K Number
K113232

Validate with FDA (Live)

Device Name
MAXMIXING BLOCK
Manufacturer
Date Cleared
2012-03-01

(120 days)

Product Code
Regulation Number
868.5330
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MaxMixing Block is designed to provide intermittent or continuous controlled flows of an air/oxygen gas mixture to infant, pediatric, and adult patients. The MaxMixing Block is intended to be used with a Maxtec Oxygen Analyzer device. The MaxMixing Block is a restricted medical device intended for use by qualified trained personnel under the direction of a physician in institutional environments where delivery and monitoring of air/oxygen mixtures is required.

Device Description

The MaxMixing Block uses the well known technology of two compensated thorpe tube flowmeter to control and measure medical grade air and oxygen at an accurate flow rate as per a calibrated reference pressure (50 psig +/- 5 psig). The two gases are then mixing block for continuous air/oxygen delivery at oxygen concentrations of 21 - 100% as prescribed. There are no new technological characteristics introduced with respect to design, materials, chemical composition or energy source for the MaxMixing Block from those of flowmeters currently sold for medical use. There are differences with respect to mixers, in that the subject device has a wider pressure differential tolerance requirement since Oxygen concentration is much less affected by pressure differential as compared to the blender predicate devices. The blender predicate devices have a single gas failure alarm which is intended to indicate narrow pressure differential between air and Oxygen gas sources and failure of a single gas source, either air or Oxygen. The subject device also has a single gas failure alam which indicates a broader pressure differential between air and Oxygen gas sources and failure of a single gas source, either air or Oxygen.

AI/ML Overview

The MaxMixing Block device is designed to provide intermittent or continuous controlled flows of an air/oxygen gas mixture. The stated accuracy for oxygen concentration measurement is at least +/- 3%.

Here's a breakdown of the acceptance criteria and study information:

1. Acceptance Criteria and Reported Device Performance:

The document mentions that both the subject device (MaxMixing Block) and predicate devices are capable of measuring oxygen concentration with an accuracy of at least +/- 3%. While the provided text doesn't contain a specific table detailing "acceptance criteria" versus "reported device performance" for the MaxMixing Block itself, it states that
"Non-clinical functional and performance tests for substantial equivalence testing and the results are noted in the attached table."
However, this "attached table" is not provided in the input text. Therefore, the specific performance data against explicit acceptance criteria cannot be directly extracted here. We can infer the primary acceptance criterion is related to oxygen concentration accuracy based on the comparison to predicate devices.

Acceptance Criteria (Inferred from Predicate Devices)Reported Device Performance
Oxygen concentration accuracy of at least +/- 3%Not explicitly detailed in the provided text, but implied to meet this based on substantial equivalence.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for Test Set: Not specified in the provided text. The document refers to "non-clinical functional and performance tests" but does not give details about the size or nature of the test set.
  • Data Provenance: Not specified. It's a non-clinical test, so there's no mention of country of origin of data or whether it's retrospective or prospective with human subjects.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

Not applicable. The study involved "non-clinical functional and performance tests" and no clinical studies were performed. Therefore, no human experts were involved in establishing ground truth for a test set in a clinical context.

4. Adjudication Method for the Test Set:

Not applicable. Since no clinical studies were performed and no human experts established ground truth, there was no adjudication method.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No. The document explicitly states: "No clinical studies were performed". Therefore, no MRMC comparative effectiveness study was conducted.

6. Standalone Performance Study:

Yes, a standalone study was done in the form of "non-clinical functional and performance tests". The device was evaluated on its own merits against expected performance, as indicated by the comparison to predicate devices.

7. Type of Ground Truth Used:

The ground truth for the non-clinical performance tests would likely be based on:

  • Reference standards/calibrated instruments: For measuring gas flow and oxygen concentration to verify the accuracy of the MaxMixing Block.
  • Engineering specifications and design requirements: To ensure the device operates within defined parameters.

8. Sample Size for the Training Set:

Not applicable. This is a medical device (mixer) employing "well known technology". Unless it utilizes Machine Learning/AI, there is typically no "training set" in the conventional sense for such devices. The document explicitly states "No clinical studies were performed" and also notes "There are no new technological characteristics introduced with respect to design, materials, chemical composition or energy source for the MaxMixing Block from those of flowmeters currently sold for medical use." This further supports the absence of a training set for an AI/ML model.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there was no training set for an AI/ML model.

{0}------------------------------------------------

6526 South Cottonwood Street • Salt Lake City, Utah 84107 • www.maxtecinc.com • 866.4.maxtec

510(k) Summary of Safety and Effectiveness 1113232

6526 South Cottonwood Street Salt Lake City, Utah 84107

Director of Regulatory and Quality

This summary of safety and effectiveness is provided as part of this Pre-Market Notification in compliance with 21 CFR Part 807, Subpart E, section 807.92, as a means of providing sufficient detail to provide understanding of the basis for a determination of substantial equivalence.

    1. Submitter's name, address, telephone number, a contact person and the date the summary was prepared.
      Maxtec. LLC

(801) 266-5300

(801) 270-5590

Maxtec, LLC

(801) 327-9870

(801) 270-5590

Tammy Lavery, RAC

Salt Lake City, UT 84107

Submitter's name/address:

Phone: Fax: Contact Name: Contact Title: Contact Address:

Phone: Fax:

Date Summary prepared:

10/19/2011

    1. Subject device information:
Device Name:Air/Oxygen Mixer
Trade Name(s):MaxMixing Block
Classification Names:Mixer, Breathing Gases, Anesthesia Inhalation
Common/Usual Name:Breathing Gas Mixer
Classification:II
Product Code:BZR
CFR Reference:21 CFR 868.5330
Classification Panel:Anesthesiology
    1. The following predicate devices were used to establish substantial equivalence for the MaxMixing Block/Analyzer System:
    • 510(k) K884973 FR6000 Flowmeter �
    • 510(k) K053232 Precision Air Oxygen Blender ●
    • 510(k) K081335 Tenacore Oxygen Blender .
    • 510(k) K973646 Bird Sentry Blender .

MAR - 1 2012

{1}------------------------------------------------

    1. Description of the subject device:
      Both the subject and predicate devices are capable of measuring oxygen concentration with an accuracy of at least +/- 3%. As per predicates listed above, the indication noted has been cleared by the FDA.
    1. Statement of indication for use:
      The MaxMixing Block is designed to provide intermittent or continuous controlled flows of an air/oxygen gas mixture to infant, pediatric, and adult patients. The MaxMixing Block is intended to be used with a Maxtec Oxygen Analyzer device. The MaxMixing Block is a restricted medical device intended for use by qualified trained personnel under the direction of a physician in institutional environments where delivery and monitoring of air/oxygen mixtures is required.
    1. Technological characteristics:
      The MaxMixing Block uses the well known technology of two compensated thorpe tube flowmeter to control and measure medical grade air and oxygen at an accurate flow rate as per a calibrated reference pressure (50 psig +/- 5 psig). The two gases are then mixing block for continuous air/oxygen delivery at oxygen concentrations of 21 - 100% as prescribed. There are no new technological characteristics introduced with respect to design, materials, chemical composition or energy source for the MaxMixing Block from those of flowmeters currently sold for medical use. There are differences with respect to mixers, in that the subject device has a wider pressure differential tolerance requirement since Oxygen concentration is much less affected by pressure differential as compared to the blender predicate devices. The blender predicate devices have a single gas failure alarm which is intended to indicate narrow pressure differential between air and Oxygen gas sources and failure of a single gas source, either air or Oxygen. The subject device also has a single gas failure alam which indicates a broader pressure differential between air and Oxygen gas sources and failure of a single gas source, either air or Oxygen.

The intended use is the same as the combined predicate devices.

  • Non-clinical functional and performance tests for substantial equivalence testing and a. the results are noted in the attached table.
  • b. No clinical studies were performed No clinical studies were performed for the UltraMaxO2 as the device represents a well known technology for a recognized indication as evidenced in Section 5.0 by comparison to the predicate devices currently cleared for sale in the US market.
    1. The conclusion drawn from the non-clinical and clinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate devices in identified in 3) above.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, representing the department's mission to protect the health of all Americans.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Tammy Lavery, RAC Director of Regulatory and Quality Maxtec, Inc 6526 South Cottonwood Street Salt Lake City, Utah 84107

Re: K113232

Trade/Device Name: MaxMixing Block Regulation Number: 21 CFR 868.5330 Regulation Name: Breathing Gas Mixer Regulatory Class: II Product Code: BZR Dated: February 20, 2012 Received: February 22, 2012

Dear Ms. Lavery:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.

MAR - 1 2012

{3}------------------------------------------------

Page 2 - Ms. Lavery

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the I ou nike of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

$\mathscr{A}$

Image /page/3/Picture/6 description: The image contains a handwritten symbol or character. It appears to be a cursive lowercase letter, possibly an 'l' or a stylized abbreviation. The stroke begins with a loop at the top and extends downward with a curved line, ending with a short horizontal stroke at the bottom.

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices

Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement I.2

510(k) Number (if known):_

Device Name: MaxMixing Block

The MaxMixing Block, a breathing gas mixer designed to provide intermittent on controlled flows of an air/oxygen gas mixture to infant, pediation . The MaxMixing Block is .
flows of an air/oxygen gas mixture to infant, pediation a costristed modical intended to be used with a Maxtec Oxygen Analyzer. The MaxMixing Block is a restricted medical intended to be used with a Maxtec Oxygen Analyzer. The direction of a physician in institutional device mended for use by qualified attitude at air/oxygen mixtures is required.

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)

Prescription Use _

AND/OR

Over-The-Counter Use

L. Schultheis

(Division Sign-Off) (Division Sign-on)
Division of Anesthesiology, General Hospital +
Division of Anesthesion Devices Division of Antonion Devices

510(k) Number: u113232

§ 868.5330 Breathing gas mixer.

(a)
Identification. A breathing gas mixer is a device intended for use in conjunction with a respiratory support apparatus to control the mixing of gases that are to be breathed by a patient.(b)
Classification. Class II (performance standards).