(277 days)
The lacrimal duct catheter is intended for use during dilation of the obstructed nasolacrimal duct.
The lacrimal duct catheter is indicated for use during dilation of the obstructed nasolacrimal duct in the following populations:
- a. The 2mm catheter is indicated for use during dilation of the obstructed nasolacrimal duct obstruction in patients over 12 months of age and under 30 months of age.
- b. The 3 mm catheter is indicated for use during dilation of the obstructed nasolacrimal duct in children over 30 months of age.
- c. The 5mm catheter is indicated for use in adults during dilation of a lacrimal duct obstruction or blocked dacryocystorhinostomy ostium as a result of the following: functional or complete nasolacrimal duct obstruction, dacryocystitis, or failed dacryocystorhinostomy.
The lacrimal duct catheter is a sterile, single use, non-pyrogenic disposable balloon catheter consisting of a semi-flexible stainless steel hypotube core and polyethylene terephthalate balloon tubing. The balloon is designed to inflate to a known diameter and length at the specific pressure. Markings are present 10 and 15 mm proximal to the beginning of the working portion of the balloon, which helps indicate when the balloon is placed correctly in the lacrimal system. The overall length of the catheter is 6 inches (15.24 cm) long. There is an opening on the distal end of the catheter hypotube to accommodate irrigation solutions. The Y-hub on the proximal end of the catheter has a luer port for inflation of the balloon catheter (labeled "inflation" with a red band) and a second luer port for irrigation through the balloon catheter (labeled "irrigation"). The balloon catheter is available in a 2 mm and 3 mm inflated diameter. The 3 mm balloon has a length of 15 mm and a deflated profile of approximately 1.1 mm. The 2 mm balloon has a length of 15 mm and a deflated profile of approximately 1.0 mm. The balloon has a 5 mm inflated diameter and a length of 10 mm. The deflated profile of the 5 mm balloon is approximately 1.2 mm.
The provided text describes the DacryoCath Lacrimal Duct Balloon Catheter and its non-clinical performance testing for substantial equivalence to a predicate device. There is no clinical study or clinical performance data described in the provided document that would establish acceptance criteria or demonstrate the device meets such criteria in a clinical setting.
The document focuses on non-clinical performance tests to show similar technological characteristics and safety.
Here's a breakdown of the information requested, based only on the provided text:
1. A table of acceptance criteria and the reported device performance:
| Test Category | Acceptance Criteria/Methodology | Reported Device Performance |
|---|---|---|
| Biocompatibility | ISO10993 (Cytotoxicity, Sensitization, Irritation) | PASS |
| ISO Guinea Pig Maximization Sensitization Test | Did not elicit a sensitization response. | PASS |
| ISO Intracutaneous Reactivity Test | Met the requirements of ISO Intracutaneous Reactivity. | PASS |
| Rabbit Pyrogen Test (Material Mediated) - ISO | Non-pyrogenic and meets ISO10993-11 guidelines. | PASS |
| ISO Acute Systemic Injection Test | Met the requirements of the ISO Acute Systemic Injection Test. | PASS |
| MEM Elution GLP Report (Cytotoxicity) | Determines cytotoxicity of extractable substances; cell monolayers examined and scored. All acceptance criteria met. | PASS |
| Functional Performance | ||
| Inflation/Deflation Time | Inflation and deflation of balloons using conventional techniques within a specified time. | PASS |
| Fatigue Testing | Repeatability of balloon inflation without failure using recommended inflation pressure. | PASS |
| Rupture Testing | Balloons will not burst at or below the maximum recommended burst pressure. | PASS |
| Tensile Testing | Bond strength at joining locations can withstand tensile forces greater than clinical use. | PASS |
| ISO 594-1 Testing (Luer Taper) | Gauging, Liquid Leakage, Air Leakage, Separation Force. (Tests conducted within ISO 594-1 guidance) | PASS |
| Sterilization & Packaging | ||
| Packaging Validation | Transportation (ASTM D169), Seal Peel Test, Dye Migration Test (ASTM F1929), Aerosol Challenge Test, Accelerated Aging Test (3.3 weeks @ 55± 4°C simulating 0.5 year shelf life). | PASS |
| EO Sterilization Validation | Microbiological challenge utilizing half-cycle (overkill) method per ISO11135. | PASS |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: The document does not specify the exact sample sizes for each individual test. It mentions "balloons of each diameter and length" for rupture testing, and "test article" for others. No specific numbers are provided for these non-clinical tests.
- Data Provenance: Not applicable for non-clinical lab tests. The tests refer to ISO and ASTM standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. These were non-clinical, laboratory-based engineering and biological safety tests. Ground truth in this context refers to the defined parameters and expected outcomes of the standardized tests, not expert interpretation of clinical data.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. Adjudication methods are typically used for clinical result interpretation, particularly when expert consensus or review is needed. These are objective engineering and biological tests with pass/fail criteria.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This document does not describe an MRMC study or any clinical comparative effectiveness study involving human readers or AI.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- No. This device is a physical medical device (balloon catheter) intended for manual use by a clinician, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- For the non-clinical tests, the "ground truth" is defined by the specific requirements and methodologies of the international and national standards (ISO, ASTM, USP) referenced for each test. For example, for biocompatibility, the ground truth is the absence of cytotoxicity, sensitization, irritation, etc., as determined by the specified assays. For functional tests, it's meeting predefined performance specifications (e.g., burst pressure, inflation time).
8. The sample size for the training set
- Not applicable. This document describes the testing of a physical medical device. There is no mention of a "training set" as would be relevant for machine learning algorithms.
9. How the ground truth for the training set was established
- Not applicable, as there is no training set described.
{0}------------------------------------------------
K113508
AUG 3 1 2012
510(k) SUMMARY
Suite 246
Armadillo Medical, LLC,
5363 Balboa Blvd.
Encino, CA 91316 TEL: 855.255.3310
510(k) Owner
Contact person
FAX: 818.783.9059 Robyn Scopis Regulatory Consultant to Armadillo Biomedical Regulatory Specialists, Inc. 3722 Ave. Sausalito Irvine, CA 92606 TEL: 949.422.3853 FAX: 949.552.2821 EMAIL: robyn@regulatoryspecialists.com
Date summary was prepared
August 20, 2012
Name of device Common Name Classification Name Regulation Product Code Predicate
DacryoCath Lacrimal duct balloon catheter Lacrimal Stents and Intubation Sets Pre-Amendment OKS K935233
Description
The lacrimal duct catheter is a sterile, single use, non-pyrogenic disposable balloon catheter consisting of a semi-flexible stainless steel hypotube core and polyethylene terephthalate balloon tubing. The balloon is designed to inflate to a known diameter and length at the specific pressure. Markings are present 10 and 15 mm proximal to the beginning of the working portion of the balloon, which helps indicate when the balloon is placed correctly in the lacrimal system. The overall length of the catheter is 6 inches (15.24 cm) long. There is an opening on the distal end of the catheter hypotube to accommodate irrigation solutions. The Y-hub on the proximal end of the catheter has a luer port for inflation of the balloon catheter (labeled "inflation" with a red band) and a second luer port for irrigation through the balloon catheter (labeled "irrigation"). The balloon catheter is available in a 2 mm and 3 mm inflated diameter. The 3 mm balloon has a length of 15 mm and a deflated profile of approximately 1.1 mm. The 2 mm balloon has a length of 15 mm and a deflated profile of approximately 1.0 mm.
{1}------------------------------------------------
K//3508
Description continued:
The balloon has a 5 mm inflated diameter and a length of 10 mm. The deflated profile of the 5 mm balloon is approximately 1.2 mm.
Intended Use
The lacrimal duct catheter is intended for use during dilation of the obstructed nasolacrimal duct.
Indications for Use
The lacrimal duct catheter is indicated for use during dilation of the obstructed nasolacrimal duct in the following populations:
- a. The 2mm catheter is indicated for use during dilation of the obstructed nasolacrimal duct obstruction in patients over 12 months of age and under 30 months of age.
- b. The 3 mm catheter is indicated for use during dilation of the obstructed nasolacrimal duct in children over 30 months of age.
- c. The 5mm catheter is indicated for use in adults during dilation of a lacrimal duct obstruction or blocked dacryocystorhinostomy ostium as a result of the following: functional or complete nasolacrimal duct obstruction, dacryocystitis, or failed dacryocystorhinostomy.
Technological Characteristics
The predicate and the DacryoCath were compared in the following areas and found to have similar technological characteristics and to be equivalent:
Material Characteristics
Both the Predicate and DacryoCath are made of Stainless Steel and Polyethylene.
Design Characteristics
Both the Predicate and DacryoCath are designed to be used as lacrimal duct balloon catheters.
Operating Characteristics
Both the Predicate and DacryoCath are operated by placing the catheter into the canaliculus, inflate the balloon to 8atm with 10cc sterile saline, and inflate for 90seconds to dilate the obstructed nasolacrimal duct.
Intended Use
Both the Predicate and DacryoCath have the same intended use: The lacrimal duct catheter is intended for use during dilation of the obstructed nasolacrimal duct
2
{2}------------------------------------------------
Technological Characteristics Similar technological characteristics continued: Balloon Length -3 mm Predicate - 15mm DacryoCath - 15mm Balloon Length - 5mm Predicate - 10mm DacryoCath - 10mm
The predicate and the DacryoCath were compared in the following areas and found to have minor different technological characteristics. The following differences have been determined. through non-clinical performance testing, to not have any impact on the safety or efficacy of the DacryoCath when used as indicated:
15/13508
Length of Catheter Predicate - 24cm DacryoCath - 15.24cm Balloon Length (2mm Balloon only) Predicate - 13mm DacryoCath - 15mm Irrigation Port -Predicate - None Available DacryoCath - Yes
The following non-clinical performance tests were conducted:
Biocompatibility to ISO10993
| Cytotoxicity | PASS |
|---|---|
| Sensitization | PASS |
| Irritation | PASS |
ISO Guinea Pig Maximization Sensitization Test
The test article did not elicit a sensitization response under the conditions of this assay.
ISO Intracutaneous Reactivity Test
The requirements of ISO Intracutaneous Reactivity have been met by the test article.
Rabbit Pyrogen Test (Material Mediated) - ISO
3
The USP 0.9% Sodium Chloride for Injection (NaCl) extract of the test article, Lot #23054 - lacrimal balloon catheters, was evaluated for its potential to produce a pyrogenic response when tested in New Zealand White rabbits. Based on the criteria of the protocol, the test article is considered non-pyrogenic and meets the requirements of the Pyrogen Test, ISO10993-11 guidelines.
{3}------------------------------------------------
Non-clinical performance tests continued:
ISO Acute Systemic Injection Test
The requirements of the ISO Acute Systemic Injection Test have been met by the test article.
MEM Elution GLP Report
The Minimal Essential Media (MEM) Elution test was designed to determine the cytotoxicity of extractable substances. An extract of the test article was added to cell monolayers and incubated. The cell monolayers were examined and scored based on the degree of cellular destruction. All test method acceptance criteria were met.
Inflation/Deflation Time PASS
Inflation/Deflation Time testing was conducted to show that the inflation and deflation of the balloons using conventional techniques can be accomplished within a specified time.
Fatigue Testing
Fatigue testing was conducted to determine the repeatability of balloon inflation without failure using the recommended inflation pressure.
Rupture Testing
PASS
PASS
15113508
Rupture testing was done on balloons of each diameter and length. Test results show that the balloons will not burst at or below the maximum recommended burst pressure.
Tensile Testing
PASS
Tensile testing was done to test the bond strength at locations where joining methods are used for bonding components of the catheter. Testing demonstrated that all bonds can withstand tensile forces greater than those that may be experienced during clinical use.
ISO 594-1 Testing
PASS
ISO 594-1 testing was conducted for the 6% (Luer) taper for syringes. Under this guidance, the following tests were conducted:
Gauging Liquid Leakage Air Leakage Separation Force
Stress Cracking
| Packaging Validation | PASS | |
|---|---|---|
| Transportation | ASTM D169 | PASS |
| Seal Peel Test | PASS | |
| Dye Migration Test | ASTM F1929 | PASS |
| Aerosol Challenge Test | PASS | |
| Accelerated Aging Test | PASS | |
| 3.3 weeks @ 55± 4°C | ||
| Simulating 0.5 year shelf life |
{4}------------------------------------------------
Non-clinical performance tests continued:
EO Sterilization Validation
PASS A microbiological challenge utilizing the half-cycle (overkill) method, using a biological indicator challenge following ISO11135.
Conclusions from non-clinical performance data
After performing non-clinical performance studies, the data shows that the DacryoCath is substantially equivalent to the predicate as a lacrimal duct balloon catheter.
Clinical performance data
A review of the published peer-reviewed literature was conducted to review appropriate uses, adverse events, and clinical experience with this device type.
5
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle or bird-like figure with flowing lines, suggesting movement or flight. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
AUG 3 1 2012
Armadillo Biomedical, LLC c/o Ms. Robyn Scopis Regulatory Specialists, Inc. 3722 Ave. Sausalito Irvine, CA 92606
Re: K113508
Trade/Device Name: DacryoCath Regulation Number: Unclassified Regulation Name: Unclassified Regulatory Class: Unclassified Product Code: OKS Dated: August 20, 2012 Received: August 29, 2012
Dear Ms. Scopis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{6}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
for
Sincerely yours,
Kesia Alexander
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: DacryoCath by Armadillo Biomedical, LLC
Indications for Use:
The lacrimal duct catheter is indicated for use during dilation of the obstructed nasolacrimal duct in the following populations:
- b. The 2mm catheter is indicated for use during dilation of the obstructed nasolacrimal duct obstruction in patients over 12 months of age and under 30 months of age.
- b. The 3 mm catheter is indicated for use during dilation of the obstructed nasolacrimal duct in children over 30 months of age.
- c. The 5mm catheter is indicated for use in adults during dilation of a lacrimal duct obstruction or blocked dacryocystorhinostomy ostium as a result of the following: functional or complete nasolacrimal duct obstruction, dacryocystitis, or failed dacryocystorhinostomy.
× AND/OR Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
CLCLy
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K113508
1 Page 1 of
N/A