(126 days)
The intended use of the PERFORMER HT is to provide isolated hyperthermic perfusion in the thoracic or peritoneal cavity by means of extra-corporeal circulation of warmed, physiologically compatible sterile solution, according to a protocol to be selected by the physician.
The device must be used by a qualified medical professional who is experienced in the operation of this or similar equipment.
The PERFORMER HT is a prescription electromechanical device intended to provide isolated hyperthermic perfusion in the thoracic or peritoneal cavity by means of extracorporeal circulation of warmed. physiologically compatible sterile solution.
The PERFORMER HT equipment is designed to:
- operate the treatment execution by means of a color touch-screen display .
- . control the temperature of solution perfused to the Patient through the extracorporeal circuit
- control the flow rates of roller pumps .
- control the volume in the extracorporeal circuit and Patient's body cavity .
- monitor Patient's body cavity temperatures .
- monitor the extracorporeal circuit pressures .
- . detect air in the circuit
- change the fluid path in the extracorporeal circuit by means of 2 two-ways pinch valves .
The Central Information Display (CID) enables the User to interact with the PERFORMER HT equipment. The touch-screen display allows the User to:
- inout Patient information .
- operate the various treatment phases
- display treatment parameters (e.g, flows, pressures, temperatures) .
- set treatment parameters .
- . set alarm limits
- display graphical trending of temperature probes connected to the HTS module
- · read alarm and alert messages
The PERFORMER HT is intended to be used with the HANG&GO HT BASIC, a disposable, single use kit intended to circulate and filtrate the clear solutions in the thoracic or abdominal cavity.
The HANG&GO HT BASIC includes all necessary lines and accessories for the treatment execution:
- · circulation line
- ·. heater bag
- · reservoir
- · table pack (Patient inlet and outlet lines) with thermal protections
The HANG&GO HT BASIC is supplied sterile, packaged in a proprietary-design container, and is intended to be used only with the PERFORMER HT.
The Performer HT is a prescription electromechanical device for isolated hyperthermic perfusion. It circulates warmed, physiologically compatible sterile solution in the thoracic or peritoneal cavity. It controls solution temperature and flow rates, monitors volumes/pressures/temperatures, and detects air. The device is intended for use by qualified medical professionals.
Here's an analysis of its acceptance criteria and supporting studies:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (What the device should do) | Reported Device Performance (What the device does) |
|---|---|
| Ability to heat and pump fluids accurately over the full range of flow rate and operating conditions. | Performed within specifications. |
| Ability to maintain fluid temperature over the full range of fluid flow rates. | Performed within specifications. |
| Ability to control pressures, volumes, and flow rate accuracy over the full range of fluid flow rate within operating temperature and pressure range. | Performed within specifications. |
| Ability to detect and alarm at unsafe or ineffective operating conditions (e.g., operator errors, sensor failures, system malfunctions). | Performed within specifications. |
| Ability to mitigate against known or predictable operator errors. | Performed within specifications. |
| Ability to store parameters and data in non-volatile memory. | Performed within specifications. |
| Compliance with various recognized standards for: - Quality management (ISO 14971, EN ISO 13485, EN ISO 14971) - Packaging and labeling (EN 980, EN 1041, ASTM D4169-04a) - Biocompatibility (AAMI ANSI ISO 10993 series) - Electrical safety (IEC 60601 series) - Sterility (AAMI ANSI ISO 11135, ISO 11607 series, ASTM F1980-07, AAMI ANSI ISO 11737) - Human Factors (IEC 62366, EN 62366) - Software (IEC 62304, EN 62304) | The PERFORMER HT is designed to, complies with, and has been tested to, the FDA Recognized Consensus Standards listed in the table. |
| Software ability to meet performance specifications. | Verified through described tests at unit, integration, and system level for all functions, actuators, and parameters. |
| User interface optimization and mitigation of user- and patient-safety risks concerning human factors. | Demonstrated that task-risks and errors have been identified, addressed, and either eliminated or reduced in a simulated use environment. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state a "sample size" for a specific test set in the context of clinical trials or retrospective studies on actual patient data. The testing primarily involved non-clinical tests conducted on the PERFORMER HT system (equipment and disposable).
Data Provenance: The tests were performed by the manufacturer, RanD S.r.l. in Italy, as part of their 510(k) submission. Therefore, the data provenance is from the manufacturer's testing environment and simulated use, and it is prospective in the sense that the tests were designed and executed to evaluate the device's performance against pre-defined specifications. It is not patient or clinical data.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. The ground truth for this medical device primarily relates to its engineering specifications, safety standards, and functional performance as verified through non-clinical testing. There is no mention of a ground truth established by human experts in a diagnostic or clinical context. The "ground truth" is defined by the technical specifications and recognized consensus standards that the device is tested against.
4. Adjudication Method for the Test Set
Not applicable in the traditional sense. Adjudication methods like 2+1 or 3+1 are typically used in studies where human readers (e.g., radiologists) interpret medical images or data, and their interpretations are compared against a consensus ground truth.
In this case, the device's performance was evaluated against predefined engineering specifications and recognized standards. The "adjudication" would involve technical experts (engineers, quality assurance personnel) verifying that the test results met these objective criteria.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
No. This is not an AI/CAD (Computer-Aided Detection/Diagnosis) device, nor is it a device that is typically evaluated through MRMC studies. The PERFORMER HT is an electromechanical device for hyperthermic perfusion.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, in essence. The performance tests described comprehensively evaluate the PERFORMER HT system (equipment and disposable) in its standalone capacity, to ensure it meets its functional and safety specifications. Even though it's an electromechanical device and not an algorithm, the tests assess its "standalone" operational capabilities without human intervention beyond initiating and monitoring the tests. The human factors study separately assessed user interaction, but the core performance tests were for the device itself.
7. The Type of Ground Truth Used
The ground truth used for the non-clinical tests is based on pre-defined engineering specifications, recognized consensus standards, and established operational parameters. For example:
- Temperature control: The ability to heat and maintain fluid temperature within a specified range defines the ground truth for this function.
- Flow rate accuracy: The specified acceptable deviation from target flow rates serves as the ground truth.
- Alarm functionality: The system's response to simulated unsafe conditions (e.g., sensor failure) according to its design is the ground truth.
- Compliance with standards: The requirements outlined in international and European standards (e.g., ISO 14971, IEC 60601-1) form the ground truth for safety, quality, and electrical compatibility.
8. The Sample Size for the Training Set
Not applicable. This device is an electromechanical system, not an AI or machine learning algorithm. Therefore, there is no "training set" in the context of data-driven model development. The device's design and functionality are based on engineering principles and validated through testing, not learned from data.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As stated above, there is no "training set" for this type of device.
{0}------------------------------------------------
K120-026
5. 510(k) Summarv
MAY - 8 2012
21 CFR 807.92(a)(1): Submitter and Manufacturing site information RanD S.r.l. via Statale 12. 62 41036 Medolla (MO) - ITAL Y Telephone: +39 0535 49283 Fax: +39 0535 660636 Contact person: Simone Ceretti
Establishment registration number: 3003793891
Date of the present summary preparation: 28th December, 2011
| 21 CFR 807.92(a)(2): Device information | |||
|---|---|---|---|
| Trade name | PERFORMER HT / HANG&GO HT BASIC | ||
| Device Classification Name | Warmer, thermal, infusion fluid | ||
| Classification Review Panel | General Hospital | ||
| Classification Product Code | LGZ | ||
| Regulatory Class | Unclassified, pre-amendment device |
21 CFR 807.92(a)(3): Predicate devices
The PERFORMER HT is substantially equivalent to the Belmont Hyperthermia Pump (K070654, dtd June 8th, 2007).
21 CFR 807.92(a)(4): Description of the device
The PERFORMER HT is a prescription electromechanical device intended to provide isolated hyperthermic perfusion in the thoracic or peritoneal cavity by means of extracorporeal circulation of warmed. physiologically compatible sterile solution.
The PERFORMER HT equipment is designed to:
- operate the treatment execution by means of a color touch-screen display .
- . control the temperature of solution perfused to the Patient through the extracorporeal circuit
- control the flow rates of roller pumps .
- control the volume in the extracorporeal circuit and Patient's body cavity .
- monitor Patient's body cavity temperatures .
- monitor the extracorporeal circuit pressures .
- . detect air in the circuit
- change the fluid path in the extracorporeal circuit by means of 2 two-ways pinch valves .
The Central Information Display (CID) enables the User to interact with the PERFORMER HT equipment. The touch-screen display allows the User to:
- inout Patient information .
- operate the various treatment phases
- display treatment parameters (e.g, flows, pressures, temperatures) .
- set treatment parameters .
- . set alarm limits
- display graphical trending of temperature probes connected to the HTS module
{1}------------------------------------------------
- · read alarm and alert messages
The PERFORMER HT is intended to be used with the HANG&GO HT BASIC, a disposable, single use kit intended to circulate and filtrate the clear solutions in the thoracic or abdominal cavity.
The HANG&GO HT BASIC includes all necessary lines and accessories for the treatment execution:
- · circulation line
- ·. heater bag
- · reservoir
- · table pack (Patient inlet and outlet lines) with thermal protections
The HANG&GO HT BASIC is supplied sterile, packaged in a proprietary-design container, and is intended to be used only with the PERFORMER HT.
21 CFR 807.92(a)(5): Intended use
The intended use of the PERFORMER HT is to provide isolated hyperthermic perfusion in the thoracic or peritoneal cavity by means of extra-corporeal circulation of warmed. physiologically compatible sterile solution, according to a protocol to be selected by the physician.
The device must be used by a qualified medical professional who is experienced in the operation of this or similar equipment.
Predicate device intended use
The intended use of the Belmont Hyperthermia Pump is to raise the temperature of the thoracic or peritoneal cavity to the desired target temperature by continuously lavaging the cavity with circulating warmed sterile solution, according to a protocol to be selected by the physician.
21 CFR 807.92(a)(6): Comparison of technological characteristics
The PERFORMER HT has the same operating principle, energy type, environmental specifications, and performance specifications of the Belmont device. It uses roller-type fluid pumps, touch screen to direct the User through set-up preparation and use, control and monitoring of fluid temperatures, and a proprietary-design disposable set including large fluid reservoir to circulate sterile fluid into and out of the body cavity. Flow from the reservoir is delivered by action of the roller pump to the heater exchanger to be heated to the desired target temperature, and then to the Patient, From the Patient, the fluid is pumped from the return line back into the reservoir. The PERFORMER HT monitors circulating flow, temperatures, pressures and volumes, detects air in the circuit and automatically responds to ineffective or unsafe operating conditions by sounding audible alarms, stopping heating and pumping in case of hazardous conditions for User/Patient.
{2}------------------------------------------------
21 CFR 807.92(b)(1): Discussion of the non-clinical tests
The PERFORMER HT is designed to, complies with, and has been tested to, the FDA Recognized Consensus Standards listed in the table below, as applicable to the equipment, the disposable lines set and the components:
| Area | US Recognized standard # | EU Recognized standard # | Title |
|---|---|---|---|
| General | ISO 14971:2007 | EN ISO 13485:2003 / AC:2007 | Medical devices -- Quality management systems -- Requirements for regulatory purposes |
| EN ISO 14971:2007 | Medical devices - Application of risk management to medical devices | ||
| Packaging andlabeling | EN 980:2008 | Symbols for use in the labelling of medical devices | |
| EN 1041:2008 | Information supplied by the manufacturer with medical devices | ||
| ASTM D4169-04a | Standard Practice for Performance Testing of Shipping Containers and Systems | ||
| Biocompatibility | AAMI ANSI ISO 10993-1:2009 | EN ISO 10993-1:2009 | Biological evaluation of medical devices - Part 1: Evaluation and testing |
| EN ISO 10993-4:2002 | Biological evaluation of medical devices - Part 4: Selection of tests for interaction with blood | ||
| AAMI ANSI ISO 10993-5:2009 | EN ISO 10993-5:2009 | Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity | |
| AAMI ANSI ISO 10993-7:2008 | EN ISO 10993-7:2009 | Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals | |
| ISO 10993-10:2002/Amd1:2006 | EN ISO 10993-10:2002/Amd 1:2006 | Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity | |
| AAMI ANSI ISO 10993-11:2006 | EN ISO 10993-11:2006 | Biological evaluation of medical devices - Part 11: Tests for systemic toxicity | |
| IEC 60601-1:1988/Amd1:1991 & Amd 2:1995 | EN 60601-1:1990/Amd1:1993 & Amd 2:1995 | Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995 | |
| Electrical safety | IEC 60601-1-1:2000 | EN 60601-1-1:2001 | Medical electrical equipment - Part 1-1: General requirements for safety -Collateral standard: Safety requirements for medical electrical systems |
| AAMI ANSI IEC 60601-1-2:2001 | EN 60601-1-2:2001 | Medical Electrical Equipment - Part 1-2: General Requirements for Safety -Collateral Standard: Electromagnetic Compatibility - Requirements and Tests | |
| IEC 60601-1-4:2000 | EN 60601-1-4:1996/Amd 1:1999 | Medical electrical equipment - Part 1-4:General requirements for safety -Collateral standard: Programmable electrical medical systems | |
| EN 60601-1-6:2004 | Medical electrical equipment -- Part 1-6:General requirements for basic safety and essential performance - Collateral Standard: Usability | ||
| Sterility | AAMI ANSI ISO 11135-1:2007 | EN ISO 11135-1:2007 | Sterilization of health care products -Ethylene oxide - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices. |
| AAMI ANSI ISO 11607-1:2006 | EN ISO 11607-1:2009 | Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems | |
| AAMI ANSI ISO 11607-2:2006 | EN ISO 11607-2:2006 | Packaging for terminally sterilizedmedical devices - Part 2: Validationrequirements for forming, sealing andassembly processes | |
| ASTM F1980-07 | Standard Guide for Accelerated Aging ofSterile Barrier Systems for MedicalDevices. | ||
| AAMI ANSI ISO 11737-1:2006 | EN ISO 11737-1:2006 | Sterilization of medical devices-Microbiological methods-Part 1:Determination of the population ofmicroorganisms on product | |
| Human factors | IEC 62366:2007 | EN 62366:2008 | Medical devices - Application of usabilityengineering to medical devices |
| Software | IEC 62304:2006 | EN 62304:2006 | Medical device software - Software lifecycle processes |
Table 1: Applicable Reference Standards list
{3}------------------------------------------------
Tests were carried out on the PERFORMER HT system (equipment and disposable) to assess:
- . The ability of the system to heat and pump fluids accurately over the full range of flow rate and operating conditions;
- The ability of the system to maintain fluid temperature over the full range of fluid . flow rates:
- . The ability of the system to control the pressures, volumes and flow rate accuracy over the full range of fluid flow rate within the operating temperature and pressure range;
- . The ability of the system to detect and alarm at unsafe or ineffective operating conditions including operator errors, the failure of the system sensors, and other internal system malfunctions;
- The ability of the system to mitigate against known or predictable operator errors; .
- The ability of the system to store parameters and data in non-volatile memory. .
The PERFORMER HT performed within the specifications in all of the above tests.
Software verification and validation
PERFORMER HT software ability to meet performance specifications has been verified through the tests described above. Software control of all functions, actuators and parameters has been verified at unit, integration and system level.
Human Factors
A human factors non-clinical study was conducted in a simulated use environment with the purpose to optimize the system's user interface through human factors analysis, testing and validation. The study evaluated user- and patient-safety risks associated with the design of the PERFORMER HT system and demonstrated that the task-risks and errors that occur during use of the device have been identified, addressed and either eliminated or reduced.
21 CFR 807.92(b) (3): Conclusions
Based upon safety and performance testing and compliance with voluntary standards, RanD believes that the PERFORMER HT is substantially equivalent to the predicate device, and does not raise any new questions of safety or effectiveness.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is oriented towards the upper right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Simone Ceretti Regulatory Affairs RanD S.r.l. Via Statle 12, 62 MEDOLLA ITALY 41036
MAY - 8 2012
Re: K120026
Trade/Device Name: PERFORMER HT System Regulation Number: Unclassified Regulation Name: Thermal Fluid Warmer Regulatory Class: II Product Code: LGZ Dated: April 24, 2012 Received: April 26, 2012
Dear Ms. Ceretti:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your (1 vir 1), it may be our of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Ceretti
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony O.m
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
4. Indications for Use Statement
510(k) Number (if known): none
Device Name: PERFORMER HT system
Indications for Use:
The intended use of the PERFORMER HT is to provide isolated hyperthermic perfusion in the thoracic or peritoneal cavity by means of extra-corporeal circulation of warmed, physiologically compatible sterile solution, according to a protocol to be selected by the physician.
The device must be used by a qualified medical professional who is experienced in the operation of this or similar equipment.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kld C. Ch 5/8/12
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K120026
4 - 1
§ 880.5725 Infusion pump.
(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).