K Number
K113844
Device Name
RADIFORCE RX240
Date Cleared
2012-02-27

(61 days)

Product Code
Regulation Number
892.2050
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The RadiForce RX240 is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The device does not support the display of mammography images for diagnosis.

Device Description

The RadiForce RX240 is a color LCD monitor for viewing medical images other than those of mammography. The matrix size (or resolution) of the panel is 1200 x 1600 pixels (2MP) with a pixel pitch of 0.270 mm. Since factory calibrated display modes, each of which is characterized by a specific tone curve (including DICOM GSDF), a specific luminance range and a specific color temperature, are stored in lookup tables within the monitor, the tone curve is e.g. DICOM compliant regardless of the display controller used. RadiCS is application software to be installed in each workstation offering worry-free quality control of the diagnostic monitors including RX240 based on several QC guidelines. The RadiCS and its subset, RadiCS LE are included in this 510(k) submission as an accessory to the RadiForce RX240.

AI/ML Overview

The provided text describes a 510(k) submission for a medical monitor, the EIZO RadiForce RX240. However, it does not contain information about acceptance criteria for device performance, nor a study proving it meets such criteria in the context of AI/human reader performance or standalone algorithm efficacy.

The document focuses on demonstrating substantial equivalence to a predicate device (RadiForce RS210) based on technological characteristics and intended use. The performance testing mentioned is "bench testing" comparing image quality characteristics of the proposed device with the predicate device, and validation of the overall design against safety and EMC standards. It explicitly states: "None of the tests revealed behaviors inconsistent with the expected performance."

Therefore, based on the provided input, I cannot fill out the requested table or sections related to acceptance criteria, AI performance studies, or ground truth establishment for such studies.

Here's a breakdown of what can be extracted and what is missing:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Not provided in the document. The document states "None of the tests revealed behaviors inconsistent with the expected performance," which is a general statement rather than specific, measurable acceptance criteria.Technological Characteristics Comparison with Predicate Device: - Matrix size: 1200 x 1600 pixels (same) - Active area: 324.0 mm x 432.0 mm (same) - DICOM calibrated luminance: 400 cd/m² (higher than predicate's 150 cd/m²) - Typical maximum luminance: 760 cd/m² (higher than predicate's 300 cd/m²) - Backlight: LED (newly employed, mercury-free, less power, slower deterioration) - Image display: In accordance with DICOM GSDF - Interface: Digital (DVI or DisplayPort only, no analog) - QC Software: Same as predicate - Backlight Sensor (BS): Same as predicate - Built-in Front Sensor (IFS): Enables automatic grayscale calibration (new feature) - Overall design: Validated in accordance with internationally recognized safety and EMC standards by independent and in-house facilities.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not applicable / Not provided. This information is relevant for studies involving medical image data and diagnostic performance (e.g., AI algorithms). The provided document describes a hardware device (monitor) and its quality control software. The performance testing mentioned is at the device specification level (e.g., luminance, resolution, compliance with standards), not clinical performance or diagnostic accuracy using a test set of medical images.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable / Not provided. As above, this pertains to clinical validation studies involving diagnostic interpretation, which is not the focus of this 510(k) submission for a medical display monitor.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable / Not provided.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. The document does not describe an MRMC study or any AI component designed to assist human readers. The device is a display monitor.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • No. The document does not describe an algorithm for standalone performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not applicable / Not provided. The "ground truth" for this device's performance relates to its technical specifications and compliance with display standards (e.g., DICOM GSDF, safety/EMC standards), not clinical diagnostic accuracy.

8. The sample size for the training set

  • Not applicable / Not provided.

9. How the ground truth for the training set was established

  • Not applicable / Not provided.

{0}------------------------------------------------

K113844

Image /page/0/Picture/1 description: The image shows the logo for EIZO. The logo consists of a stylized, geometric shape to the left of the word "EIZO". The geometric shape is a square with a checkered pattern inside. The word "EIZO" is written in a bold, sans-serif font.

EIZO NANAO CORPORATION, 153 Shimokashiwano, Hakusan, Ishikawa

U.S. Food and Drug Administration Center for Devices and Radiological Health Document Mail Center - WO66-G609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002

Name Department

Hiroaki Hashimoto Medical System Standards

Telephone Fax E-Mail

+81 (76) 274-2468 +81 (76) 274-2484 hiro@nanao.co.jp

Date

December 19th, 2011

Traditional 510(k) Summary (in accordance with 21 CFR 807.92)

    1. Date of Summary
      December 19th, 2011
    1. Company EIZO NANAO CORPORATION 153 Shimokashiwano, Hakusan Ishikawa 924-8566 Japan
    1. Authorized Contact Person Hiroaki Hashimoto

4. Device Information

  • RadiForce RX240 Trade Name/Model: .
  • 2MP Color LCD Monitor Common Name: .
  • Classification Name: System, Image Processing, Radiological .
  • 21 CFR 892.2050, Product Code LLZ Classification Number: .

5. Predicate Device

  • Color LCD Monitor, RadiForce RS210 (K092613) �

6. Device Description

The RadiForce RX240 is a color LCD monitor for viewing medical images other than those of mammography. The matrix size (or resolution) of the panel is 1200 x 1600 pixels (2MP) with a pixel pitch of 0.270 mm.

Since factory calibrated display modes, each of which is characterized by a specific tone curve (including DICOM GSDF), a specific luminance range and a specific color temperature, are stored in lookup tables within the monitor, the tone curve is e.g. DICOM compliant regardless of the display controller used.

FEB 2 7 2012

{1}------------------------------------------------

RadiCS is application software to be installed in each workstation offering worry-free quality control of the diagnostic monitors including RX240 based on several QC guidelines. The RadiCS and its subset, RadiCS LE are included in this 510(k) submission as an accessory to the RadiForce RX240.

7. Intended Use

The RadiForce RX240 is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The device does not support the display of mammography images for diagnosis.

8. Technological Characteristics

The RadiForce RX240 can be said to have at least the same display performances as those of the predicate device by default due to the following reasons:

  • The matrix sizes (1200 x 1600) and the active area sizes (324.0 mm x 432.0 mm) of a. the LCD panels used by the both devices are the same.
  • The DICOM calibrated luminance (400 cd/m²) and the typical maximum luminance b. (760 cd/m2) is higher than that of the predicate device (150 cd/m2, 300 cd/m2). The higher luminance to be maintained constantly was realized by the employment of LED backlight deteriorating more slowly than conventional CCFL backlights.
  • The LED backlight was newly employed instead of CCFL backlight because it is C. mercury-free, consumes less power and deteriorates more slowly. We have not recognized any adverse effects of the LED backlight on the quality of displayed images. Refer to section 18 "Performance Testing - Bench" where several image quality characteristics of the proposed device are compared with those of the predicate device.
  • d. The both devices display images in accordance with DICOM GSDF by default utilizing the factory calibrated display mode stored in lookup tables inside of them.
  • Analog video interface is not supported by the proposed device. The quality of e. displayed images is usually better via digital interfaces like DVI or DisplayPort than via analog ones.

As for the maintenance, the same QC software is used for the both devices and the implementation of the Backlight Sensor (BS) stabilizing the backlight is also the same.

As for built-in sensors, in addition to BS common to the both devices, RX240 has three kinds of sensors. However, only the Built-in Front Sensor (IFS) has something to do with the maintenance or the calibration; the Presence Sensor (PS) detects the absence of the user to trigger the power saving mode of the monitor and the Ambient Light Sensor (ALS) is used to measure the ambient light by lx. The IFS enables automatic grayscale calibration by measuring the luminance at the screen surface. Without IFS, the grayscale calibration process requires human intervention and the use of an external sensor. The accuracy data of the calibration with

{2}------------------------------------------------

external sensors and that with the IFS is included in section 16.9 "Verification and Validation Documentation".

The overall design of the RadiForce RX240 was validated in accordance with internationally recognized safety and EMC standards by independent testing facilities and in-house ones. Besides, EIZO NANAO CORPORATION performed a range of system and performance tests to ensure that the RadiForce RX240 performs in accordance with its specifications. None of the tests revealed behaviors inconsistent with the expected performance.

9. Conclusion

The 2MP Color LCD Monitor, RadiForce RX240 is substantially equivalent to the predicate device with respect to technical characteristics, application and intended use. The specifications of the primary component employed by the proposed device are the same as those of the predicate device and other differences have been independently validated. Any differences between the devices do not affect safety or effectiveness.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure with three arms or lines extending upwards, representing health and human services. To the left of the figure is a circular inscription that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in capital letters.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

FEB 2 7 2012

Mr. Hiroaki Hashimoto Manager EIZO NANAO Corporation 153 Shimokashiwano 924-8566 HAKUSAN ISHIKAWA JAPAN

Re: K113844

Trade/Device Name: 2MP Color LCD Monitor, RadiForce RX240 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 26, 2011 Received: December 28, 2011

Dear Mr. Hashimoto:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{4}------------------------------------------------

medical device-related adverse-events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Form

510(k) Number (if known):

2MP Color LCD Monitor, RadiForce RX240 Device Name:

Indications for Use:

The RadiForce RX240 is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The device does not support the display of mammography images for diagnosis.

X Over-The-Counter Use Prescription Use AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Mary Slaster

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K113844

Page 1 of 1

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).