K Number
K113284

Validate with FDA (Live)

Date Cleared
2012-03-07

(121 days)

Product Code
Regulation Number
892.1710
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FUJIFILM Biopsy Positioner (FDR-1000BPY) is an optional accessory for the Aspire HD (FDR MS-1000) digital mammography system. It is designed to allow for accurate determination of three dimensional lesion locations in the breast using information extracted from stereotactic pairs of two-dimensional images. It is intended to provide guidance for interventional purposes (such as biopsy, presurgical localization or treatment devices).

Device Description

The FUJIFILM Biopsy Positioner is an optional accessory for the Aspire HD (FDR MS-1000) full field digital mammography system and future Fujifilm digital mammography exposure stands. It is designed for positioning the needle when performing stereotactic biopsies, fine needle aspirations, core needle, and vacuum assisted biopsies in an upright position. The Biopsy Positioner uses a stereo pair of images, taken with the tube arm angled to +1 5° from vertical (orthogonal to the plane of the image receptor). The position of the three coordinates (X, Y, and Z) is computed using the position of the object on each of the two images, as indicated by the operator, and the known geometry of the system.

The Biopsy Positioner is composed of the following elements:

  • The Positioner that supports and positions the needle .
  • The Operation panel which displays the distance between the compression plate and target ● pathology, the distance between the target pathology and the needle, and also electrically drives the positioner in the X, Y and Z directions.
  • The Positioner Control Unit which supplies power to the positioner
AI/ML Overview

The provided text describes a 510(k) premarket notification for the FUJIFILM Biopsy Positioner (FDR-1000BPY) but does not contain a specific study proving the device meets quantitative acceptance criteria for performance. The document focuses on demonstrating substantial equivalence to a predicate device and compliance with various safety and electrical standards.

However, based on the information provided regarding the device's function, we can infer a key performance metric and a potential acceptance criterion for its primary function.

Primary Function of the Device:
The device's core purpose is to calculate the three-dimensional coordinates of a target pathology with position accuracy. The document states: "The software calculates the three-dimensional coordinates of the target pathology with position accuracy of -/+1.0mm or less." This statement itself implies an acceptance criterion.

Here's a breakdown of the requested information based on the provided text, with an acknowledgment of missing details:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Position accuracy of ±1.0 mm for 3D coordinates of target pathology.The software calculates the three-dimensional coordinates of the target pathology with position accuracy of -/+1.0mm or less.

Missing Information: The document does not explicitly state other quantitative acceptance criteria, nor does it provide a detailed report of a study directly demonstrating the device's performance against this (or any other) criterion. The statement "The software calculates... with position accuracy of -/+1.0mm or less" is a claim about the device's capability rather than a report of a verification study result with a measured mean or standard deviation.


2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not specified.
  • Data Provenance: Not specified. The document outlines the device's design and intended use but does not describe a clinical or technical test set used to verify its accuracy.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Number of Experts: Not applicable, as no described test set or ground truth establishment process is detailed for performance verification.
  • Qualifications of Experts: Not applicable.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable, as no described test set or adjudication process is detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? No. The document does not describe an MRMC study. The device is a "Biopsy Positioner," which is an accessory for guidance, not a diagnostic AI system directly interpreting images alongside human readers.

6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop)

  • Was a standalone study done? The document states: "The software calculates the three-dimensional coordinates of the target pathology with position accuracy of -/+1.0mm or less." While this implies an internal validation of the software's accuracy, the text does not describe the methodology or results of a standalone study to prove this claim. It's a statement of capability rather than a report of a study's findings.

7. Type of Ground Truth Used

  • Type of Ground Truth: Not explicitly stated for performance verification. For a device that measures physical dimensions and positions, the ground truth would typically be established through precise metrological measurements using phantom objects or calibrated systems, not clinical pathology or outcomes data. However, the document does not detail how the stated accuracy was verified.

8. Sample Size for the Training Set

  • Sample Size: Not applicable. The "Biopsy Positioner" is described as a system that calculates coordinates based on known geometry and operator input from stereo images. It is not presented as an AI/ML device that requires a training set in the typical sense (e.g., for image interpretation). Its function is more akin to precise geometric calculation rather than pattern recognition learning.

9. How Ground Truth for the Training Set Was Established

  • How Ground Truth Was Established: Not applicable. As noted above, the device's functionality does not appear to involve a "training set" in the context of machine learning. Its accuracy would be verified based on its computational precision relative to physical measurements rather than learned patterns.

{0}------------------------------------------------

MAR - 7 2012

FUJIFILM

FUJIFILM MEDICAL SYSTEMS U.S.A.

419 WEST AVENUE STAMFORD, CT 06902 203-324-2000 PHONE· TOLL-FREE: 800-431-1850 203-353-0926

510(k) Summary

Date Prepared: November 4, 2011

Submitter's Information

FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, Connecticut 06902 Telephone: (203) 602-3665 Facsimile: (203) 251-7863 Contact: Kimerly A. Sharp

Device Name and Classification:

Product Name: Model Number: Classification Name: Classification Panel: CFR Section: Device Class: Product Code:

Biopsy Positioner FDR-1000BPY Mammographic X-ray System Radiology 21 CFR 892.1710 Class II IZH

Substantial Equivalence/Predicate Device:

Hologic Affirm Breast Biopsy Guidance System (K103512)

The FUJIFIM Biopsy Positioner (FDR-1000BPY) is substantially equivalent to the Hologic Affirm Breast Biopsy Guidance System (used with Selenia Dimensions 2D FFDM) which is a stereotactic lesion localization system that has the application of localizing, and then giving a physician the capability of performing fine needle aspiration or core biopsy of lesions determined to be suspicious through prior mammographic examination

Description of the Device:

The FUJIFILM Biopsy Positioner is an optional accessory for the Aspire HD (FDR MS-1000) full field digital mammography system and future Fujifilm digital mammography exposure stands. It is designed for positioning the needle when performing stereotactic biopsies, fine needle aspirations, core needle, and vacuum assisted biopsies in an upright position. The Biopsy Positioner uses a stereo pair of images, taken with the tube arm angled to +1 5° from vertical (orthogonal to the plane of the image receptor). The position of the three coordinates (X, Y, and Z) is computed using the position of the object on each of the two images, as indicated by the operator, and the known geometry of the system.

The Biopsy Positioner is composed of the following elements:

  • The Positioner that supports and positions the needle .
  • The Operation panel which displays the distance between the compression plate and target ● pathology, the distance between the target pathology and the needle, and also electrically drives the positioner in the X, Y and Z directions.
  • The Positioner Control Unit which supplies power to the positioner

{1}------------------------------------------------

FUJKFILM

FUJIFILM Biopsy Positioner (FDR-1000BPY) Traditional 510(k) Premarket Notification

The FUJIFILM Biopsy Positioner slides onto the Aspire HD image receptor. The cable from the positioner control cabinet has two connectors which are plugged into the lower part on the front of the exposure stand. The operation panel is then attached to one of the armrests. Stereotactic images are acquired at +/- 15°. The biopsy software processes and displays the images on the Acquisition Workstation (AWS) for targeting procedure. Targeting procedure involves checking the same object on each image which results in setting the target mark. The software calculates the threedimensional coordinates of the target pathology with position accuracy of -/+1.0mm or less. When targeting is completed, the three-dimensional coordinates of the target pathology are calculated based on the target mark at one position of each stereo image. The positioner can be moved by operating the dials for X, Y and Z axes or via the electric drive buttons on the operation panel (Z axis movements are manually applied when needle is connected).

Safety Features include:

  • An audible alarm will sound from the operation panel under the following conditions: (Unless appropriate action is taken, the alarm will continue to sound)
  • The Positioner is not locked to the exposure unit.
  • Needle end is positioned less than 4.0mm from exposure table surface prevents collision.
  • The needle holder is attached and the positioner is within 20mm from the top end in the Z direction (prevents the needle and the tube from interfering with each other).
    • The Stereo Biopsy Software displays information on the selected needle, the impossibility of a study, precautions for executing a study, etc.
    • Target Position is shown via diagram on the display screen

Intended Use:

The FUJIFILM Biopsy Positioner (FDR-1000BPY) is an optional accessory for the Aspire HD (FDR MS-1000) digital mammography system. It is designed to allow for accurate determination of three dimensional lesion locations in the breast using information extracted from stereotactic pairs of two-dimensional images. It is intended to provide guidance for interventional purposes (such as biopsy, pre-surgical localization or treatment devices).

Safety Information:

The Biopsy Positioner introduces no new safety or efficacy issues other than those already identified with the predicate devices. The results of the Hazard Analysis combined with the appropriate preventive measures taken indicate that the device is of moderate concern as per the May 11, 2005 issue of the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." And consistent with that indicated in the "Class II Special Controls Guidance Document: Full-Field Digital Mammography System" document issued on: November 5, 2010.

Summary of Testing:

This equipment conforms to the following:

  • IEC 60601-1-2:2001+A1:2004 .
  • . IEC 60601-1:1988 + A1:1991 +A2:1995
  • . IEC 60601-1-1:2000
  • IEC 60601-2-45: 2001 .
  • IEC 62304: 2006

{2}------------------------------------------------

FUJIFILM

FUJIFILM Biopsy Positioner (FDR-1000BPY) Traditional 510(k) Premarket Notification

  • IEC 60601-2-45: 2001 •
  • . DICOM (Version 3)

Conclusion:

This 510(k) premarket notification submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. We conclude the subject device to be as safe and effective as the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

MAR - 7 2012

Ms. Kimerly A. Sharp RA/QA Specialist FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue STAMFORD CT 06902

Re: K113284

Trade/Device Name: Biopsy Positioner (FDR-1000BPY) Regulation Number: 21 CFR 892.1710 Regulation Name: Mammographic x-ray system Regulatory Class: II Product Code: IZH Dated: January 20, 2012 Received: January 23, 2012

Dear Ms. Sharp:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{4}------------------------------------------------

Page 2

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirences as begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): (3284

Device Name: Biopsy Positioner (FDR-1000BPY)

Indications for Use:

The FUJIFILM Biopsy Positioner (FDR-1000BPY) is an optional accessory for the Aspire HD (FDR MS-1000) digital mammography system. It is designed to allow for accurate determination of three dimensional lesion locations in the breast using information extracted from stereotactic pairs of two-dimensional images. It is intended to provide guidance for interventional purposes (such as biopsy, presurgical localization or treatment devices).

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)Division of Radiological DevicesOffice of In Vitro Diagnostic Device Evaluation and Safety
510K.K113284

ge 1 of 1

§ 892.1710 Mammographic x-ray system.

(a)
Identification. A mammographic x-ray system is a device intended to be used to produce radiographs of the breast. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.