K Number
K113341

Validate with FDA (Live)

Date Cleared
2012-05-03

(171 days)

Product Code
Regulation Number
892.1750
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SCENARIA Whole-body X-ray CT System with guideShot Option is a x-ray imaging device that produce cross-sectional images of the body at different angles. The system reconstructs, processes, displays, and stores the collected images. The guideShot Option adds a remote inroom display and controls to support interventional imaging. The device output can provide an aid to diagnosis when used by a qualified physician.

Device Description

The SCENARIA Whole-body X-ray CT System is a multi-slice computed tomography system that uses x-ray data to produce cross-sectional images of the body at various angles. The quideShot Option adds a remote monitor and controls at the patient table to allow the operator to initiate data collection and to view resulting CT images in order to support interventional procedures.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the "guideShot Option, SCENARIA Whole-body X-ray CT System." The core of the submission is to demonstrate substantial equivalence to a predicate device, not necessarily to prove specific performance criteria against a clinical ground truth.

Here's an analysis based on your request:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific CriteriaReported Device Performance Statement
Non-Clinical EvaluationsDose Profile"The evaluation results of the SCENARIA Whole-body X-ray CT System with guideShot Option were comparable to the predicate device."
Image Noise"The evaluation results of the SCENARIA Whole-body X-ray CT System with guideShot Option were comparable to the predicate device."
Modulation Transfer Function (MTF)"The evaluation results of the SCENARIA Whole-body X-ray CT System with guideShot Option were comparable to the predicate device."
Slice Thickness and Sensitivity Profile"The evaluation results of the SCENARIA Whole-body X-ray CT System with guideShot Option were comparable to the predicate device."
Slice Plane Location"The evaluation results of the SCENARIA Whole-body X-ray CT System with guideShot Option were comparable to the predicate device."
CT Dose Index"The evaluation results of the SCENARIA Whole-body X-ray CT System with guideShot Option were comparable to the predicate device."
Technological EquivalenceConcept"Technologically equivalent in concept... to the predicate device."
Function"Technologically equivalent in... function... to the predicate device."
Performance"Technologically equivalent in... performance to the predicate device."
Clinical EquivalenceAids diagnosis when used by a qualified physician (Implied from Intended Use Statement)The device "can provide an aid to diagnosis when used by a qualified physician," similar to previous CT systems. This is an intended use statement, not directly a performance criterion proven by this study.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a sample size or data provenance for human-read images or clinical studies. The performance comparison mentioned is based on non-clinical evaluations of physical and image quality characteristics, comparing the new device to its predicate. These evaluations are likely performed in a controlled lab setting using phantoms or test objects.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This information is not provided in the document. The performance evaluation was primarily non-clinical, comparing technical specifications to a predicate device. Clinical ground truth, derived from expert consensus on patient images, would not be directly relevant to the described testing.

4. Adjudication Method for the Test Set

This information is not provided. Given the non-clinical nature of the performance evaluations, an adjudication method for clinical ground truth would not be applicable.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done or described in this submission. The document focuses on technological and physical equivalence to a predicate device, not on demonstrating improved human reader performance with the new device or any AI assistance (which is not part of this device).

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

No, a standalone algorithm-only performance study was not done or described. The device is a CT imaging system with an add-on option for remote in-room display and controls. It does not appear to incorporate an AI algorithm that would have standalone performance in interpreting images.

7. The Type of Ground Truth Used

The ground truth used for the performance evaluations was based on physical and technical measurements of image quality (e.g., dose, noise, MTF, slice thickness) against established specifications or comparisons to the predicate device's measured performance. There is no mention of expert consensus, pathology, or outcomes data being used as ground truth for the evaluations described.

8. The Sample Size for the Training Set

This information is not applicable and not provided. The device is a CT scanner, not a machine learning algorithm that requires a training set in the conventional sense. The "guideShot Option" is a hardware/software addition for remote control and display during interventional procedures.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable and not provided for the same reasons as point 8.

{0}------------------------------------------------

510(k) Summary

K 113241

Submitter Information

Submitter: Hitachi Medical Systems America, Inc. 1959 Summit Commerce Park Twinsburg, Ohio 44087-2371 ph: (330) 425-1313 fax: (330) 963-0749 Douglas J. Thistlethwaite Contact: Date: November 4, 2011

Device Name

Classification Name:Computed tomography x-ray system
Classification Number:90-IYN, 892.1550
Trade/Proprietary Name:guideShot Option, SCENARIA Whole-body X-ray CT System
Predicate Device(s):guideShot Option, ECLOS Computed Tomography X-ray System(K091103)

Device Intended Use

The SCENARIA Whole-body X-ray CT System with guideShot Option is a x-ray imaging device that produce cross-sectional images of the body at different angles. The system reconstructs, processes, displays, and stores the collected images. The guideShot Option adds a remote inroom display and controls to support interventional imaging. The device output can provide an aid to diagnosis when used by a qualified physician.

Device Description

Function

The SCENARIA Whole-body X-ray CT System is a multi-slice computed tomography system that uses x-ray data to produce cross-sectional images of the body at various angles. The quideShot Option adds a remote monitor and controls at the patient table to allow the operator to initiate data collection and to view resulting CT images in order to support interventional procedures.

Scientific Concepts

The SCENARIA Whole-body X-ray CT System uses 64-slice CT technology, where the X-ray tube and detector assemblies are mounted on a frame that rotates continuously around the patient using slip ring technology. The solid-state detector assembly design collects up to 64 slices of data simultaneously. The X-ray sub-system features a high frequency generator, X-ray tube, and collimation system that produces a fan beam X-ray output. The system can operate in a helical (spiral) scan mode where the patient table moves during scanning. As the X-ray tube/detector assembly rotates around the patient, data is collected at multiple angles.

The collected data is then reconstructed into cross-sectional images by a high-speed reconstruction sub-system. The images are displayed on a Computer Workstation, stored, printed, and archived as required. The workstation is based on current PC technology using the Windows™ operating system.

MAY - 3 2012

{1}------------------------------------------------

Physical and Performance Characteristics

The SCENARIA Whole-body X-ray CT System consists of a gantry, operator's workstation, patient table, high-frequency x-ray generator, and accessories. The guideShot option adds a remote monitor and footswitch to control acquisitions and display images in the scan room. The system performance is similar to the predicate device.

Performance Comparison

Because the SCENARIA Whole-body X-ray CT System with quideShot Option and the predicate device are both Hitachi designs, they were subjected to the same non-clinical evaluations as stipulated in 21 CFR 1020.33(c). Evaluations include: dose profile, image noise, modulation transfer function (MTF), slice thickness and sensitivity profile, slice plane location, and CT dose index.

The evaluation results of the SCENARIA Whole-body X-ray CT System with guideShot Option were comparable to the predicate device and support our conclusion that the system is substantially equivalent.

Device Technological Characteristics

The SCENARIA Whole-body X-ray CT System with guideShot Option acquires data in the same manner as the predicate device. Physically, the SCENARIA Whole-body X-ray CT System with guideShot Option is very similar to the predicate device. The key differences are the ability to initiate a scan and display data in the scan room.

The ability to collect and display image on an in-room monitor does not change the essential characteristics of the finished images. The operation of the system is virtually identical to the predicate because guideShot simply adds remote display and controls to the predicate device. The SCENARIA Whole-body X-ray CT System operating system software is essentially the same, as well as the user interface. The patient table design and gantry controls are unchanged.

In conclusion, the SCENARIA Whole-body X-ray CT System with guideShot Option is technologically equivalent in concept, function, and performance to the predicate device.

Conclusions

It is the opinion of Hitachi Medical Systems America, Inc. the SCENARIA Whole-body X-ray CT System with guideShot Option is substantially equivalent with respect to hardware, base elements of the software, safety, effectiveness, and functionality to the ECLOS Computed Tomography X-ray System with guideShot Option (K091103).

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image contains the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it, representing medicine and healing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular fashion around the caduceus symbol. The text is in all caps and is in a sans-serif font.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Mr. Doug Thistlethwaite Manager, Regulatory Affairs Hitachi Medical Systems America, Inc. 1959 Summit Commerce Park TWINSBURG OH 44087

MAY - 3 2012

Re: K113341

Trade/Device Name: guideShot Option, SCENARIA Whole-body X-ray CT System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: April 13, 2012 Received: April 18, 2012

Dear Mr. Thistlethwaite:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{3}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: guideShot Option, SCENARIA Whole-body X-ray CT System

Indications For Use:

The SCENARIA Whole-body X-ray CT System with guideShot Option is a x-ray imaging device that produce cross-sectional images of the body at different angles. The system reconstructs, processes, displays, and stores the collected images. The guideShot Option adds a remote in-room display and controls to support interventional imaging. The device output can provide an aid to diagnosis when used by a qualified physician.

Prescription Use (Part 21 CFR 801 Subpart D)

810K

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Sarety

Page 1 of

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.