(74 days)
The ACL TightRope and ACL TightRope Double Bundle are to be used for fixation of bone to bone or soft tissue to bone, and are intended as fixation posts, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair. Specifically, Arthrex will be offering these for ACL/PCL repair and reconstruction.
The ACL TightRope consists of an adjustable nonabsorbable suture loop and titanium button. The ACL TightRope Double Bundle consists of nonabsorbable suture loops, titanium button, and a PEEK Femoral Wedge. The ACL TightRope and ACL TightRope Double Bundle is to be used for fixation of bone to bone or soft tissue to bone, and is intended as fixation posts, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair. Specifically, Arthrex will be offering this for Anterior Cruciate Ligament (ACL) and Posterior Cruciate Ligament (PCL) Repair and Reconstruction.
The provided document K112990 is a 510(k) summary for the Arthrex ACL TightRope and ACL TightRope Double Bundle devices, seeking to expand their indications to include Posterior Cruciate Ligament (PCL) repair and reconstruction. This document does not describe a study involving an algorithm or AI. Instead, it focuses on the substantial equivalence of the subject devices to existing predicate devices based on design and mechanical testing.
Therefore, many of the requested sections regarding AI/algorithm performance and study methodology are not applicable to this submission.
Here's an analysis based on the information available in the document:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Mechanical strength: Substantial equivalence to predicate devices (K100652, K101837, K110123) by demonstrating similar safety and effectiveness. | "Bench testing was conducted to determine the tensile load to failure strength of the ACL TightRope devices." The submission concludes that "the ACL TightRope and ACL TightRope Double Bundle with expanded indications are substantially equivalent to the currently marketed predicate devices." This implies that the tensile load to failure strength met comparable levels to the predicate devices, satisfying the acceptance criterion for mechanical strength and safety for the expanded PCL indication. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: Not explicitly stated for the "bench testing" mentioned. The document only refers to "tensile load to failure strength of the ACL TightRope devices" without providing specific numbers of samples tested.
- Data Provenance: Not applicable in terms of country of origin of patient data, as this was mechanical bench testing, not a clinical study. The testing was conducted by or for Arthrex, Inc., based in Naples, Florida, USA.
- Retrospective/Prospective: Not applicable. This was mechanical laboratory testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. This submission concerns mechanical testing of medical devices, not an AI or diagnostic algorithm requiring expert ground truth for interpretation of images or patient data. The "ground truth" here is the physical measurement of tensile strength.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. As described above, this was mechanical testing, not a clinical or diagnostic study requiring adjudication of expert opinions.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, a multi-reader, multi-case comparative effectiveness study was not done. This submission is for a physical medical device, not an AI or software algorithm.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. No algorithm or AI device is being submitted or studied here.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- For the tensile load to failure strength testing, the "ground truth" would be the objectively measured physical properties (e.g., maximum load at failure) of the devices under controlled laboratory conditions, according to established mechanical testing standards.
8. The sample size for the training set
- Not applicable. There is no AI or algorithm mentioned, hence no training set.
9. How the ground truth for the training set was established
- Not applicable. There is no AI or algorithm mentioned, hence no training set or ground truth establishment for it.
Summary of Device Acceptance Criteria and Study:
The acceptance criteria for Arthrex ACL TightRope and ACL TightRope Double Bundle for expanded PCL indications were based on demonstrating substantial equivalence to existing predicate devices (K100652, K101837, K110123). This means the device, with its expanded indication, must be as safe and effective as the already cleared predicate devices.
The study that proves the device meets these criteria was bench testing. This testing specifically "determined the tensile load to failure strength of the ACL TightRope devices." The report concludes that based on this testing, the subject devices are "substantially equivalent" to predicate devices, implying that their mechanical performance (tensile strength) was comparable and therefore considered safe and effective for the expanded PCL indication. The document does not provide specific numerical results or sample sizes for this bench testing, but the FDA's clearance indicates satisfaction with the presented data.
{0}------------------------------------------------
K112990 (pg. 1 of 2)
Arthrex S TRADITIONAL 510(k): Arthrex ACL TightRope Devices, expand indications to include POL
510(k) Summary of Safety and Effectiveness
| Date Summary Prepared | October 4, 2011 |
|---|---|
| Manufacturer/Distributor/Sponsor | Arthrex, Inc.1370 Creekside BoulevardNaples, FL 34108-1945 USA |
| 510(k) Contact | Christina FloresRegulatory Affairs SpecialistArthrex, Inc.1370 Creekside BoulevardNaples, FL 34108-1945 USATelephone: 239/643.5553, ext. 1819Fax: 239/598.5508Email: Christina.flores@arthrex.com |
| Trade Name | ACL TightRope; ACL TightRope Double Bundle |
| Common Name | Pin, fixation, smoothSuture, Nonabsorbable, synthetic, polyethylene |
| Product Code -ClassificationNameCFR | 21 CFR 888.3040: Smooth or threaded metallic bonefixation fastenerHTY - Smooth or threaded metallic bone fixationfastener21 CFR 878.5000: Nonabsorbable poly(ethyleneterephthalate) surgical sutureGAT - Nonabsorbable poly(ethylene terephthalate)surgical suture |
| Predicate Device | K100652Arthrex ACL TightRopeK101837Arthrex ACL TightRope Double BundleK110123Arthrex PCL TightRope |
| Purpose of Submission | This traditional 510(k) premarket notification issubmitted to obtain clearance for the addition of thePCL indication to the ACL TightRope and the ACLTightRope Double Bundle. |
| Device Description and IntendedUse | The ACL TightRope consists of an adjustablenonabsorbable suture loop and titanium button.TheACL TightRope Double Bundle consists ofnonabsorbable suture loops, titanium button, and aPEEK Femoral Wedge. The ACL TightRope andACL TightRope Double Bundle is to be used forfixation of bone to bone or soft tissue to bone, and isintended as fixation posts, a distribution bridge, or fordistributing suture tension over areas of ligament ortendon repair. Specifically, Arthrex will be offering |
{1}------------------------------------------------
K112990 (pg. 2 of 2)
this for Anterior Cruciate Ligament (ACL) and Posterior Cruciate Ligament (PCL) Repair and Reconstruction. ---------Substantial Equivalence The ACL TightRope and ACL TightRope Double Summary Bundle with expanded indications are substantially equivalent to the existing ACL TightRope and ACL TightRope Double Bundle as no changes have been made to the devices. The ACL TightRope and ACL TightRope Double Bundle with expanded indications are substantially equivalent to the PCL TightRope device in which the basic design features and intended uses are very similar. Any design differences between the subject devices (ACL TightRope and ACL TightRope Double Bundle with expanded indications) to the PCL TightRope are considered minor and do not raise questions concerning safety and effectiveness. Bench testing was conducted to determine the tensile load to failure strength of the ACL TightRope devices. Based on the information submitted, Arthrex, Inc. has determined that the ACL TightRope and ACL TightRope Double Bundle with expanded indications are substantially equivalent to the currently marketed predicate devices.
Arthrex. 6 TRADITIONAL 510(k): Arthrex ACL TightRope Devices, expand indications to include PCL
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -W066-G609 Silver Spring, MD 20993-0002
Arthrex, Incorporated % Ms. Christina Flores Regulatory Affairs Specialist 1370 Creekside Boulevard Naples, Florida 34108
DEC 1 9 2011
Re: K112990
Trade/Device Name: ACL TightRope and ACL Tightrope Double Bundle Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HTY, GAT Dated: October 4, 2011 Received: October 6, 2011
Dear Ms. Flores:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Christina Flores
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours
For Ploth
KEI L.U.M
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Arthreat B TRADITIONAL 510(k): Arthrex ACL TightRope Devices, expand indications to include PCL
2 Indications for Use Form
| 510(k) Number: | K112990 |
|---|---|
| Device Name: | ACL TightRope and ACL Tightrope Double Bundle |
The ACL TightRope and ACL TightRope Double Bundle are to be used for fixation of bone to bone or soft tissue to bone, and are intended as fixation posts, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair. Specifically, Arthrex will be offering these for ACL/PCL repair and reconstruction.
Prescription Use _X_AND/OR Over-The-Counter Use
(Per 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
l of l
Muhel Idsm
fur
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K112990
5
§ 888.3040 Smooth or threaded metallic bone fixation fastener.
(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.