K Number
K111914

Validate with FDA (Live)

Date Cleared
2012-08-21

(412 days)

Product Code
Regulation Number
864.1860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VS800 system is an automated digital slide creation, management, and viewing system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape.

The VS800HER2 Manual Read (MR) of digital slide application is intended for use as an aid to the pathologist in the detection and semi-quantitative measurement of HER2 by manual examination of the digital slide of formalin-fixed, paraffin-embedded and neoplastic tissue IHC stained for HER2 receptors on a computer monitor. HER2 results are indicated for use as an aid in the management, prognosis and prediction of therapy outcomes of breast cancer.

The VS800HER2 MR of digital slide application is intended for use as an accessory to the DakoHercepTest to aid the pathologist in the detection and semi-quantitative measurement of HER2 by manual examination of the digital slide of formalin-fixed, paraffin-embedded and neoplastic tissue immunohistochemically stained for HER2 receptors on a computer monitor. When used with the Dako Hercep Test, it is indicated for use as an aid in the assessment of breast cancer patients for whom HERCEPTIN® (Trastuzumab) treatment is being considered.

Note: The actual correlation of the Dako Hercep Test to the Herceptin® clinical outcome has not been established.

Device Description

The VS800 System is an automated digital slide creation, management and viewing system. The VS800 System components consist of an automated digital microscope slide scanner (VS800-SS) which include a computer, keyboard and mouse, operating monitor (VS800-MTR) and VS Viewer software (VS2-ASW-IDB). The system capabilities include digitizing microscope slides at high resolution, storing and managing the resulting digital slide images, retrieving and displaying digital slides, including support for remote access over wide-area networks, providing facilities for annotating digital slides and editing metadata associated with digital slides, and facilities for image analysis of digital slides. The remote digital slide viewing capabilities of the system support reading digital slides on a computer monitor, enabling Pathologists to make clinically relevant decisions analogous to those they make using a conventional microscope. Specifically, the system supports the pathologist in the detection of HER2/neu by manual examination of the digital slide of formalin-fixed, paraffin-embedded normal and neoplastic tissue immunohistochemically stained for HER2 receptors on a computer monitor.

The VS800-SS (an automated digital microscope slide scanner) creates high resolution, color digital slide images of entire glass slides in a matter of minutes. High numeric aperture 20x objectives, specially designed for VS800-SS optical system and real time contrast auto focus system (AF) are used to produce high-quality images. VS800-SS employs a 2D CCD imager for fine image acquisition which is same technologies used in conventional microscope imaging system. VS800-SS captured image is as same as conventional microscope image.

The VS-ASW-IDB (VS Viewer software) is a full-featured digital pathology information management system. The software runs on a server computer, which stores digital slide images on disk storage such as a RAID array, and which hosts an SQL database that contains digital slide metadata. The VS-ASW-IDB includes a web application and services which encapsulate database and digital slide image access for other computers. The VS-ASW-IDB also includes support for locally or remotely connected Image Server, which run digital slide viewing software provided as part of VS-ASW-IDB.

The laboratory technician or operator of VS800-SS loads glass microscope slides into a specially designed slide carrier with a capacity up to 100 slides per carrier (300 total). The scanning process begins when the operator starts the VS800-SS scanner and finishes when the scanner has completed scanning of all loaded slides. As each glass slide is processed, the system automatically stores stitched images as a single digital slide image, which represents a histological reconstruction of the entire tissue section. When the slide scanning finished, then operator of scanner will confirms the image quality and records to the database. When the images are recorded, pathologists or authorized parsons can observe these images to access the VS-ASW-IDB.

AI/ML Overview

Here's a summary of the acceptance criteria and study details for the Olympus VS800HER2 MR Application, based on the provided 510(k) summary:

Acceptance Criteria and Device Performance

Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance (Mean %)Reported Device Performance (Range %)
Agreement with Manual Microscopy Reads (Trichotomous HER2 Scores: 0,1+; 2+; 3+)(Individual Pathologist % Agreements are shown in the tables below.)
Site 1, Pathologist 1
HER2 0, 1+N/A (Comparison study, not a specific threshold for acceptance)90.91%(75.67%, 98.08%) CI
HER2 2+N/A88.24%(72.55%, 96.70%) CI
HER2 3+N/A96.97%(84.24%, 99.92%) CI
Site 1, Pathologist 2
HER2 0, 1+N/A91.18%(76.32%, 98.14%) CI
HER2 2+N/A90.91%(75.67%, 98.08%) CI
HER2 3+N/A96.97%(84.24%, 99.92%) CI
Site 1, Pathologist 3
HER2 0, 1+N/A60.00%(38.67%, 78.87%) CI
HER2 2+N/A97.22%(85.47%, 99.93%) CI
HER2 3+N/A87.18%(72.57%, 95.70%) CI
Site 2, Pathologist 1
HER2 0, 1+N/A85.19%(66.27%, 95.81%) CI
HER2 2+N/A80.95%(65.88%, 91.40%) CI
HER2 3+N/A100%(88.78%, 100%) CI
Site 2, Pathologist 2
HER2 0, 1+N/A96.67%(82.78%, 99.92%) CI
HER2 2+N/A78.38%(61.79%, 90.17%) CI
HER2 3+N/A100%(89.42%, 100%) CI
Site 2, Pathologist 3
HER2 0, 1+N/A63.89%(46.22%, 79.18%) CI
HER2 2+N/A80.65%(62.53%, 92.55%) CI
HER2 3+N/A93.94%(79.77%, 99.26%) CI
Precision Study (Overall Agreements for Manual Digital Reads)
Intra-Instrument (Intra-Pathologist)N/A (Comparison study, not a specific threshold for acceptance)100%(95.98%, 100%) CI
Inter-Instruments (Intra-Pathologist)N/A95.6%(89.01%, 98.78%) CI

Note: The study describes percentages of agreement without predefined acceptance thresholds for substantial equivalence in the document. The statistical analysis is presented as Percent Agreement (PA) with a 95% Confidence Interval (CI) between manual microscopy reads and manual digital reads, and for precision studies.


Study Details for the Olympus VS800HER2 MR Application

  1. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: 100 slides per clinical site, so a total of 200 slides were used for the comparison study (100 slides at Site 1, 100 slides at Site 2).
    • For the precision study, a subset of 30 slides from the comparison study was used.
    • Data Provenance: Retrospective. The slides were "selected from archive." The country of origin is not explicitly stated, but the study was conducted at "two clinical sites," implying local (likely within the US, given FDA submission context) data.
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • Number of Experts: Six pathologists in total for the comparison study (three at each of two clinical sites). One pathologist for the precision study (who repeated reads).
    • Qualifications of Experts: Described as "Pathologists." Specific years of experience or sub-specialty certifications are not provided in the summary.
  3. Adjudication Method for the Test Set:

    • Adjudication Method: None for establishing a single "ground truth." The study compared each pathologist's "manual digital reads" against their own "manual microscopy reads." This is a paired comparison, where the conventional microscopy read by the same pathologist is considered the reference for that pathologist's digital read. The summary doesn't describe an external or consensus ground truth for the comparison study itself; rather, it assesses agreement between two reading methods by the same individual.
    • For the precision study, there was also no external adjudication; it assessed agreement of repeated reads by a single pathologist.
  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This was a Multi-Reader Multi-Case (MRMC) comparative effectiveness study of digital pathology reads vs. conventional microscopy reads, but without AI assistance. The VS800HER2 MR Application is a "Manual Read" application, meaning the pathologist manually interprets the digital image.
    • Therefore, there is no AI component, and no effect size regarding human readers improving with AI assistance is reported or applicable to this specific application. The study focuses on the agreement between conventional microscopy and manual reading of digital slides.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No standalone algorithm-only performance study was done. The device (VS800HER2 MR Application) is explicitly described as for "Manual Read (MR) of digital slide application," intended "as an aid to the pathologist in the detection and semi-quantitative measurement of HER2 by manual examination of the digital slide." It is a display and management system for pathologists to manually review digital slides, not an automated AI-driven diagnostic algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the comparison study: The "ground truth" or reference standard for each pathologist's digital read was their own prior manual microscopy read of the physical glass slide. This is a paired comparison, where the same pathologist acts as their own control.
    • For the precision study: The reference was the pathologist's own repeated manual digital reads from the same or different instruments.
  7. The sample size for the training set:

    • The 510(k) summary does not mention a training set for an algorithm, as the VS800HER2 MR Application is a manual read application. The study described is entirely a clinical validation/test set.
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no mention of a training set or an algorithm being developed (which would typically require a training set with established ground truth). The device acts as a digital visualization and management system for manual pathologist interpretation.

{0}------------------------------------------------

K111914

AUG 2 1 2012

510(k) SUMMARY

Virtual Slide System OLYMPUS VS800 System, VS800HER2 MR Application

General Information 1

. Applicant:

. Contact Person:

1 Manufacturer:

Device Identification 2

■ Device Trade Name:

. Regulation Number:

■ Regulation Name:

Requlatory Class: ■

Classification Panel: .

Product Code:

OLYMPUS CORPORATION Shinjuku Monolith 2-3-1 Nishi shinjuku Shinjuku ku, Tokyo, 163-0914 JAPAN Establishment Registration No.: 8010047

Sheri L. Musgnung Associate Manager, Regulatory Affairs Olympus America Inc. 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610 Phone: 484-896-3147 FAX: 484-896-7131 Sheri.Musgnung@olympus.com Email: Establishment Registration No .: 2429304

OLYMPUS CORPORATION 5128 Nishi-machi Ina-shi Nagano prefecture 396-0026, Japan Establishment Registration No .: 9614612

Virtual Slide System, OLYMPUS VS800 System, VS800 HER2 MR Application

21CFR 864.1860

Immunohistochemistry reagents and kits.

II

Pathology

OEO

{1}------------------------------------------------

3 Predicate Device Information

Device Name510(k) No.Manufacturer
ScanScope XT SystemK071671Aperio Technologies, Inc.

Device Description ব

The VS800 System is an automated digital slide creation, management and viewing system. The VS800 System components consist of an automated digital microscope slide scanner (VS800-SS) which include a computer, keyboard and mouse, operating monitor (VS800-MTR) and VS Viewer software (VS2-ASW-IDB). The system capabilities include digitizing microscope slides at high resolution, storing and managing the resulting digital slide images, retrieving and displaying digital slides, including support for remote access over wide-area networks, providing facilities for annotating digital slides and editing metadata associated with digital slides, and facilities for image analysis of digital slides. The remote digital slide viewing capabilities of the system support reading digital slides on a computer monitor, enabling Pathologists to make clinically relevant decisions analogous to those they make using a conventional microscope. Specifically, the system supports the pathologist in the detection of HER2/neu by manual examination of the digital slide of formalin-fixed, paraffin-embedded normal and neoplastic tissue immunohistochemically stained for HER2 receptors on a computer monitor.

The VS800-SS (an automated digital microscope slide scanner) creates high resolution, color digital slide images of entire glass slides in a matter of minutes. High numeric aperture 20x objectives, specially designed for VS800-SS optical system and real time contrast auto focus system (AF) are used to produce high-quality images. VS800-SS employs a 2D CCD imager for fine image acquisition which is same technologies used in conventional microscope imaging system. VS800-SS captured image is as same as conventional microscope image.

The VS-ASW-IDB (VS Viewer software) is a full-featured digital pathology information management system. The software runs on a server computer, which stores digital slide images on disk storage such as a RAID array, and which hosts an SQL database that contains digital slide metadata. The VS-ASW-IDB includes a web application and services which encapsulate database and digital slide image access for other computers. The VS-ASW-IDB also includes support for locally or remotely connected Image Server, which run digital slide viewing software provided as part of VS-ASW-IDB.

The laboratory technician or operator of VS800-SS loads glass microscope slides into a specially designed slide carrier with a capacity up to 100 slides per carrier (300 total). The scanning process begins when the operator starts the VS800-SS scanner and finishes when the scanner has completed scanning of all loaded slides. As each glass slide is processed, the system automatically stores stitched images as a single digital slide image, which represents a histological reconstruction of the entire tissue section. When the slide scanning finished, then operator of scanner will confirms the image quality and records to the database. When the images are recorded, pathologists or authorized parsons can observe these images to access the VS-ASW-IDB.

Indications for Use ട്

The VS800 system is an automated digital slide creation, management, and viewing system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size,

{2}------------------------------------------------

pattern and shape.

The VS800HER2 Manual Read (MR) of digital slide application is intended for use as an aid to the pathologist in the detection and semi-quantitative measurement of HER2 by manual examination of the digital slide of formalin-fixed, paraffin-embedded and neoplastic tissue IHC stained for HER2 receptors on a computer monitor. HER2 results are indicated for use as an aid in the management, prognosis and prediction of therapy outcomes of breast cancer.

The VS800HER2 MR of digital slide application is intended for use as an accessory to the DakoHercepTest to aid the pathologist in the detection and semi-quantitative measurement of HER2 by manual examination of the digital slide of formalin-fixed, paraffin-embedded and neoplastic tissue immunohistochemically stained for HER2 receptors on a computer monitor. When used with the Dako Hercep Test, it is indicated for use as an aid in the assessment of breast cancer. patients for whom HERCEPTIN® (Trastuzumab) treatment is being considered.

Note: The actual correlation of the Dako Hercep Test to the Herceptin® clinical outcome has not been established.

Comparison of Technological Characteristics ರಿ

A summary of the technological characteristics of the VS800 System, subject device, in comparison to the predicate device follows:

Method of cell detection: The method of cell detection is by colorimetric pattern recognition by microscopic examination of prepared cells by size, shape, hue and intensity as observed by a computer-automated, microscopic digital slide scanner system and/or by visual observation by a health care professional.

System Components: The system components comprising the VS800 System, subject device, are substantially equivalent to those in the predicate device; i.e., a computer-automated digital microscope slide scanner, computer, color monitor, and keyboard.

Energy Source: The electrical service is 100 - 120vAC / 220 - 240vAC, 50Hz/60 Hz, 5.8A MAX, which is similar to the predicate device electrical service requirements.

Summary of Non-Clinical Testing 7

In support of the non-clinical tests, a risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971. The design verification tests and their acceptance criteria were identified and performed as a result of the risk analysis assessment. In addition, electrical safety and electromagnetic compatibility testing was performed and carried out in accordance to IEC61010-2-101:2002 and EN 61326-2-6:2006.

The software validation activities were performed in accordance with the FDA Guidance for the Contents of Premarket Submissions for Software Contained in Medical Devices." Olympus has accessed that our proposed VS800 System has a Moderate Level of concern, and the

{3}------------------------------------------------

documents within the 510(k) support the level of concern.

Summary of Clinical Testing 8

A prospective clinical study was conducted to validate the performance for reading images of slides on a computer monitor (manual digital reads) in comparison to reading the glass-slides with a conventional microscope (manual microscopy reads). The VS800 system was used to digitize, manage and view the slides. The specific application is limited to Formalin-Fixed, Paraffin-Embedded (FFPE) breast tissue specimens ImmunoHistoChemically (IHC) stained with Human Epidermal growth factor Receptor 2 (HER2) using DakoCytomation's FDA approved i HercepTest™ kit.

A study was conducted at two clinical sites to compare the performance of manual digital reads with manual microscopy reads. One clinical laboratory provided two clinical sites with different slide sets of one hundred (100) slides each. The slides were selected from archive to provide an equal distribution of HER2 scores in the trichotomous categorization of the HER2 scores combining 0 and 1+ and leaving 2+ and 3+ uncombined, and then re-cut and re-stained. At each clinical site, three different pathologists performed first a blinded read of the glass-slides using a conventional microscope. Then, the slides were digitized using a different instrument at each clinical site, and after a wash-out period of at least one week and randomization, the same three pathologists at each site performed another blinded read of the slides, but this time they read the images of the slides on their computer monitors.

The statistical analysis is presented across all slides comparatively between manual microscopy reads and manual digital reads using Percent Agreement (PA) with a 95% Confidence Interval (CI). The statistical analysis is provided for a trichotomous agreement of the HER2 scores combining 0 and 1+, and leaving 2+ and 3+ uncombined, as suggested by FDA.

Tables 1, 2, 3, 4, 5, and 6 show the 4x4 tables for the agreements between the manual microscopy reads and manual digital reads using column-wise Percent Agreement (PA; i.e., percent of manual digital reads that agree with a given manual microscopy read) with a Exact 95% Confidence Interval (CI) for each of the three pathologists at site 1 and site 2. For example, there were 33 (10 + 23) slides with manual microscopy read either 0 or 1+ and 30 (4 + 0 + 6 +20) of these were called 0 or 1+ by manual digital read, resulting in a PA of 90.91% (30/33).

{4}------------------------------------------------

Pathologist 1
01+2+3+TOTAL
ManualDigitalRead040004
1+6203029
2+0330134
3+0013233
TOTAL10233433100
ScorePAExact 95% CI
0, 1+90.91%(75.67%,98.08%)
2+88.24%(72.55%,96.70%)
3+96.97%(84.24%,99.92%)

Table 1:

Manual Microscopy Reads vs. Manual Digital Reads – Site 1 4x4 Tables for Pathologist 1

Manual Microscope Read
Pathologist 201+2+3+TOTAL
ManualDigitalRead081009
1+5173025
2+0330134
3+0003232
TOTAL13213333100
ScorePAExact 95% CI
0, 1+91.18%(76.32%,98.14%)
2+90.91%(75.67%,98.08%)
3+96.97%(84.24%, 99.92%)

Table 2: Manual Microscopy Reads vs. Manual Digital Reads – Site 1 4x4 Tables for Pathologist 2

{5}------------------------------------------------

Manual Microscope Read
Pathologist 301+2+3+TOTAL
030003
ManualDigitalRead1+0121013
2+01035550
3+0003434
TOTAL3223639100
ScorePAExact 95% CI
0, 1+60.00%(38.67%,78.87%)
2+97.22%(85.47%,99.93%)
3+87.18%(72.57%,95.70%)

Table 3:

and the comments of the comments of

. . . . . .

Manual Microscopy Reads vs. Manual Digital Reads – Site 1 4x4 Tables for Pathologist 3

. . .

Manual Microscope Read
01+2+3+TOTAL
Pathologist 1920011
ManualDigitalRead2101013
1334038
0073138
TOTAL12154231100
ScorePAExact 95% CI
0, 1+85.19%(66.27%,95.81%)
2+80.95%(65.88%,91.40%)
3+100%(88.78%, 100%)

Table 4:

Manual Microscopy Reads vs. Manual Digital Reads – Site 2 4x4 Tables for Pathologist 1

:

:

:

.

{6}------------------------------------------------

Manual Microscope Read
Pathologist 201+2+3+TOTAL
ManualDigitalRead01501016
1+2124018
2+0129030
3+0033336
TOTAL17133733100
ScorePAExact 95% CI
0, 1+96.67%(82.78%,99.92%)
2+78.38%(61.79%,90.17%)
3+100%(89.42%, 100%)

Table 5:

Manual Microscopy Reads vs. Manual Digital Reads - Site 2 4x4 Tables for Pathologist 2

Manual Microscope Read
01+2+3+TOTAL
ManualDigitalRead01110012
1+380011
2+01325240
3+0063137
TOTAL14223133100
ScorePAExact 95% CI
0, 1+63.89%(46.22%,79.18%)
2+80.65%(62.53%,92.55%)
3+93.94%(79.77%,99.26%)

Table 6: Manual Microscopy Reads vs. Manual Digital Reads - Site 2 4x4 Tables for Pathologist 3

In addition, a precision study was conducted at one clinical site to assess the intra-instrument and inter-instruments precision for manual digital reads. A subset of thirty (30) slides from the comparison with conventional microscopy study was used. The slides were selected to provide an equal distribution of HER2 scores in the trichotomous categorization of the HER2 scores combining 0 and 1+, and leaving 2+ and 3+ uncombined. For comparison, a pathologist reads first the thirty (30) glass-slides three times using the same conventional microscope. Then, the same pathologist reads the images of the thirty (30) glass-slides three times on a computer monitor, re-scanned each times on the same instrument. Finally, the same pathologist reads the images of the thirty (30) glass-slides another three times on a computer monitor, this time the slides were re-scanned each time on a different instrument. The wash-out period between reads were at least one week and the slides were randomized between reads.

{7}------------------------------------------------

The statistical analysis is presented for intra- and inter-instruments variations, using Percent Agreement (PA) with a 95% Confidence Interval (CI). The statistical analysis is provided for a trichotomous categorization of the HER2 scores combining 0 and 1+, and leaving 2+ and 3+ uncombined. Table 7 shows the overall agreements across all slides and all pair wise comparisons of the different reads.

PA95% CI
Intra-Instrument100%(95.98, 100)
Inter-Instruments95.6%(89.01, 98.78)

Table 7: Overall Agreements for the Precision of Manual Digital Reads.

The intra-instrument intra-pathologist agreement for manual digital reads is 100% with a 95% confidence interval from 95.98% to 100%. The inter-instruments intra-pathologist agreement for manual digital reads is 95.6% with a 95% confidence interval from 89.01% to 98.78%.

ਰੇ Conclusion

In summary, as results of the above evaluations for comparing to the predicate device, the subject device, OLYMPUS VS800 System do not incorporate any significant changes that could affect the safety or effectiveness of the device.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird with outstretched wings, rendered in a simple, abstract design.

10903 New Hampshire Avenue Silver Spring, MD 20993

AUG 2 1 2012

Olympus America, Inc. c/o Ms. Sheri L. Musgnung Associate Manager, Regulatory Affairs 3500 Corporate Parkway P.O. Box 610 Center Valley, PA 18034-0610

Re: K111914

Trade/Device Name: Olympus Virtual Slide System, VS800 System, VS800 HER2 Manual Read (MR) application Regulation Number: 21 CFR 864.1860 Regulation Name: Immunohistochemistry reagents and kits Regulatory Class: Class II Product Code: OEO Dated: August 13, 2012 Received: August 15, 2012

Dear Ms. Musgnung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{9}------------------------------------------------

Page 2 - Ms. Sheri L. Musgnung

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mana M. Chan

Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

Indications for Use

1111914 510(k) Number:

Virtual Slide System, Olympus VS800 System, VS800 HER2 MR Application Device Name: Indications for use:

The VS800 system is an automated digital slide creation, management, and viewing system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape.

The VS800HER2 Manual Read (MR) of digital slide application is intended for use as an aid to the pathologist in the detection and semi-quantitative measurement of HER2 by manual examination of the digital slide of formalin-fixed, paraffin-embedded and neoplastic tissue IHC stained for HER2 receptors on a computer monitor. HER2 results are indicated for use as an aid in the management, prognosis and prediction of therapy outcomes of breast cancer.

The VS800HER2 MR of digital slide application is intended for use as an accessory to the DakoHercepTest to aid the pathologist in the detection and semi-quantitative measurement of HER2 by manual examination of the digital slide of formalin-fixed, paraffin-embedded and neoplastic tissue immunohistochemically stained for HER2 receptors on a computer monitor. When used with the Dako Hercep Test, it is indicated for use as an aid in the assessment of breast cancer patients for whom HERCEPTIN® (Trastuzumab) treatment is being considered.

Note: The actual correlation of the Dako Hercep Test to the Herceptin® clinical outcome has not been established.

Prescription Use_ (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use_ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Senk R. Slin

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K111914

§ 864.1860 Immunohistochemistry reagents and kits.

(a)
Identification. Immunohistochemistry test systems (IHC's) are in vitro diagnostic devices consisting of polyclonal or monoclonal antibodies labeled with directions for use and performance claims, which may be packaged with ancillary reagents in kits. Their intended use is to identify, by immunological techniques, antigens in tissues or cytologic specimens. Similar devices intended for use with flow cytometry devices are not considered IHC's.(b)
Classification of immunohistochemistry devices. (1) Class I (general controls). Except as described in paragraphs (b)(2) and (b)(3) of this section, these devices are exempt from the premarket notification requirements in part 807, subpart E of this chapter. This exemption applies to IHC's that provide the pathologist with adjunctive diagnostic information that may be incorporated into the pathologist's report, but that is not ordinarily reported to the clinician as an independent finding. These IHC's are used after the primary diagnosis of tumor (neoplasm) has been made by conventional histopathology using nonimmunologic histochemical stains, such as hematoxylin and eosin. Examples of class I IHC's are differentiation markers that are used as adjunctive tests to subclassify tumors, such as keratin.(2) Class II (special control, guidance document: “FDA Guidance for Submission of Immunohistochemistry Applications to the FDA,” Center for Devices and Radiologic Health, 1998). These IHC's are intended for the detection and/or measurement of certain target analytes in order to provide prognostic or predictive data that are not directly confirmed by routine histopathologic internal and external control specimens. These IHC's provide the pathologist with information that is ordinarily reported as independent diagnostic information to the ordering clinician, and the claims associated with these data are widely accepted and supported by valid scientific evidence. Examples of class II IHC's are those intended for semiquantitative measurement of an analyte, such as hormone receptors in breast cancer.
(3) Class III (premarket approval). IHC's intended for any use not described in paragraphs (b)(1) or (b)(2) of this section.
(c)
Date of PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the Federal Food, Drug, and Cosmetic Act is required for any device described in paragraph (b)(3) of this section before this device may be commercially distributed. See § 864.3.