Intended Use

The Signal Gear Urethral Catheter Electrode is intended for mucosal surface stimulation/recording from the external urinary sphincter for use in conjunction with urodynamic evaluation of the patient.

Device Description

The Signal Gear Urethral Catheter Electrode is a single patient use, disposable sterile device. It consists of an adhesive, highly flexible, polyurethane film substrate that wraps around the urethral (Foley) catheter. The conductive portion of the electrode is printed onto the polyurethane film substrate in a pattern that, when the electrode is wrapped around the catheter, leaves two conductive contact surfaces forming the two electrical contacts of the electrode. The remainder of the conductive pattern is overwrapped by an additional layer of polyurethane film during manufacture in order to insulate the traces. The conductive ink used is silver, with the exception of the two electrical contacts of the electrode, which have an additional printed layer of gold ink.

Lead wires are attached that terminate in a safety connector that cannot be connected to an AC power outlet. The Electrode connects to the user's electrodiagnostic equipment. The electrode is to be used under the supervision of a physician. When the urethral catheter with the applied electrode is inserted in the human urethra, the exposed two electrical contacts of the electrode contact the mucosal lining of the urethra in the location of the external urinary sphincter muscle, and the end of the electrode with leadwire attachments are outside of the urethral meatus.

AI/ML Overview

The provided 510(k) summary for the Neurovision Medical Products Inc. Signal Gear Urethral Catheter Electrode (K113771) describes a non-clinical study to support substantial equivalence. It does not present acceptance criteria or detailed performance data for clinical use, nor does it describe a study involving human subjects or AI.

Here's an analysis based on your request, highlighting what information is not present in the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

This information is not present in the provided document. The 510(k) summary focuses on non-clinical testing to demonstrate substantial equivalence to a predicate device, rather than defining specific performance metrics or acceptance criteria related to its intended physiological function (mucosal surface stimulation/recording) in a clinical setting.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not applicable, as this was a non-clinical study.
  • Data Provenance: Not applicable, as this was a non-clinical study.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

Not applicable, as this was a non-clinical study. The "ground truth" in this context would be engineering specifications and standards for electrical, biocompatibility, and sterilization testing.

4. Adjudication Method for the Test Set

Not applicable, as this was a non-clinical study.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study was not done and is not mentioned. This type of study is relevant for diagnostic devices where human readers interpret images or data, often with AI assistance. The Signal Gear Urethral Catheter Electrode is a medical instrument for stimulation/recording, not a diagnostic imaging device.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

No, a standalone (algorithm only) study was not done and is not mentioned. The device is an electrode for use in conjunction with urodynamic evaluation, implying human operation and interpretation of data. There is no mention of an "algorithm" as a component of this device's performance.

7. Type of Ground Truth Used

For the non-clinical testing, the "ground truth" was based on:

  • Engineering specifications and quality standards for manufacturing.
  • Established international and industry standards for biocompatibility (ISO 10993-5:2009, ISO 10993-10:2010).
  • Standards for accelerated aging (ASTM F1980-07).
  • Standards for packaging validation (UNI EN ISO 868-5, 11607-1, DIN 58953-6).
  • Standards for bioburden and sterility (UNI EN ISO 11737-1, 11737-2).

8. Sample Size for the Training Set

Not applicable. This device does not involve a "training set" in the context of machine learning or AI.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As there is no training set mentioned, the ground truth for it was not established.


Summary of the Study that Proves the Device Meets Acceptance Criteria (Non-Clinical):

The Neurovision Medical Products Inc. Signal Gear Urethral Catheter Electrode underwent several non-clinical tests to demonstrate its safety and performance, and thus its substantial equivalence to the predicate device (Dantec 13L40 St Mark's Pudendal Electrode).

The study involved:

  • Inspection for defects: Manufacturing, workmanship, and packaging.
  • Electrical continuity testing: To ensure proper electrical function.
  • Lead wire pull testing: To assess the mechanical strength of the lead wire attachment.
  • Biocompatibility testing:
    • In vitro cytotoxicity: According to ISO 10993-5:2009.
    • Irritation and delayed-type hypersensitivity: According to ISO 10993-10:2010.
  • Accelerated aging: Conducted according to ASTM F1980-07 to predict shelf life.
  • Packaging validation tests: Conducted according to UNI EN ISO 868-5, 11607-1, and DIN 58953-6 to ensure package integrity and maintain sterility.
  • Bioburden and sterility tests: Conducted according to UNI EN ISO 11737-1 and 11737-2 to confirm the device is sterile when packaged.

The document indicates that these non-clinical tests were sufficient to demonstrate that "No new questions of safety or effectiveness are raised," supporting its substantial equivalence to the predicate device. However, specific acceptance criteria for each test (e.g., minimum pull strength, maximum cytotoxicity score) and the quantitative results are not provided in this summary. The FDA's 510(k) clearance implies that the agency found these non-clinical tests adequate for demonstrating substantial equivalence based on the provided data.

{0}------------------------------------------------

K113771

PAGE 1 OF 2

Image /page/0/Picture/1 description: The image shows the logo for Neurovision Medical. The word "neurovision" is in large, bold, black letters. Below it, there are two horizontal lines with a small EKG waveform in between, followed by the word "MEDICAL" in smaller, bold, black letters.

Traditional 510k -Signal gear Signal Gear Urethral Catheter Electrode

5. 510(k) Summary

In accordance with Title 21 of the Code of Federal Regulations, Part 807 and in particular 21 CFR §807.92, the following summary of information is provided:

Applicant Information

Christine Vergely Neurovision Medical Products Inc. 2225 Sperry Ave., Suite 1000 Ventura, CA 93003 Telephone Number: 866-815-6999 Fax: 413-410-4579 Email Address: christie@neurovisionmedical.com

Device Identification

Trade or Proprietary Name: Signal Gear Urethral Catheter Electrode Device classification: 21 CFR 876.1620, Urodynamics Measurement System Device class: Class II Product code: FAP

Predicate Devices

The Signal Gear Urethral Catheter Electrode is substantially equivalent to the following predicate device previously distributed commercially in the U. S .:

  • K874758 Dantec Electronic, Inc. St. Mark's Pudendal Electrode .

Device Description

The Signal Gear Urethral Catheter Electrode is a single patient use, disposable sterile device. It consists of an adhesive, highly flexible, polyurethane film substrate that wraps around the urethral (Foley) catheter. The conductive portion of the electrode is printed onto the polyurethane film substrate in a pattern that, when the electrode is wrapped around the catheter, leaves two conductive contact surfaces forming the two electrical contacts of the electrode. The remainder of the conductive pattern is overwrapped by an additional layer of polyurethane film during manufacture in order to insulate the traces. The conductive ink used is silver, with the exception of the two electrical contacts of the electrode, which have an additional printed layer of gold ink.

Lead wires are attached that terminate in a safety connector that cannot be connected to an AC power outlet. The Electrode connects to the user's electrodiagnostic equipment. The electrode is to be used under the supervision of a physician. When the urethral catheter with the applied electrode is inserted in the human urethra, the exposed two electrical contacts of the electrode contact the mucosal lining of the urethra in the location of the external urinary sphincter muscle, and the end of the electrode with leadwire attachments are outside of the urethral meatus.

Intended Use

The Signal Gear Urethral Catheter Electrode is intended for mucosal surface stimulation/recording from the external urinary sphincter for use in conjunction with urodynamic evaluation of the patient.

DEC 1 3 2012

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for Neurovision Medical. The word "neurovision" is in large, bold, black letters. Below the word "neurovision" is a black line with a small EKG waveform in the middle, followed by the word "MEDICAL" in smaller, black letters.

atheter Electrode Traditional 510k -Signal gear Signal Gear Urethral

This device is indicated for use in adults, as well as in pediatric patients ages 2 and older.

Technological Characteristics of Device in Relation to Predicate Devices

The Signal Gear Urethral Catheter Electrode is an attachment electrode designed to attach to a urethral (Foley) catheter. It is a printed electrode on a flexible base designed for surface stimulation of and surface recording from physiologic tissue adjacent to the urethra such as the external urinary sphincter muscle. It is self adhesive and is attached during manufacture by wrapping around a urethral (Foley) catheter. It is disposable for single use.

The predicate Dantec 13L40 St Mark's Pudendal Electrode is an attachment electrode, printed on a flexible base designed to stimulate the pudendal nerve and to record from the anal sphincter muscle. It is self adhesive and attaches by wrapping around the finger of a surgical glove. It is disposable for single use.

The technological characteristics of the Signal Gear Urethral Catheter Electrode are substantially equivalent to the predicate device. No new questions of safety or effectiveness are raised.

Summary of Non-Clinical Testing in Support of Substantial Equivalence

Nonclinical testing included inspection for defects in manufacturing, workmanship and packaging, electrical continuity testing and lead wire pull testing.

Biocompatibility was tested according to ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity and ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and delayed-type Hypersensitivity.

Accelerated aging was conducted according to ASTM F1980-07.

Packaging validation tests were conducted according to the methods described in Standards UNI EN ISO 868-5 and 11607-1 and European Standard DIN 58953-6.

The bioburden and sterility tests were conducted according to the methods described in Standards UNI EN ISO 11737-1 and 11737-2.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 13, 2012

Neurovision Medical Products, Inc. % Ms. Christine Vergély Regulatory Manager 2225 Sperry Avenue, Suite 1000 VENTURA CA 93003

Re: K113771

Trade/Device Name: Signal Gear Urethral Catheter Electrode Regulation Number: 21 CFR 876.1620 Regulation Name: Urodynamics measurement system Regulatory Class: II Product Code: FAP Dated: November 29, 2012 Received: November 29, 2012

Dear Ms. Vergély:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{3}------------------------------------------------

Page 2 – Ms. Christine Vergély

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please i go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Herbert R. Lerner

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Neurovision Medical. The word "neurovision" is written in bold, black letters. Below the word, there are two horizontal lines with a small graphic between them, and the word "MEDICAL" is written in smaller, black letters to the right of the graphic.

Traditional 510k -Signal gear Signa r Electrode

Indications for Use

510(k) Number: K113771 (Urethral Surface Electrode)

Device Name: Signal Gear Urethral Catheter Electrode

Indications for Use:

The Signal Gear Urethral Catheter Electrode is intended for mucosal surface stimulation/recording from the external urinary sphincter for use in conjunction with urodynamic evaluation of the patient.

This device is indicated for use in adults, as well as in pediatric patients ages 2 and older.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert P. Lerner

(Division Sign-On)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K113771

§ 876.1620 Urodynamics measurement system.

(a)
Identification. A urodynamics measurement system is a device used to measure volume and pressure in the urinary bladder when it is filled through a catheter with carbon dioxide or water. The device controls the supply of carbon dioxide or water and may also record the electrical activity of the muscles associated with urination. The device system may include transducers, electronic signal conditioning and display equipment, a catheter withdrawal device to enable a urethral pressure profile to be obtained, and special catheters for urethral profilometry and electrodes for electromyography. This generic type of device includes the cystometric gas (carbon dioxide) device, the cystometric hydrualic device, and the electrical recording cystometer, but excludes any device that uses air to fill the bladder.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.