(40 days)
Not Found
No
The description focuses on the mechanical and functional aspects of a catheter for aortic occlusion during cardiopulmonary bypass, with no mention of AI or ML technologies.
No
This device is used to occlude and vent the ascending aorta and deliver cardioplegia during cardiopulmonary bypass. While it assists in a medical procedure, it does not directly treat or cure a disease or condition itself, which is the primary characteristic of a therapeutic device.
No
The device is designed for occlusion and venting of the ascending aorta, delivery of cardioplegia, and monitoring aortic root pressure during cardiopulmonary bypass. While it monitors pressure, its primary function is therapeutic (occlusion, venting, delivery) rather than diagnostic interpretation or analysis of disease.
No
The device description clearly details a physical catheter with lumens, a balloon, and other hardware components. There is no mention of software as the primary or sole component.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostic devices are used to examine specimens (like blood, urine, or tissue) taken from the human body to provide information for diagnosis, monitoring, or screening.
- Device Function: The IntraClude Intra-Aortic Occlusion Catheter is a physical device used within the body during surgery (cardiopulmonary bypass). Its function is to physically occlude the aorta, deliver cardioplegia, and monitor pressure. It does not analyze biological specimens.
The description clearly indicates a surgical tool used for a specific procedure, not a device for laboratory testing of samples.
N/A
Intended Use / Indications for Use
The IntraClude Intra-Aortic Occlusion Catheter is indicated for use in patients undergoing cardiopulmonary bypass. The IntraClude Intra-Aortic Occlusion Catheter occludes and vents the ascending aorta when the balloon is inflated. The device's central lumen allows delivery of cardioplegia to arrest the heart. The pressure lumen allows monitoring of the aortic root pressure.
Product codes
DXC
Device Description
The IntraClude Intra-Aortic Occlusion Device is a 9 Fr (3mm), triple-lumen, 100-cm-long catheter with an elastomeric balloon near its distal tip designed to occlude the ascending aorta in order to partition the aortic root from arterial circulation during cardiopulmonary bypass (CPB). The balloon expands to occlude a range of aorta sizes from 20 to 40 mm. The shaft is provided with an extended strain relief, which tapers from 10.5 Fr to the remaining 9 Fr catheter, and is designed to prevent kinking and to avoid compressing the shaft when the hemostasis valve of an arterial cannula introducer sheath is firmly closed around the catheter body.
The large central lumen of the IntraClude catheter serves three functions: accommodating the guidewire, delivering cardiopledia solution to the aortic root, and venting fluid and air from the aortic root. The two remaining lumens serve as conduits for balloon inflation/deflation and aortic root pressure monitoring. The hub has two flexible extension tubes with an integrated luer connection to provide access for accessories. The shaft is provided with markers to indicate the insertion depth. A Clamp-Lock™ device, provided on the extended strain relief portion, allows the IntraClude catheter to be locked in position to prevent balloon migration during aortic occlusion. The devices are provided sterile and non-pyrogenic; they are intended for single use only.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
ascending aorta, aortic root
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
Functional/Safety Testing: The following functional tests were performed. All data met pre-determined acceptance criteria: Biocompatibility Per ISO 10993-1 for External communicating device, direct circulating blood path, duration ≤ 24 hours. Tensile testing. Flow and Venting Rates. Catheter Bending. Dynamic Pressure. Balloon testing. Design Validation through animal and cadaver studies.
Key Metrics
Not Found
Predicate Device(s)
K974175: Heartport Endoaortic Clamp Catheter
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.4450 Vascular clamp.
(a)
Identification. A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.(b)
Classification. Class II (performance standards).
0
510(k) Summary | |
---|---|
Submitter: | Edwards Lifesciences® LLC |
Contact Person: | Spencer Walker, Regulatory Affairs Associate II |
12050 Lone Peak Pkwy | |
Draper, UT 84020 | |
(801) 565-6100 | |
Date Prepared: | October 27, 2011 |
Trade Name: | Edwards Lifesciences® IntraClude™ Intra-Aortic Occlusion Catheter |
Classification Name: | Vascular Clamp |
21 CFR Part 870.4450, Product Code DXC, Class II | |
Predicate Device: | K974175: Heartport Endoaortic Clamp Catheter |
Device Description:
The IntraClude Intra-Aortic Occlusion Device is a 9 Fr (3mm), triple-lumen, 100-cm-long catheter with an elastomeric balloon near its distal tip designed to occlude the ascending aorta in order to partition the aortic root from arterial circulation during cardiopulmonary bypass (CPB). The balloon expands to occlude a range of aorta sizes from 20 to 40 mm. The shaft is provided with an extended strain relief, which tapers from 10.5 Fr to the remaining 9 Fr catheter, and is designed to prevent kinking and to avoid compressing the shaft when the hemostasis valve of an arterial cannula introducer sheath is firmly closed around the catheter body.
The large central lumen of the IntraClude catheter serves three functions: accommodating the guidewire, delivering cardiopledia solution to the aortic root, and venting fluid and air from the aortic root. The two remaining lumens serve as conduits for balloon inflation/deflation and aortic root pressure monitoring. The hub has two flexible extension tubes with an integrated luer connection to provide access for accessories. The shaft is provided with markers to indicate the insertion depth. A Clamp-Lock™ device, provided on the extended strain relief portion, allows the IntraClude catheter to be locked in position to prevent balloon migration during aortic occlusion. The devices are provided sterile and non-pyrogenic; they are intended for single use only.
Indications For Use:
The IntraClude Intra-Aortic Occlusion Catheter is indicated for use in patients undergoing cardiopulmonary bypass. The IntraClude Intra-Aortic Occlusion Catheter occludes and vents the ascending aorta when the balloon is inflated. The device's central lumen allows delivery of cardioplegia to arrest the heart. The pressure lumen allows monitoring of the aortic root pressure.
Comparative Analysis:
It has been demonstrated that the IntraClude catheter is comparable to the predicate device in intended use, fundamental scientific technology, design, principles of operation and functional performance evaluations. The IntraClude catheter has been fully assessed within the Edwards' Risk Management and Design Controls systems. All necessary verification steps met pre-determined acceptance criteria to confirm safety and efficacy.
1
Functional/Safety Testing: -
The following functional tests were performed. All data met pre-determined acceptance criteria.
- Biocompatibility Per ISO 10993-1 for External communicating device, direct circulating blood . path, duration ≤ 24 hours.
- Tensile testing Confirmation of the bond strength of the catheter, lumens, and connected . accessories.
- . Flow and Venting Rates - Inspection of catheter, lumens, and accessories for cardioplegia flow and aortic root venting rates.
- . Catheter Bending - Inspection of catheter, lumens, and accessories to confirm functionality after manipulation of the catheter.
- Dynamic Pressure Confirmation of device integrity after exposure to dynamic water flow and . manipulation of the catheter, lumens, and accessories.
- Balloon testing Testing for inflation times, balloon diameter, shape, insertion/retraction . force, and burst pressures.
- Design Validation Design was validated through animal and cadaver studies. .
Conclusion:
The IntraClude Intra-Aortic Occlusion Catheter is substantially equivalent to the cited predicate device. Additionally, the IntraClude catheter met all pre-determined acceptance criteria to confirm safety and efficacy.
2
Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular and contains an image of a bird in the center. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged around the perimeter of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DEC - 7 2011
Edwards LifeSciences, LLC c/o Mr. Spencer Walker Regulatory Affairs Associate 12050 Lone Peak Parkway Draper, UT 84020
Re: K113182
IntraClude Intra-Aortic Occlusion Device Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II (two) Product Code: DXC Dated: October 27, 2011 Received: October 28, 2011
Dear Mr. Walker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act (Act that do not require approval of the Federal Food, Drug,
You may, therefore, market the device, subject to the approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misting and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations or class III (PMA)
can be found in the Code of Fodarel Procellations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
3
Page 2 – Mr. Spencer Walker
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements not the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the election c product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
M. A. Willeheme
C - Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
4
Edwards Lifesciences LLC Special 510{k) Premarket Notification IntraClude Intra-Aortic Occlusion Catheter
Statement of Indications for Use
Indications for Use
KITSIBL 510(k) Number (if known):
Device Name: Edwards Lifesciences IntraClude Intra-Aortic Occlusion Catheter
The IntraClude Intra-Aortic Occlusion Catheter is indicated for use in patients undergoing cardiopulmonary bypass. The IntraClude Intra-Aortic Occlusion Catheter occludes and vents the ascending aorta when the balloon is inflated. The device's central lumen allows delivery of cardioplegia to arrest the heart. The pressure lumen allows monitoring of the aortic root pressure.
Prescription Use × (Per 21 CFR 801.109)
OR Over-The-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
te of CDRH, Office Of Device Evaluation (ODE)
on Sign. Division of Cardiovascular Devices
510(k) Number_KII 3182
PROPRIETARY DATA: This document and the information contained herein may not be reproduced, used, or disclosed without written permission from Edwards Lifesciences, LLC.