K Number
K113683
Manufacturer
Date Cleared
2012-01-12

(28 days)

Product Code
Regulation Number
880.5440
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

LIFESHIELD™ VISION™ Infusion sets are intended for the delivery of fluids including but not limited to blood and blood products from a container into a patients vascular system.

Device Description

The LIFESHIELD™ VISION™ infusion sets, are intended for use as gravity sets or with dedicated Hospira Infusion Pumps. These devices are obtainable in custom lengths and component options according to facility needs and physician preference, Hospira infusion sets are disposable devices for single patient use, which incorporate various set configurations and components. The components include Pre-pierced access ports (Pre-pierced Male Adapter Plug, Pre-pierced Y-site, and Pre-pierced T-connector) which are the subject of this submission, may be shared across Hospira set families may be used with sets for

AI/ML Overview

The provided document is a 510(k) summary for a medical device called LIFESHIELD™ VISION™ Infusion sets with Pre-pierced Reseal Modification. This document describes the device, its intended use, and the non-clinical tests performed to demonstrate its substantial equivalence to previously cleared predicate devices.

Here's an analysis of the acceptance criteria and study information provided, structured as requested:

1. Table of Acceptance Criteria and Reported Device Performance

Characteristic / TestAcceptance Criteria (Standard/Test Method)Reported Device Performance
Biocompatibility
CytotoxicityISO 10993-5: 2009Pass
SensitizationISO 10993-10: 2010Pass
Irritation / Intracutaneous ReactivityISO 10993-10: 2010Pass
Systemic Toxicity (Acute)ISO 10993-11: 2006Pass
HemocompatibilityISO 10993-4: 2006Pass
Sterility (SAL 10-6)ISO 11137-2: 2006Pass
Dimensional Conformance and Connection CompatibilityISO 594-2 1998 (Conical fittings with a 6% (LUER) taper for syringes, needles and certain other equipment)Pass
Other Performance CharacteristicsNot explicitly detailed as criteria, but mentioned as having met requirementsParticulate, Leakage, Tensile Strength, Filter characteristics: Met requirements

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the exact sample sizes (number of devices or tests performed) for each specific test in the "Performance Test Summary-New Device" table. It generally states that "The LIFESHIELD™ VISION™ Infusion sets with Pre-pierced reseal access ports have passed all specified test requirements."

The data provenance is from non-clinical tests conducted by the manufacturer, Hospira, Incorporated, in the United States (based on the company's address). These tests are not human studies (clinical trials), so the concepts of retrospective or prospective human data do not apply directly. These are laboratory/bench tests.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable to this submission. The tests performed are non-clinical, bench testing (e.g., cytotoxicity, sterility, dimensional conformance) against established international standards (ISO). There is no "ground truth" derived from human expert consensus or interpretation of medical images/data in this context. The "ground truth" is adherence to the specified technical standards.

4. Adjudication Method for the Test Set

This information is not applicable. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies involving interpretation of medical data by multiple experts. For non-clinical bench testing, results are typically determined by adherence to pre-defined pass/fail criteria of the standard.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There was no MRMC comparative effectiveness study and no mention of AI assistance. This submission pertains to a modification of a physical medical device (infusion set), not an AI-powered diagnostic or assistive technology.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

There was no standalone (algorithm only) performance study. This is a physical device, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the tests performed is based on:

  • International Standards (ISO): Compliance with the specified parameters and methodologies outlined in standards like ISO 10993 (Biocompatibility), ISO 11137 (Sterility), and ISO 594-2 (Luer fittings).
  • Pre-defined Pass/Fail Criteria: Each standard has specific criteria that must be met for a "Pass" result. The device's physical attributes and performance were confirmed to meet these requirements.

8. The Sample Size for the Training Set

This information is not applicable. There is no "training set" as this is not a machine learning or AI algorithm. The device's design and manufacturing processes are validated against established engineering and quality control principles.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as there is no training set.

{0}------------------------------------------------

K113683

LIFESHIELD™ VISION™ Pre-Pierced Reseal Modification Special 510(k) Date: December 14, 2011

JAN 1 2 2012

! t

イ .

Section 6: 510(k) SUMMARY

A summary of 510(k) safety and effectiveness information in accordance with the requirements of
21 CFR 807.92. LIFESHIELD" VISION" Infusion sets with Pre-pierced Reseal modi

Submitter Information
NameHospira, Incorporated
AddressD-393, Bldg. H2275 North Field DriveLake Forest, IL. 60046
Phone number(224) 212-5316
Fax number224-212-5401
EstablishmentRegistration NumberOwner/Operator #9063339
Name of contact personKaren Keener
Date preparedDecember 14, 2011
Name of device
Trade or proprietarynameLIFESHIELD™ VISION™ Infusion sets with Pre-pierced Reseal Access Port
Common or usual nameFluid Delivery Tubing
Classification nameInfusion Sets
Classification panelClass II
Regulation21-CFR Part 880.5440
Product Code(s)80-FPA
Legally marketed device(s) towhich equivalence is claimedK941214 LIFESHIELD™ Extension SetK052722 LIFESHIELD™ Latex Free Microbore Extension SetK101677 Hospira Infusion Blood Sets10/03/199411/02/200508/11/2010
The changes addressed in this submission include:Material change of the pre-pierced housing access port from an acrylic multipolymer to polycarbonate Addition of a Septum Holder using a medical grade material Minor Dimensional change to the pre-pierced access ports Use of BD plastic blunt cannula for port access
Device descriptionThe LIFESHIELD™ VISION™ infusion sets, are intended for use as gravity sets orwith dedicated Hospira Infusion Pumps. These devices are obtainable in customlengths and component options according to facility needs and physicianpreference, Hospira infusion sets are disposable devices for single patient use,which incorporate various set configurations and components.The components include Pre-pierced access ports (Pre-pierced Male Adapter Plug,Pre-pierced Y-site, and Pre-pierced T-connector) which are the subject of thissubmission, may be shared across Hospira set families may be used with sets for

{1}------------------------------------------------

Section 6:510(k) SUMMARY
Intended use of the deviceInfusion sets are intended for the delivery of fluids including but not limited toblood and blood products from a container into a patient's vascular system. gravity or other pump platforms.
Summary of the technological characteristics of the device compared to the predicate device
CharacteristicProposed DevicePredicate
Intended UseSameSame
Set FunctionalitySameSame
Visual characteristicsColorless polycarbonate of similar sizeGreen Acrylic Multipolymer
white septum holderDoes not contain septum holder.
Bonding processesSameSame
BiocompatibilitySameSame
Principle of OperationSameSame
Port AccessPlastic Blunt Cannula and needles≤18guageMetal Cannula and needles
PERFORMANCE DATA
SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE*
Performance Test Summary-New Device
CharacteristicStandard/Test MethodStandard / Test TitleDevice Performance
BiocompatibilityISO 10993-5: 2009CytotoxicityPass
BiocompatibilityISO 10993-10: 2010SensitizationPass
BiocompatibilityISO 10993-10: 2010Irritation / IntracutaneousReactivityPass
BiocompatibilityISO 10993-11:2006Systemic Toxicity (Acute)Pass
BiocompatibilityISO 10993-4:2006HemocompatibilityPass
SAL 10-6ISO 11137-2:2006SterilityPass
Dimensional conformanceandConnection compatibilityISO 594-2 1998Conical fittings with a 6%(LUER) taper for syringes,needles and certain otherequipmentPass

Summary discussion of Bench Performance Data

The LIFESHIELD" VISION" Infusion sets with Pre-pierced reseal access ports have passed all specified test requirements.

The validation and verification testing have confirmed these devices meet user needs and design inputs for an infusion set.

Testing also confirmed physical attributes and device performance meet requirements of the standards listed in the "Performance Test Summary-New Devices" table above. These starility, biocompatibility, particulate, leakage, tensile strength, and filter characteristics.

{2}------------------------------------------------

Section 6: 510(k) SUMMARY

CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA

Statement of Safety and Efficacy:

The LIFESHIELD" VISION" Infusion sets with pre-pierced reseal access port meet the functional claims, and intended use as described in the product labeling. The safety and effectiveness, are substantially equivalent to the predicate Hospira Infusion sets cleared in K941214 LIFESHIELD " Extension Set, K052722 LIFESHIELD" Latex Free Microbore Extension Set and K101677 Hospira Infusion Blood Sets.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three stripes forming its wing and body. The eagle is enclosed in a circle of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JAN 1 2 2012

Ms. Karen Keener Senior Specialist Hospira, Incorporated 375 N. Field Drive Lake Forest, Illinois 60045

Re: K113683

Trade/Device Name: LIFESHIELD™ VISION™ Infusion sets Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: December 14, 2011 Received: December 15, 2011

Dear Ms. Keener:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Keener

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony D. Watson

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Section 5: Indications for Use

510(k) Number (unknown at this time) & 113683

Device Name: LIFESHIELD™ VISION™ Infusion sets

Indications for Use:

LIFESHIELD™ VISION™ Infusion sets are intended for the delivery of fluids including but not limited to blood and blood products from a container into a patients vascular system.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Khd C. A
(Division Sign-Off) 1/9/12

Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K113683

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.