K Number
K111982
Date Cleared
2011-11-17

(128 days)

Product Code
Regulation Number
872.3690
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Indications for Use: Post cementation and core build-up

Device Description

The subject device is a dual-cured (light-cured with self-curing property), radiopaque two-component core build-up material supplied in an automix delivery system which can mix equal amount of two components, and is available in two shades, Dentin and White.

AI/ML Overview

This document describes a 510(k) premarket notification for a dental device, not an AI/ML medical device. Therefore, much of the requested information regarding AI/ML device studies, such as sample sizes for test/training sets, ground truth establishment by experts, adjudication methods, and MRMC studies, is not applicable or available within this document.

The document primarily focuses on demonstrating substantial equivalence to predicate devices based on intended use, chemical ingredients/safety, and physical/mechanical properties.

Here's an attempt to answer the questions based on the provided text, while acknowledging the limitations:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Standard)Reported Device Performance (Summary)
ISO 4049: 2009 RequirementsComplies with the requirements of ISO 4049: 2009 (Dentistry - Polymer-based restorative materials).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The document states that physical and mechanical properties were evaluated according to ISO 4049: 2009, but it does not specify the sample size used for these evaluations or the data provenance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable as this is a dental material device, not an AI/ML diagnostic device requiring expert interpretation for ground truth. The evaluation focused on compliance with a technical standard (ISO 4049).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable for the same reason as above. There was no "test set" in the context of expert judgment for ground truth.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. This is a dental material device, not an AI/ML diagnostic or assistive device that would involve human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. This is a physical dental material, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" in this context refers to the requirements of the ISO 4049: 2009 standard for physical and mechanical properties of polymer-based restorative materials. The device's performance was compared against these predefined physical and mechanical thresholds.

8. The sample size for the training set

This information is not provided and is not applicable as this is not an AI/ML device requiring a training set.

9. How the ground truth for the training set was established

This is not applicable as this is not an AI/ML device requiring a training set or its associated ground truth establishment.

{0}------------------------------------------------

[CLEARFIL DC CORE PLUS, KURARAY MEDICAL INC.]
Section & Section & Section & C

Kllllag82

NOV 1 7 2011

Section 3: Summary Date: July 6, 2011

:

. . .. . .

. ·

:

.. . ון

510(k) Summary

  • 3-1. 510(k) owner (submitter)
  1. Name

  2. Address

  3. Contact person

  4. Contact person in US

KURARAY MEDICAL INC.

1621 Sakazu, Kurashiki, Okayama 710-0801, Japan
Mira

Michio Takigawa Quality Assurance Department

Kiyoyuki Arikawa KURARAY AMERICA INC. 600 Lexington Avenue, 26th Floor New York, NY 10022 Tel: 1002014, 11 110022
Tel: (212)-986-2230 (Ext. 115) or (800)-879-1676
Fax: (212)-867-3543 Fax: (212)-867-3543

3-2. Name of Device

l ) Trade / Proprietary name

    1. Classification name
  1. Common name

3-3. Predicate device

б

I) CLEARFIL DC GO

CLEARFIL DC CORE PLUS

Tooth shade resin material (21 CFR section 872.3690. Product code: EBF)
(21 CFR section 872.3690. Product code: EBF)
Dual

Dual-cured core build-up material

1) CLEARFIL DC CORE AUTOMIX510(k) Number:Product Code:21 CFR Section:Applicant:K043177EBF872.3690KURARAY MEDICAL INC.
2) CLEARFIL SA CEMENT510(k) Number:Product Code:21 CFR Section:Applicant:K081583EMA872.3275KURARAY MEDICAL INC.
3) CLEARFIL ESTHETIC CEMENT EX510(k) Number:Product Code:21 CFR Section:Applicant:K062410EMA872.3275(b)KURARAY MEDICAL INC.
4) CLEARFIL MAJESTY Flow510(k) Number:Product Code:21 CFR Section:Applicant:K063593EBF872.3690KURARAY MEDICAL INC.
5) CLEARFIL AP-X510(k) Number:Product Code:21 CFR Section:Applicant:K012740EBF872.3690KURARAY MEDICAL INC.
6) ESTENIA C&B510(k) Number:Product Code:21 CFR Section:Applicant:K042929EBF and EBG872.3690KURARAY MEDICAL INC.

{1}------------------------------------------------

3-4. Device Description

The subject device is a dual-cured (light-cured with self-curing property), radiopaque two-component core build-up material supplied in an automix delivery system which can mix equal amount of two components, and is available in two shades, Dentin and White.

3-5. Substantial Equivalence Discussion

l) Intended use

The Intended use of the subject device was written up based on that of CLEARFIL DC CORE AUTOMIX. Therefore, the intended use of the subject device is substantially same as that of the predicate device.

2) Chemical ingredients / Safety

Except for a new ingredients in the subject device have been used in the predicate devices. Regarding the predicate devices, there have not been any reported problems or recalls according to the post market adverse event reporting requirements in regions where they have been sold.

And the new ingredient has been widely used for many years in the food industry as a food addictive. There have already been many reports that show its biological safety. As the result, it was concluded that the new ingredient was biologically safe.

In conclusion, it can be said that the safety of the subject device is substantially equivalent to that of the predicate devices.

3 ) Effectiveness / Performance

Physical and mechanical properties of the subject device have been evaluated according to the applicable FDA recognized consensus standard, ISO 4049: 2009 (Dentistry - Polymer-based restorative materials) which is applicable to dental composite resin. The predicate device devoted for the comparative study was CLEARFIL DC CORE AUTOMIX that had the same structure and application as the subject device.

The study results indicate that the subject device and the predicate device comply with the requirements of ISO 4049: 2009. From this, it can be said that the subject device is effective as well as the predicate device.

3-6. Biocompatibility

Except for a new ingredients in the subject device have been used in the predicate devices as listed on the tables of "7-4 Chemical ingredients": in "Section 7: Substantial Equivalence Discussion". Regarding the predicate devices, there have not been any reported problems or recalls according to the post market adverse event reporting requirements in regions where they have been sold.

And the new ingredient has been widely used for many vears in the food industry as a food addictive. There have already been many reports that show its biological safety. As the result, it was concluded that the new ingredient was biologically safe.

Therefore, it was concluded that the biological safety of the subject device could be assured.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: the department's name arranged in a circular fashion around the left side, and a stylized symbol on the right. The symbol is composed of three curved lines that resemble a human figure embracing or supporting something, and the text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

KURARARY MEDICAL Incorporated C/O Mr. Kiyoyuki Arikawa General Manager Dental Material Division 600 Lexington Avenue, 26th Floor New York, New York 10022

NOV 1 7 2011

Re: K111982

Trade/Device Name: CLEARFIL DC CORE PLUS Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: II Product Code: EBF Dated: October 24, 2011 Received: October 24, 2011

Dear Mr. Arikawa:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Nh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital; Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):__K111982

Device Name: CLEARFIL DC CORE PLUS

Indications for Use: Post cementation and core build-up

Prescription Use ____ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ N/A (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Runne

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K111982

§ 872.3690 Tooth shade resin material.

(a)
Identification. Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.(b)
Classification. Class II.