K Number
K120019

Validate with FDA (Live)

Device Name
COLLAGEN RIBBON
Date Cleared
2012-02-23

(51 days)

Product Code
Regulation Number
878.3300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Collagen Ribbon Tissue Matrix is intended to reinforce soft tissue where weakness exists, specifically, for the reinforcement of soft tissue repaired by sutures or suture anchors during tendon repair surgery, including reinforcement of the rotator cuff, patellar, Achilles, biceps, quadriceps, peroneal, posterial tibial, and other tendons. The Collagen Ribbon Tissue Matrix is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair. Sutures used to repair the tear and suture or bone anchors used to attach the tissue to the bone provide biomechanical strength for the tendon repair.

Device Description

The subject Collagen Ribbon Tissue Matrix is a narrow, ribbon-like version of the predicate WMT Collagen Dermal Matrix (rebranded BIOTAPE XM® Tissue Matrix - K073219). The ribbon-like dimensions of the subject device allow for new surgical techniques compared to the predicate without changing the intended use-reinforcement of soft tissue in orthopedic applications. The subject Collagen Ribbon is the identical source material to the predicate and is manufactured and sterifized in the same manner.

AI/ML Overview

The provided text describes a medical device called the "Collagen Ribbon Tissue Matrix" and its substantial equivalence to a predicate device (WMT Collagen Dermal Matrix, K073219). The submission focuses on non-clinical evidence to demonstrate this equivalence. Therefore, the questions related to clinical studies, human reader performance, training and test set ground truth, and expert adjudication are not applicable as no such studies were performed or described for this 510(k) submission.

Here's a breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance:

The document primarily focuses on demonstrating substantial equivalence to a predicate device rather than establishing new acceptance criteria for a novel device. The "acceptance criteria" are implied by the performance of the predicate device and the new device's ability to maintain similar characteristics.

Acceptance Criteria (Implied by Predicate)Reported Device Performance (Collagen Ribbon Tissue Matrix)
Biocompatibility: Absence of cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, implantation reactions (2wk & 13wk), subchronic & chronic toxicity, carcinogenicity, hemolysis, genotoxicity.Passed all described biocompatibility testing.
Material Composition: Decellularized, intact collagen matrix.Histology results indicated an identical decellularized, intact collagen matrix.
Mechanical Properties: Suitable for orthopedic applications (tensile and suture retention testing).Maintains the same minimum performance characteristics as the predicate device (confirmed through tensile and suture retention testing).
Rehydration Properties: Maintenance of performance after rehydration.Rehydration testing confirms the same minimum performance characteristics as the predicate device.
Sterilization: Same as predicate.Manufactured and sterilized in the same manner as the predicate.
Packaging: No alteration in performance due to packaging changes.Functional and in vivo performance testing confirms that the addition of a second Tyvek barrier to packaging does not alter performance of the material.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: Not explicitly stated for specific tests. The document refers to "biocompatibility testing," "animal results," "histology results," "bench testing," "mechanical testing," and "rehydration testing." However, the number of samples or animals used for each test is not detailed.
  • Data Provenance: The data primarily comes from non-clinical testing conducted by the manufacturer, Wright Medical Technology, Inc., for both the subject device and the predicate device (K073219). The tests were done in compliance with U.S. Food and Drug Administration regulations.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This submission relies on non-clinical laboratory and bench testing, not on human expert interpretations of medical images or data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This submission relies on non-clinical laboratory and bench testing, not on adjudicated expert opinions.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. No MRMC study was conducted or described, as this is a material science device, not an AI or imaging device involving human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithmic device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the non-clinical tests would be the established scientific and regulatory standards for material performance and safety. For example:

  • Biocompatibility: Compliance with ISO standards and FDA regulations (21 CFR, Part 58) for various toxicity and reactivity assessments.
  • Material Characteristics: Histological analysis confirming specific cellular and matrix structures.
  • Mechanical Performance: Standard test methods (e.g., ASTM) for tensile strength, suture retention, etc.

8. The sample size for the training set

Not applicable. No training set is mentioned as this device does not involve machine learning or AI.

9. How the ground truth for the training set was established

Not applicable. No training set is mentioned.

{0}------------------------------------------------

FEB 2 3 2012

K120019 Page 1/2

510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of the Collagen Ribbon Tissue Matrix.

(a)(1). Submitted By:Wright Medical Technology, Inc.
5677 Airline Road
Arlington, TN 38002
Date:December 29, 2011
Contact Person:Leslie Fitch
Regulatory Affairs Specialist
(901) 867-4120
(a)(2). Proprietary Name:Collagen Ribbon Tissue Matrix
Common Name:Animal-derived, surgical mesh
Classification Name and Reference:21 CFR 878.3300 - Class II
Device Product Code and Panel Code:FTM: Surgical Mesh
(a)(3). Predicate Device:K073219 - WMT Collagen Ribbon

(a)(4). Device Description

The subject Collagen Ribbon Tissue Matrix is a narrow, ribbon-like version of the predicate WMT Collagen Dermal Matrix (rebranded BIOTAPE XM® Tissue Matrix - K073219). The ribbon-like dimensions of the subject device allow for new surgical techniques compared to the predicate without changing the intended use-reinforcement of soft tissue in orthopedic applications. The subject Collagen Ribbon is the identical source material to the predicate and is manufactured and sterifized in the same manner.

(a)(5). Intended Use

The Collagen Ribbon Tissue Matrix is intended to reinforce soft tissue where weakness exists, specifically, for the reinforcement of soft tissue repaired by sutures or suture anchors during tendon repair surgery, including reinforcement of the rotator cuff, patellar, Achilles, biceps, quadriceps, peroneal, posterial tibial, and other tendons. The Collagen Ribbon Tissue Matrix is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair. Sutures used to repair the tear and suture or bone anchors used to attach the tissue to the bone provide biomechanical strength for the tendon repair.

The indications for Collagen Ribbon are the same as the legally marketed predicate device with the addition of 2 additional specific tendons. The specific tendons were evaluated for safety and effectiveness through a clinical literature review. A summary of the literature review is provided in the Premarket Notification.

(a)(6). Technological Characteristics Comparison

{1}------------------------------------------------

K120019 Page 2/2

The technological characteristics of the Collagen Ribbon Tissue Matrix are substantially equivalent to technological characteristics of the predicate identified in this 510(k) submission. The subject Collagen Ribbon is the identical source material to the predicate and is manufactured and sterilized in the same manner. The ribbon-like dimensions of the subject device allow for new surgical techniques compared to the predicate without changing the intended use-reinforcement of soft tissue in orthopedic applications.

(b)(1). Substantial Equivalence - Non-Clinical Evidence

Due to the similarities of the subject device with the predicate, biocompatibility, animal. histological, and bench testing results from 510(k) K073219 are applicable to this submission. The subject material has passed all biocompatibility testing including: cvtotoxicity, maximization test, intracutaneous reactivity, systemic toxicity implantation (2wk & 13 wk), pyrogenicity, subchronic toxicity & chronic toxicity, carcinogenicity, hemolysis, and genotoxicity. The biocompatibility testing that was previously conducted was in compliance with U.S. Food and Drug Administration regulations set forth in 21 CFR, Part 58. Exempt from Good Laboratory Practice were the characterization and stability of the test article, according to 21 CFR, Parts 58.105 and 58.113. Animal results indicated cellular infiltration and revascularization characteristics. Histology results indicated that the Collagen Ribbon is a decellularized, intact collagen matrix.

A Design Failure Modes and Effects Analysis was used to identify appropriate testing to demonstrate specification conformance for Collagen Ribbon material. Mechanical testing presented in K073219 characterized the suitability of this material in orthopedic applications. Functional and in vivo performance testing confirms that the addition of a second Tyvek barrier to packaging does not alter performance of the material. In addition, mechanical and rehydration testing of Collagen Ribbon demonstrate that Collagen Ribbon maintains the same minimum performance characteristics as the predicate device.

(b)(2). Substantial Equivalence - Clinical Evidence

N/A

(b)(3). Substantial Equivalence - Conclusions

The subject Collagen Ribbon is the identical source material to the predicate and is manufactured and sterilized in the same manner. Substantial equivalence is confirmed through tensile and suture retention testing, rehydration testing, and surgical technique validation studies. The data and evidence presented in this Premarket Notification, the subject device maintains the same minimum performance characteristics as the predicate device.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter of the circle. Inside the circle is an abstract symbol that resembles an eagle or bird with outstretched wings. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-W066-G609 Silver Spring, MD 20993-0002

FEB 2 3 2012

Wright Medical Technology, Inc. % Ms. Leslie Fitch 5677 Airline Road Arlington, Tennessee 38002

Re: K120019

Trade/Device Name: Collagen Ribbon Tissue Matrix Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTM Dated: December 29, 2011 Received: January 03, 2012

Dear Ms. Fitch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the-Federal Register-

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

· Page 2 - Ms. Leslie Fitch

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set de rice related duverse overals) (2) regulation (21 CFR Part 820); and if applicable, the electronic form in the quality by events (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n you desire spective ad resorted to read on the securities CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part note the regulations regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Tou may overn of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,
Mark McMillan

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): K120019

Device Name: Collagen Ribbon Tissue Matrix

Indications for Use:

The Collagen Ribbon Tissue Matrix is intended to reinforce soft tissue where weakness exists, specifically, for the reinforcement of soft tissue repaired by sutures or suture anchors during tendon repair surgery, including reinforcement of the rotator cuff, patellar, Achilles, dailing tendon ropan ourgal, posterial tibial, and other tendons. The Collagen Ribbon Tissue Marrix is not intended to replace normal body structure or provide the full mechanical . strength to support tendon repair. Sutures used to repair the tear and suture or bone anchors used to attach the tissue to the bone provide biomechanical strength for the tendon repair.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1

Daniel Kune lu MXM

(Division Sign-Oil) (Division Sign-O-1). Orthopedic, and Restorative Devices

510(k) Number K120019

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.