K Number
K112930

Validate with FDA (Live)

Device Name
MOBILE MIM
Manufacturer
Date Cleared
2011-12-02

(60 days)

Product Code
Regulation Number
892.2050
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Mobile MIM software program is used for the viewing, registration, fusion, and/or display for diagnosis of medical images from the following modalities: SPECT, PET, CT, MRI, X-ray and Ultrasound.

Mobile MIM can be used to review images, contours, DVH, and isodose curves from radiation treatment plans. Mobile MIM can be used to approve these plans.

Mobile MIM provides wireless and portable access to medical images. This device is not intended to replace full workstations and should be used only when there is no access to a workstation.

This device is not to be used for mammography.

Device Description

Mobile MIM (RT) extends the Mobile MIM software application previous cleared under K103785. In addition to SPECT. PET. CT. MRI modalities. Mobile MIM can be used for the viewing and/or display for diagnosis of X-ray and Ultrasound medical images.

It also provides functionality for the review of medical images, contours. DVH, and isodose curves from radiation treatment plans. In addition. Mobile MIM (RT) will allow permitted users the ability to approve reviewed radiation treatment plans.

AI/ML Overview

The provided text describes the 510(k) summary for the Mobile MIM (RT) software, which is an extension of the existing Mobile MIM software. The document details the device's intended use, substantial equivalence to predicate devices, and performance data.

Here's an analysis of the provided information regarding acceptance criteria and the studies conducted:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state quantitative (numerical) acceptance criteria. Instead, it describes general qualitative goals for the performance studies. The reported device performance is also qualitative, affirming that the software met these general expectations.

Acceptance Criteria (Inferred from Study Descriptions)Reported Device Performance
Image quality for diagnostic viewing of X-ray and Ultrasound modalities under different environmental conditions must be acceptable."Results of these studies affirm the diagnostic image viewing capabilities of Mobile MIM when used as indicated."
Display quality of isodose curves, DVH graphs, and contours for radiation therapy plan review must be acceptable."The results indicated the display quality for of isodose curves, DVH graphs, and contours was of acceptable quality for review and approval of radiation therapy plans, and were equivalent to those viewed on a full workstation."
Software must pass performance requirements and meet specifications through verification, validation, and functional testing."In all cases, the software passed its performance requirements and met specifications." Note: This refers to internal software quality assurance processes rather than clinical performance studies described above.

2. Sample Size Used for the Test Set and Data Provenance

  • X-ray and Ultrasound Diagnostic Viewing Study:

    • Sample Size: Not specified. The document states "multiple studies" were performed by "qualified radiologists" evaluating "the two additional modalities of X-ray and Ultrasound."
    • Data Provenance: Not specified. It does not mention the country of origin or if the data was retrospective or prospective.
  • Radiation Therapy Plan Review Study:

    • Sample Size: Not specified. It mentions evaluating "multiple areas of treatments" and using "plan data from 3 major vendors."
    • Data Provenance: Not specified. It does not mention the country of origin or if the data was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

  • X-ray and Ultrasound Diagnostic Viewing Study:

    • Number of Experts: Not specified, but referred to as "qualified radiologists."
    • Qualifications of Experts: "qualified radiologists." No further details on experience (e.g., "10 years of experience") are provided.
  • Radiation Therapy Plan Review Study:

    • Number of Experts: Not specified, but referred to as "trained medical professionals."
    • Qualifications of Experts: "dosimetrists and radiation oncologists." No further details on experience are provided.

4. Adjudication Method for the Test Set

The document does not describe any specific adjudication method (e.g., 2+1, 3+1, none) for either study. The studies appear to be based on expert evaluations, but the mechanism for resolving disagreements or establishing a consensus ground truth is not detailed.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

No MRMC comparative effectiveness study is explicitly mentioned. The studies described are focused on assessing the standalone performance or equivalence of the Mobile MIM (RT) display capabilities to a full workstation, rather than measuring human reader improvement with AI assistance. Therefore, no effect size for human reader improvement with AI vs. without AI assistance is provided.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The performance studies described in the document are primarily focused on the human-in-the-loop performance, where radiologists, dosimetrists, and radiation oncologists are using the Mobile MIM (RT) software to evaluate images and plans. The studies assess the software's ability to display information adequately for human interpretation and decision-making. No "algorithm only" standalone performance is discussed, as the device is a viewing and review software, not an AI diagnostic algorithm in the modern sense.

7. The Type of Ground Truth Used

  • X-ray and Ultrasound Diagnostic Viewing Study: The ground truth for evaluating "diagnostic image viewing capabilities" would likely be the consensus opinion of the qualified radiologists regarding the diagnostic quality of the images displayed on the Mobile MIM (RT), presumably compared to "gold standard" interpretations or full workstation views. This is an inferred-expert consensus.
  • Radiation Therapy Plan Review Study: The ground truth for evaluating "display quality for of isodose curves, DVH graphs, and contours" and "review and approval of radiation therapy plans" would be the consensus opinion of the trained medical professionals (dosimetrists and radiation oncologists), compared to views on a "full workstation." This is an inferred-expert consensus and comparison to a recognized standard (full workstation).

8. The Sample Size for the Training Set

The document does not describe the development or training of an AI algorithm or machine learning model. Mobile MIM (RT) is presented as an image viewing and review software application. Therefore, there is no mention of a "training set" in the context of an AI model. The "performance data" section focuses on testing the usability and display quality of the software for its stated functions.

9. How the Ground Truth for the Training Set Was Established

As there is no mention of a training set for an AI algorithm, this information is not applicable. The device is software for viewing and interacting with medical images and plans, not an AI system that learns from a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo for MIM Software. The logo consists of two overlapping squares with a circle in the middle on the left side. To the right of the squares are the letters "mim" in a stylized font, with the word "SOFTWARE" below it in smaller letters.

Image /page/0/Picture/1 description: The image shows the text "K112930 Page 1 of 3". The text appears to be handwritten in a dark ink. The text "Page 1 of 3" is located below the other text and is smaller in size.

DEC - 2 2011

510(k) Summary of Safety and Effectiveness

(The following information is in conformance with 21 CFR 807.92)

Submitter:

MIM Software Inc. 25200 Chagrin Blvd. Suite 200 Cleveland, OH 44122

Phone:216-455-0600
Fax:216-455-0601
Contact Person:Lynn Hanigan
Date Summary Prepared:Sept 30, 2011

Device Name

Trade Name:Mobile MIM (RT)
Common Name:Medical Imaging Software
Classification Name:System, Imaging Processing, Radiological

Predicate Device

K103785Mobile MIMMIM Software Inc.
K042956VisionVarian Medical System

Intended Use / Indications for Use

The Mobile MIM software program is used for the viewing, registration, fusion, and/or display for diagnosis of medical images from the following modalities: SPECT, PET, CT, MRI, X-ray and Ultrasound.

Mobile MIM can be used to review images, contours, DVH, and isodose curves from radiation treatment plans. Mobile MIM can be used to approve these plans.

Mobile MIM provides wireless and portable access to medical images. This device is not intended to replace full workstations and should be used only when there is no access to a workstation.

This device is not to be used for mammography.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for MIM Software. The logo consists of a black square with a white circle in the center, followed by the text "mim" in a bold, sans-serif font. Below the text "mim" is the word "SOFTWARE" in a smaller, sans-serif font.

Device Description

Mobile MIM (RT) extends the Mobile MIM software application previous cleared under K103785. In addition to SPECT. PET. CT. MRI modalities. Mobile MIM can be used for the viewing and/or display for diagnosis of X-ray and Ultrasound medical images.

It also provides functionality for the review of medical images, contours. DVH, and isodose curves from radiation treatment plans. In addition. Mobile MIM (RT) will allow permitted users the ability to approve reviewed radiation treatment plans.

Substantial Equivalence

Mobile MIM is substantially equivalent to Mobile MIM software (K103785) and portions of the Vision product (K042956). It extends Mobile MIM functionality by adding 2 additional image modalities to its indication and having the capability to serve as a mobile reviewing device for radiation treatment plans.

Performance Data

MIM Software Inc. has performed multiple studies with qualified radiologists. dosimetrists and radiation oncologists. Radiologists tested Mobile MIM by evaluating the image quality of the two additional modalities of X-ray and Ultrasound under different environmental conditions. Results of these studies affirm the diagnostic image viewing capabilities of Mobile MIM when used as indicated.

MIM Software also orchestrated radiation therapy plan review tests evaluating multiple areas of treatments by trained medical professionals using plan data from 3 major vendors, and using both smaller format (iPhone and/or iPod touch) and larger format (iPad) devices. The results indicated the display quality for of isodose curves, DVH graphs, and contours was of acceptable quality for review and approval of radiation therapy plans, and were equivalent to those viewed on a full workstation.

Furthermore, MIM Software Inc. has conducted verification, validation, and functional testing on the Mobile MIM software. In all cases, the software passed its performance requirements and met specifications.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for MIM Software. The logo consists of a graphic on the left and the text "mim SOFTWARE" on the right. The graphic is made up of two overlapping squares with a circle in the center. The text is in a sans-serif font, with "mim" in a larger font size than "SOFTWARE".

Conclusion

Therefore, from all evidence gathered, it is our belief that Mobile MIM (RT) provides a safe and effective diagnostic viewer of the following medical imaging modalities: SPECT, PET, CT, MRI, X-ray and Ultrasound. It also is safe and effective in that it is substantially equivalent to the radiation treatment plan review functionality of Vision (K042956), allowing for portable device characteristics and accessibility when there is no access to a full workstation.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white and has a simple, clean design. The caduceus is a common symbol associated with healthcare and medicine.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

DEC - 2 2011

Ms. Lynn Hanigan Quality Manager MIM Software, Inc. 25200 Chagrin Blvd. Suite 200 CLEVELAND OH 44122

Re: K112930 Trade/Device Name: Mobile MIM Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ & MUJ Dated: September 30, 2011 Received: October 3, 2011

Dear Ms. Hanigan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

\

{4}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Mary S. Patel

Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K112930
P. 1 of 1

Indications for Use

510(k) Number (if known): TBD KII2930

Device Name: Mobile MIM

Indications for Use:

The Mobile MIM software program is used for the viewing, registration, fusion, and/or display for diagnosis of medical images from the following modalities: SPECT, PET, CT, MRI, X-ray and Ultrasound.

Mobile MIM can be used to review images, contours, DVH, and isodose curves from radiation treatment plans. Mobile MIM can be used to approve these plans.

· Mobile MIM provides wireless and portable access to medical images. This device is not intended to replace full workstations and should be used only when there is no access to a workstation.

This device is not to be used for mammography.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K112930

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).