K Number
K112629

Validate with FDA (Live)

Date Cleared
2011-10-11

(32 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Onset Medical Access Catheter System is indicated for the introduction of interventional/diagnostic devices into the peripheral, coronary and neuro vasculature.

Device Description

The Onset Access Catheter System is a two-catheter system comprised of a Sheath and a Dilator. The Sheath has a distal region which expands when the Dilator is advanced. The Onset Access Catheter System can be used with hydrophilic guidewires up to.038 inches in diameter to access the desired anatomy. The proximal ends of the Sheath and the Dilator have a luer fitting to allow attachment of accessories and infusion of liquids through the catheter. The catheter is offered in various sizes to accommodate physician preferences and patient anatomy. A split sheath is provided in the package to provide support and facilitate the introduction of the Sheath's distal expandable region into the catheter introducer sheath. The Catheter System is provided sterile, non-pyrogenic, and is intended for single use only.

AI/ML Overview

The provided document describes the "Onset Access Catheter System," a medical device. However, it does not contain the specific information required to answer the questions about acceptance criteria for an AI/ML powered device, as this is a traditional medical device (a catheter). The document focuses on demonstrating substantial equivalence to a predicate device through biocompatibility, sterilization, and bench-top testing.

The questions you've asked are typically related to the validation of AI/ML-powered medical devices, which involve performance metrics like accuracy, sensitivity, and specificity, and studies to establish ground truth. This document describes a physical medical device and its general performance and safety.

Therefore, I cannot extract the following information from the provided text:

  1. A table of acceptance criteria and the reported device performance (for AI/ML metrics)
  2. Sample size used for the test set and the data provenance (for AI/ML models)
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (for AI/ML models)
  4. Adjudication method (for AI/ML models)
  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance (for AI/ML models)
  6. If a standalone study was done (for AI/ML models)
  7. The type of ground truth used (for AI/ML models)
  8. The sample size for the training set (for AI/ML models)
  9. How the ground truth for the training set was established (for AI/ML models)

What the document does provide in relation to "acceptance criteria" and "study":

The document details a series of bench-top (design verification) tests and biocompatibility tests to show that the Onset Access Catheter System meets established criteria for substantial equivalence to a predicate device.

Acceptance Criteria and Reported Device Performance (Bench Tests):

Bench TestsAcceptance Criteria / Result
Dimensional and Visual InspectionMet established criteria
Guidewire CompatibilityMet established criteria
Torque ResponseMet established criteria
Torque StrengthMet established criteria
System Deployment Cycle TestMet established criteria
Kink ResistanceMet established criteria
Flexibility TestMet established criteria
Tensile StrengthMet established criteria
Catheter Leak Test (Liquid Leakage)Met established criteria
Catheter Leak Test (Air Leakage)Met established criteria
Dynamic Pressure TestMet established criteria
Static Burst TestMet established criteria
Aspiration TestMet established criteria
Hub GaugingMet established criteria
Corrosion ResistanceMet established criteria
USP Particulate TestMet established criteria
Navigation and Accessibility Capabilities in vitroMet established criteria
In vivo Tests
System Deliverability, Compatibility, Visibility and Aspiration PerformanceMet established criteria
Acute histopathology of treated vesselsMet established criteria
Biocompatibility testingMet established criteria

Biocompatibility Testing:

TestMethodResult
CytotoxicityL929 MEM Elution TestPassed
SensitizationKligman MaximizationPassed
Intracutaneous Reactivity (Irritation)Intracutaneous Injection TestPassed
Systemic Toxicity (Acute)ISO Acute Systemic Injection TestPassed
HaemocompatibilityComplement ActivationPassed
HemolysisPassed
Inactivated Partial Thromboplastin Time TestPassed
In vivo thrombogenicityPassed
PyrogenicityUSP Material Mediated Rabbit Pyrogen TestPassed

Study Information (for this traditional device):

  • Sample sizes: Not explicitly detailed for each test (e.g., number of catheters tested for torque response), but the document states "All testing was performed on units which were twice sterilized."
  • Data provenance: Not applicable in the context of clinical AI/ML data; these are laboratory and animal (in vivo) tests.
  • Experts/Ground Truth/Adjudication: Not applicable in the context of clinical AI/ML data. The "ground truth" for these tests is defined by established engineering and biological standards.
  • MRMC / Standalone / Training Set: Not applicable as this is not an AI/ML device.

In summary, the provided text describes the performance validation of a physical medical device through standard engineering and biocompatibility testing, not the performance validation of an AI/ML algorithm.

{0}------------------------------------------------

4. 510(k) Summary according to 807.92

Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Onset Medical Corporation is providing the summary of Substantial Equivalence for the Onset Access Catheter System.

4.1 Sponsor /Applicant Name and Address

Onset Medical Corporation 13900 Alton Parkway Suite 120 Irvine, CA 92618

4.2 Sponsor Contact Information

Joseph Bishop President, CEO Onset Medical Corporation Phone: 949-716-1100 Fax: 949-716-1101 email: jbishop@Onsetmedical.com

4.3 Date of Preparation of 510(k) Summary

September 8, 2011

4.4 Device Trade or Proprietary Name

Onset Access Catheter System

4.5 Device Common/Usual or Classification Name

Catheter, Percutaneous (Product Code: DQY)

4.6 Identification of the Legally Marketed Devices to which Equivalence is Being Claimed:

Name of Predicate DeviceName of Manufacturer(Town, State)510(k) Number
Reverse RePort™ GuideCatheter SystemReverse Medical CorporationIrvine, CA 92618K102418

OCT 1 1 2011

11 112629

{1}------------------------------------------------

4.7 Device Description

The Onset Access Catheter System is a two-catheter system comprised of a Sheath and a Dilator. The Sheath has a distal region which expands when the Dilator is advanced. The Onset Access Catheter System can be used with hydrophilic guidewires up to.038 inches in diameter to access the desired anatomy.

The proximal ends of the Sheath and the Dilator have a luer fitting to allow attachment of accessories and infusion of liquids through the catheter. The catheter is offered in various sizes to accommodate physician preferences and patient anatomy. A split sheath is provided in the package to provide support and facilitate the introduction of the Sheath's distal expandable region into the catheter introducer sheath. The Catheter System is provided sterile, non-pyrogenic, and is intended for single use only.

4.8 Intended Use

The Onset Access Catheter System is indicated for the introduction of interventional/diagnostic devices into the peripheral, coronary and neuro vasculature.

Onset Access Catheter SystemRePort™ Guide Catheter System
510(k) NumberTBDK102418
ClassificationSAMEClass II, DQY
IndicationSAMEThe RePort Catheter is indicated for theintroduction of interventional /diagnostic devices into the peripheral,coronary, and neuro vasculature.
Material
- Shaft MaterialsSAMEPTFE lined nylon/polyurethane withstainless steel and nitinol braid support
- Proximal EndConfigurationSAMELuer Hub
- Radiographic markers /RadiopacitySAME• Sheath: Platinum/tungsten bandsembedded at junction of expandablesection; gold marker at tip ofexpandable section;• Dilator: shaft is radiopaque throughout
- PackagingSAMECatheter in polyethylene hoopattached to packaging card insidePET/PE/Tyvek pouch inside SBScarton
SterilizationSAMEEtO

4.9 Comparison to Predicate Devices

{2}------------------------------------------------

4.10 Summary of Non-clinical Data

4.10.1 Biocompatibility and Sterilization

The Onset Access Catheter System is classified as Externally Communicating Devices, Circulating Blood, Limited Contact (≤ 24 hours). Results of testing demonstrate that the blood contacting materials are biocompatible.

Blood contacting materials were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO 10993-1 guidelines "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing". The Catheter System successfully passed all of the following biocompatibility tests:

TestMethod
CytotoxicityL929 MEM Elution Test
SensitizationKligman Maximization
Intracutaneous Reactivity (Irritation)Intracutaneous Injection Test
Systemic Toxicity (Acute)ISO Acute Systemic Injection Test
HaemocompatibilityComplement Activation
Hemolysis
Inactivated Partial Thromboplastin Time Test
In vivo thrombogenicity
PyrogenicityUSP Material Mediated Rabbit Pyrogen Test

Sterilization conditions have been validated according to ANSI / AAMI / ISO 11135, Sterilization of Health Care Products-Part 1: Requirement, validation and routine control of a sterilization process for medical devices to provide a Sterility Assurance Level (SAL) of 10 °.

4.10.2 Design Verification (Bench-Top Testing)

The physical, mechanical and performance testing of the Catheter System demonstrated that the product is substantially equivalent to currently marketed predicate devices. Design Verification testing was conducted to evaluate the physical and mechanical properties of the Catheter System. All testing was performed on units which were twice sterilized and met all inspection criteria. Tests on the Catheter System included:

{3}------------------------------------------------

yBench TestsResult
Dimensional and Visual InspectionMet established criteria
Guidewire CompatibilityMet established criteria
Torque ResponseMet established criteria
Torque StrengthMet established criteria
System Deployment Cycle TestMet established criteria
Kink ResistanceMet established criteria
Flexibility TestMet established criteria
Tensile StrengthMet established criteria
Catheter Leak Test (Liquid Leakage)Met established criteria
Catheter Leak Test (Air Leakage)Met established criteria
Dynamic Pressure TestMet established criteria
Static Burst TestMet established criteria
Aspiration TestMet established criteria
Hub GaugingMet established criteria
Corrosion ResistanceMet established criteria
USP Particulate TestMet established criteria
Navigation and Accessibility Capabilities in vitroMet established criteria
In vivo TestsResult
System Deliverability, Compatibility, Visibility andAspiration PerformanceMet established criteria
Acute histopathology of treated vesselsMet established criteria
Biocompatibility testingMet established criteria

Verification and Test Summary

The physical, mechanical and performance testing of the subject Catheter System demonstrated that the product is safe and effective for its labeled indications and is Substantially Equivalent to currently marketed predicate devices.

4.11 Substantial Equivalence

The Onset Access Catheter System is the same device as the RePort™ Guide Catheter System. The only difference is that it is manufactured and marketed by Onset Medical rather than Reverse Medical. Therefore, it is equivalent in intended use, design, technology/principles of operation, materials and performance to the predicate device.

The data presented in K102418 demonstrated that the RePort™ Guide Catheter System was safe and effective for its labeled indications and Substantially Equivalent to currently marketed predicate devices. The Onset Access Catheter is the same Catheter System and is, therefore, also safe and effective for its labeled indications and Substantially Equivalent to currently marketed predicate devices.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized symbol resembling a bird or a wing, composed of three curved lines. The symbol is positioned to the right of the text, which is arranged in a circular pattern around the symbol. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Onset Medical Corporation c/o Mr. Joseph Bishop President. CEO 13900 Alton Parkway Suite 120 Irvine, CA 92618

OCT 1 1 2011

Re: K112629

Trade Name: Onset Access Catheter System Regulation Number: 21 CFR 870.1250 Regulation Name: Catheter, Percutaneous Regulatory Class: 11 (two) Product Code: DQY Dated: September 7, 2011 Received: September 9. 2011

Dear Mr. Bishop:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Joseph Bishop

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutfDA/CentersOffices/CDRH/CDRHOffices/ucm113809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Mu c

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosures

{6}------------------------------------------------

3. Indications for Use

6112629 510(k) Number (if known): _

Device Name: Onset Medical Access Catheter System

Indications for Use:

The Onset Medical Access Catheter System is indicated for the introduction of interventional/diagnostic devices into the peripheral, coronary and neuro vasculature.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over the Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

on Sign-Off) Division of Cardiovascular Devices

510(k) Number K112629

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).