K Number
K112916
Date Cleared
2012-08-21

(323 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TysonBio MD100 Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in capillary whole blood samples may be drawn from the fingertip, palm and forearm. The TysonBio MD100 Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The TysonBio MD100 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The TysonBio MD100 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly). The TysonBio MD100 Test Strips are for use with the TysonBio MD100 Blood Glucose Meter to quantitatively measure qlucose in venous whole blood sample and fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The TysonBio MD100 Control Solutions is for use on the TysonBio MD100 Blood Glucose Monitoring Systems to check that the meter and test strips are working together properly and providing accurate results.

The TysonBio MD100 Pro Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in capillary and venous whole blood samples may be drawn from the fingertip, palm and forearm. The TysonBio MD100 Pro Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. This system should only be used with auto-disabling, single-use lancing devices. The TysonBio MD100 Pro Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly). The TysonBio MD100 Pro Blood Glucose Test Strips are for use with the TysonBio MD100 Pro Blood Glucose Meter to quantitatively measure glucose in venous whole blood sample and fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The TysonBio MD100 Control Solutions is for use on the TysonBio MD100 Pro Blood Glucose Monitoring Systems to check that the meter and test strips are working together properly and providing accurate results.

Device Description

The TysonBio MD100 Blood Glucose Monitoring System consists of three main products: the meter, test strip, and control solutions. These products have been designed, tested, and proven to work together as a system accurate blood glucose test results. Use only TysonBio MD100 Test Strips and TysonBio MD100 Control Solution with the TysonBio MD100 Blood Glucose Monitoring System.

The TysonBio MD100 Pro Blood Glucose Monitoring System consists of three main products: the meter, test strip, and control solutions. These products have been designed, tested, and proven to work together as a system accurate blood glucose test results. Use only TysonBio MD100 Pro Test Strips and TysonBio MD100 Control Solution with the TysonBio MD100 Pro Blood Glucose Monitoring System.

The TysonBio MD100 / MD100 Pro Blood Glucose Monitoring System consists of:
a. TysonBio MD100/MD100 Pro Blood Glucose Meter
b. TysonBio MD100/MD100 Pro Test Strips
TysonBio MD100 Control Solution

AI/ML Overview

The provided text does not contain specific acceptance criteria or an analytical study that proves the device meets those criteria. It is a 510(k) summary document that describes the device, its intended use, and states that performance studies were conducted and the device meets its intended use, but it does not detail these studies or their results in a way that allows for the extraction of the requested information.

Therefore, many of the requested details cannot be provided from this document.

However, based on the information that is present, I can provide the following:

1. Table of Acceptance Criteria and Reported Device Performance:

This information is not provided in the document. The document states "The performance of TysonBio MD100 / TysonBio MD100 Pro Blood Glucose Monitoring System was studied in the laboratory and in clinical settings by healthcare professionals and lay users. The studies demonstrated that the performance of this system meets its intended use." However, it does not specify what those acceptance criteria were or what the actual reported performance metrics were against those criteria (e.g., accuracy percentages, bias, precision data).

2. Sample Size Used for the Test Set and Data Provenance:

This information is not provided in the document.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not provided in the document. The document mentions "healthcare professionals" were involved in clinical settings, but does not specify their number or qualifications for establishing ground truth.

4. Adjudication Method for the Test Set:

This information is not provided in the document.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size:

This information is not provided in the document. The device is a blood glucose monitoring system, which typically involves direct use by individuals or healthcare professionals rather than interpretation of medical images by multiple readers. Therefore, an MRMC study is highly unlikely for this type of device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

The device is a Blood Glucose Monitoring System, which by its nature is a standalone device that provides direct measurements to the user or healthcare professional. There is no "human-in-the-loop performance" that would be separate from the device's own measurement. So, in essence, the device's performance is its standalone performance. However, the document does not provide the details of this standalone performance.

7. The Type of Ground Truth Used:

The document implies that the ground truth for glucose measurements would be established through a reference method or laboratory analyzer, as is standard for glucose monitoring systems. While not explicitly stated, the phrase "quantitatively measure glucose" and the nature of the device suggest that comparison would be made against a more precise, established glucose measurement method.

8. The Sample Size for the Training Set:

This information is not provided in the document.

9. How the Ground Truth for the Training Set was Established:

This information is not provided in the document.

{0}------------------------------------------------

K 112916

Image /page/0/Picture/1 description: The image shows the logo for TysonBio. The logo is black and white and features the word "Tyson" in bold, sans-serif font. The word "Bio" is attached to the end of Tyson and is surrounded by a circular design with two dots. The logo is simple and modern.

YSON BIORESEARCH, INC #22 Ke F Road III Science-Based Industrial Park aiwan, R.O.C. Postal Code: 350 Website: www.tys 01 +886-37-585988 86-37-585996

510(k) SUMMARY

AUG 2 1 2012

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

    1. Date prepared August 20, 2012
    1. Sponsor information

Tyson Bioresearch, Inc.

5 F., # 22, Ke E. Road III., science based industrial park Chun-Nan, Miao-Li county, Taiwan 350

Correspondent:

WEN-HAI TSAI

Phone: 886-37-585988 EXT 720 Facsimile: 886-37-585996

3. Device name and classification

Device Name:

TysonBio MD100 Blood Glucose Monitoring System TysonBio MD100 Pro Blood Glucose Monitoring System TysonBio MD100 Control Solution TysonBio MD100 Test Strip TysonBio MD100 Pro Test Strip

Common Names:

Blood Glucose Monitoring System

Classification:

Classification Regulation: 21 CFR 862.1345 and 21CFR 862.1660

Classification: Class II (Glucose Test System)

Class I (Quality control material )

Product Code: LFR (Glucose Dehydrogenase, Glucose)

NBW (System, Test, Blood Glucose, Over The Counter)

  • JJX, Single (specified) analyte controls (assayed and unassayed)
    Panel - Clinical Chemistry and Toxicology

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for TysonBio. The logo is black and white and features the word "TysonBio" in a bold, sans-serif font. A curved line extends over the top of the word "TysonBio" and circles around the letters. The logo is simple and modern.

TYSON BIORESEARCH. INC. 5F. #22, Ke E. Road III Science-Based Industrial Park, Chu-Nan, Taiwan, R.O.C. Postal Code: 350 Website: www.tysonbjo.com Ol +886-37-585988

4. Device description

The TysonBio MD100 Blood Glucose Monitoring System consists of three main products: the meter, test strip, and control solutions. These products have been designed, tested, and proven to work together as a system accurate blood glucose test results. Use only TysonBio MD100 Test Strips and TysonBio MD100 Control Solution with the TysonBio MD100 Blood Glucose Monitoring System.

The TysonBio MD100 Pro Blood Glucose Monitoring System consists of three main products: the meter, test strip, and control solutions. These products have been designed, tested, and proven to work together as a system accurate blood glucose test results. Use only TysonBio MD100 Pro Test Strips and TysonBio MD100 Control Solution with the TysonBio MD100 Pro Blood Glucose Monitoring System.

The TysonBio MD100 / MD100 Pro Blood Glucose Monitoring System consists of:

  • a. TysonBio MD100/MD100 Pro Blood Glucose Meter
  • TysonBio MD100/MD100 Pro Test Strips b.
  • TysonBio MD100 Control Solution

5. Intended use

TysonBio MD100 Blood Glucose Monitoring System :

The TysonBio MD100 Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in capillary whole blood samples may be drawn from the fingertip, palm and forearm. The TysonBio MD100 Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

The TysonBio MD100 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The TysonBio MD100 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The TysonBio MD100 Test Strips are for use with the TysonBio MD100 Blood Glucose Meter to quantitatively measure qlucose in venous whole blood sample and fresh capillary whole blood samples drawn from the fingertips, forearm or palm.

The TysonBio MD100 Control Solutions is for use on the TysonBio MD100 Blood Glucose Monitoring Systems to check that the meter and test strips are working together properly and providing accurate results.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for TysonBio. The logo is black and features the word "TysonBio" in a bold, sans-serif font. The "o" in "Tyson" and the "o" in "Bio" are stylized as circles, and a curved line connects the two circles, creating a visual link between the two parts of the name. The overall design is simple and modern.

TYSON BIORESEARCH. INC. 5F. #22, Ke E. Road III Science-Based Industrial Park, Chu-Nan; Taiwan, R.O.C. Postal Code: 350 Website: www.tysonbio.com O] +886-37-585988 F1 +886-37-585996

TysonBio MD100 Pro Blood Glucose Monitoring System :

The TysonBio MD100 Pro Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in capillary and venous whole blood samples may be drawn from the fingertip, palm and forearm. The TysonBio MD100 Pro Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. This system should only be used with auto-disabling, single-use lancing devices. The TysonBio MD100 Pro Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alfernative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The TysonBio MD100 Pro Blood Glucose Test Strips are for use with the TysonBio MD100 Pro Blood Glucose Meter to quantitatively measure glucose in venous whole blood sample and fresh capillary whole blood samples drawn from the fingertips, forearm or palm.

The TysonBio MD100 Control Solutions is for use on the TysonBio MD100 Pro Blood Glucose Monitoring Systems to check that the meter and test strips are working together properly and providing accurate results.

6. Test principle

The test principle is based on electrochemical biosensor technology using glucose dehydrogenase. There has been no change to the fundamental scientific technology.

7. Predicate device

ItemPredicate Device (K062058)Proposed Device
Ascensia ContourBlood Glucose Monitoring SystemTysonBio MD100 /MD100 ProBlood Glucose Monitoring System
Meter size77 mm (H) x 57 mm (W) x 19 mm (T)94 mm (H) x 63 mm (W) x 25 mm (T)
Meter weight47.5 grams60 grams without battery
Sample volume0.6uL0.5uL
ASTPalm, forearm and heel (neonates)Palm and forearm
Measuring Range10-600mg/dL20-600mg/dL
Hematocrit range0~70 %10-70%
OperatingTemperatureRange:5 to 45 degrees C (41-113°F)10 to 40 degrees C (50-104°F)
Humidity10-93%10-90%
Memory FeatureStores most recent 480 test resultsStores most recent 500 test results
Battery Type:Two 3-volt lithium batteries(DL2032 or CR2032)Two AAA batteries
Battery Life:Approximately 1000 tests(1 yr. average use)Approximately 1000 tests
MarkerMeal and log bookMeal
Reminder alarmPost-Meal Test Alarm4 user setting alarms
ButtonThree operating button(M, up and down)Three operating button(M, up and down)One ejection button

Ascensia CONTOUR Blood Glucose Monitoring System (K062058)

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for TysonBio. The logo is in black and white, with the word "Tyson" in bold letters. The word "Bio" is also in bold letters and is connected to the word "Tyson" by a curved line that forms a circle around the letters "sonBi".

TYSON BIORESEARCH, INC.

5F. #22, Ke E. Road III Science-Based Industrial Park, Chu-Nan, Taiwan, R.O.C. Postal Code: 350 Website: www.tysonbio.com 0] +886-37-585988 F1 +886-37-585996

Similarities:

ItemPredicate Device (K062058)Proposed Device
Ascensia ContourBlood Glucose Monitoring SystemTysonBio MD100 / MD100 ProBlood Glucose Monitoring System
Test principleElectrochemical biosensor withglucose dehydrogenase (FAD)Electrochemical biosensor with glucosedehydrogenase (FAD)
Test SampleWhole blood(Capillary and venous blood)Whole blood(Capillary and venous blood)
MeasuringTime5 seconds5 seconds
CodingAuto coding by inserting the test stripAuto coding by inserting the test strip
Hypoglycemicandhyperglycemicalarm2 user setting alarms2 user setting alarms
Average result7,14 and 30days7,14 and 30days

8. Performance characteristic summary

The performance of TysonBio MD100 / TysonBio MD100 Pro Blood Glucose Monitoring System was studied in the laboratory and in clinical settings by healthcare professionals and lay users. The studies demonstrated that the performance of this system meets its intended use.

9. Conclusion

The TysonBio MD100 / MD100 Pro Blood Glucose Monitoring System is substantially equivalent to the predicate device.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration

Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with stylized feathers. The symbol is black and appears to be in motion.

Tyson Bioresearch, Inc. c/o Wen-Hai Tsai 5 F., #22, Ke E. Road III, Science-Based Industrial Park Chu-Nan, Miao-Li County China (Taiwan) 350

10903 New Hampshire Avenue Silver Spring, MD 20993

AUG 2 1 2012

Re: K112916

Trade Name: TysonBio MD100 Blood Glucose Monitoring System TysonBio MD100 Pro Blood Glucose Monitoring System Regulation Number: 21 CFR §862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Codes: NBW, LFR, JJX Dated: August 9, 2012 Received: August 13, 2012

Dear Wen-Hai Tsai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical

Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm

Sincerely yours,

N

Couriney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indication for Use

510(k) Number (if known):

K112916

Device Name: TysonBio MD100 Pro Blood Glucose Monitoring System

Indication for Use:

The TysonBio MD100 Pro Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in capillary and venous whole blood samples. Capillary samples may be drawn from the fingertip, palm and forearm. The TysonBio MD100 Pro Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. This system should only be used with auto-disabling, single-use lancing devices. The TysonBio MD100 Pro Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The TysonBio MD100 Pro Blood Glucose Test Strips are for use with the TysonBio MD100 Pro Blood Glucose Meter to quantitatively measure glucose in venous whole blood sample and fresh capillary whole blood samples drawn from the fingertips, forearm or palm.

The TysonBio MD100 Control Solutions is for use on the TysonBio MD100 Pro Blood Glucose Monitoring Systems to check that the meter and test strips are working together properly and providing accurate results.

X Prescription Use (21 CFR Part 801 Subpart D) And/Or Over the Counter Use X (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Raul Luus

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K112916

Page 1 of 2

{7}------------------------------------------------

Indication for Use

510(k) Number (if known):

K112916

Device Name: TysonBio MD100 Blood Glucose Monitoring System

Indication for Use:

The TysonBio MD100 Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in capillary and venous whole blood samples. Capillary samples may be drawn from the fingertip, palm and forearm. The TysonBio MD100 Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

The TysonBio MD100 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The TysonBio MD100 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The TysonBio MD100 Test Strips are for use with the TysonBio MD100 Blood Glucose Meter to quantitatively measure glucose in venous whole blood sample and fresh capillary whole blood samples drawn from the fingertips, forearm or palm.

The TysonBio MD100 Control Solutions is for use on the TysonBio MD100 Blood Glucose Monitoring Systems to check that the meter and test strips are working together properly and providing accurate results.

Prescription Use _ (21 CFR Part 801 Subpart D) And/Or Over the Counter Use x (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign Off

Division Sign Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K112916

Page Lof 2

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.