K Number
K120023
Date Cleared
2012-05-02

(120 days)

Product Code
Regulation Number
878.4340
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CoolSculpting System is a skin cooling or heating device. The device is indicated for cold-assisted lipolysis of the abdomen, as well as the flank, or "love handles" in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of the abdomen and the flank. Cooling with the device may also be used to minimize pain and thermal injury during laser and dermatological treatments and act as a local anesthetic for procedures that induce minor local discomfort.

The CoolSculpting System is also indicated for use to provide localized thermal therapy (hot or cold) to minimize pain post-trauma and post-surgery, and for temporary relief of minor aches, pains, and muscle spasms. The optional massage function can also be used for the temporary relief of minor muscle aches, pain, and spasm, for temporary improvement in local circulation and temporary reduction in the appearance of cellulite.

The ZELTIQ Gelpad facilitates thermal contact of the device with a patient's skin by mitigating minor variances in device-to-skin contact.

Device Description

The CoolSculpting System is a thermoelectric cooling and heating device that applies controlled cooling or heating to a treatment site. The device includes an optional massage feature.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the ZELTIQ CoolSculpting device, based on the provided text:

ZELTIQ CoolSculpting Device Acceptance Criteria and Study Details

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Primary Endpoint)Reported Device Performance
Photographic evaluation by three independent blinded evaluators to correctly identify the baseline image at 16-week follow-upEvaluators correctly identified the baseline image 85% of the time. (This result was statistically significant.)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Sixty (60) subjects.
  • Data Provenance: The study was a "prospective multicenter clinical study." The country of origin is not explicitly stated, but as the submission is to the FDA, it's highly likely to have included US sites or followed international regulatory standards acceptable to the FDA. It is prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Number of Experts: Three (3)
  • Qualifications of Experts: They are described as "independent blinded evaluators." No specific professional qualifications (e.g., dermatologists, plastic surgeons, radiologists) or years of experience are provided in the document.

4. Adjudication Method for the Test Set

The adjudication method is not explicitly detailed beyond "photographic evaluation by three independent blinded evaluators." It implies a consensus or majority agreement was used to determine the correct identification of the baseline image, but the specific rule (e.g., 2 out of 3, all 3, etc.) is not mentioned. It is not explicitly stated as 2+1, 3+1, or none.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

There is no MRMC comparative effectiveness study described in the provided text. The study evaluates the device's performance in achieving its intended effect, not in comparing human reader performance with or without AI assistance. The evaluators are assessing the outcome of the treatment, not diagnosing or interpreting images with AI.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

This information is not applicable as the device is a physical cooling/heating device for lipolysis, not an AI or algorithm-based imaging diagnostic tool. The "performance data" refers to the effect of the physical device on patients, not an algorithm's output.

7. Type of Ground Truth Used

The primary ground truth for the primary endpoint was expert consensus (three independent blinded evaluators judging photographic images).

Additional corroborating evidence was used:

  • Patient-reported outcomes (overall positive patient survey results).
  • Objective measurements (statistically significant ultrasound results for fat layer reduction).
  • Safety data (lipid profile and liver function tests, absence of serious adverse events).

8. Sample Size for the Training Set

The document does not mention a training set as this study is a clinical trial evaluating the therapeutic effect of a physical device, not an AI/machine learning algorithm that requires training data.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as there is no mention of a training set for an AI/machine learning algorithm.

{0}------------------------------------------------

K12.0023 1/2

MAY - 2 2012

510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

APPLICANT:ZELTIQ™ Aesthetics, Inc.4698 Willow RoadPleasanton, CA 94588
TRADE NAME:ZELTIQ CoolSculpting
COMMON NAME:Skin Cooling Device
CLASSIFICATIONNAME:Contact Cooling System for Aesthetic Use
DEVICECLASSIFICATION:Class II, 21 CFR §878.4340
PRODUCT CODE:OOK
PREDICATEDEVICE:The ZELTIQ CoolSculpting device is substantially equivalent to theZELTIQ Dermal Cooling Device, cleared as K080521, and also knownas the ZELTIQ Lipolysis System and the ZELTIQ System (K090094).

SUBSTANTIALLY EQUIVALENT TO:

The ZELTIQ CoolSculpting System is substantially equivalent to the ZELTIQ Dermal Cooling Device (K080521), also known as the ZELTIQ CoolSculpting System, which has been cleared for the indication of cold-assisted lipolysis of the flank (love handle). It is also substantially equivalent to the ZELTIQ System (K090094) which has been cleared for the indication of a skin cooling device to minimize pain and thermal injury during laser and dermatological treatments.

DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:

The CoolSculpting System is a thermoelectric cooling and heating device that applies controlled cooling or heating to a treatment site. The device includes an optional massage feature.

INDICATION FOR USE:

The CoolSculpting System is a skin cooling or heating device. The device is indicated for cold-assisted lipolysis of abdomen, as well as the flank, or "love handles" in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of the abdomen and the flank. Cooling with the device may also be used to minimize pain and thermal injury during laser and dermatological treatments and act as a local anesthetic for procedures that induce minor local discomfort.

{1}------------------------------------------------

K120623 2/2

The CoolSculpting System is also indicated for use to provide localized thermal therapy (hot or cold) to minimize pain post-trauma and post-surgery, and for temporary relief of minor aches, pains, and muscle spasms. The optional massage function can also be used for the temporary relief of minor muscle aches, pain, and spasm, for temporary improvement in local circulation and temporary reduction in the appearance of cellulite. The ZELTIQ Gelpad facilitates thermal contact of the device with a patient's skin by mitigating minor variances in device-to-skin contact.

TECHNICAL CHARACTERISTICS:

The CoolSculpting System is a thermoelectric cooling and heating device that applies controlled cooling or heating to a treatment site. The device also provides an optional massage feature.

PERFORMANCE DATA:

Animal data has been used to establish device safety and preferential selectivity of the CoolSculpting System for fat cells without damaging surrounding normal skin and muscle.

CLINICAL PERFORMANCE:

A prospective multicenter clinical study was conducted to evaluate the use of the ZELTIO System for subcutaneous fat reduction in the abdominal area. Sixty (60) subjects were treated with the ZELTIQ System at CIF 42 (a Cooling Intensity Factor setting of 42 corresponds to 72.9 mW/cm² of heat extracted from the body) over 60 minutes. The optional massage feature was not used as part of this study.

The primary endpoint was the photographic evaluation by three independent blinded evaluators at the 16 week follow-up. The evaluators correctly identified the baseline image 85% of the time and the result was statistically significant. Fat layer reduction in the abdomen was further confirmed by overall positive patient survey results and statistically significant ultrasound results. Lipid profile and liver function tests conducted at several points post-treatment showed no discernible difference from baseline. No serious adverse events were reported. Results from the clinical study support safe and effective use of the device for cold-assisted lipolysis of the abdomen.

BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE:

Clinical testing has demonstrated the ability of the CoolSculpting System to cause lipolysis of the subcutaneous fat in the abdomen in the same way that lipolysis occurs when the device is used in the flank. The device is used in the same way and with equivalent settings and use parameters as when used for treatment of the flank.

There has been no substantial change to the device design or operating principles from those of the previously cleared device for a similar indication. For that reason, bench testing was not required to demonstrate that the device is functionally equivalent to the predicate devices regarding design or operating principles.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The eagle is composed of three curved lines that form the shape of the bird's body and wings. The text is in a simple, sans-serif font.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

ZELTIO Aesthetics, Inc. % Mr. Louis-Pierre Marcoux Director of Regulatory Affairs 4698 Willow Road Pleasanton, California 94115

JUN - 1 2012

Re: K120023

Trade/Device Name: ZELTIQ CoolSculpting Regulation Number: 21 CFR 878.4340 Regulation Name: Contact cooling system for aesthetic use Regulatory Class: II Product Code: OOK Dated: April 20, 2012 Received: April 23, 2012

Dear Mr. Marcoux:

This letter corrects our substantially equivalent letter of May 2, 2012.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

Page 2 - Mr. Louis-Pierre Marcoux

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

y yours,

Melkerson

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

4. INDICATIONS FOR USE STATEMENT

510(k) Number (if known): K120023

Device Name: ZELTIO CoolSculpting

Indications for Use:

The CoolSculpting System is a skin cooling or heating device. The device is indicated for cold-assisted lipolysis of the abdomen, as well as the flank, or "love handles" in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of the abdomen and the flank. Cooling with the device may also be used to minimize pain and thermal injury during laser and dermatological treatments and act as a local anesthetic for procedures that induce minor local discomfort.

The CoolSculpting System is also indicated for use to provide localized thermal therapy (hot or cold) to minimize pain post-trauma and post-surgery, and for temporary relief of minor aches, pains, and muscle spasms. The optional massage function can also be used for the temporary relief of minor muscle aches, pain, and spasm, for temporary improvement in local circulation and temporary reduction in the appearance of cellulite.

The ZELTIO Gelpad facilitates thermal contact of the device with a patient's skin by mitigating minor variances in device-to-skin contact.

Prescription Use x (Part 21 CFR 801 Subpart D)

· 510(k) Number

AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil R. Ogden for mxm
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices

§ 878.4340 Contact cooling system for aesthetic use.

(a)
Identification. A contact cooling system for aesthetic use is a device that is a combination of a cooling pad associated with a vacuum or mechanical massager intended for the disruption of adipocyte cells intended for non-invasive aesthetic use.(b)
Classification. Class II (special controls). The special controls for this device is FDA's “Guidance for Industry and FDA Staff; Class II Special Controls Guidance Document: Contact Cooling System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.