(182 days)
Not Found
No
The description focuses on standard electrochemical glucose measurement technology and basic device functionality (data storage, user interface). There is no mention of AI/ML algorithms for data analysis, prediction, or interpretation.
No
Explanation: The device is intended for quantitative measurement of glucose to aid in monitoring the effectiveness of diabetes control. It is explicitly stated that it is "not intended for the diagnosis of or screening for diabetes." Therefore, it is a diagnostic device used for monitoring rather than a therapeutic device that treats or prevents a condition.
Yes
The device measures glucose levels in fresh capillary whole blood, which provides information to aid in monitoring the effectiveness of diabetes control. While it explicitly states it is "not intended for the diagnosis of or screening for diabetes," the measurement of glucose levels for monitoring purposes, particularly for a chronic condition like diabetes, falls under the umbrella of diagnostic or monitoring tools. It provides data that informs health management decisions, which is a key function of diagnostic devices, even if it's not for initial diagnosis.
No
The device description clearly outlines a "hand-held testing device" which is a hardware component, along with test strips and control solutions. This is not a software-only device.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use explicitly states that the device is for "quantitative measurement of glucose in fresh capillary whole blood" and "for self testing outside the body". This clearly indicates testing of a biological sample (blood) outside of the body to obtain diagnostic information (glucose levels).
- Device Description: The description of the test strips mentions a "reaction layer" and "glucose-enzyme" which are components involved in a chemical reaction with the blood sample to measure glucose. This is characteristic of in vitro diagnostic tests.
- Control Solutions: The inclusion and description of control solutions for "quality control check to verify the accuracy of blood glucose test results" further supports its classification as an IVD. Control solutions are used to ensure the proper functioning and accuracy of in vitro diagnostic devices.
- Explicit Statement: The description of the Nova Max One Glucose Test Strips explicitly states they are "for testing outside the body (in vitro diagnostic use only)".
All these points align with the definition of an In Vitro Diagnostic device, which is a medical device intended for use in vitro for the examination of specimens derived from the human body solely or principally for the purpose of providing information concerning a physiological or pathological state, or concerning a congenital abnormality, or to monitor therapeutic measures.
N/A
Intended Use / Indications for Use
The Nova Max One Blood Glucose Monitor is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. It is intended for single-patient home use and should not be used for testing multiple patients. It is intended for self testing outside the body by people with diabetes mellitus as an aid to monitor the effectiveness of diabetes control. The Nova Max One Blood Glucose Monitor is specifically indicated for the quantitative measurement of glucose in fresh whole blood capillary samples obtained from the fingertip or alternative site testing (AST) on the forearm. AST on the forearm can be used only during steady-state blood glucose conditions. It is not intended for the diagnosis of or screening for diabetes, and it is not intended for use on newborns.
Nova Max One Glucose Test Strips are intended for use only with the Nova Max One Blood Glucose Monitor for quantitative tests. The Glucose Monitor is intended to quantitatively measure glucose (sugar) in fresh capillary whole blood obtained from the finger tip or alternative site testing (AST) on the forearm. AST can be used only during steady-state blood glucose conditions. The Glucose Monitor is calibrated to provide plasma equivalent results to laboratory methods. Nova Max One Glucose Test Strips are for testing outside the body (in vitro diagnostic use only). The Monitor should only be used by a single user in the home and should not be shared by users. It is not intended for the diagnosis of or screening for diabetes, and it is not intended for use on newborns.
Nova Max Control Solutions are intended for use with the Nova Max Blood Glucose Monitoring Systems as a quality control check to verify the accuracy of blood glucose test results. There are three levels of controls, (Levels 1, 2, 3).
Product codes (comma separated list FDA assigned to the subject device)
NBW, LFR, JJX
Device Description
Nova Max One Blood Glucose Monitor
The monitor is a hand-held testing device that works in conjunction with Nova Max One glucose-test strips to measure glucose in a whole blood sample. Monitor operation is self-prompting using three user interface buttons. In addition to measuring glucose, the monitor also stores patient test and quality control test data.
The self-prompting menu system is navigated by means of a three-button keypad. It offers audible feedback for user inputs, and audible and/or visual feedback for prompts and user alerts.
A "battery low" warning will alert the user to change the batteries. Battery charge state information is available on the "monitor status screen". The user can select the auto shutoff option to conserve power when the monitor is not in use. Test data and monitor setup information will be stored in a non-volatile format to prevent data loss.
Nova Max One Blood Glucose Test Strips
The test strips contain a reaction layer that contains a glucose-enzyme (greater than 1.0 IU) and ferricyanide as a mediator and will utilize glucose dehydrogenase flavin-adenine dinucleotide (GDH-FAD) chemistry. The test strip is touched to a drop of blood to initiate the test process. The strip is designed such that when a drop of blood is touched to the end of the strip, the blood is drawn into the reaction space via capillary action. A simple one-step process provides a blood glucose result. Ten test strips will be provided with the meter kit and will also be available separately in vials of 25 strips.
Historical data (on file at Nova Biomedical) shows minimal differences between lots of test strips produced over 2 years. As a result, test strips will now be manufactured to meet a single calibration code requirement, allowing for the elimination of a calibration step for the user when using the Nova Max One Blood Glucose Monitor.
Control Solutions
The control solutions are aqueous assayed solutions, containing buffered D-Glucose, viscosity-adjusting agent, preservatives and other non-reactive ingredients (dye). They contain no products of human origin. There are three levels of controls (1, 2 and 3). One level of control (Level 2) will be supplied with the monitor kit and all three levels will be available for sale separately from the monitor. These controls are manufactured by Nova Biomedical and identical to those cleared for market with the predicate Nova Max Blood Glucose Monitor System (K070255).
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
fingertip or alternative site testing (AST) on the forearm
Indicated Patient Age Range
The device is not intended for use on newborns.
Intended User / Care Setting
single-patient home use, self testing outside the body by people with diabetes mellitus
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The performance of the Nova Max One Blood Glucose Monitor was studied in the laboratory and in clinical settings by healthcare professionals and lay users. The studies demonstrated that lay users can obtain blood glucose results that are substantially equivalent to the current methods for blood glucose measurements.
Results of laboratory and clinical testing demonstrate that the performance of the Nova Max One Blood Glucose Monitor has the same intended uses, with similar technological characteristics and can produce results that are substantially equivalent to results obtained on the predicate devices. The system performs as intended and raises no new safety or effectiveness issues.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 862.1345 Glucose test system.
(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
0
MAR - 9 2012
510(k) Summary
510(K) Owner: | Nova Biomedical Corporation |
---|---|
Registration | |
Number: | 1219029 |
Address: | 200 Prospect St. |
Waltham, MA 02454 USA | |
Phone: | 781-894-0800 |
Fax Number: | 784-891-4806 |
Contact Person: | Paul W. MacDonald |
Date Prepared: | 12/19/2011 |
Proprietary Name: | Nova Max One Blood Glucose Monitoring System |
Common Or Usual | |
Name: | Blood Glucose Monitor |
Classification | |
Name: | System, Test, Blood Glucose, Over the Counter |
Product Codes: | NBW, LFR, JJX |
Predicate Device: | K070255 - Nova Max Blood Glucose Monitor System |
K051839 - FreeStyle Freedom Blood Glucose Monitoring System | |
K092638 - FreeStyle Blood Glucose Test Strips | |
Device | |
Description: | Nova Max One Blood Glucose Monitor |
The monitor is a hand-held testing device that works in conjunction with Nova Max | |
One glucose-test strips to measure glucose in a whole blood sample. Monitor | |
operation is self-prompting using three user interface buttons. In addition to | |
measuring glucose, the monitor also stores patient test and quality control test | |
data. | |
The self-prompting menu system is navigated by means of a three-button keypad. | |
It offers audible feedback for user inputs, and audible and/or visual feedback for | |
prompts and user alerts. | |
A "battery low" warning will alert the user to change the batteries. Battery charge | |
state information is available on the "monitor status screen". The user can select | |
the auto shutoff option to conserve power when the monitor is not in use. Test data | |
and monitor setup information will be stored in a non-volatile format to prevent data | |
loss. | |
Nova Max One Blood Glucose Test Strips | |
The test strips contain a reaction layer that contains a glucose-enzyme (greater | |
than 1.0 IU) and ferricyanide as a mediator and will utilize glucose dehydrogenase | |
flavin-adenine dinucleotide (GDH-FAD) chemistry. The test strip is touched to a | |
drop of blood to initiate the test process. The strip is designed such that when a | |
drop of blood is touched to the end of the strip, the blood is drawn into the reaction | |
space via capillary action. A simple one-step process provides a blood glucose | |
result. Ten test strips will be provided with the meter kit and will also be available | |
separately in vials of 25 strips. | |
Historical data (on file at Nova Biomedical) shows minimal differences between lots | |
of test strips produced over 2 years. As a result, test strips will now be | |
manufactured to meet a single calibration code requirement, allowing for the | |
elimination of a calibration step for the user when using the Nova Max One Blood | |
Glucose Monitor. | |
Control Solutions | |
The control solutions are aqueous assayed solutions, containing buffered D- | |
Glucose, viscosity-adjusting agent, preservatives and other non-reactive | |
ingredients (dye). They contain no products of human origin. There are three | |
levels of controls (1, 2 and 3). One level of control (Level 2) will be supplied with | |
the monitor kit and all three levels will be available for sale separately from the | |
monitor. These controls are manufactured by Nova Biomedical and identical to | |
those cleared for market with the predicate Nova Max Blood Glucose Monitor | |
System (K070255). | |
Intended Use: | The Nova Max One Blood Glucose Monitor is intended to be used for the |
quantitative measurement of glucose in fresh capillary whole blood. It is intended | |
for single-patient home use and should not be used for testing multiple patients. It | |
is intended for self testing outside the body by people with diabetes mellitus as an | |
aid to monitor the effectiveness of diabetes control. The Nova Max One Blood | |
Glucose Monitor is specifically indicated for the quantitative measurement of | |
glucose in fresh whole blood capillary samples obtained from the fingertip or | |
alternative site testing (AST) on the forearm. AST on the forearm can be used only | |
during steady-state blood glucose conditions. It is not intended for the diagnosis of | |
or screening for diabetes, and it is not intended for use on newborns. | |
Nova Max One Glucose Test Strips are intended for use only with the Nova Max | |
One Blood Glucose Monitor for quantitative tests. The Glucose Monitor is intended | |
to quantitatively measure glucose (sugar) in fresh capillary whole blood obtained | |
from the finger tip or alternative site testing (AST) on the forearm. AST can be used | |
only during steady-state blood glucose conditions. The Glucose Monitor is | |
calibrated to provide plasma equivalent results to laboratory methods. Nova Max | |
One Glucose Test Strips are for testing outside the body (in vitro diagnostic use | |
only). The Monitor should only be used by a single user in the home and should not | |
be shared by users. It is not intended for the diagnosis of or screening for diabetes, | |
and it is not intended for use on newborns. | |
Nova Max Control Solutions are intended for use with the Nova Max Blood Glucose | |
Monitoring Systems as a quality control check to verify the accuracy of blood | |
glucose test results. There are three levels of controls, (Levels 1, 2, 3). |
1
.
، "،
2
| Summary of
Technological
Characteristics: | The Nova Max One Blood Glucose Monitor has the same fundamental scientific
technology and the same intended use as the current on-market Nova Max Blood
Glucose Monitor (K070255). The Nova Max One Blood Glucose Monitor is
substantially equivalent to the Nova Max Glucose Monitor System.
The Nova Max One Blood Glucose Monitor measures glucose electrochemically as
described in K070255 (Nova Max Glucose Monitor System). In the same manner,
the magnitude of the current is proportional to the amount of glucose present in the
sample, providing a quantitative measure of glucose in whole blood and control
solutions. |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison to
Predicate Devices: | The Nova Max One Blood Glucose Monitor uses the same fundamental scientific
technology and has the same intended use as the predicate Nova Max Blood
Glucose Monitor (K070255). |
| Performance
Studies: | The performance of the Nova Max One Blood Glucose Monitor was studied in the
laboratory and in clinical settings by healthcare professionals and lay users. The
studies demonstrated that lay users can obtain blood glucose results that are
substantially equivalent to the current methods for blood glucose measurements. |
| Conclusion: | Results of laboratory and clinical testing demonstrate that the performance of the
Nova Max One Blood Glucose Monitor has the same intended uses, with similar
technological characteristics and can produce results that are substantially
equivalent to results obtained on the predicate devices. The system performs as
intended and raises no new safety or effectiveness issues. |
.
.
: :
. . . . . . .
the was an a subscription of the subscription in the subscription in
:
3
Comparison of Predicate Devices and Proposed device
Characteristic | Measuring Range | Operating Principle | Intended Use | Characteristic | Hematocrit Range | Sample type | Sample size | Glucose Units | Sample application | Handheld meter? | Data storage | Analysis Time | Insulin Tracking | Power source | Accessories: | Controls: | Test Strips Active reagent: | Test Strip Calibration Coding |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Predicate | ||||||||||||||||||
K070255 - Nova Max Blood | ||||||||||||||||||
Glucose Monitor System | 20-600 mg/dL | Coulometric Electro-chemical | ||||||||||||||||
Sensor | The Nova Max Blood Glucose | |||||||||||||||||
Monitor is intended to be used for | ||||||||||||||||||
the quantitative measurement of | ||||||||||||||||||
glucose in whole blood. It is | ||||||||||||||||||
intended for use by people with | ||||||||||||||||||
diabetes mellitus in the home as | ||||||||||||||||||
an aid to monitor the effectiveness | ||||||||||||||||||
of diabetes control. it is not | ||||||||||||||||||
intended for use in the diagnosis | ||||||||||||||||||
of or screening for diabetes | ||||||||||||||||||
mellitus and is not intended for | ||||||||||||||||||
use on neonates. The Nova Max | ||||||||||||||||||
Blood Glucose Monitor is | ||||||||||||||||||
specifically indicated for the | ||||||||||||||||||
quantitative measurement of | ||||||||||||||||||
glucose in whole blood samples | ||||||||||||||||||
obtained from the fingertip, | ||||||||||||||||||
forearm and palm. | Predicate | |||||||||||||||||
K070255 - Nova Max Blood | ||||||||||||||||||
Glucose Monitor System | 25% to 60% | Capillary blood: fingertip, forearm, palm | 0.30 uL | mg/dL | Test strip capillary draw | Yes | Up to 400 blood glucose and | |||||||||||
control solution tests | 5 seconds | No | 3 volt coin cell battery | Liquid | Glucose Oxidase | No User Input of Calibration code required | ||||||||||||
Predicate | ||||||||||||||||||
K051839- Abbott FreeStyle | ||||||||||||||||||
Freedom Blood Glucose | ||||||||||||||||||
Monitoring System | 20-500 mg/dL | Coulometric Electro-chemical | ||||||||||||||||
Sensor | The FreeStyle Freedom Blood | |||||||||||||||||
Glucose Monitoring System is | ||||||||||||||||||
intended for use in the | ||||||||||||||||||
quantitative measurement of | ||||||||||||||||||
glucose in capillary whole | ||||||||||||||||||
blood from the finger, upper | ||||||||||||||||||
arm and palm, and venous | ||||||||||||||||||
whole blood. It is intended for | ||||||||||||||||||
use by healthcare | ||||||||||||||||||
professionals and people with | ||||||||||||||||||
diabetes mellitus at home as | ||||||||||||||||||
an aid in monitoring the | ||||||||||||||||||
effectiveness of a diabetes | ||||||||||||||||||
control program. It is not | ||||||||||||||||||
intended for the diagnosis of | ||||||||||||||||||
or screening for diabetes | ||||||||||||||||||
mellitus, and is not intended | ||||||||||||||||||
for use on neonates or arterial | ||||||||||||||||||
blood. | Predicate | |||||||||||||||||
K051839 – Abbott FreeStyle | ||||||||||||||||||
Freedom Blood Glucose | ||||||||||||||||||
Monitoring System | 15% to 65% | Capillary blood: forearm, | ||||||||||||||||
upper arm, hand, thigh, calf, | ||||||||||||||||||
or fingers | 0.30 uL | mg/dL | Test strip capillary draw | Yes | Up to 400 blood glucose and | |||||||||||||
control solution tests | 5 seconds | No | 3 volt coin cell battery | Liquid | Glucose Dehydrogenase- | |||||||||||||
PQQ | No User Input of Calibration | |||||||||||||||||
coding required | ||||||||||||||||||
Predicate | ||||||||||||||||||
K092638 - Abbott FreeStyle | ||||||||||||||||||
Blood Glucose Test Strips | 20-500 mg/dL | Coulometric Electro-chemical | ||||||||||||||||
Sensor | The FreeStyle Lite Blood | |||||||||||||||||
Glucose Monitoring System is | ||||||||||||||||||
intended for use in the | ||||||||||||||||||
quantitative measurement of | ||||||||||||||||||
glucose in capillary whole | ||||||||||||||||||
blood from the finger, upper | ||||||||||||||||||
arm and palm, and venous | ||||||||||||||||||
whole blood. It is intended for | ||||||||||||||||||
use by healthcare | ||||||||||||||||||
professionals and people with | ||||||||||||||||||
diabetes mellitus at home as | ||||||||||||||||||
an aid in monitoring the | ||||||||||||||||||
effectiveness of a diabetes | ||||||||||||||||||
control program. It is not | ||||||||||||||||||
intended for the diagnosis of | ||||||||||||||||||
or screening for diabetes | ||||||||||||||||||
mellitus, and is not intended | ||||||||||||||||||
for use on neonates or | ||||||||||||||||||
arterial blood. | Predicate | |||||||||||||||||
K092638 – Abbott FreeStyle | ||||||||||||||||||
Blood Glucose Test Strips | 15% to 65% | Capillary blood: forearm, | ||||||||||||||||
upper arm, hand, thigh, calf, | ||||||||||||||||||
or fingers | 0.30 uL | mg/dL | Test strip capillary draw | Yes | Up to 400 blood glucose and | |||||||||||||
control solution tests | 4 seconds | No | 3 volt coin cell battery | Liquid | Glucose Dehydrogenase - | |||||||||||||
FAD | Calibration coding required | |||||||||||||||||
Proposed | ||||||||||||||||||
Nova Max One Blood Glucose | ||||||||||||||||||
Monitor System | 20-600 mg/dL | Coulometric Electro-chemical | ||||||||||||||||
Sensor | The Nova Max One Blood | |||||||||||||||||
Glucose Monitor is intended to | ||||||||||||||||||
be used for the quantitative | ||||||||||||||||||
measurement of glucose in | ||||||||||||||||||
fresh capillary whole blood. It is | ||||||||||||||||||
intended for single-patient | ||||||||||||||||||
home use and should not be | ||||||||||||||||||
used for testing multiple | ||||||||||||||||||
patients. It is intended for self | ||||||||||||||||||
testing outside the body by | ||||||||||||||||||
people with diabetes mellitus | ||||||||||||||||||
as an aid to monitor the | ||||||||||||||||||
'effectiveness of diabetes | ||||||||||||||||||
control. The Nova Max One | ||||||||||||||||||
Blood Glucose Monitor is | ||||||||||||||||||
specifically indicated for the | ||||||||||||||||||
quantitative measurement of | ||||||||||||||||||
glucose in fresh whole blood | ||||||||||||||||||
capillary samples obtained | ||||||||||||||||||
from the fingertip or alternative | ||||||||||||||||||
site testing (AST) on the | ||||||||||||||||||
forearm. AST on the forearm | ||||||||||||||||||
can be used only during | ||||||||||||||||||
steady-state blood glucose | ||||||||||||||||||
conditions. It is not intended for | ||||||||||||||||||
the diagnosis of or screening | ||||||||||||||||||
for diabetes, and it is not | ||||||||||||||||||
intended for use on newborns | Proposed | |||||||||||||||||
Nova Max One Blood Glucose | ||||||||||||||||||
Monitor System | 25% to 60% | Capillary blood: fingertip, | ||||||||||||||||
forearm | 0.40 uL | mg/dL | Test strip capillary draw | Yes | Up to 400 blood glucose and | |||||||||||||
control solution tests | 4 seconds | No. | 3 volt coin cell battery | Liquid | Glucose Dehydrogenase - FAD | No User Input of Calibration | ||||||||||||
code required |
Page 4 of 5
4
.. ... . . . . . . .
Page 5 of 5
.
·
5
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem that resembles an abstract representation of a human form or a bird in flight. The emblem is composed of several curved lines that create a sense of movement and dynamism.
10903 New Hampshire Avenue Silver Spring, MD 20993
Nova Biomedical Corporation c/o Paul W. MacDonald Chief Quality and Regulatory Affairs Officer 200 Prospect Street Waltham, MA 02454
MAR 0 9 2012
Re: K112638
Trade/Device Name: Nova Max One Blood Glucose Monitoring System Regulation Number: 21 CFR §862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW, LFR, JJX Dated: January 30, 2012 Received: February 1, 2012
Dear Dr. MacDonald:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); lan the Ace oregallerinents, medical device reporting (reporting of medical device-related laberse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
6
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical
Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm
Sincerely yours,
Counney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
7
Indications for Use
510(k) Number (if known): K112638
Device Name:
Nova Max One Blood Glucose Monitoring System
Indications for Use:
The Nova Max One Blood Glucose Monitor is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. It is intended for single-patient home use and should not be used for testing multiple patients. It is intended for self testing outside the body by people with diabetes mellitus as an aid to monitor the effectiveness of diabetes control. The Nova Max One Blood Glucose Monitor is specifically indicated for the quantitative measurement of glucose in fresh whole blood capillary samples obtained from the fingertip or alternative site testing (AST) on the forearm. AST on the forearm can be used only during steady-state blood glucose conditions. It is not intended for the diagnosis of or screening for diabetes, and it is not intended for use on newborns.
Nova Max One Glucose Test Strips are intended for use only with the Nova Max One Blood Glucose Monitor for quantitative tests. The Glucose Monitor is intended to quantitatively measure glucose (sugar) in fresh capillary whole blood obtained from the finger tip or alternative site testing (AST) on the forearm. AST can be used only during steady-state blood glucose conditions. The Glucose Monitor is calibrated to provide plasma equivalent results to laboratory methods. Nova Max One Glucose Test Strips are for testing outside the body (in vitro diagnostic use only). The Monitor should only be used by a single user in the home and should not be shared by users. It is not intended for the diagnosis of or screening for diabetes, and it is not intended for use on newborns.
Nova Max Control Solutions are intended for use with the Nova Max Blood Glucose Monitoring Systems as a quality control check to verify the accuracy of blood glucose test results. There are three levels of controls, (Levels 1, 2, 3).
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K112638
Page 1 of