(82 days)
The Joint Biological Agent Identification and Diagnostic System (JBAIDS) Influenza A & B Detection Kit is intended for use on the JBAIDS instruments, for the in vitro qualitative detection of Influenza A and Influenza B viral nucleic acids isolated and purified from nasopharyngeal swab (NPS) and nasopharyngeal wash (NPW) specimens from human patients with signs and symptoms of respiratory infection. The JBAIDS Influenza A & B Detection Kit contains reverse transcriptase real-time polymerase chain reaction (rRT-PCR) assays that target the Matrix protein gene of Influenza A viruses, and the Non-structural protein gene of Influenza B viruses. This kit is not intended to detect Influenza C viruses.
Test results are to be used in conjunction with other clinical and epidemiological information. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions.
Performance characteristics for detection of influenza A were established when 2009 H1N1 Influenza, Influenza A H1N1, and Influenza A H3N2 were the predominant influenza A viruses in circulation. Due to low seasonal prevalence, performance characteristics for detection of seasonal Influenza A/H1 were established primarily with retrospective and surrogate clinical specimens. When other influenza A viruses are present, performance characteristics may vary.
All users, analysts, and any person reporting diagnostic results from use of this device should be trained to perform and interpret the results from this procedure by JBAIDS instructors or designees prior to use. Use of this device is limited to designated Department of Defense (DoD) laboratories equipped with the JBAIDS instruments.
If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a biosafety laboratory (BSL) 3+ facility is available to receive and culture specimens.
The JBAIDS Influenza A & B Detection Kit is a rRT-PCR test kit, which, when used with the JBAIDS instrument and software, allows the qualitative in vitro detection of influenza A and B viral RNA. These two assays have been optimized as freeze-dried assays with primer and fluorescent-probe sets for the detection of influenza A and B viral RNA. In particular, the influenza A assay targets a region of the matrix gene specific to the influenza A virus genera, and the influenza B assay targets a region of the non-structural gene specific to the influenza B virus genera. The tests are performed using the previously FDA-cleared JBAIDS instrument and software. A human gene target assay (RNAseP target) will be used as an inhibition and extraction control.
Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
For the JBAIDS Influenza A & B Detection Kit, the primary performance metrics are Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) compared to a reference method. The document does not explicitly state pre-defined acceptance criteria (e.g., "PPA must be >90%"). However, the reported performance is presented to demonstrate substantial equivalence, implying that these agreement percentages were considered acceptable for clearance.
| Metric | Sample Type | Purification Kit | Reported Device Performance |
|---|---|---|---|
| Influenza A - PPA | NPW | Platinum Path | 100.0% (65/65) |
| NPW | MagNA Pure | 100.0% (39/39) | |
| NPW | Combined | 100.0% (104/104) | |
| NPS | Platinum Path | 100.0% (42/42) | |
| NPS | MagNA Pure | 100.0% (20/20) | |
| NPS | Combined | 100.0% (62/62) | |
| Influenza A - NPA | NPW | Platinum Path | 99.5% (213/214) |
| NPW | MagNA Pure | 98.8% (170/172) | |
| NPW | Combined | 99.2% (383/386) | |
| NPS | Platinum Path | 100.0% (91/91) | |
| NPS | MagNA Pure | 99.4% (160/161) | |
| NPS | Combined | 99.6% (251/252) | |
| Influenza B - PPA | NPW | Platinum Path | 95.1% (39/41) |
| NPW | MagNA Pure | 93.5% (29/31) | |
| NPW | Combined | 94.4% (68/72) | |
| NPS | Platinum Path | 94.7% (18/19) | |
| NPS | MagNA Pure | 100.0% (6/6) | |
| NPS | Combined | 96.0% (24/25) | |
| Influenza B - NPA | NPW | Platinum Path | 99.6% (237/238) |
| NPW | MagNA Pure | 98.3% (177/180) | |
| NPW | Combined | 99.0% (414/418) | |
| NPS | Platinum Path | 100.0% (114/114) | |
| NPS | MagNA Pure | 100.0% (175/175) | |
| NPS | Combined | 100.0% (289/289) | |
| Influenza A/H1 (Archived) - PPA | NPS | Combined | 100% (30/30) |
| Influenza A/H1 (Archived) - NPA | NPS | Combined | 100% (21/21) |
| Influenza A (Surrogate) - PPA | NPW | Not Specified | 98.1% (53/54) |
| Influenza A (Surrogate) - NPA | NPW | Not Specified | 100.0% (8/8) |
| Influenza A (Surrogate) - PPA | NPS | Not Specified | 100.0% (59/59) |
| Influenza A (Surrogate) - NPA | NPS | Not Specified | 100.0% (7/7) |
| Influenza B (Surrogate) - NPA | NPW | Not Specified | 98.5% (66/67) |
| Influenza B (Surrogate) - NPA | NPS | Not Specified | 100.0% (65/65) |
| Reproducibility (Flu A, ≥ LoD) | NPS | Both Kits | 99% (179/180) |
| NPW | Both Kits | 99% (179/180) | |
| Reproducibility (Flu B, ≥ LoD) | sNPS | Both Kits | 100% (180/180) |
| sNPW | Both Kits | 100% (180/180) |
2. Sample Sizes and Data Provenance
- Clinical Performance (Prospective Study):
- Test Set Sample Size: 804 valid specimens (314 NPS, 490 NPW).
- Data Provenance: Prospective study conducted at 5 geographically separated military clinical sites in the U.S. during the 2010-2011 influenza season (December 2010 to April 2011).
- Testing of Preselected Archived Specimens:
- Test Set Sample Size: 51 NPS specimens (30 known positive seasonal Influenza A/H1, 21 influenza-negative).
- Data Provenance: Retrospective, pre-selected archived clinical NPS specimens. Country of origin not explicitly stated but implied to be US-based given the context of US military sites.
- Testing of Surrogate Clinical Specimens:
- Test Set Sample Size: 136 individual influenza-negative clinical specimens (68 NPS, 68 NPW) that were spiked. 128 produced valid test results for analysis.
- Data Provenance: Contrived clinical samples created by spiking residual influenza-negative NPS and NPW samples. Origin of the original negative specimens is clinical, likely from the US military sites or similar sources.
- Analytic Studies (LoD, Inclusivity, Exclusivity, Reproducibility):
- These studies used distinct sample sizes, generally involving multiple replicates (e.g., 20 replicates for LoD, 30-90 results per spike level for reproducibility). The samples were either live, quantified virus strains or simulated NPS/NPW matrices spiked with specific organisms.
3. Number of Experts and Qualifications for Ground Truth
The document does not mention the use of human experts to establish ground truth for the test sets in the clinical performance study.
- Clinical Performance (Prospective, Archived, and Surrogate Studies): The ground truth was established by a comparator method, the CDC rRT-PCR Flu Panel influenza A and influenza B assays. This is a laboratory-based reference standard, not a human expert consensus.
- Analytic Studies: The ground truth for these (e.g., LoD, inclusivity, exclusivity) was based on the known presence and concentration of specific viruses or bacteria in the spiked samples.
4. Adjudication Method
There is no mention of an adjudication method in the studies described, as the comparator method (CDC rRT-PCR Flu Panel) served as the direct reference standard. The results of the JBAIDS kit were compared to this objective method.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study was performed or described. This device is an in-vitro diagnostic (IVD) assay designed for laboratory use, not for direct interpretation by human readers in the same way an image-based diagnostic AI might be. Therefore, the concept of human readers improving with AI assistance is not applicable in this context. The device's results are intended to be interpreted by trained laboratory personnel.
6. Standalone Performance
Yes, a standalone performance study was done. The entire set of clinical and analytical performance evaluations (PPA, NPA, LoD, inclusivity, exclusivity, reproducibility) directly measures the performance of the JBAIDS Influenza A & B Detection Kit algorithm and associated testing procedure in isolation (without human-in-the-loop performance influencing the assay's output). The device's software automatically analyzes fluorescence amplification curves and reports results (positive, negative, or uncertain).
7. Type of Ground Truth Used
- Clinical Performance (Prospective, Archived, Surrogate): The ground truth was based on the results from a legally marketed comparator device: the CDC rRT-PCR Flu Panel influenza A and influenza B assays. This is a highly sensitive and specific laboratory-based molecular diagnostic method, often considered a "gold standard" for viral nucleic acid detection.
- Analytical Studies (LoD, Inclusivity, Exclusivity, Reproducibility): The ground truth was based on known concentrations of specific viral or bacterial strains in the spiked samples.
8. Sample Size for the Training Set
The document does not specify a separate training set for the JBAIDS Influenza A & B Detection Kit. As an IVD based on rRT-PCR, the "training" analogous to machine learning models is typically inherent in the assay design and optimization process (e.g., primer and probe design, reaction condition optimization), rather than a distinct data-driven training phase on labeled samples as seen in AI/ML products. The mentioned studies focus on the validation of the optimized kit.
9. How the Ground Truth for the Training Set was Established
As no explicit training set is described in the context of an AI/ML model, this question is not directly applicable. If one considers the development and optimization of the rRT-PCR assays as an analogous process, the "ground truth" during this phase would involve using well-characterized viral isolates and clinical samples with confirmed presence/absence of influenza to optimize the assay's sensitivity and specificity parameters.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains a sequence of handwritten characters, which appear to be alphanumeric. The sequence starts with a capital letter 'K', followed by two instances of the number '1'. After that, the number '7' appears twice, and the sequence concludes with the number '5'. The characters are written in a simple, clear style, and they are the primary focus of the image.
SEP 1 3 2011
510(k) Summary JBAIDS Influenza A & B Detection Kit
- According to the requirements of 21 CFR 807.92, the following information Introduction: provides sufficient detail to understand the basis for a determination of substantial equivalence.
- U.S. Army Medical Materiel Development Activity Submitted by: Division of Regulated Activities and Compliance 1430 Veterans Drive Fort Detrick, MD
- Robert Miller Ph.D., RAC Primary Contact: Director, Division of Regulated Activities and Compliance Telephone: 301-619-0317 Facsimile: 301-619-0197 e-mail: usamrmcregulatoryaffairs@us.army.mil
- Secondary Contact: Patricia Beverly, RAC Division of Regulated Activities and Compliance Telephone: 301-619-2980 Facsimile: 301-619-0197
e-mail: patricia.m.beverly@us.army.mil or usamrmcregulatoryaffairs(@amedd.army.mil
Beth Lingenfelter Technical Contact: Director of Regulatory Affairs, Idaho Technology, Inc. Telephone: 801-736-6354, ext 407 Facsimile: 801-588-0507 e-mail: bethl@idahotech.com
Date Prepared: August, 2011
Device Name: Trade Name:
JBAIDS Influenza A & B Detection Kit
Common Name:
Real-time PCR assay for detection of Influenza A and Influenza B
Classification Name:
Respiratory Viral Panel Multiplex Nucleic Acid Assay (CFR 866.3980)
{1}------------------------------------------------
Intended Use
The Joint Biological Agent Identification and Diagnostic System (JBAIDS) Influenza A & B Detection Kit is intended for use on the JBAIDS instruments, for the in vitro qualitative detection of Influenza A and Influenza B viral nucleic acids isolated and purified from nasopharyngeal swab (NPS) and nasopharyngeal wash (NPW) specimens from human patients with signs and symptoms of respiratory infection. The JBAIDS Influenza A & B Detection Kit contains reverse transcriptase real-time polymerase chain reaction (rRT-PCR) assays that target the Matrix protein gene of Influenza A viruses, and the Non-structural protein gene of Influenza B viruses. This kit is not intended to detect Influenza C viruses.
Test results are to be used in conjunction with other clinical and epidemiological information. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions.
Performance characteristics for detection of influenza A were established when 2009 H1N1 Influenza, Influenza A H1N1, and Influenza A H3N2 were the predominant influenza A viruses in circulation. Due to low seasonal prevalence, performance characteristics for detection of seasonal Influenza A/H1 were established primarily with retrospective and surrogate clinical specimens. When other influenza A viruses are present, performance characteristics may vary.
All users, analysts, and any person reporting diagnostic results from use of this device should be trained to perform and interpret the results from this procedure by JBAIDS instructors or designees prior to use. Use of this device is limited to designated Department of Defense (DoD) laboratories equipped with the JBAIDS instruments.
If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a biosafety laboratory (BSL) 3+ facility is available to receive and culture specimens.
Device Description
The JBAIDS Influenza A & B Detection Kit is a rRT-PCR test kit, which, when used with the JBAIDS instrument and software, allows the qualitative in vitro detection of influenza A and B viral RNA. These two assays have been optimized as freeze-dried assays with primer and fluorescent-probe sets for the detection of influenza A and B viral RNA. In particular, the influenza A assay targets a region of the matrix gene specific to the influenza A virus genera, and the influenza B assay targets a region of the non-structural gene specific to the influenza B virus genera. The tests are performed using the previously FDA-cleared JBAIDS instrument and software. A human gene target assay (RNAseP target) will be used as an inhibition and extraction control.
{2}------------------------------------------------
Assay Principle
Before testing. NPS or NPW specimens are purified using Idaho Technology's 1-2-3™ Platinum Path Sample Purification Kit or the Roche MagNA Pure Compact Nucleic Acid Isolation Kit I. The resulting purified sample is added to Unknown reagent vials and a Sample Control reagent vial, along with reconstitution buffer. When viral RNA is present, a fragment of influenza A or B viral RNA is transcribed and amplified. The amplicon is detected by fluorescence using a specific hydrolysis probe. Each probe is labeled on one end with a fluorescent reporter morety (6-carboxyfluorescein (6-FAM)) and elsewhere with a quencher moiety (carboxy tetramethylrhodamine (TAMRA)). When the probe is intact, the quencher absorbs the light emitted by the reporter moiety. During PCR, the probe hybridizes to the target sequence before the exonuclease activity of Tag polymerase hydrolyzes the probe, separating the fluorophore from the quencher and permitting detection of the fluorescent signal generated by the reporter. The fluorescent signal increases as additional templates are amplified and more probes are hydrolyzed.
JBAIDS Software analyzes the fluorescence amplification curves and reports results as positive, negative, or uncertain. A failure of the Positive or Negative Control will result in the entire run being called invalid. Failure of the Sample Control yields a result of "sample control failure" when the associated sample has a negative result for the target assay. Retesting is required to resolve uncertain, invalid or sample control failure results.
Substantial Equivalence
The JBAIDS Influenza A & B Detection System is substantially equivalent to other products in commercial distribution intended for similar use. The JBAIDS instrument has been previously cleared under K051713.
The JBAIDS Influenza A & B Detection Kit is substantially equivalent to the CDC Human Influenza Virus real-time RT-PCR Detection and Characterization Panel, which was cleared on September 30, 2008 under 510k# K080570.
The CDC rRT-PCR Flu Panel is a panel of oligonucleotide primers and dual-labeled hydrolysis probes for the qualitative detection and differentiation of influenza viruses. The panel can be used to test upper respiratory tract specimens (including NPS, NS, TS, NA, NW, NPS/TS) and lower respiratory tract specimens (including BAL, BW, TA, sputum, and lung tissue) or virus culture.
Prior to testing, the samples are purified with one of four commercially available sample purification methods (see chart below). A PCR master mix is prepared by combining the appropriate quantities of primers and probes from the rRT-PCR Flu Panel with a commercially available reverse transcription enzyme master mix. The PCR master mix is aliquoted into a 96 well plate followed by addition of the purified samples and controls. The prepared plates are placed on an ABI 7500 Fast Dx Real-Time PCR instrument and thermocycled according to the cycling conditions described in the rRT-PCR Flu Panel product insert.
Each run on the ABI includes a no-template control (nuclease-free water), a seasonal influenza virus control (influenza A/H1, A/H3 and influenza B with cultured human cells), an A/H5 virus
{3}------------------------------------------------
control (noninfectious reassortant influenza A/H5 virus with cultured human cells) and a human specimen control (cultured human cells). The human specimen control is extracted with the test samples and is intended to ensure that the extraction process was properly performed. The notemplate, seasonal influenza virus and A/H5 virus control templates are added to the PCR plate prior to thermocycling.
At the conclusion of the run, the operator is required to set a baseline for all assays included in the run. The operator is then required to interpret the test results for each sample based on the results of the control and the Ct values for each sample.
| Element | JBAIDS Influenza A & B Detection Kit | CDC rRT-PCR Flu Panel (K080570) |
|---|---|---|
| Technology | Real-time PCR using hydrolysis probes | Same |
| Viruses Detected | Qualitative in vitro detection of influenza Aand influenza B RNA | SameSee below for differences |
| Specimen Types | Nasopharyngeal swabs | SameSee below for differences |
| ExtractionMethods | Roche MagNA Pure Compact Nucleic AcidIsolation Kit I | SameSee below for differences |
Table 1. Similarities Between the JBAIDS Influenza A & B Detection Kit and the CDC rRT-PCR Flu Panel
Table 2. Differences Between the JBAIDS Influenza A & B Detection Kit and the CDC rRT-PCR Flu Panel
| Element | JBAIDS Influenza A & B Detection Kit | CDC rRT-PCR Flu Panel (K080570) |
|---|---|---|
| Viruses Detected | Does not subtype Influenza A viruses | Differentiation of Influenza A/H1, A/H3 andA/H5 (Asian lineage) |
| Specimen Types | Nasopharyngeal washes | Upper respiratory tract specimens (includingNPS, NS, TS, NA, NW, NPS/TS) and lowerrespiratory tract specimens (including BAL,BW, TA, sputum, and lung tissue) and virusculture |
| RequiredInstrumentation | JBAIDS instrument | Applied Biosystems 7500 Fast Dx Real-timePCR instrument with SDS software v 1.4 |
| Interpretation ofTest Results | Automated analysis of test results and controls | User required to interpret test and controlresults |
| Enzyme MasterMix | Assays come in freeze-dried single use vialsthat include all components of master mix | Invitrogen SuperScript™ III Platinum® One-Step Quantitative RT-PCR Kits |
| Reagent Storage | Reagents are stored at room temperature | Reagents are stored at ≤ -15°C |
| ExtractionMethods | IT 1-2-3TM Platinum Path Sample PurificationKitRoche MagNA Pure Compact Nucleic AcidIsolation Kit I | QIAamp® Viral RNA Mini Kit, QIAGENRNeasy® Mini Kit, or Roche MagNA PureTNA Kit |
{4}------------------------------------------------
Summary of Performance Data
Clinical Performance
The clinical performance of the JBAIDS Influenza A & B Detection Kit was evaluated during a prospective study at 5 geographically separated military clinical sites over the 2010-2011 influenza season (December 2010 to April 2011). Subjects with signs and/or symptoms of influenza-like illness were enrolled. Upon obtaining informed consent, NPS and NPW specimens were collected for JBAIDS and comparator testing. A total of 804 valid specimens were analyzed at the five study sites; 314 NPS and 490 NPW specimens. Table 3 provides a summary of demographic information for the 804 subjects that participated in the prospective study.
| Overall | Site 1 | Site 2 | Site 3 | Site 4 | Site 5 | ||
|---|---|---|---|---|---|---|---|
| NPS | 314 | 51 | 207 | 56 | 0 | 0 | |
| NPW | 490 | 325 | 0 | 0 | 119 | 46 | |
| Total | 804 | 376 | 207 | 56 | 119 | 46 | |
| Sex | Female | 410 (51%) | 192 (50.9%) | 122 (58.9%) | 23 (41.1%) | 57 (47.9%) | 16 (34.8%) |
| Male | 394 (49%) | 185 (49.1%) | 85 (41.1%) | 33 (58.9%) | 62 (52.1%) | 30 (65.2%) | |
| Age | Avg. | 26.5 | 23.4 | 24.5 | 30.3 | 23.1 | 31.0 |
| Median | 24.0 | 24.0 | 18.0 | 27.5 | 17.0 | 27.0 | |
| Min | 0.5 | 0.5 | 0.5 | 2.0 | 0.5 | 18.0 | |
| Max | 92.0 | 92.0 | 69.0 | 81.0 | 68.0 | 62.0 |
Table 3. Demographic Summary for the JBAIDS Influenza A & B Detection Kit Prospective Study
8 0.5 was used for all ages under 1 year for these calculations.
Of the 804 prospective specimens, successful results were obtained for 97% (778/804) of these specimens on the first attempt (Site 1: 356/376 =95%; Site 2: 203/207 =98%; Site 3: 56/56 =100%; Site 4: 118/119 =99%; Site 5: 45/46 =98%). The remaining 3% (26/804) required retesting: "Invalid"(16/26), "Uncertain"(0/26), "SC Failure"(5/26), positive for both the Flu A and Flu B assays (2/26), or were re-extracted and retested due to user labeling error (3/26) (20 samples from Site 1; 4 samples from Site 2; 0 samples from Site 3; 1 sample from Site 4; and 1 sample from Site 5). All 26 samples were resolved upon a subsequent retest.
June, 2011
{5}------------------------------------------------
Nucleic acid from each specimen was isolated using either the IT I-2-3 Platinum Path Sample Purification Kit (manual sample processing) or the Roche MagNA Pure Compact Nucleic Acid Isolation Kit I (automated sample processing) and tested with the JBAIDS Influenza A & B Detection Kit. The performance of the JBAIDS Influenza A & B Detection Kit was evaluated by comparing the JBAIDS test results with the comparator/reference method. The reference method was the CDC rRT-PCT Flu Panel influenza A and influenza B assays. Specimens collected at each study site were split into single use aliquots for testing with the JBAIDS Influenza A & B Detection Kit (performed at each site) and the CDC rRT-PCR Flu Panel influenza A and influenza B assays (performed at Idaho Technology) to determine Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA). PPA was calculated as 100% x (TP / (TP + FN)) and NPA was calculated as 100% x (TN / (TN + FP)). The exact binomial twosided 95% confidence interval was calculated for both performance measures. The results are summarized in Table 1.
| InfluenzaAssay | SampleType | PurificationKit | PPA | 95% CI | NPA | 95% CI |
|---|---|---|---|---|---|---|
| Flu A | NPW | Platinum Path | 65/65100.0% | 94.5-100% | 213/21499.5% | 97.4-100% |
| NPW | MagNA Pure | 39/39100.0% | 90.8-100% | 170/17298.8% | 95.9-99.9% | |
| NPW | Combined | 104/104100.0% | 96.5-100% | 383/38699.2% | 97.8-99.8% | |
| Flu A | NPS | Platinum Path | 42/42100.0% | 91.6-100% | 91/91100.0% | 96.0-100% |
| NPS | MagNA Pure | 20/20100.0% | 83.2-100% | 160/16199.4% | 95.6-100% | |
| NPS | Combined | 62/62100.0% | 94.2-100% | 251/25299.6% | 97.8-100% | |
| Flu B | NPW | Platinum Path | 39/4195.1% | 83.5-99.4 | 237/23899.6% | 97.7-100% |
| NPW | MagNA Pure | 29/3193.5% | 78.6-99.2% | 177/18098.3% | 95.2-99.7% | |
| NPW | Combined | 68/7294.4% | 86.4-98.5% | 414/41899.0% | 97.6-99.7% | |
| Flu B | NPS | Platinum Path | 18/1994.7% | 74.0-99.9% | 114/114100.0% | 96.8-100% |
| NPS | MagNA Pure | 6/6100.0% | 54-100% | 175/175100.0% | 97.9-100% | |
| NPS | Combined | 24/2596.0% | 79.7-99.9% | 289/289100.0% | 98.7-100% |
Table 1. JBAIDS Influenza A & B Detection Kit Prospective Clinical Performance Summary
Seasonal influenza A/H1 virus was not circulating during the 2010-2011 influenza season (http://www.cdc.gov/flu/) and was not detected during the prospective clinical study of the JBAIDS Influenza A & B Detection Kit. To supplement the results of the clinical study, an evaluation of preselected archived samples was performed. Due to the limited availability of archived specimens, the clinical performance was further supplemented with surrogate clinical contrived specimens.
Testing of Preselected Archived Specimens
Additional testing of pre-selected archived clinical NPS specimen was performed at two different clinical study sites to supplement the prospective clinical testing data. Because it is possible that the archived samples had been misidentified or had degraded during storage or previous handling, the presence or absence of Influenza A/H1 viral RNA was confirmed using "validation" PCR assays. The validation PCR assays were identical to
{6}------------------------------------------------
the comparator assays that were used for the prospective clinical evaluation study. A total of 51 NPS specimens were obtained and confirmed for testing: 30 known to be positive seasonal Influenza A/H1 specimens and 21 influenza-negative specimens. The specimens were split evenly for purification with the Platinum Path or MagNA Pure purification kits and then randomized such that the users performing the JBAIDS Influenza A & B Detection Kit testing were blinded as to the expected test result.
Table 5 presents the PPA and NPA for the archived clinical specimens. Data from both extraction kits are combined due to identical performance.
| A MILL SO T TOT TAX TECHNICA FI PERSONAL P | Computers programs and the state one of the comments of the contribution | ||||||
|---|---|---|---|---|---|---|---|
| InfluenzaAssay | SampleType | PPA | Percent | 95% CI | NPA Percent | 95% Cl | |
| Flu A | NPS | 30/30 | 100% | 88.4-100% 21/21 | i 100% | 83.4-100% . |
Table 2. Performance Summary of Seasonal Influenza A/H1 Archived Clinical Specimens
Due to the absence of seasonal Influenza A/H1 virus in circulation during the 2010-2011 influenza seasonal (http://www.cdc.gov/flu/) and lack of availability of archived NPW specimens for seasonal Influenza A/H1, contrived clinical samples (residual influenza negative NPS and NPW samples spiked with a known concentration of seasonal Influenza A/H1 virus) were used as a surrogate to further evaluate the performance of the JBAIDS Influenza A & B Detection Kit.
Testing of Surrogate Clinical Specimens
A total of 136 individual influenza-negative clinical specimens (68 NPS samples and 68 NPW samples) were spiked at a range of concentrations, including near the system limit of detection (LoD), as well as un-spiked, then randomized, and sent to two different clinical trial study sites for testing. Of the 136 surrogate samples included in this study, a valid JBAIDS test result was obtained for 128 samples (62 NPW and 66 NPS). The remaining 8 samples with invalid results could not be retested due to insufficient sample volume, and were not included in the analysis.
Table 3 presents the PPA and NPA for the surrogate clinical specimens. Half of the samples were extracted using the Platinum Path purification kit and half using the MagNA pure kit.
| InfluenzaAssay | SampleType | PPA | NPA | ||||
|---|---|---|---|---|---|---|---|
| TP/(TP+FN) | Percent | 95% CI | TN/(TN+FP) | Percent | 95% CI | ||
| Flu A | NPW | 53/54 | 98.1% | 90.1-100% | 8/8 | 100.0% | 63.1-100% |
| Flu A | NPS | 59/59 | 100.0% | 93.9-100% | 7/7 | 100.0% | 59.0-100% |
| Flu B | NPW | 0/0 | - | - | 66/67 | 98.5% | 92.0-100% |
| Flu B | NPS | 0/0 | - | - | 65/65 | 100.0% | 94.5-100% |
Table 3. Performance Summary of Seasonal Influenza A/H1 Surrogate Clinical Specimens
Out of 67 valid sample results, one (1) false positive Flu B result was obtained, almost certainly due to switching of the Flu A and Flu B capillaries in the run setup.
{7}------------------------------------------------
Analyses of the clinical data set, preselected archived specimens, and surrogate clinical specimens demonstrate that the JBAIDS Influenza A & B Detection Kit is a sensitive and specific test system for the detection of influenza A and influenza B viruses.
Selected Analytic Studies
Limit of Detection
The analytical sensitivity or Limit of Detection (LoD) for each target assay (Flu A and Flu B) was determined using both NPS and NPW samples spiked with live, quantified virus strains. The LoD is defined as the lowest concentration at which the target is consistently detected (detection is ≥95% of samples tested). Independent specimens from independent donors were spiked with each live virus strain and 20 replicates were tested at the LoD concentration. The LoD levels for representative virus strains detected by the JBAIDS Influenza A & B Detection Kit are listed in Table 4.
| Assay | Influenza Type | Strain | LoD(EID50/mL) |
|---|---|---|---|
| Flu A | Influenza A H1N1 | A/New Caledonia/20/1999 | 50 |
| Influenza A H3N2 | A/New York/55/2004 | 5 | |
| 2009 Influenza A H1N1 | A/New York/18/2009 | 50 | |
| Flu B | Influenza B | B/Ohio/1/ 2005 | 5 |
| Influenza B | B/Florida/ 7/2004 | 10 |
Table 4. LoD Concentrations for Representative Virus Strains Detected by the JBAIDS Influenza A & B Detection Kit
Inclusivity
The reactivity of the JBAIDS Influenza A & B Detection Kit assays was evaluated with an inclusivity panel consisting of 25 influenza A and 9 influenza B strains or isolates that represent the genetic, temporal, and geographic diversity of the influenza analytes. Each organism was tested in a simulated NPS sample matrix at or near the system LoD. Higher concentrations were tested if the analyte was not detected at the LoD. Each of the 34 influenza strains tested in this study was reactive with the appropriate JBAIDS influenza assays. Results from the inclusivity tested are presented in Table Error! No text of specified style in document, and Table 5. The concentration at which each strain was detected is indicated.
{8}------------------------------------------------
| InfluenzaType/Subtype | Strain | Lowest ConcentrationDetected |
|---|---|---|
| H2N2 (Avian) | A/chicken/Pennsylvania/298101-4/2004 | 0.5 TCID50/mL |
| H3N8 (Avian) | A/MAL/ALB/16/87 | 50 TCID50/mL |
| H4N8 (Avian) | A/chicken/Alabama/1975 | 5 EID50/mL |
| H5N1 (Avian) | A/DK/PA/4560069-9/06 | 50 TCID50/mL |
| H5N1 (Avian-Human Recombinant) | A/Vietnam/1203/2004(H5N1)-PR8 | 0.5 EID50/mL |
| H6N2 (Avian) | A/Chicken/CA/32213-1/2000 | 0.5 EID50/mL |
| H7N3 (Avian) | A/TY/UT/24721-10/95 | 500 TCID50/mL |
| H9N2 (Avian) | A/Turkey/Wisconsin/1966 | 0.5 EID50/mL |
| H3N8 (Canine) | A/canine/Florida/43/2004 | 5000 TCID50/mL |
| H3N8 (Equine) | A/Equine/Ohio/01/2009 | 50 TCID50/mL |
| H1N1 (Swine) | A/SW/GB/19582/92 | 50 TCID50/mL |
| H1N1 (Swine) | A/swine/Wisconsin/125/1997 | 500 TCID50/mL |
| H3N2 (Swine) | A/SW/IA/1/99 | 0.5 TCID50/mL |
| H1N1 (Human of swine lineage) | A/lowa/1/2006 | 500 TCID50/mL |
| H1N1 (Human of swine lineage) | A/Maryland/12/1991 | 5000 TCID50/mL |
| H7N2 (Human) | A/New York/107/2003 | 1 × 10 $^{-10}$ Dilution of Stock |
| Sesonal H1N1 (Human) | A/NWS/33 | 0.5 TCID50/mL |
| A/PR/8/34 | 5 TCID50/mL | |
| A/1/Denver/1/57 | 0.5 TCID50/mL | |
| Seasonal H3N2 (Human) | A/Aichi/2/68 | 50 TCID50/mL |
| A/Hong Kong/8/68 | 5 TCID50/mL | |
| A/Victoria/3/75 | 5 TCID50/mL | |
| 2009 swine lineage H1N1 (Human) | A/England/195/2009 | 5 TCID50/mL |
| A/Mexico/4108/2009 | 500 EID50/mL | |
| A/New York/18/2009 | 50 EID50/mL |
Table Error! No text of specified style in document.. Results of Influenza A Inclusivity
Table 5. Results of Influenza B Inclusivity
| Strain | Lowest ConcentrationDetected |
|---|---|
| B/Lee/40 | 5 TCID50/mL |
| B/Allen/45 | 50 TCID50/mL |
| B/GL/1739/54 | 0.5 TCID50/mL |
| B/Maryland/1/59 | 5 TCID50/mL |
| B/Taiwan/2/62 | 0.5 TCID50/mL |
| B/Hong Kong/5/72 | 0.5 TCID50/mL |
| B/Malaysia/2506/04 | 50 TCID50/mL |
| B/FL/04/06 | 50 TCID50/mL |
| B/Brigit | 0.5 TCID50/mL |
Exclusivity
.
1
The analytical specificity, or exclusivity, of the JBAIDS influenza assays was evaluated by testing simulated NPS samples spiked with high concentrations of influenza viruses (tens to thousands-fold higher than LoD). No cross-reactivity was observed when high
{9}------------------------------------------------
concentration of one influenza strain (e.g., influenza A) was tested with the alternate influenza assay (e.g., Flu B assay) at the concentrations listed in Table 6.
| JBAIDSInfluenza AssayTested | Type/Subtype (Host) | Strain | ConcentrationTested |
|---|---|---|---|
| Flu A | Influenza B (Human) | B/Lee/40 | 7.36E+03 TCID50/mL |
| B/Allen/45 | 1.00E+05 TCID50/mL | ||
| B/GL/1739/54 | 7.36E+03 TCID50/mL | ||
| B/Maryland/1/59 | 7.36E+03 TCID50/mL | ||
| B/Taiwan/2/62 | 4.54E+04 TCID50/mL | ||
| B/Hong Kong/5/72 | 7.36E+03 TCID50/mL | ||
| B/Malaysia/2506/04 | 5.09E+03 TCID50/mL | ||
| B/FL/04/06 | 1.50E+04 TCID50/mL | ||
| B/Brigit | 3.14E+04 TCID50/mL | ||
| Flu B | H2N2 (Avian) | A/chicken/Pennsylvania/298101-4/2004 | 3.16E+07 TCID50/mL |
| H3N8 (Avian) | A/MAL/ALB/16/87 | 1.72E+03 TCID50/mL | |
| H4N8 (Avian) | A/chicken/Alabama/1975 | 1.00E+08 EID50/mL | |
| H5N1 (Avian-HumanRecombinant) | A/Vietnam/1203/2004(H5N1)-PR8 | 3.16E+07 EID50/mL | |
| H5N1 (Avian) | A/DK/PA/4560069-9/06 | 1.00E+05 TCID50/mL | |
| H7N3 (Avian) | A/TY/UT/24721-10/95 | 3.06E+04 TCID50/mL | |
| H6N2 (Avian) | A/Chicken/CA/32213-1/2000 | 1.26E+07 EID50/mL | |
| H9N2 (Avian) | A/Turkey/Wisconsin/1966 | 5.60E+07 EID50/mL | |
| H3N8 (Canine) | A/canine/Florida/43/2004 | 1.00E+05 TCID50/mL | |
| H3N8 (Equine) | A/Equine/Ohio/01/2009 | 1.00E+05 TCID50/mL | |
| H1N1 (Swine) | A/swine/Wisconsin/125/1997 | 1.00E+05 TCID50/mL | |
| H1N1 (Swine) | A/SW/GB/19582/92 | 5.64E+03 TCID50/mL | |
| H3N2 (Swine) | A/SW/IA/1/99 | 1.41E+03 TCID50/mL | |
| H1N1 (Human of swinelineage) | A/Maryland/12/1991 | 1.00E+05 TCID50/mL | |
| H1N1 (Human of swinelineage) | A/Iowa/1/2006 | 1.00E+05 TCID50/mL | |
| H7N2 (Human) | A/New York/107/2003 | 30 µl of an unknownconcentration into 1mL | |
| Seasonal H1N1 (Human) | A/Brisbane/59/07 | 1.00E+05 TCID50/mL | |
| A1/FM/1/47 | 4.24E+03 TCID50/mL | ||
| A/PR/8/34 | 1.00E+05 TCID50/mL | ||
| A/NWS/33 | 4.24E+03 TCID50/mL | ||
| A/1/Denver/1/57 | 4.24E+03 TCID50/mL | ||
| A/Solomon Islands/3/2006 | 1.25E+04 TCID50/mL | ||
| A/Weiss/43 | 4.24E+03 TCID50/mL | ||
| A/Mal/302/54 | 1.25E+04 TCID50/mL | ||
| Seasonal H3N2 (Human) | A/Port Chalmers/1/73 | 5.10E+03 TCID50/mL | |
| A/Victoria/3/75 | 4.24E+03 TCID50/mL | ||
| JBAIDSInfluenza AssayTested | Type/Subtype (Host) | Strain | ConcentrationTested |
| A/Aichi/2/68 | 1.00E+05 TCID50/mL | ||
| A/Hong Kong/8/68 | 1.00E+05 TCID50/mL | ||
| A/Alice (VR-776) | 4.24E+03 TCID50/mL | ||
| A/MRC-2 recomb (VR-777) | 7.36E+03 TCID50/mL | ||
| A/Brisbane/10/07 | 7.36E+03 TCID50/mL | ||
| Swine NY/02/2009 | 1.25E+04 TCID50/mL | ||
| Swine NY/03/2009 | 7.36E+03 TCID50/mL | ||
| Swine NY/01/2009 | 3.78E+04 TCID50/mL | ||
| 2009 swine lineage H1N1(Human) | A/Mexico/4108/2009 | 1.00E+05 EID50/mL | |
| A/California/8/2009 | 1.00E+05 TCID50/mL | ||
| A/California/04/2009 | 1.00E+05 TCID50/mL | ||
| A/Texas/48/2009 | 1.00E+05 TCID50/mL | ||
| A/Washington/29/2009 | 1.00E+05 TCID50/mL | ||
| A/South Carolina/18/2009 | 1.00E+05 TCID50/mL | ||
| A/England/195/2009 | 4.74E+04 TCID50/mL | ||
| A/North Carolina/39/2009 | 1.00E+05 TCID50/mL |
Table 6. Results of Testing for Cross-Reactivity with Influenza A and B Strains
U.S. Army Office of the Surgeon General. 510(k) JBAIDS Influenza A Subtyping System
June, 2011
{10}------------------------------------------------
The non-influenza exclusivity panel consisted of 17 bacteria, 18 viruses, and 1 fungus, which were selected based on the relatedness to JBAIDS influenza analytes, clinical relevance (cause respiratory symptoms or represent nasopharyngeal flora), or high prevalence within the population (e.g. Herpes Simplex Virus). Simulated NPS samples were spiked with bacteria or fungi at a concentration of 10° CFU/mL or TCID50/mL and viruses at a concentration between 103 - 105 copies/mL or TCID50/mL. The JBAIDS Influenza A and B assays did not cross-react with any of the organisms tested in the the exclusivity panel at the test concentrations listed in Table 7.
{11}------------------------------------------------
| Virus | Concentration Tested | Bacteria/Fungi | Concentration Tested |
|---|---|---|---|
| Adenovirus | 1.00E+05 TCID50/mL | Bordetella pertussis | 1.00E+06 CFU/mL |
| Bocavirus | 4.20E+07 copies/mL | Candida albicans | 1.00E+06 CFU/mL |
| Coronavirus 229E | 7.35E+03 TCID50/mL | Corynebacterium diptheriae | 1.00E+06 CFU/mL |
| Coronavirus OC43 | 6.57E+04 TCID50/mL | Escherichia coli | 1.00E+06 CFU/mL |
| Coronavirus NL63 | 5.10E+03 TCID50/mL | Haemophilus influenza | 7.80E+04 CFU/mL |
| Coronavirus HKU1 | 1.00E+05 copies/mL | Lactobacillus plantarum | 1.00E+06 CFU/mL |
| Cytomegalovirus (CMV) | 1.50E+04 TCID50/mL | Legionella pneumophila | 1.00E+06 TCID50/mL |
| Enterovirus | 1.00E+05 TCID50/mL | Moraxella catarrhalis | 1.00E+06 CFU/mL |
| Epstein-Barr Virus (EBV) | 1.00E+05 copies/mL | Mycobacterium tuberculosis | 1.00E+06 CFU/mL |
| Human Metapneumovirus | 7.35E+03 TCID50/mL | Mycoplasma pneumonia | 1.69E+05 TCID50/mL |
| Human Rhinovirus | 5.10E+03 TCID50/mL | Neisseria elongata | 1.00E+06 CFU/mL |
| Measles Virus (Rubeola) | 1.00E+05 TCID50/mL | Neisseria meningitidis | 1.00E+06 CFU/mL |
| Mumps | 4.53E+04 TCID50/mL | Pseudomonas aeruginosa | 1.00E+06 CFU/mL |
| Parainfluenza virus 1 | 1.25E+04 TCID50/mL | Staphylococcus aureus | 1.00E+06 CFU/mL |
| Parainfluenza virus 2 | 1.50E+04 TCID50/mL | Staphylococcus epidermidis | 1.00E+06 CFU/mL |
| Parainfluenza virus 3 | 1.00E+05 TCID50/mL | Streptococcus pneumonia | 1.00E+06 CFU/mL |
| Parainfluenza virus 4 | 1.00E+05 TCID50/mL | Streptococcus pyogenes | 1.00E+06 CFU/mL |
| Respiratory Syncytial Virus | 1.25E+04 TCID50/mL | Streptococcus salivarius | 7.59E+05 CFU/mL |
Non Influonso Exclucivity Panob
Reproducibility
A multicenter study was performed to determine overall system reproducibility. Reproducibility testing occurred at three test sites, utilizing four total panels. Panels of NPS and NPW samples, each, were spiked with a representative influenza A virus (A/New Caledonia/20/1999). Panels of simulated NPS and simulated NPW samples, each, were spiked with a representative influenza B virus (B/Ohio/1/2005). Samples in each panel consisted of three samples that were spiked below LoD (high negative, LoD/20), at a low concentration of virus (low positive, LoD), and at a medium concentration of virus (medium positive, 3×LoD) for a total of 9 samples per panel. Each panel was tested twice daily at each site for five days for a total of 30 results per sample and 90 results per spike level. The detection rate was ≥ 98% for samples containing influenza yirus spiked at or above the LoD. As expected, samples spiked below LoD had variable results. Results are shown in Table 8 and Table 9.
{12}------------------------------------------------
| SampleType | IT 1-2-3 Platinum PathSample Purification Kit | Roche MagNA Pure CompactNucleic Acid Isolation Kit I | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| VirusSpikeLevel | Number Positive Samples/Total Samples(% Positive Detection) | Number Positive Samples/Total Samples(% Positive Detection) | Both Kits,All Sites | 95% CI | |||||||
| Site 1 | Site 2 | Site 3 | OverallforAll Sites | Site 1 | Site 2 | Site 3 | OverallforAll Sites | ||||
| NPS | 3×LoD | 15/15(100%) | 15/15(100%) | 15/15(100%) | 45/45(100%) | 15/15(100%) | 14/15(93%) | 15/15(100%) | 44/45(98%) | 89/90(99%) | 94.0-99.9 |
| LoD | 15/15(100%) | 15/15(100%) | 15/15(100%) | 45/45(100%) | 15/15(100%) | 15/15(100%) | 15/15(100%) | 45/45(100%) | 90/90(100%) | 96.7-99.9 | |
| Detection≥ LoD | 30/30(100%) | 30/30(100%) | 30/30(100%) | 90/90(100%) | 30/30(100%) | 29/30(97%) | 30/30(100%) | 89/90(99%) | 179/180(99%) | 96.9-99.9 | |
| LoD/20 | 9/15(60%) | 11/15(73%) | 11/15(73%) | 31/45(69%) | 13/15(87%) | 14/15(93%) | 15/15(100%) | 42/45(93%) | 73/90(81%) | 71.5-88.6 | |
| Detectionall Levels | 39/45(87%) | 41/45(91%) | 41/45(91%) | 121/135(90%) | 43/45(96%) | 43/45(96%) | 45/45(100%) | 131/135(97%) | 252/270(93%) | 89.7-96.0 | |
| NPW | 3×LoD | 15/15(100%) | 15/15(100%) | 15/15(100%) | 45/45(100%) | 15/15(100%) | 15/15(100%) | 15/15(100%) | 45/45(100%) | 90/90(100%) | 96.7-99.9 |
| LoD | 15/15(100%) | 15/15(100%) | 15/15(100%) | 45/45(100%) | 15/15(100%) | 14/15(93%) | 15/15(100%) | 44/45(98%) | 89/90(99%) | 94.0-99.9 | |
| Detection≥ LoD | 30/30(100%) | 30/30(100%) | 30/30(100%) | 90/90(100%) | 30/30(100%) | 29/30(97%) | 30/30(100%) | 89/90(99%) | 179/180(99%) | 96.9-99.9 | |
| LoD/20 | 14/15(93%) | 11/15(73%) | 13/15(87%) | 38/45(84%) | 15/15(100%) | 13/15(87%) | 15/15(100%) | 43/45(96%) | 81/90(90%) | 81.9-95.3 | |
| Detectionall Levels | 44/45(98%) | 41/45(91%) | 43/45(96%) | 128/135(95%) | 45/45(100%) | 42/45(93%) | 45/45(100%) | 132/135(98%) | 260/270(96%) | 93.3-98.2 |
Table 8. Summary of Reproducibility Testing for the Flu A Assay (Agreement with Expected Positive Results)
·
{13}------------------------------------------------
| SampleType | VirusSpikeLevel | IT 1-2-3 Platinum PathSample Purification Kit | Roche MagNA Pure CompactNucleic Acid Isolation Kit I | Both Kits,All Sites | 95% CI | ||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Site 1 | Site 2 | Site 3 | OverallforAll Sites | Site 1 | Site 2 | Site 3 | OverallforAll Sites | ||||
| sNPS | 3×LoD | 15/15 | 15/15 | 15/15 | 45/45 | 15/15 | 15/15 | 15/15 | 45/45 | 90/90(100%) | 96.7-99.9 |
| (100%) | (100%) | (100%) | (100%) | (100%) | (100%) | (100%) | (100%) | ||||
| LoD | 15/15 | 15/15 | 15/15 | 45/45 | 15/15 | 15/15 | 15/15 | 45/45 | 90/90(100%) | 96.7-99.9 | |
| (100%) | (100%) | (100%) | (100%) | (100%) | (100%) | (100%) | (100%) | ||||
| Detection≥ LoD | 30/30(100%) | 30/30(100%) | 30/30(100%) | 90/90(100%) | 30/30(100%) | 30/30(100%) | 30/30(100%) | 90/90(100%) | 180/180(100%) | 98.3-99.9 | |
| LoD/20 | 15/15(100%) | 14/15(93%) | 15/15(100%) | 44/45(98%) | 15/15(100%) | 14/15(93%) | 15/15(100%) | 44/45(98%) | 88/90(98%) | 92.2-99.7 | |
| DetectionAll Levels | 45/45(100%) | 44/45(98%) | 45/45(100%) | 134/135(99%) | 45/45(100%) | 44/45(98%) | 45/45(100%) | 134/135(99%) | 268/270(99%) | 97.4-99.9 | |
| sNPW | 3×LoD | 15/15 | 15/15 | 15/15 | 45/45 | 15/15 | 15/15 | 15/15 | 45/45 | 90/90(100%) | 96.7-99.9 |
| (100%) | (100%) | (100%) | (100%) | (100%) | (100%) | (100%) | (100%) | ||||
| LoD | 15/15 | 15/15 | 15/15 | 45/45 | 15/15 | 15/15 | 15/15 | 45/45 | 90/90(100%) | 96.7-99.9 | |
| (100%) | (100%) | (100%) | (100%) | (100%) | (100%) | (100%) | (100%) | ||||
| Detection≥ LoD | 30/30(100%) | 30/30(100%) | 30/30(100%) | 90/90(100%) | 30/30(100%) | 30/30(100%) | 30/30(100%) | 90/90(100%) | 180/180(100%) | 98.3-99.9 | |
| LoD/20 | 14/15(93%) | 15/15 | 12/15 | 41/45(91%) | 15/15(100%) | 15/15(100%) | 15/15(100%) | 45/45(100%) | 86/90(96%) | 89.0-98.8 | |
| (100%) | (80%) | ||||||||||
| DetectionAll Levels | 44/45(98%) | 45/45(100%) | 42/45(93%) | 131/135(97%) | 45/45(100%) | 45/45(100%) | 45/45(100%) | 135/135(100%) | 266/270(99%) | 96.3-99.6 |
Table 9. Summary of Reproducibility Testing for the Flu B Assay (Agreement with Expected Positive Results)
.
.
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized depiction of an eagle or other bird-like figure, with its wings spread and its body forming a flowing, abstract shape.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
U.S. Army Medical Materiel Development Activity c/o Robert E. Miller. Ph.D.. RAC Director. Division of Regulated Activities and Compliance 1430 Veterans Drive Fort Detrick, MD 21702-9232
SEP 1 3 201
Re: K111775
| Trade/Device Name: | JBAIDS Influenza A & B Kit |
|---|---|
| Regulation Number: | 21 CFR §866.3980 |
| Regulation Name: | Respiratory Viral Panel Multiplex Nucleic Acid Assay |
| Regulatory Class: | Class II |
| Product Codes: | OCC, OOI |
| Dated: | June 20, 2011 |
| Received: | June 23, 2011 |
Dear Dr. Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{15}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/Abouti7DA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.
Sincerely yours,
Uwe Scheef for
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{16}------------------------------------------------
Indications for Use
510(k) Number: K111775 Device Name: JBAIDS Influenza A & B Detection Kit
Indications for Use:
The Joint Biological Agent Identification and Diagnostic System (JBAIDS) Influenza A & B Detection Kit is intended for use on the JBAIDS instruments, for the in vitro qualitative detection of Influenza A and Influenza B viral nucleic acids isolated and purified from nasopharyngeal swab (NPS) and nasopharyngeal wash (NPW) specimens from human patients with signs and symptoms of respiratory infection. The JBAIDS Influenza A & B Detection Kit contains reverse transcriptase real-time polymerase chain reaction (rRT-PCR) assays that target the Matrix protein gene of Influenza A viruses, and the Non-structural protein gene of Influenza B viruses. This kit is not intended to detect Influenza C viruses.
Test results are to be used in conjunction with other clinical and epidemiological information. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions.
Performance characteristics for detection of influenza A were established when 2009 H1N1 Influenza, Influenza A H1N1, and Influenza A H3N2 were the predominant influenza A viruses in circulation. Due to low seasonal prevalence, performance characteristics for detection of seasonal Influenza A/H1 were established primarily with retrospective and surrogate clinical specimens. When other influenza A viruses are present, performance characteristics may vary.
All users, analysts, and any person reporting diagnostic results from use of this device should be trained to perform and interpret the results from this procedure by JBAIDS instructors or designees prior to use. Use of this device is limited to designated Department of Defense (DoD) laboratories equipped with the JBAIDS instruments.
If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a biosafety laboratory (BSL) 3+ facility is available to receive and culture specimens.
{17}------------------------------------------------
X_____________________________________________________________________________________________________________________________________________________________________________ Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices Evaluation and Safety (OIVD)
Tamara Feldberg
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) /// 775
§ 866.3980 Respiratory viral panel multiplex nucleic acid assay.
(a)
Identification. A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B;
(2) Influenza A subtype H1 and Influenza A subtype H3;
(3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B;
(4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus;
(5) Human Metapneumovirus;
(6) Rhinovirus; and
(7) Adenovirus.
(b)
Classification. Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.