K Number
K113323
Manufacturer
Date Cleared
2012-02-06

(88 days)

Product Code
Regulation Number
866.3980
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The artus® Infl A/B RG RT-PCR Kit is a multiplex real time PCR in vitro diagnostic test for the qualitative detection and identification of Influenza A and Influenza B virus RNA in nasopharyngeal swab specimens using the Rotor-Gene® Q MDx instrument. The test is intended for use as an aid in the differential diagnosis of Influenza A and Influenza B viral infections in patients symptomatic for respiratory tract infection in conjunction with clinical and epidemiological risk factors. It is not intended to detect Influenza C virus.

Negative results do not preclude respiratory virus infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

Performance characteristics for Influenza A were established during the 2009/2010 and 2010/2011 flu seasons when Influenza A (H3N2) and Influenza A/2009 (H1N1) were the predominant Influenza A viruses in circulation. When other Influenza A viruses emerge, performance characteristics may vary.

If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Device Description

The artus Infl A/B RG RT-PCR Kit contains reagents and instructions for the detection and differentiation of Influenza A and Influenza B viral RNA in nasopharyngeal swabs of symptomatic patients.

The assay utilizes the EZ1 Advanced XL instrument with the EZ1 Advanced XL DSP Virus Card v. 1.0. and the EZ1 DSP Virus Kit (QIAGEN) for viral nucleic acid extraction. The Rotor-Gene Q MDx instrument with the artus Influenza Assay Software Package (QIAGEN) is used for amplification and detection.

Pathogen detection by the reverse transcription polymerase chain reaction (RT-PCR) is based on the reverse transcription of the RNA into complementary DNA (cDNA) and subsequent amplification of specific regions of the pathogen genome. In real-time PCR the amplified product is detected via fluorescent dyes. These are linked to oligonucleotides that bind specifically to the amplified product. Monitoring the fluorescence intensities during the PCR run (i.e., in real time) allows the detection of the accumulating product without having to re-open the reaction tubes after the PCR run.

The Influenza A/B Master contains primers, probes, enzymes, and other reaction components (except Mg solution) needed for the specific amplification of a 141 bp region of the Influenza A virus genome and a 95 bp region of the influenza virus B genome, and for the direct detection of the specific amplicons in two fluorescence channels of the Rotor-Gene O MDx instrument. The primers are complementary to highly conserved regions of the Matrix gene locus within the influenza B virus genome. The probes are dual-labeled with a reporter dye attached to the 5'-end and a quencher dye attached to the 3'-end. Detection is performed on the Rotor-Gene O MDx instrument at wavelengths listed in Table 5.1.

In addition, the Master contains a second heterologous primer/probe set to detect the Influenza A/B Internal Control (IC). The IC result identifies possible failure of RNA extraction or the presence of PCR inhibition. The Internal Control is detected in a third fluorescence channel.

An Influenza A Control and an Influenza B Control comprised of in vitro transcripts representing the amplified regions of the Influenza A virus genome and the Influenza B virus genome, respectively, are provided. PCR grade water is provided as a negative (notemplate) control.

The procedure consists of four consecutive steps:

  1. Sample collection: Collect nasopharyngeal swab specimens from symptomatic patients using a polyester, nylon, or rayon swab and place it into virus transport medium.
  2. Nucleic acid extraction: Add the Influenza IC to the carrier RNA before starting the extraction procedure. Extract viral RNA using the EZ1 DSP Virus Kit in combination with the EZ1 Advanced XL instrument.
  3. Real-time RT-PCR: Add the extracted RNA and positive and negative control material to Influenza A/B Master mix. Perform real-time RT-PCR using the Rotor-Gene O MDx instrument.
  4. Result interpretation: The Influenza Assay Package evaluates the results of the positive and negative controls to determine if the run is valid. If the run is valid, the internal control and target-specific results of each specimen are evaluated.
AI/ML Overview

Here's an analysis of the acceptance criteria and supporting studies for the artus® Infl A/B RG RT-PCR Kit, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria (e.g., "Sensitivity must be >90%"). Instead, it presents the "performance characteristics" observed in the clinical evaluations. The interpretation below assumes that achieving these reported performance levels was sufficient for acceptance.

Metric (Target)Influenza A Reported PerformanceInfluenza B Reported PerformanceStudy Type
Sensitivity (Prospective Clinical)100% (95% CI: 80-100%)100% (95% CI: 92-100%)Prospective Clinical Evaluation
Specificity (Prospective Clinical)95.0% (95% CI: 91-97%)94.7% (95% CI: 91-97%)Prospective Clinical Evaluation
Positive Agreement (Retrospective Clinical 1)100% (No CI provided, but 175/175 detected)100% (95% CI: 90-100%)Retrospective Clinical Evaluation (vs. FDA cleared molecular test)
Negative Agreement (Retrospective Clinical 1)N/A (implied from 0 false negatives)98.9% (95% CI: 96-100%)Retrospective Clinical Evaluation (vs. FDA cleared molecular test)
Positive Agreement (Retrospective Clinical 2)97.9% (95% CI: 93-99%)100% (95% CI: 79-100%)Prospectively collected & archived clinical evaluation (vs. FDA cleared molecular test)
Negative Agreement (Retrospective Clinical 2)94.8% (95% CI: 92-97%)99.6% (95% CI: 98-100%)Prospectively collected & archived clinical evaluation (vs. FDA cleared molecular test)
Overall Reproducibility (across 3 sites)99.4% (overall agreement for all panel members)99.4% (overall agreement for all panel members)Non-clinical Reproducibility Study
Limit of Detection (LoD)Ranged from 10e0.2 to 10e1.1 TCID50/ml for various A strains (≥95% detection rate)Ranged from 10e-0.1 to 10e0.9 TCID50/ml for various B strains (≥95% detection rate)Non-clinical LoD Study
Analytical ReactivityAll 18 Influenza A strains detected at near LoD concentrationsAll 6 Influenza B strains detected at near LoD concentrationsNon-clinical Reactivity Study
Interfering SubstancesNo interference observed for tested substances, except for one instance with Mupirocin at 10 mg/ml for Influenza B (resolved at 2 mg/ml)No interference observed for tested substances, except for one instance with Mupirocin at 10 mg/ml for Influenza B (resolved at 2 mg/ml)Non-clinical Interfering Substances Study
Analytical SpecificityNo false positives for 21 non-target respiratory viruses. Invalid results for Bordetella pertussis and Streptococcus pneumoniae at high concentrations (resolved at lower concentrations).No false positives for 21 non-target respiratory viruses. Invalid results for Bordetella pertussis and Streptococcus pneumoniae at high concentrations (resolved at lower concentrations).Non-clinical Analytical Specificity Study
Competitive InhibitionInhibition observed for A/California/7/09-like virus reference strain when B/Florida/4/2006-like virus or B/Brisbane/60/2008-like virus were present at 10e5 TCID50/ml (resolved at 10e4 TCID50/ml)No inhibition observed for the reference influenza B strain in the presence of 7 competing Influenza A strains at 10e5 TCID50/mlNon-clinical Competitive Inhibition Study
Carryover/Cross ContaminationNo evidence of carryover or cross-contaminationNo evidence of carryover or cross-contaminationNon-clinical Carryover/Cross Contamination Study

2. Sample Size Used for the Test Set and Data Provenance

  • Prospective Clinical Evaluation:
    • Sample Size: 254 nasopharyngeal (NP) swab specimens.
    • Data Provenance: Prospective, collected at 3 clinical laboratories in the United States during the 2010-11 respiratory virus season.
  • Retrospective Clinical Evaluation 1:
    • Sample Size: 212 specimens.
    • Data Provenance: Retrospective, collected from the 2009 and 2010 respiratory virus season. Country of origin not explicitly stated but implied to be US clinical laboratories where initial testing occurred.
  • Prospectively Collected and Archived Clinical Evaluation (Retrospective Clinical 2):
    • Sample Size: 462 well-characterized, de-identified, residual specimens.
    • Data Provenance: Retrospective (archived), collected from August 2009 to May 2011 respiratory virus season. Country of origin not explicitly stated but implied to be US clinical settings.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the "number of experts" or their "qualifications" for establishing the ground truth in the clinical studies.

  • Prospective Clinical Evaluation: The reference method was "standard viral culture (shell vial) followed by direct fluorescent antibody (DFA) screening and identification." This typically involves trained laboratory personnel (e.g., microbiologists, lab technologists) rather than a panel of clinical experts for interpretation. Bidirectional sequencing was used to confirm some discrepant results.
  • Retrospective Clinical Evaluation 1 & 2: The ground truth was established using "an FDA cleared molecular test" or "FDA cleared molecular tests for routine patient management." This implies reliance on the validated results from these predicate devices, not on a panel of human experts re-interpreting raw data.

4. Adjudication Method for the Test Set

The document does not describe a formal adjudication method (e.g., 2+1, 3+1) involving multiple experts for the test set.

  • In the prospective study, discrepant results (artus positive, culture negative) were investigated using "bidirectional sequencing." This is a molecular confirmation, not an expert panel adjudication.
  • In the retrospective studies, the "FDA cleared molecular tests" were used as the reference, suggesting agreement with those established results was the primary measure, not a separate adjudication process.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

No, an MRMC comparative effectiveness study was not done. This device is a molecular diagnostic test (RT-PCR kit), not an imaging or AI-assisted diagnostic tool that would typically involve human "readers" or assess improvement with AI assistance. The performance is entirely based on the kit's analytical and clinical accuracy.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the performance reported for the artus® Infl A/B RG RT-PCR Kit is inherently standalone (algorithm only, without human-in-the-loop performance in the sense of interpretative assistance). The device outputs a qualitative positive/negative result for Influenza A and B. While human operators are involved in sample preparation and running the assay, the "results interpretation" step involves the "Influenza Assay Package evaluating the results of the positive and negative controls to determine if the run is valid" and then evaluating the "internal control and target-specific results of each specimen." This is an automated interpretation by the device's software package based on pre-defined thresholds.

7. The Type of Ground Truth Used

  • Prospective Clinical Evaluation: The primary ground truth was standard viral culture (shell vial) followed by direct fluorescent antibody (DFA) screening and identification. For discrepant results, bidirectional sequencing was used as a confirmatory method.
  • Retrospective Clinical Evaluation 1 & 2: An FDA cleared molecular test was used as the ground truth (reference method).

8. The Sample Size for the Training Set

The document does not specify a "training set" or its sample size. For molecular diagnostic kits like this, the "development" or "training" phase usually refers to the analytical studies (e.g., determining optimal primer/probe concentrations, establishing LoD, testing specificity and reactivity) and initial method validation, rather than a distinct 'training set' for an algorithm in the way it's used for AI/machine learning. The provided data focuses on the performance of the finalized device.

9. How the Ground Truth for the Training Set Was Established

Since no specific "training set" is identified for the device in the context of an algorithm, the concept of establishing ground truth for it is not applicable here. The ground truth for the performance evaluation (test sets) was established as described in section 7. The analytical parameters of the assay (e.g., target genes, detection channels, controls) are designed and experimentally validated, not "trained" on a dataset with external ground truth in the AI sense.

{0}------------------------------------------------

510(k) SUMMARY

General Information

Submitted by:QIAGEN GmbHQiagen Strasse 1Hilden, Germany D-40724
Contact Person:Volker RiemenschneiderQIAGEN GmbHQiagen Strasse 1Hilden, Germany D-40724
Phone: +49 2103-29-0Fax: +49 2103-29-2177Email: Volker.riemenschneider@qiagen.com
Date Prepared:February 1, 2012
Device Name:
Trade Name:artus ® Infl A/B RG RT-PCR Kit
Common Name:Influenza A/B Test
Classification Name:Respiratory Viral Panel Multiplex Nucleic Acid Assay, (21 CFR 866.3980, Product Code OCC)

Predicate Device

ManufacturerProduct Name510(k) No.
Gen-Probe Prodesse, Inc.ProFlu+ AssayK110968

Device Description

The artus Infl A/B RG RT-PCR Kit contains reagents and instructions for the detection and differentiation of Influenza A and Influenza B viral RNA in nasopharyngeal swabs of symptomatic patients.

The assay utilizes the EZ1 Advanced XL instrument with the EZ1 Advanced XL DSP Virus Card v. 1.0. and the EZ1 DSP Virus Kit (QIAGEN) for viral nucleic acid extraction. The Rotor-Gene Q MDx instrument with the artus Influenza Assay Software Package (QIAGEN) is used for amplification and detection.

Pathogen detection by the reverse transcription polymerase chain reaction (RT-PCR) is based on the reverse transcription of the RNA into complementary DNA (cDNA) and subsequent amplification of specific regions of the pathogen genome. In real-time PCR the amplified product is detected via fluorescent dyes. These are linked to

{1}------------------------------------------------

oligonucleotides that bind specifically to the amplified product. Monitoring the fluorescence intensities during the PCR run (i.e., in real time) allows the detection of the accumulating product without having to re-open the reaction tubes after the PCR run.

The Influenza A/B Master contains primers, probes, enzymes, and other reaction components (except Mg solution) needed for the specific amplification of a 141 bp region of the Influenza A virus genome and a 95 bp region of the influenza virus B genome, and for the direct detection of the specific amplicons in two fluorescence channels of the Rotor-Gene O MDx instrument. The primers are complementary to highly conserved regions of the Matrix gene locus within the influenza B virus genome. The probes are dual-labeled with a reporter dye attached to the 5'-end and a quencher dye attached to the 3'-end. Detection is performed on the Rotor-Gene O MDx instrument at wavelengths listed in Table 5.1.

In addition, the Master contains a second heterologous primer/probe set to detect the Influenza A/B Internal Control (IC). The IC result identifies possible failure of RNA extraction or the presence of PCR inhibition. The Internal Control is detected in a third fluorescence channel.

ComponentTargetTargetGeneDetection Range(nm)
Influenza A/B MasterInfluenza A VirusMatrix510 +/- 5
Influenza B VirusMatrix710 +/- 5
Internal ControlSyntheticsequence610 +/- 5

Table 5.1. Target genes and detection channels

An Influenza A Control and an Influenza B Control comprised of in vitro transcripts representing the amplified regions of the Influenza A virus genome and the Influenza B virus genome, respectively, are provided. PCR grade water is provided as a negative (notemplate) control.

The procedure consists of four consecutive steps:

    1. Sample collection: Collect nasopharyngeal swab specimens from symptomatic patients using a polyester, nylon, or rayon swab and place it into virus transport medium.
    1. Nucleic acid extraction: Add the Influenza IC to the carrier RNA before starting the extraction procedure. Extract viral RNA using the EZ1 DSP Virus Kit in combination with the EZ1 Advanced XL instrument.

{2}------------------------------------------------

    1. Real-time RT-PCR: Add the extracted RNA and positive and negative control material to Influenza A/B Master mix. Perform real-time RT-PCR using the Rotor-Gene O MDx instrument.
    1. Result interpretation: The Influenza Assay Package evaluates the results of the positive and negative controls to determine if the run is valid. If the run is valid, the internal control and target-specific results of each specimen are evaluated.

Intended Use

The artus® Infl A/B RG RT-PCR Kit is a multiplex real time PCR in vitro diagnostic test for the qualitative detection and identification of Influenza B virus RNA in nasopharyngeal swab specimens using the Rotor-Gene® O MDx instrument. The test is intended for use as an aid in the differential diagnosis of Influenza A and Influenza B viral infections in patients symptomatic for respiratory tract infection in conjunction with clinical and epidemiological risk factors. It is not intended to detect Influenza C virus.

Negative results do not preclude respiratory virus infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

Performance characteristics for Influenza A were established during the 2009/2010 and 2010/2011 flu seasons when Influenza A (H3N2) and Influenza A/2009 (H1N1) were the predominant Influenza A viruses in circulation. When other Influenza A viruses emerge, performance characteristics may vary.

If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Comparison of the artus® Infl A/B RG RT-PCR Kit and the Predicate Device

The artus® Infl A/B RG RT-PCR Kit is substantially equivalent to the predicate device:

  • · K110968; ProFlu+ Assay, Gen-Probe Prodesse, Inc.
    Similarities and differences between the artus® Infl A/B RG RT-PCR Kit and the predicate device are shown in Table 5.2.

{3}------------------------------------------------

Table 5.2. Comparison of the artus® Infl A/B RG RT-PCR Kit with the predicate
device.
CharacteristicDevicePredicate
Nameartus® Infl A/B RG RT-PCR KitProFlu+TM Assay
510(k) No.K113323K110968
Regulation866.3980866.3980
Product CodeOCCOCC
Device ClassIIII
Similarities
Intended UseThe artus® Infl A/B RG RT-PCRThe ProFlu" + Assay is a
Kit is a multiplex real time PCR inmultiplex Real-Time PCR (RT-
vitro diagnostic test for thePCR) in vitro diagnostic test for
qualitative detection andthe rapid and qualitative detection
identification of influenza A andand discrimination of Influenza A
influenza B virus RNA inVirus, Influenza B Virus, and
nasopharyngeal swab specimensRespiratory Syncytial Virus
using the Rotor-Gene® Q MDx(RSV) nucleic acids isolated and
instrument. The test is intended forpurified from nasopharyngeal
use as an aid in the differential(NP) swab specimens obtained
diagnosis of influenza A andfrom symptomatic patients. This
influenza B virus infections intest is intended for use to aid in
patients symptomatic forthe differential diagnosis of
respiratory tract infection inInfluenza A, Influenza B and
conjunction with clinical andRSV viral infections in humans
epidemiological risk factors. It isand is not intended to detect
not intended to detect influenza CInfluenza C.
virus.Negative results do not preclude
influenza or RSV virus infection
Negative results do not precludeand should not be used as the sole
respiratory virus infection andbasis for treatment or other
should not be used as the solemanagement decisions.
basis for diagnosis, treatment orConversely, positive results do
other patient managementnot rule out bacterial infection or
decisions.co-infection with other viruses.
The agent detected may not be the
Performance characteristics fordefinite cause of disease. The use
influenza A were establishedof additional laboratory testing
during the 2009/2010 andand clinical presentation must be
2010/2011 flu seasons whenconsidered in order to obtain the
influenza A (H3N2) and influenzafinal diagnosis of respiratory viral
A/2009 (HIN1) were theinfection.
predominant influenza A virusesPerformance characteristics for
circulation. When otherinInfluenza A Virus were
Influenza A viruses emerge,established when Influenza A/H3
performance characteristics mayand A/H1 were the predominant

{4}------------------------------------------------

CharacteristicDevicePredicate
vary.If infection with a novel influenzaA virus is suspected based oncurrent clinical andepidemiological screening criteriarecommended by public healthauthorities, specimens should becollected with appropriateinfection control precautions fornovel virulent influenza virusesand sent to state or local healthdepartments for testing. Viralculture should not be attempted inthese cases unless a BSL 3+facility is available to receive andculture specimens.Influenza A viruses in circulation(2006 - 2007 respiratory season).Performance characteristics forInfluenza A were confirmed whenInfluenza A/Hl, Influenza A/H3,and Influenza A/2009 H1N1 werethe predominant Influenza Aviruses in circulation (2008 and2009). When other Influenza Aviruses are emerging,performance characteristics mayvary.If infection with a novel InfluenzaA virus is suspected based oncurrent clinical andepidemiological screening criteriarecommended by public healthauthorities, specimens should becollected with appropriateinfection control precautions fornovel virulent Influenza virusesand sent to state or local healthdepartment for testing. Viralculture should not be attempted inthese cases unless a BSL 3+facility is available to receive andculture specimens.
Specimen TypeNasopharyngeal swabNasopharyngeal swab
Assay TargetsInfluenza A, Influenza BInfluenza A, Influenza B, RSV
Amplificationand DetectionTechnologyMultiplex real-time PCRMultiplex real-time PCR
Assay ControlsInfluenza A Control, Influenza BControl, Influenza A/B InternalControl, each prepared from invitro transcripts.Influenza A RNA Control,Influenza B RNA Control, RSV ARNA Control, RSV B RNAControl and Internal RNAControl, each prepared from invitro transcripts.
Influenza AVirus TargetMatrix geneMatrix gene
Differences
Influenza BVirus TargetMatrixNon-structural NS1 and NS2
CharacteristicDevicePredicate
Nucleic AcidExtractionAutomated extraction using theEZ1 DSP Virus Kit with the EZ1Advanced XL instrumentAutomated extraction using theRoche MagNA Pure LC Systemwith the MagNA Pure TotalNucleic Acid Isolation Kit or thebioMerieux NucliSENSeasyMAG System with theAutomated Magnetic ExtractionReagents.
Amplificationand DetectionInstrumentSystemRotor-Gene® Q MDxCepheid SmartCycler® II

{5}------------------------------------------------

Clinical Studies

Prospective clinical evaluation

The performance of the artus Infl A/B RG RT-PCR test was established during a prospective study at 3 clinical laboratories in the United States during the 2010-11 respiratory virus season. Nasopharyngeal (NP) swab specimens were collected for routine influenza testing by each site. In this prospective study the reference method was standard viral culture (shell vial) followed by direct fluorescent antibody (DFA) screening and identification. Demographic details for this patient population are summarized in Table .

SexNumber of Subjects
Male138 (54%)
Female116 (46%)
Age (years)
<590 (35.4%)
≥5 and ≤21106 (41.7%)
≥22 and ≤5951 (20.1%)
≥607 (2.8%)

Table 5.3 General Demographic Data for Prospectively Collected Specimens

A total of 254 NP swab specimens were evaluated. Of these, 60 specimens were positive by viral culture for influenza: 15 were positive for influenza A and 45 were positive for influenza B. The artus Infl A/B RG RT-PCR test identified all 15 influenza A specimens as positive for influenza A and all 45 influenza B specimens as positive for influenza B. There were no dual specimens identified by either the artus Infl A/B RG RT-PCR test or the reference test in the prospective study.

{6}------------------------------------------------

Reference method
+-Total
artus Infl A/B RG+1512a27
RT-PCR test-0227227
Total15239254
Sensitivity = 100% (95% confidence interval: 80-100%)
Specificity = 95.0% (95% confidence interval: 91-97%)

Table 5.4 Influenza A results from prospective clinical evaluation

a The artus Infl A/B RG RT-PCR test also identified an additional 12 culture-negative specimens as positive for influenza A. Ten of these 12 specimens were confirmed by bidirectional sequencing

Table 5.5 Influenza B results from prospective clinical evaluation
Reference method
+-Total
artus Infl A/B RGRT-PCR test4511a56
0198198
Total45209254
Sensitivity = 100% (95% confidence interval: 92-100%)
Specificity = 94.7% (95% confidence interval: 91-97%)

4 An additional 11 culture-negative specimens were identified as positive for influenza B. Six of these 11 specimens could not be confirmed by bidirectional sequencing. However, 3 of these specimens were positive in a validated, independent, real-time PCR reference test for influenza B.

Retrospective clinical evaluation

A retrospective analyses was conducted. A total of 212 specimens from the 2009 and 2010 respiratory virus season were tested at 3 sites. After testing with the artus Infl A/B RG RT-PCR test, aliquots of all specimens were sent to a central reference laboratory for reference testing using an FDA cleared molecular test.

A total of 208 specimens were positive for influenza by the reference test: 175 were positive for influenza A and 33 were positive for influenza B. The artus Infl A/B RG RT-PCR test identified all 175 influenza A specimens as positive for influenza A and all 33 influenza B specimens as positive for influenza B.

{7}------------------------------------------------

Reference method
+-Total
artus Infl A/B RG+1752a177
RT-PCR test-03535
Total17537212

Table 5.6 Influenza A results from retrospective clinical evaluation 1

a The artus Infl A/B RG RT-PCR test also identified an additional 2 specimens as positive for influenza A that were called negative or "No Call" by the molecular test.

Reference method
+-Total
artus Infl A/B RG+332a35
RT-PCR test-0177177
Total33179212
Positive agreement = 100% (95% confidence interval: 90-100%)
Negative agreement = 98.9% (95% confidence interval: 96-100%)

4 An additional 2 specimens were identified as positive for influenza B that were called negative by the molecular test .

Prospectively collected and archived clinical evaluation

A total of 462 well-characterized, de-identified, residual specimens were collected from August 2009 to May 2011 respiratory virus season. All specimens had been previously tested with FDA cleared molecular tests for routine patient management. These clinical results were used as the reference method. Demographic details for this patient population are summarized in Table 5.8

SexNumber of Subjects
Male190 (41%)
Female272 (59%)
Age (years)
<57 (1.5%)
≥5 and ≤2143 (9.3%)
≥22 and ≤59327 (70.8%)
≥6085 (18.4%)

Table 5.8 General Demographic Data for Archived Specimens

A total of 110 specimens were positive for influenza by the reference method: 96 were positive for influenza A and 14 were positive for influenza B. The artus Infl A/B RG RT-

{8}------------------------------------------------

PCR test identified 94 of the 96 influenza A specimens as positive for influenza A and all 14 influenza B specimens as positive for influenza B.

Reference method
+-Total
artus Infl A/B RG +9419a113
RT-PCR test -2347349
Total96366462
Positive agreement = 97.9% (95% confidence interval: 93-99%)
Negative agreement = 94.8% (95% confidence interval: 92-97%)

Table 5.9 Influenza A results from retrospective clinical evaluation 2

a The artus Infl A/B RG RT-PCR test also identified an additional 19 specimens as positive for influenza A that were reported as negative during routine clinical testing.

Table 5.10 Influenza B results from retrospective clinical evaluation 2

Reference method
+-Total
artus Infl A/B RG+142a16
RT-PCR test-0446446
Total14448462
Positive agreement = 100% (95% confidence interval: 79–100%)
Negative agreement = 99.6% (95% confidence interval: 98-100%)

4 An additional 2 specimens were identified as positive for influenza B that were called negative during routine clinical testing.

Non-clinical Studies

Reproducibility

The reproducibility of the artus Infl A/B RG RT-PCR test was evaluated using 3 investigational sites. A panel of 10 simulated specimens was provided for testing. Five of the specimens contained influenza A and the other 5 specimens contained influenza B. Each half of the panel included duplicate low-positive specimens at approximately 2 times the limit of detection (Low Pos #1 and #2), duplicate moderately positive specimens at approximately 10 times the limit of detection (Mod Pos #1 and #2), and one specimen well below the limit of detection for each analyte (Neg). The 10-member panel plus 3 controls were tested by 2 different technologists each day for 6 days. Therefore, 10 samples plus 3 controls, times 2 technologists, for 6 days, at 3 sites equals 468. The overall percent agreement for the artus Infl A/B RG RT-PCR test was 99.4% (Table 5. , Table 5., Table 5. , Table 5.).

{9}------------------------------------------------

Table 5. 11 Reproducibility - site 1

Panel memberAgreement withexpected resultAverage CTCV%
Influenza A Neg12/1229.542.2
Influenza A Low Pos #112/1231.380.9
Influenza A Low Pos #212/1231.471.5
Influenza A Mod Pos #112/1229.200.9
Influenza A Mod Pos #212/1229.171.2
Influenza B Neg12/1229.552.2
Influenza B Low Pos #112/1230.351.2
Influenza B Low Pos #212/1230.421.1
Influenza B Mod Pos #112/1228.211.1
Influenza B Mod Pos #212/1228.271.1
Influenza A Control12/1234.352.2
Influenza B Control12/1230.851.9
Negative control12/1227.191.4

.

{10}------------------------------------------------

Table 5.3 Reproducibility - site 2
Agreement with
Panel memberexpected resultAverage СтCV%
Influenza A Neg11/1229.892.5
Influenza A Low Pos #112/1231.241.5
Influenza A Low Pos #212/1231.351.2
Influenza A Mod Pos #112/1229.171.4
Influenza A Mod Pos #212/1229.161.4
Influenza B Neg12/1229.772.5
Influenza B Low Pos #112/1230.391.2
Influenza B Low Pos #212/1230.311.1
Influenza B Mod Pos #112/1228.291.0
Influenza B Mod Pos #212/1228.281.1
Influenza A Control12/1233.830.9
Influenza B Control12/1231.062.1
Negative control12/1227.482.1

Table 5. 4 Reproducibility – site 3

Agreement with
Panel memberexpected resultAverage СтCV%
Influenza A Neg10/1229.891.7
Influenza A Low Pos #I12/1231.450.8
Influenza A Low Pos #212/1231.350.6
Influenza A Mod Pos #112/1229.261.1
Influenza A Mod Pos #212/1229.452.4
Influenza B Neg12/1230.021.4
Influenza B Low Pos #112/1230.541.3
Influenza B Low Pos #212/1230.521.3
Influenza B Mod Pos #112/1228.461.8
Influenza B Mod Pos #212/1228.401.3
Influenza A Control12/1234.582.6
Influenza B Control12/1231.532.4
Negative control12/1227.611.4

Table 5.5 Reproducibility - combined (sites 1 - 3)

{11}------------------------------------------------

Panel memberAgreementwith expectedresultAverageCTCV%95%confidenceinterval
Influenza A Neg33/3629.762.278–97%
Influenza A Low Pos #172/7231.371.195–100%
Influenza A Mod Pos #172/7229.231.595–100%
Influenza B Neg36/3629.782.190–100%
Influenza B Low Pos #172/7230.421.295–100%
Influenza B Mod Pos #172/7228.321.295–100%
Influenza A Control36/3634.252.290–100%
Influenza B Control36/3631.142.390–100%
Negative control36/3627.421.890–100%

Limit of Detection (LoD)

The limit of detection (LoD) of the artus Infl A/B RG RT-PCR Kit was determined and confirmed for six influenza A strains (two strains representing each of the influenza A subtypes of H1N1, H3N2, and 2009 H1N1) and two influenza B strains. Samples were prepared from cultured virus diluted in nasopharyngeal clinical matrix. The LoD of each strain was initially determined by limiting dilution testing of three replicates per dilution level. The result was confirmed by testing an additional 20 replicates at the LoD concentration. The LoD, defined as the level of virus that yields at least a 95% (19/20) detection rate with the artus Infl A/B RG RT-PCR Kit, ranged from 10e1.1 to 10e-0.1 TCID50/ml (Table 5.).

{12}------------------------------------------------

Table 5.6 Limit of detection

Influenza strainLoD concentration (TCID50/ml*)
A/New Caledonia/20/1999 (H1N1)10e0.5
A/Brisbane59/2007-like virus (H1N1)10e1.1
A/Hong Kong/8/68 (TC-adapted) (H3N2)10e0.2
A/Wisconsin/67/2005 (H3N2)10e0.4
A/California/7/09-like virus (2009 H1N1)10e0.9
A/Hamburg/05/09 (2009 H1N1)10e0.4
B/Brisbane/60/2008-like virus10e-0.1
B/Florida/4/2006-like virus10e0.9
  • Based on re-titered virus stock concentrations.

Reactivity

The analytical reactivity of the artus Infl A/B RG RT-PCR Kit was demonstrated by testing 18 strains of influenza A (including four strains originally identified in non-human species) and six strains of influenza B at concentrations near the limit of detection (LoD) of the test. Samples were prepared from cultured virus diluted in nasopharyngeal clinical matrix. Each viral strain was extracted and tested in triplicate. A "+" results indicates the concentration was detected positive in three of three replicates; a "-" result indicates a negative response in each of the replicates (Table 5.).

Table 5.7 Analytical reactivity

Influenza strainConcentrationInfluenza AInfluenza B
A/Virginia/ATCC2/2009(2009 H1N1)1 x 102 TCID50/ml+-
A/PR/8/34 (H1N1)1 x 101 TCID50/ml+-
A/FM/1/47 (H1N1)1 x 101 CEID50/ml+-
A/Solomon Islands/3/2006like virus (H1N1)1 x 101 TCID50/ml+-
A/Mal/302/54 (H1N1)1 x 101 CEID50/ml+-
A/New Jersey/8/76 (H1N1)1 x 101 CEID50/ml+-
A/NWS/33 (H1N1)1 x 102 CEID50/ml+-
A1/Denver/1/57 (H1N1)1 x 103 CEID50/ml+-
A/Weiss/43 (H1N1)1 x 104 CEID50/ml+-
A/Victoria x187 (H3N2)1 x 101 TCID50/ml+-
A2/Aichi2/68 (H3N2)1 x 101 CEID50/ml+-

{13}------------------------------------------------

Influenza strainConcentrationInfluenza AInfluenza B
A/Victoria/3/75 (H3N2)1 x 10¹ CEID50/ml+-
A/Alice (H3N2)1 x 10² EID50/ml+-
A/MRC2 (H3N2)1 x 10² CEID50/ml+-
A/Turkey/Germany 176/95 -like (H9N2)1 x 10⁰ TCID50/ml+-
A/Duck/Potsdam 2243/84-like (H5N6)1 x 10⁰ TCID50/ml+-
A/Swine/Iowa/15/30 (H1N1)1 x 10² CEID50/ml+-
A/Equine/2/Miami/63-like(H3N8)1 x 10⁴ CEID50/ml+-
B/Lee/401 x 10¹ TCID50/ml-+
B/Allen/451 x 10¹ CEID50/ml-+
B/Taiwan/2/621 x 10¹ CEID50/ml-+
B/Hong Kong/5/721 x 10¹ CEID50/ml-+
B/Maryland/1/591 x 10¹ CEID50/ml-+
B/Malaysia/2506/20041 x 10¹ TCID50/ml-+

Interfering Substances

The potential for blood or medications that might be present in a nasopharyngeal swab specimen to interfere with the detection of low levels of influenza A or influenza B by the artus Infl A/B RG RT-PCR test was evaluated. Human blood and substances representing the active ingredient in over-the-counter or prescription medications were tested. Each potentially interfering substance was evaluated in triplicate with each of two reference influenza strains, one influenza A strain and one influenza B strain. The reference influenza strains were successfully detected in all samples except for one aliquot containing the reference influenza B strain and Mupirocin at 10 mg/ml. The reference influenza B strain was successfully detected in three of three aliquots with Mupirocin at 2 mg/ml. (Mupirocin is an antibiotic ointment used to treat skin infections. The product tested contains Mupirocin at 20mg/g.). Substances tested are shown in Table 5..

SourceDescription /Active IngredientConcentration
Blood (human)Blood (human)†5% v/v
VisadronPhenylephrine10% v/v(125 µg/ml)
Table 5.8 Potentially interfering substances

{14}------------------------------------------------

SourceDescription /Active IngredientConcentration
Beclomet NasalAqua (nasal spray)Beclomethasonedipropionate61.73 µg/ml
Luffa opperculataLuffa opperculata4.5 mg/ml
Galphimia glaucaGalphimia glauca4.5 mg/ml
MentholMenthol5 mg/ml
RelenzaZanamivir3 mg/ml
Infectopyoderm(ointment)Mupirocin10 mg/ml,2mg/ml
TobramaxinTobramycin0.3 mg/ml
  • The human blood sample was stored frozen before testing.

Analytical Specificity

The analytical specificity of the artus Infl A/B RG RT-PCR Kit was evaluated by testing a panel of respiratory pathogens consisting of 21 non-target respiratory viruses and 18 bacterial strains. The pathogens were tested at medically relevant levels. Samples were prepared by diluting the cultured virus or bacteria in Universal Transport Medium. Each strain was extracted once and tested in triplicate. There were no false positive or invalid results among the 21 non-target respiratory viruses tested. Of the 18 bacterial strains evaluated, 16 were negative for influenza A and influenza B; two strains, Bordetella pertussis (at 104 cfu/ml) and Streptococcus pneumoniae (at 2 x 10° cfu/ml), were invalid in three of three replicates. Each strain was prepared at a one-log lower concentration, extracted once, and tested in triplicate. At the lower concentration each strain generated valid negative results for influenza A and influenza B in three of three replicates (Table 5.).

StrainConcentrationInfluenza AInfluenza B
RSVA VR-2610e5.0 TCID50/ml--
RSVB VR-140010e4.45 TCID50/ml--
PIV1 VR-9410e3.45 TCID50/ml--
PIV2 VR-9210e5.0 TCID50/ml--
PIV3 VR-9310e5.0 TCID50/ml--
PIV4a VR-137810e3.95 TCID50/ml--
ADVE 4 VR-157210e5.0 TCID50/ml--
StrainConcentrationInfluenza AInfluenza B
ADVB 3 VR-310e5.0 TCID50/ml--
ADVC 5 VR-510e5.0 TCID50/ml--
Echovirus 11 VR-4110e5.0 TCID50/ml--
Rhinovirus 1a VR-155910e5.0 TCID50/ml--
Rhinovirus 39 VR-34010e5.0 TCID50/ml--
Coxsackie B1 VR-2810e5.0 TCID50/ml--
229E VR-74010e4.2 TCID50/ml--
OC43 VR-155810e4.45 TCID50/ml--
CMV VR-53810e4.95 TCID50/ml--
HSV VR-26010e5.0 TCID50/ml--
Varicella-zoster virusVR-136710e3.95 TCID50/ml--
EBV VR-60310e3.0 cfu/ml--
Measels VR-2410e3.2 TCID50/ml--
Mumps virus VR-10610e5.0 TCID50/ml--
Bordetella pertussis10e3.0 cfu/mlInvalid*Invalid*
Bordetella pertussis10e2.0 cfu/ml--
Chlamydophilapneumoniae10e5.0 TCID50/ml--
Corynebacterium sp.10e3.0 cfu/ml--
Escherichia coli10e6.0 cfu/ml--
Hemophilus influenzae10e6.0 cfu/ml--
Lactobacillus sp.10e3.0 cfu/ml--
Legionella spp.9 x 10e5.0 bacteria/ml--
Moraxella catarrhalis10e3.0 cfu/ml--
Mycobacteriumtuberculosis avirulent10e2.0 cfu/ml--
Mycoplasmapneumoniae10e3.0 cfu/ml--
Neisseria ssp. #1468510e2.0 cfu/ml--
StrainConcentrationInfluenza AInfluenza B
Neisseria meningitis10e6.0 bacteria/ml--
Pseudomonasaeruginosa10e6.0 cfu/ml--
Staphylococcus aureusProtein A producer10e6.0 bacteria/ml--
Staphylococcusepidermidis10e6.0 bacteria/ml--
Streptococcuspneumoniae2 x 10e5.0 cfu/mlInvalid*Invalid*
Streptococcuspneumoniae2 x 10e4 cfu/ml--
Streptococcus pyogenes10e3.0 cfu/ml--
Streptococcus salivarius10e6.0 cfu/ml--

Table 5.9 Analytical specificity

{15}------------------------------------------------

{16}------------------------------------------------

  • Samples were invalid because influenza A and influenza B results were negative and the Cr of the IC was outside of the acceptance range for specimen validity. A second concentration was tested successfully and is presented below the "invalid" sample.

Competitive Inhibition

Competitive inhibition of the artus Infl A/B RG RT-PCR test was evaluated by testing samples containing reference strains A/California/7/09-like virus or B/Florida/4/2006like virus at the limit of detection (LoD) concentration of 10e1 TCIDsoml. Samples containing the influenza A reference strain also contained high levels of one of three competing influenza B strains while samples containing the influenza B reference strain also contained high levels of one of seven competing influenza A strains. Samples were extracted once and tested in triplicate with the artus Infl A/B RG RT-PCR Kit.

Detection of the reference influenza A strain was inhibited when the competing influenza B strains B/Florida/4/2006-like virus or B/Brisbane/60/2008-like virus was present at 10e5 TCIDso/ml; no inhibition was observed when the same competing virus strains were present at 10e4 TCID50ml. No inhibition was observed with the competing influenza B strain B/Malaysia/2506/2004. The reference influenza B strain was successfully detected in the presence of each of the seven competing Influenza A strains present at 10e5 TCID50/ml (Table 5.).

{17}------------------------------------------------

Positive replicates/total replicates
Competing virus strainConcentration(TCID50/ml)Influenza AreferencestrainInfluenza Breferencestrain
A/California/7/09-like virus10e5n/a*3/3
A/Hong Kong/8/68 (TCadapted)10e5n/a3/3
A/PR/8/3410e5n/a3/3
A2/Wisconsin/67/200510e5n/a3/3
A/Solomon Islands/3/2006(H1N1)-like virus10e5n/a3/3
A/Duck/Potsdam 2243/8410e5n/a3/3
A/Brisbane/59/2007(H1N1)-like virus10e5n/a3/3
B/Florida/4/2006-like virus10e51/3n/a
B/Florida/4/2006-like virus10e43/3n/a
B/Brisbane/60/2008-likevirus10e52/3n/a
B/Brisbane/60/2008-likevirus10e43/3n/a
B/Malaysia/2506/200410e53/3n/a

Table 5.10 Competitive inhibition

Carryover / Cross Contamination

The artus Infl A/B RG RT-PCR test (including extraction using the EZ1 DSP Virus Kit with the EZ1 Advanced XL) showed no evidence of carryover or cross-contamination when 5 runs of a panel of 6 members of mock samples containing influenza A at a concentration just below the limit of detection of the assay were extracted and tested in alternating order with a panel of 6 members of mock samples of the same strain present at a high concentration.

{18}------------------------------------------------

Image /page/18/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three overlapping wings, representing the department's commitment to health, human services, and science. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.

10903 New Hampshire Avenue Silver Spring, MD 20993

FEB - 6 2012

Oiagen GmbH c/o Hina Patel Associate Director, Regulatory and Clinical Affairs 1201 Clopper Road Gaithersburg, MD 20878

Re: K113323

Trade/Device Name: artus® Infl A/B RG RT-PCR Kit Regulation Number: 21 CFR§ 866.3980 Regulation Name: Respiratory viral panel multiplex nucleic acid assay Regulatory Class: Class II Product Code: OCC, OOI, JJH Dated: October 24, 2011 Received: November 10, 2011

Dear Ms. Patel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket

{19}------------------------------------------------

Page 2 - Hina Patel

notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Salla Agnis

Sally A. Hojvat, M.Sc., I Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{20}------------------------------------------------

Indications for Use

510(k) Number (if known): K113323

Device Name: artus® Infl A/B RG RT-PCR Kit

Indications for Use:

The artus® Infl A/B RG RT-PCR Kit is a multiplex real time PCR in vitro diagnostic test for the qualitative detection and identification of Influenza A and Influenza B virus RNA in nasopharyngeal swab specimens using the Rotor-Gene® Q MDx instrument. The test is intended for use as an aid in the differential diagnosis of Influenza A and Influenza B viral infections in patients symptomatic for respiratory tract infection in conjunction with clinical and epidemiological risk factors. It is not intended to detect Influenza C virus.

Negative results do not preclude respiratory virus infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

Performance characteristics for Influenza A were established during the 2009/2010 and 2010/2011 flu seasons when Influenza A (H3N2) and Influenza A/2009 (H1N1) were the predominant Influenza A viruses in circulation. When other Influenza A viruses emerge, performance characteristics may vary.

If infection with a novel Influenza A virus is suspected based on current clinical and evidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Prescription Use X (21 CFR 801 Subpart D)

And/Or

Over-The-Counter Use | | (21 CFR 807 Subpart C)

(Do Not Write Below This Line - Continue on Another Page if Needed)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Tauran V. Feldberg

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K 113323

Page 1 of 1

10/24/2011

§ 866.3980 Respiratory viral panel multiplex nucleic acid assay.

(a)
Identification. A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B;
(2) Influenza A subtype H1 and Influenza A subtype H3;
(3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B;
(4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus;
(5) Human Metapneumovirus;
(6) Rhinovirus; and
(7) Adenovirus.
(b)
Classification. Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.