K Number
K111692
Date Cleared
2011-11-08

(145 days)

Product Code
Regulation Number
878.4300
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TUMARK® Flex is intended for radiographically and radiologically percutaneous marking of soft tissue, especially breast tissue, via a clip marker.

The TUMARK® Flex is not indicated to be used with magnetic resonance imaging (MRI) techniques.

Device Description

The TUMARK® Flex is a sterile, single use, preloaded tissue site marking system consisting of a non-absorbable Nitinol clip-marker, a guide wire or tube and a handle with ejection mechanism. The guide wire is composed of a flexible tube, a distal ramp made of surgical high-grade steel with an opening for releasing the clip marker and a depth stopper with snap-in tip. The guide tube is composed of a tube section, a distal ramp made of surgical steel with an opening for releasing the clip marker, and a marking line, which shows the orientation of the ejection port for the clip marker. The handle is provided with a slider by means of which the clip can be released. The clip marker is situated in the distal ramp. TUMARK® Flex can be used together with, e.g. ultrasound and stereotactic X-ray imaging procedures.

The TUMARK® Flex is not indicated to be used in Magnetic Resonance Tomography (MRT). However, the clip marker placed in the patient can be exposed to a magnetic field of up to 3.0 Tesla, for instance in follow-up examinations.

AI/ML Overview

The provided document is a 510(k) summary for the SOMATEX® TUMARK® Flex Tissue Site Marking System. It primarily focuses on demonstrating substantial equivalence to predicate devices and does not contain detailed information about acceptance criteria or a specific study proving device performance against those criteria. The submission is not for a software or AI-driven device, but a physical medical device. Therefore, many of the requested items (e.g., sample size for test set, number of experts, adjudication method, AI assistance studies, training set details) are not applicable to this type of submission.

Here's a summary of what can be extracted based on the provided text, and where information is not available:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state quantitative acceptance criteria or report device performance against such criteria. The submission focuses on demonstrating substantial equivalence to legally marketed predicate devices, implying that the new device performs acceptably because it is "identical or similar in technology, design and material," with "minor technological differences" that "raise no new questions of safety and effectiveness."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable. This is a submission for a physical medical device. The document does not describe a "test set" in the context of data used for algorithm validation or a clinical study with a specified sample size for proving performance against acceptance criteria.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This is a submission for a physical medical device, not a diagnostic or interpretive algorithm requiring expert ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is a submission for a physical medical device, not a diagnostic or interpretive algorithm requiring adjudicated ground truth.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a tissue site marker, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable. The concept of "ground truth" as typically applied to diagnostic AI/software is not relevant to the approval of this physical medical device. Device performance is generally demonstrated through engineering testing, biocompatibility testing, material safety, and substantial equivalence to existing devices.

8. The sample size for the training set

Not applicable. This is a physical medical device, not a machine learning model requiring a training set.

9. How the ground truth for the training set was established

Not applicable. This is a physical medical device, not a machine learning model.

{0}------------------------------------------------

NOV - 8 2011 K111692

1/3

SOMATEX® Medical Technologies GmbH Rheinstrasse 7 d D-14513 Teltow / Germany

Summary of Safety and Effectiveness

・.

Titel: TUMARK® Flex

May 30 , 2011

SubmitterSOMATEX®Medical Technologies GMBHRheinstrasse 7 dD-14513 TeltowGermanyFrank KniepGeneral ManagerTelefon: ++49 3328 / 30 76 0FAX: ++49 3328 / 30 76 99E-mail: info@somatex.com
Application correspondent(consultant)Harald JungGeneral ManagerJUNG ConsultingUnterer Winkel 3D-78573 WurmlingenGermanyTelefon +49 7461-96 92 36FAX +49 7461-96 92 37E-mail: hjung@harald-jung.de
Trade NameTUMARK® Flex
Common NameTissue Site Marking System
Classification NameRadiographic Implantable Marker,21 C.F.R. 878.4300
Regulatory Class:II
Product Code :NEU
Performance StandardsNo performance standards have beenestablished under Section 514 of the Food, Drugand Cosmetic Act for Tissue Site MarkingSystems.

{1}------------------------------------------------

2/2

SOMATEX® Medical Technologies GmbH Rheinstrasse 7 d D-14513 Teltow / Germany

Legally marketed (unmodified) device:

The predicate device is the TUMARK® Professional Tissue Site Marking System, [K073095, March 19, 2008] and the TUMARK® Professional / MRJ TUMARK® Professional Tissue Site Marking System, [K093064, February 17, 2010]

Device Description (new devices):

The TUMARK® Flex is a sterile, single use, preloaded tissue site marking system consisting of a non-absorbable Nitinol clip-marker, a guide wire or tube and a handle with ejection mechanism. The guide wire is composed of a flexible tube, a distal ramp made of surgical high-grade steel with an opening for releasing the clip marker and a depth stopper with snap-in tip. The guide tube is composed of a tube section, a distal ramp made of surgical steel with an opening for releasing the clip marker, and a marking line, which shows the orientation of the ejection port for the clip marker. The handle is provided with a slider by means of which the clip can be released. The clip marker is situated in the distal ramp. TUMARK® Flex can be used together with, e.g. ultrasound and stereotactic X-ray imaging procedures.

The TUMARK® Flex is not indicated to be used in Magnetic Resonance Tomography (MRT). However, the clip marker placed in the patient can be exposed to a magnetic field of up to 3.0 Tesla, for instance in follow-up examinations.

Indications for Use:

The TUMARK® Flex is intended for radiographically and radiologically percutaneous marking of soft tissue, especially breast tissue, via a clip marker.

The TUMARK® Flex is not indicated to be used with magnetic resonance imaging (MRI) techniques.

{2}------------------------------------------------

Page REV 03. B 3

Summary of Safety and Effectiveness

SOMATEX® Medical Technologies GmbH Rheinstrasse 7 d D-14513 Teltow / Germany

Comparison to cleared devices (Substantial Equivalence):

The TUMARK® Flex is substantial equivalent to the TUMARK® Professional and MRI TUMARK® Professional, also manufactured by Somatex® Medical Technologies GmbH that have been cleared by FDA (K093064 and K073095). The TUMARK® Flex and TUMARK® Professional and MRI TUMARK® Professional are intended to attach a marker to soft tissue at the surgical site during a percutaneous procedure. All devices are indicated for use to radiographically and radiologically mark the surgical location in breasts following a percutaneous procedure.

In addition, both the proposed devices and the predicate devices are identical or similar in technology, design and material. In particular, the proposed devices and the predicate devices consist of the same primary components and the component materials of the proposed devices and the predicate devices are substantially equivalent and/or standard materials for invasive medical techniques. The minor technological differences between the proposed device and the predicate devices raise no new questions of safety and effectiveness.

K111692

3/2

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines and a wavy base.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

S9OMATEX® Medical Technologies GmbH % JUNG Consulting Mr. Harald Jung Unterer Winkel 3 Wurmlingen 78573 Germany

Re: K111692

Trade/Device Name: TUMARK® Flex Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: II Product Code: NEU Dated: September 30, 2011 Received: October 17, 2011

Dear Mr. Jung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

NOV - 8 2011

{4}------------------------------------------------

Page 2 - Mr. Harald Jung

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm far the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR, Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolloffree mobile (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely your

R

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: TUMARK® Flex

Indications for Use:

The TUMARK® Flex is intended for radiographically and radiologically percutaneous marking of soft tissue, especially breast tissue, via a clip marker.

The TUMARK® Flex is not indicated to be used with magnetic resonance imaging (MRI) techniques.

Prescription Use YES (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use NO (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel Kune for MXM
(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K111692
Page REV 03. A 3

§ 878.4300 Implantable clip.

(a)
Identification. An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.