(131 days)
The Glucommander System is a glycemic management tool intended to evaluate current and cumulative patient blood glucose values and coupled with patient information including age, weight and height, and, based on the aggregate of these measurement parameters, whether one or many, recommend an IV dosage of insulin, glucose or saline or a subcutaneous basal and bolus insulin dosing recommendation to adjust and maintain the blood glucose level towards a configurable clinician determined target range.
The Glucommander System is indicated for use in adult and pediatric (ages 2 - 17 years) patients.
The Glucommander System logic is not a substitute for, but rather an adjunct to clinical reasoning. The measurements and calculations generated are intended to be used by qualified and trained medical personnel in evaluating patient conditions in conjunction with clinical history, symptoms, and other diagnostic measurements, as well as the medical professional's clinical judgment. No medical decision should be based solely on the recommended guidance provided by this software program.
The Glytec, LLC Glucommander System is a software device used to evaluate current and cumulative patient blood glucose values, and, based on the aggregate of those measurements, whether one or many, recommend an IV dosage of insulin, glucose or saline to direct the blood glucose level towards a predetermined target range. Once that target blood glucose range has been reached, the system's function is to recommend a titration of insulin, glucose, and saline for the purpose of maintaining the patient's blood glucose level in that target range. The system is programmed to provide intravenous dosing of glucose, saline, and insulin, as well as subcutaneous dosing of insulin for both pediatric (ages 2 - 17 years) and adult patients.
Acceptance Criteria and Study for Glytec LLC, Glucommander™ System (K113853)
This 510(k) submission (K113853) is for a modification to an existing device, the GlyTec, LLC G+ System (K101344). The modification is the addition of a pediatric protocol that considers the weight of the patient. As such, the submission focuses on the performance of the modified software and its substantial equivalence to the predicate device, rather than establishing new acceptance criteria or conducting a comprehensive clinical study to prove general device performance.
Based on the provided document, the following can be inferred regarding acceptance criteria and supporting studies:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document does not explicitly state specific quantitative acceptance criteria or performance metrics for the modified Glucommander™ System (K113853) in a table format.
Instead, the submission relies on demonstrating substantial equivalence to the predicate device (K101344) and emphasizes that the modifications introduced (pediatric protocol) do not raise new safety or effectiveness concerns.
The reported "performance" is qualitative, focusing on functional aspects:
Acceptance Criterion (Inferred) | Reported Device Performance |
---|---|
Functional Equivalence to predicate device (K101344) for adult patients | "Other user screen labeling such as the function of screen buttons, the ability to add additional patient data and user information, the ability to password protect information and other similar security features, data handling performance, user controlled automatic system shutdown and software update notifications are unchanged." |
Correct Calculation of Initial Multiplier for pediatric patients based on weight | "An additional input validation on the date of birth field identifies pediatric patients and calculates the patient's initial multiplier based on the patient's weight..." |
Appropriate User Interface Design for adult/pediatric identification | "Initial starting parameters and the user interface design were modified to allow for the easy identification and calculation of initial multiplier based on a pediatric patient's weight that clearly indicate the order set as either adult or pediatric." |
Software Functionality as intended | "The results of usability testing of representative users of the device, software testing and performance testing of the device demonstrate the device functions as intended." |
No New Safety or Effectiveness Concerns compared to predicate device | "Glytec, Inc. believes no differences exist between this system and the predicate system that raise any new safety.or effectiveness concerns." |
2. Sample Size for the Test Set and Data Provenance
The document mentions "usability testing of representative users of the device, software testing and performance testing of the device." However, it does not specify the sample size used for these tests or the data provenance (e.g., country of origin, retrospective/prospective nature of data).
Given the nature of the submission (modification to an existing device for substantial equivalence), it's likely these tests were focused on verifying the correct implementation of the new pediatric protocol and ensuring no regressions in existing functionalities, rather than a large-scale clinical validation study.
3. Number of Experts and Qualifications for Ground Truth
The document does not mention the use of experts to establish a "ground truth" for a test set. The evaluation appears to be based on internal testing (usability, software, performance) to confirm the new features work as designed and that overall functionality remains equivalent to the predicate.
4. Adjudication Method for the Test Set
Since there is no mention of external experts establishing ground truth for a test set, there is no information on an adjudication method.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study was done or reported. The submission focuses on substantial equivalence based on technical modifications rather than a comparative study of human reader performance with and without AI assistance.
6. Standalone Performance Study
A standalone performance study focused on the algorithm only (without human-in-the-loop performance) was implicitly done through "software testing and performance testing of the device." These tests would assess the internal logic and calculations of the Glucommander™ System, including the new pediatric protocol's ability to correctly calculate the initial multiplier. However, the specific metrics and results of such a standalone study are not detailed.
7. Type of Ground Truth Used
For the "software testing and performance testing," the ground truth would likely be based on:
- Engineering specifications and design documents: Verifying that the software calculations and logic conform to the intended design, especially for the new pediatric weight-based multiplier.
- Predetermined correct outputs: For specific test cases, the expected correct IV dosage recommendations or multiplier calculations would be known and used to validate the device's output.
- Clinical rationale: The pediatric protocol's design would be based on clinical understanding of pediatric glycemic management.
The document does not explicitly state that pathology, outcomes data, or expert consensus (external to the development process) were used as "ground truth" in a formal sense for this specific submission.
8. Sample Size for the Training Set
The document does not provide information on the sample size for a training set. As this is a software device for recommending insulin dosages based on blood glucose values and patient parameters, it's likely built on established algorithms and clinical guidelines, rather than relying on a large "training set" in the context of machine learning for image classification, for example. If any machine learning elements were involved, details are not provided.
9. How the Ground Truth for the Training Set Was Established
Given the lack of information on a training set, the method for establishing its ground truth is also not described.
§ 868.1890 Predictive pulmonary-function value calculator.
(a)
Identification. A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.(b)
Classification. Class II (performance standards).