning | Positron Emission | ||||
image registration, | "fused" (overlaid | SPECT and | and performing | Tomography | ||||
and fusion of | in the same spatial | MR scanners. | image-guided | (PET), Single | ||||
medical images. | orientation) | UniSyn creates | interventional | Photon Emission | ||||
Additional | format to provide | multi-planar | procedures. | Computed | ||||
software features | ||||||||
include database | combined | reformat and | Supported imaging | Tomography | ||||
functional and | maximum | modalities include | (SPECT), | |||||
management, | anatomical data | intensity | Magnetic | Fluoroscopy, | ||||
communication. | providing | projection | Resonance (MR), | Endoscopy and | ||||
surface rendering, | different angular | displays of the | Ultrasound (US), | other imaging | ||||
segmentation, ROI | perspectives for | data and | Single Photon | systems. it | ||||
delineation, | interpretation by | provides | Emission | displays the | ||||
measurements, and | trained | measurements | Computed | simulated image | ||||
reporting. | professionals | such as area, | Tomography | of a tracked | ||||
volume and | (SPECT), | insertion tool | ||||||
Standard | Computed | such as a biopsy | ||||||
Uptake Values | Tomography (CT), | needle, guidewire | ||||||
for user defined | Positron | or probe on a | ||||||
regions on the | Emission | computer monitor | ||||||
image | Tomography | screen that shows | ||||||
(PET), | images of the | |||||||
Fluoroscopy and | target organs and | |||||||
Endoscopy. Images | the current and the | |||||||
and data are | projected future | |||||||
received from | path of the | |||||||
various imaging | interventional | |||||||
systems and other | instrument taking | |||||||
sources such as | into account | |||||||
calibrated spatial | movements of the | |||||||
positioning | patient. This is | |||||||
devices. | intended for | |||||||
This device | treatment | |||||||
provides the | planning and | |||||||
capability to | intra-operative | |||||||
overlay annotations | guidance for | |||||||
on 2D or 3D | surgical | |||||||
medical image | procedures. | |||||||
displays. | The device also | |||||||
These annotations | supports an | |||||||
may represent the | image-free mode | |||||||
position of | in which the | |||||||
instruments | proximity of the | |||||||
including but not | interventional | |||||||
limited to biopsy | device is | |||||||
needles, guidance | displayed relative | |||||||
wires, imaging | to another device. | |||||||
probes or other | The device is | |||||||
tracked devices. | intended to be | |||||||
This device is | used in clinical | |||||||
intended to assist | interventions and | |||||||
skilled medical | for anatomical | |||||||
professionals in | structures where | |||||||
skilled medical | ||||||||
professionals in | ||||||||
clinical screening | ||||||||
and | clinical screening | for anatomical | ||||||
structures where | ||||||||
imaging is | ||||||||
currently used for | imaging is |
.
510(k) Summary
.
EIGEN PREMARKET NOTIFICATION – MULTI-MODALITY IMAGE FUSION
・
・
.
. . . . .
,
.
3
510(k) Summary | |||||
---|---|---|---|---|---|
interventions, for | |||||
anatomical | |||||
structures where | |||||
imaging is | |||||
currently used for | |||||
visualizing such | |||||
structures, | |||||
including head and | |||||
neck, breast, | |||||
thoracic, and | |||||
abdominal | |||||
applications | |||||
(including | visualizing such | ||||
procedures. The | |||||
device is also | |||||
intended for use | |||||
in clinical | |||||
interventions to | |||||
determine the | |||||
proximity of | |||||
one device | |||||
relative to | |||||
another. | |||||
pelvis). | |||||
Lossy compressed | |||||
mammographic | |||||
images and | |||||
digitized film | |||||
screen images must | |||||
not be | |||||
reviewed for | |||||
primary image | |||||
interpretations. | |||||
Mammographic | |||||
images may only | |||||
be interpreted | |||||
using | |||||
a FDA approved | |||||
monitor that offers | |||||
at least 5 MPixel | |||||
resolution and | |||||
meets other | |||||
technical | |||||
specifications | |||||
reviewed and | |||||
accepted by the | |||||
FDA. | |||||
Where | |||||
Used | Office settings in | ||||
clinic or hospital | Office settings in | ||||
clinic or hospital | Office settings | ||||
in clinic or | |||||
hospital | Not specified | clinical | |||
interventions | |||||
Software | |||||
Device | Yes | Yes | Yes | Yes | Yes |
Image | |||||
Registratio | |||||
n | Multi-modality | ||||
image registration | Multi-modality | ||||
image registration | Multi-modality | ||||
image | |||||
registration | Multi-modality | ||||
image registration | Yes | ||||
Image | |||||
Fusion | |||||
/Overlay | |||||
Display | Fused overlay of | ||||
images from | |||||
multiple modalities | Fused overlay of | ||||
images from | |||||
multiple | |||||
modalities | Fusion overlay | ||||
of images from | |||||
multiple | |||||
modalities | overlay of images | ||||
from multiple | |||||
modalities | Yes | ||||
Opacity | |||||
Control | Yes | Yes | Yes | Not Specified | Not Specified |
3-D | |||||
Rendering | Yes | Yes | Yes | Yes | Yes |
Surface | |||||
Rendering | Yes | Yes | Yes | Yes | Yes |
Regions of | |||||
interest | Yes | Yes | Yes | Yes | Yes |
Configura | Yes | Yes | Yes | Yes | Yes |
EIGEN
.
:
DIGE."
PREMARKET NOTIFICATION – MULTI-MODALITY IMAGE FUSION
·
4
510(k) Summary
ble image layouts | |||||
---|---|---|---|---|---|
3D | |||||
Contouring | Yes | Not Specified | Yes | Yes | Yes |
Export of | |||||
3D contours | |||||
for | |||||
planning | Yes | Not Specified | Yes | Not Specified | Yes |
Image | |||||
Storage | |||||
and | |||||
Communication | DICOM, JPEG | DICOM (check) | DICOM | DICOM | DICOM |
Modalities | CT, MRI, | ||||
Ultrasound | Positron Emission | ||||
Tomography | |||||
(PET) and | |||||
X-ray Computed | |||||
Tomography | |||||
(CT). | CT, PET, | ||||
SPECT and | |||||
MR scanners | MRI, US, SPECT, | ||||
CT, PET, | |||||
Fluoroscopy, | |||||
Endoscopy and | |||||
others | Computed | ||||
Tomography | |||||
(CT), Magnetic | |||||
Resonance, (MR), | |||||
Ultrasound (US), | |||||
Positron Emission | |||||
Tomography | |||||
(PET), Single | |||||
Photon Emission | |||||
Computed | |||||
Tomography | |||||
(SPECT), | |||||
Fluoroscopy, | |||||
Endoscopy and | |||||
other imaging |
The product's technical features are substantially equivalent to Syntegra (K041182). UniSyn (K081987), Aegis Navigation (K093672) and Abaris (K053610). The Multi-Modality Image Fusion is a software product that runs on PC-based workstations. Image data is input to the devices and used to generate 3-D views and perform image processing. Like the four predicate devices, the software has image measurement, multi-planar reformatting, segmentation and image registration abilities, fusion of images from different modalities, image storage and retrieval, as well as patient information management functions.
The Syntegra and UniSyn are software products that accept multiple image data types including magnetic resonance, and computed tomography. Aegis Navigation and Abaris are systems that accept multiple modalities that include magnetic resonance, computed tomography and ultrasound. The Multi-Modality Image Fusion has been designed for processing medical images in standard DICOM format. The system is composed of two modules: an offline and an online module for image fusion for planning a procedure.
EIGEN PREMARKET NOTIFICATION – MULTI-MODALITY IMAGE FUSION
5
Testing and Performance Data:
All product and engineering specifications were verified and validated. Test phantoms incorporating simulated prostates were developed and were used to verify system performance.
The device has been designed and manufactured to conform to the following standards:
ACRINEMA PS3.1-3.18 Digital Imaging and Communications in Medicine (DICOM)
ISO 14971 Application of Risk Management to Medical Devices
ISO 13485 Medical Devices - Quality Management Systems - Requirements For Regulatory purposes
The following tests were performed:
MMIF Hardware verification Test Procedure
MMIF Online Verification Test Procedure
MMIF Offline Software Application Verification Test Procedure .
MMIF Risk Mitigation Verification Test Procedure
Registration Accuracy for Clinical Data (bench test)
Phantom Volume Measurements (bench test)
Phantom Registration Accuracy (bench test)
All the above mentioned tests passed. Theses testing activities were conducted to establish the performance, safety, effectiveness, functionality usability, and reliability characteristics of the new device.
The purpose of the performance testing (bench testing) was to ensure the performance of the new MMIF software by verifying the accuracy of the specifications and simulating real customer data collected from the hospital setting.
The conducted verification tests have confirmed static and dynamic performance of the complete system consisting of both the online and off line software installed on specified hardware. Furthermore these tests establish compliance to the specification under conditions simulation real environment of use.
Conclusion:
The results of comparing the intended use, function, technological characteristics, mode of operation and specifications of the Multi-Modality Image Fusion with those of the two predicate devices demonstrate that the Multi-Modality Image Fusion is substantially
- EIGEN
PREMARKET NOTIFICATION – MULTI-MODALITY IMAGE FUSION
6
equivalent to existing products on the market today.
. . . .
.
ﺎ ﺳﮯ ﺍ
1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 -
·
The verification and validation testing included bench performance tests, functionality tests, and system test. All tests were successfully completed.
. '·
'
.
: :
EIGEN PREMARKET NOTIFICATION – MULTI-MODALITY IMAGE FUSION
. .
7
Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
APR - 6 2012
IGT LCC dba Eigen % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services 1394 25th Street NW BUFFALO MN. 55313
Re: K120187
Trade/Device Name: Multi-Modality Image Fusion Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications System Regulatory Class: II Product Code: LLZ Dated: March 1, 2012 Received: March 2 2012
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls: Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or uny 1 vict all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
8
Page 2
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
9
Indications for Use Form
510(k) Number (if known):
Device Name: Multi-Mogality Image Fusion
Indications for Use: Multi-Modality Image Fusion is a software application to be used by physicians in the clinic or hospital for 2-D and 3-D visualization, image registration, and fusion of MRI, CT and Ultrasound imaging modalities for mapping planning information across modalities. Additional software features include database management, data communication, surface rendering, segmentation, regions of interest (ROI) delineation, volumetric measurements, and data reporting.
Prescription Use (Part 21 CFR 801 Subpart D) | X |
---|---|
AND/OR |
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE OF ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Devices Evaluation and Safety
510(k) K120187
1-2