The summary describes a standard pulse oximetry sensor and does not mention any AI or ML components or capabilities.
No
The device is for monitoring vital signs and does not directly provide therapy or treatment.
Yes.
The device measures functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate, which are physiological parameters used to assess a patient's condition for potential diagnosis or monitoring of effectiveness of treatment.
No
The device description explicitly states that the device is a pulse oximetry sensor that measures tissue oxygenation non-invasively through infrared emitters and detectors, indicating it is a hardware device.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The description clearly states that the Masimo LNOP/M-LNCS/LNCS Multisite-L Oximetry Sensors are used for continuous noninvasive monitoring of oxygen saturation and pulse rate. This is done by measuring tissue oxygenation through infrared emitters and detectors applied to the patient's skin.
- No Sample Analysis: The device does not analyze samples taken from the body. It directly measures physiological parameters on the body.
Therefore, this device falls under the category of non-invasive medical devices rather than in vitro diagnostics.
N/A
Intended Use / Indications for Use
The Masimo LNOP/M-LNCS/LNCS Multisite-L Oximetry Sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoqlobin (SpO2) and pulse rate (measured by an SpO2 sensor) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.
Product codes
DQA
Device Description
The Masimo LNOP/M-LNCS/LNCS Multisite-L Oximetry Sensors are pulse oximetry sensors. They measure tissue oxygenation non-invasively through infrared emitters and detectors. They are fully compatible for use with Masimo SET and Masimo Rainbow SET instruments.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
adult, pediatric, infant, and neonatal patients
Intended User / Care Setting
hospitals, hospital-type facilities, mobile, and home environments.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Functional/Safety Testing: The following non-clinical testing was conducted to verify that the Masimo LNOP/M-LNCS/LNCS Multisite-L Oximetry Sensors met all design specifications: Ambient light rejection, noise rejection, electrostatic discharge, voltage breakdown, bend testing, off-patient detection, tape strip life cycle, current transfer ratio, light piping, electrocautery noise rejection, sensor skin temperature, moisture resistance and cleaning, drop, storage and operating conditions, bend and pull.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.2700 Oximeter.
(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).
0
11888
Masimo M-LNCS/LNCS/LNOP Multisite-L Sensors
Image /page/0/Picture/2 description: The image shows the Masimo Corporation logo and contact information. The logo is a stylized "M" with the word "Masimo" next to it. The text "K111888 - Additional Information" is at the top of the image. The address "Forty Parker, Irvine, CA 92618" is below the company name.
SEP 2 3 2011
510(k) Summary
| Submitter: | Masimo Corporation
40 Parker
Irvine, CA 92618
(949) 297-7000
Fax: (949) 297-7592 |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | David Collette
Senior Manager, Regulatory Affairs |
| Date Prepared: | August 24, 2011 |
| Trade name: | Masimo LNOP/M-LNCS/LNCS Multisite-L Oximetry Sensors |
| Classification
Name: | Oximeter Sensor, (21 CFR 870.2700) |
| Predicate
Device: | K101896, LNCS/M-LNCS Oximetry Sensors |
| Device
Description: | The Masimo LNOP/M-LNCS/LNCS Multisite-L Oximetry
Sensors are pulse oximetry sensors. They measure tissue
oxygenation non-invasively through infrared emitters and
detectors. They are fully compatible for use with Masimo
SET and Masimo Rainbow SET instruments. |
| Indications for
Use: | The Masimo LNOP/M-LNCS/LNCS Multisite-L Oximetry
Sensors are indicated for the continuous noninvasive
monitoring of functional oxygen saturation of arterial
hemoqlobin (SpO2) and pulse rate (measured by an SpO2
sensor) for use with adult, pediatric, infant, and neonatal
patients during both no motion and motion conditions, and
for patients who are well or poorly perfused in hospitals,
hospital-type facilities, mobile, and home environments. |
:
1
Masimo M-LNCS/LNCS/LNOP Multisite-L Sensors
. . .
K111888 - Additional Information
| | MASIMO CORPORATION
Forty Parker
Irvine, CA 92618 |
--------------------------------- | -------------------------------------------------------- |
---|
.
.
:
Comparative Analysis: | The Masimo LNOP/M-LNCS/LNCS Multisite-L Oximetry Sensors are substantially equivalent to the predicate device in the design, principles of operation, and performance. The subject and predicate device operate on identical principles of non-invasive optical assessment of tissue oxygenation using emitters and detectors. |
---|---|
Functional/Safety Testing: | The following non-clinical testing was conducted to verify that the Masimo LNOP/M-LNCS/LNCS Multisite-L Oximetry Sensors met all design specifications: Ambient light rejection, noise rejection, electrostatic discharge, voltage breakdown, bend testing, off-patient detection, tape strip life cycle, current transfer ratio, light piping, electrocautery noise rejection, sensor skin temperature, moisture resistance and cleaning, drop, storage and operating conditions, bend and pull. |
Conclusion: | The Masimo LNOP/M-LNCS/LNCS Multisite-L Oximetry Sensors are substantially equivalent to the predicate device. |
:
. . . .
.
2
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines forming the staff and a snake winding around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MID-20993-0002
Mr. Anil Bhalani Director. Regulatory Affairs Masimo Corporation 40 and 50 Parker Irvine, California 92618
2 3 2011
Re: K111888
Trade/Device Name: Masimo LNOP/M-LNCS/LNCS Multisite-L Oximetry Sensors Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: August 24. 2011 Received: August 25, 2011
Dear Mr. Bhalaní:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
3
Page 2 - Mr. Bhalani
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality svstems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /uem115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safetv /ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices /ResourcesforYou/Industry/default.htm.
Sincerely yours,
Rh for
Anthony D. Watson. B.S.. M.S.. M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
4
Appendix 2 -- Indications for Use Statement
510(k) Number (if known):
Masimo LNOP/M-LNCS/LNCS Multisite-L Oximetry Sensors Device Name:
Indications for Use:
The Masimo LNOP/M-LNCS/LNCS Multisite-L Oximetry Sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use_ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L Schutten
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K211188