(232 days)
Not Found
No
The summary describes a standard pulse oximeter with data storage, display, alarm, and Bluetooth transmission functions. There is no mention of AI, ML, or any advanced algorithms beyond basic signal processing for SpO2 and pulse rate calculation. The performance studies focus on standard clinical testing and compliance with relevant standards, not on the evaluation of AI/ML model performance.
No
The device is described as a pulse oximeter for continuous monitoring and spot-checking of SpO2 and pulse rate. Its function is to measure and monitor physiological parameters, not to treat or cure a disease or condition.
Yes
The device measures functional pulse oxygen saturation (SpO2) and pulse rate (PR), which are physiological parameters used to assess a patient's health status and can aid in identifying potential medical conditions.
No
The device description explicitly states it is a "handheld pulse oximeter" and describes physical components like a "dual light source and photo detector" and an "external sensor," indicating it is a hardware device with integrated software.
No, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. They are used to examine these samples outside of the body.
- Device Function: The description clearly states that the device works by applying an external sensor to a pulsating arteriolar vascular bed (e.g., a finger). It measures functional pulse oxygen saturation (SpO2) and pulse rate (PR) by shining light through the tissue and detecting the absorption. This is a non-invasive measurement performed on the body, not on a sample taken from the body.
Therefore, based on the provided information, the MD300M122/MD300M222/MD300M322 handheld pulse oximeter is a medical device, but it is not an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The MD300M122/MD300M222/MD300M322 handheld pulse oximeter is intended for continuous monitoring, spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of single adult and pediatric patients in hospitals and home care.
Product codes (comma separated list FDA assigned to the subject device)
DQA
Device Description
The applicant device MD300M122 Handheld Pulse Oximeter is integrated with Bluetooth® technology allowing the user to transfer measurement data any time and anywhere. The oximeter is designed with the measurement, storage, review, visible and audible alarms and data transmission function.
The MD300M122 Handheld Pulse Oximeter works by applying a external sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660nm, which is red light; the other is 940nm, which is infrared light.
The applicant device of MD300M122 Handheld Pulse Oximeter is the handheld equipment, which mainly function are measurement, display, alarm, data storage & replay etc.
Measurement function
That's the mainly function of the device, which use the method described above to measure the SpO2 value and Pulse Rate value of user.
Display function
The display function of the device display the SpO2&PR value, waveform synchronously, battery capacity indicator, alarm state indication and time.
Alarm function
The applicant device has physiological alarm function as SpO2 and PR parameter and technology alarm function as Low Battery Voltage, probe off and Finger out.
Each alarm of device contains VISUAL and AUDIBLE alarm. And you can turn on or off the alarm and set the alarm limits.
There are three-level alarm priorities in Oximeter. All the three priorities divided by built-in module and can't be changed by user.()High priority: "Di- Di ----- Di - Di" indicates the patient is in the very dangerous situation. @Medium priority: "Di - Di" indicates the warning-attention should be paid.(3)Low priority: "Di" indicates the technical alarm caused by the device itself.
Data storage, replay & transmission function
The measured record is stored automatically every four seconds. The monitor can store 72 hours records.
The device includes the function as "record review" on the Oximeter equipment. And you can review "Data Browse", "SpO2 Trend", "PR Trend" and "Data Clear" by this function. The data stored in the device can be transmitted through Bluetooth or USB cable.
Power
The applicant device uses three AA alkaline batteries or rechargeable batteries for power supply. The device has charge function by AC/DC power adapter, the specification of which is: 100-240 (VAC), 47/63 (Hz) input; 5Vd.c. output.
The device can measure normally during the process of charging, but we won't recommend to do so.
The applicant device is not for life-supporting or life-sustaining, not for implant. The device is not sterile and does not need sterilization or re-sterilization. The device is for prescription. The device does not contain drug or biological product.
The device is software-driven and the software validation is provided in Section 10 Software.
. MD300M222 Handheld Pulse Oximeter
MD300M222 is the same as MD300M122 except the appearance.
● MD300M222 Handheld Pulse Oximeter
MD300M322 is the same as MD300M122 except the appearance.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
adult and pediatric patients
Intended User / Care Setting
hospitals and home care.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
The Clinical Tests following ISO 9919:2005, Annex EE.2 are conducted in the lab of Beijing Military General Hospital. The study protocol is subjected to ISO 9919:2005 Annex EE. Procedures of testing required in EE.2 are adopted. It can be determined from the result of the study that the accuracy of the proposed device is compliance to the specification claimed by the manufacturer compared with "Golden Standard" Co-Oximeter.
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The proposed device was performed the tests according to the following standards:
IEC 60601-1:1988+A1:1991+A2:1995, Medical Electrical Equipment - Partl: General requirements for safety.
IEC 60601-1-2:2007, Medical Electrical Equipment - Part 1: General requirements for safety-2, Collateral Standard: Electromagnetic compatibility - Requirements and tests.
FCC PART 15, SUBPART C: 2008 (Senction 15.249)
The Clinical Tests following ISO 9919:2005, Annex EE.2 are conducted in the lab of Beijing Military General Hospital. The study protocol is subjected to ISO 9919:2005 Annex EE. Procedures of testing required in EE.2 are adopted. It can be determined from the result of the study that the accuracy of the proposed device is compliance to the specification claimed by the manufacturer compared with "Golden Standard" Co-Oximeter.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.2700 Oximeter.
(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).
0
Additional Information K111494
Exhibit I 510(K) Summary
As required by 21 CFR 807.92 The Assigned 510(K) Number is: K111494
- Date Prepared: December 16, 2011
2. Sponsor Information
Beijing Choice Electronic Technology Co., Ltd. Bailangyuan Building B 1127-1128, Fuxing Road, A36 Beijing 100039, China
Mr. Lei Chen North Building 3F, No.9 Shuangyuan Road, Bdachu Hi-tech Zone, Shijingshan District Beijing China 100041 Phone:+86-10-88790480-6018 Email:cc@choicemmed.com
3. Submission Correspondent
Ms. Sunny Wang Beijing Choice Electronic Technology Co., Ltd. Floor 4, Jingyang Building No. 15, Xijing Rd., Shijingshan District, Beijing 100041, P.R. China Phone: +86-10-88204630-8176 Fax: +86-10-88204632 Email: wangsuqin@choicemmed.com & cc@choicemmed.com
4. US Agent
Wei Huang Mid-Link International INC 307 Eagle Heights Apt E, Madison, Wisconsin, 53705, United States Phone: 608-3355979 Fax: 760-4665084 Email:info@mid-link.net
1
5. Proposed Device Information
Device Common or Usual Name: Pulse Oximeter Device Trade or Proprietary Name: Handheld Pulse Oximeter Model: MD300M122, MD300M222, MD300M322 Classification Name: Oximeter Regulation Number: 21 CFR 870.2700 Product Code: DQA Panel: Anesthesiology
6. Predicate Device
Fingertip Pulse Oximeter MD300C318 (K092620) Beijing Choice Electronic Technology Co., Ltd. Room 1127-1128 Building B, Bailangyuan Fuxing Road, No.A36 Beijing, CHINA 10039
7. Device Description
● MD300M122 Handheld Pulse Oximeter
The applicant device MD300M122 Handheld Pulse Oximeter is integrated with Bluetooth® technology allowing the user to transfer measurement data any time and anywhere. The oximeter is designed with the measurement, storage, review, visible and audible alarms and data transmission function.
The MD300M122 Handheld Pulse Oximeter works by applying a external sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660nm, which is red light; the other is 940nm, which is infrared light.
The applicant device of MD300M122 Handheld Pulse Oximeter is the handheld equipment, which mainly function are measurement, display, alarm, data storage & replay etc.
Measurement function
That's the mainly function of the device, which use the method described above to measure the SpO2 value and Pulse Rate value of user.
Display function
The display function of the device display the SpO2&PR value, waveform synchronously, battery capacity indicator, alarm state indication and time.
2
Additional Information K111494
Alarm function
The applicant device has physiological alarm function as SpO2 and PR parameter and technology alarm function as Low Battery Voltage, probe off and Finger out.
Each alarm of device contains VISUAL and AUDIBLE alarm. And you can turn on or off the alarm and set the alarm limits.
There are three-level alarm priorities in Oximeter. All the three priorities divided by built-in module and can't be changed by user.()High priority: "Di- Di ----- Di - Di" indicates the patient is in the very dangerous situation. @Medium priority: "Di - Di" indicates the warning-attention should be paid.(3)Low priority: "Di" indicates the technical alarm caused by the device itself.
Data storage, replay & transmission function
The measured record is stored automatically every four seconds. The monitor can store 72 hours records.
The device includes the function as "record review" on the Oximeter equipment. And you can review "Data Browse", "SpO2 Trend", "PR Trend" and "Data Clear" by this function. The data stored in the device can be transmitted through Bluetooth or USB cable.
Power
The applicant device uses three AA alkaline batteries or rechargeable batteries for power supply. The device has charge function by AC/DC power adapter, the specification of which is: 100-240 (VAC), 47/63 (Hz) input; 5Vd.c. output.
The device can measure normally during the process of charging, but we won't recommend to do so.
The applicant device is not for life-supporting or life-sustaining, not for implant. The device is not sterile and does not need sterilization or re-sterilization. The device is for prescription. The device does not contain drug or biological product.
The device is software-driven and the software validation is provided in Section 10 Software.
. MD300M222 Handheld Pulse Oximeter
MD300M222 is the same as MD300M122 except the appearance.
● MD300M222 Handheld Pulse Oximeter
MD300M322 is the same as MD300M122 except the appearance.
8. Intended use
3
Additional Information K111494
The MD300M122/MD300M222/MD300M322 handheld pulse oximeter is intended for continuous monitoring, spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of single adult and pediatric patients in hospitals and home care.
9. Substantial Equivalence
The applicant devices MD300M222/MD300M222/MD300M322 Handheld Pulse Oximeter have the same classification, same intended use, same design principle, similar specifications and same safety performance as the predicate device. These are no obvious differences to influence the effectiveness and safety of the device.
The proposed device is Substantially Equivalent (SE) to the predicate device which is US legally market device. Therefore, the subject device is determined as safe and effectiveness.
10. Test Conclusion
Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The proposed device was performed the tests according to the following standards:
IEC 60601-1:1988+A1:1991+A2:1995, Medical Electrical Equipment - Partl: General requirements for safety.
IEC 60601-1-2:2007, Medical Electrical Equipment - Part 1: General requirements for safety-2, Collateral Standard: Electromagnetic compatibility - Requirements and tests.
FCC PART 15, SUBPART C: 2008 (Senction 15.249)
The Clinical Tests following ISO 9919:2005, Annex EE.2 are conducted in the lab of Beijing Military General Hospital. The study protocol is subjected to ISO 9919:2005 Annex EE. Procedures of testing required in EE.2 are adopted. It can be determined from the result of the study that the accuracy of the proposed device is compliance to the specification claimed by the manufacturer compared with "Golden Standard" Co-Oximeter.
4
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Sunny Wang Correspondent Beijing Choice Electronic Technology Company, Limited Floor 4. Jungyang Bldg, No.15 Xiiing Road Shijingshan District, Beijing CHINA 100041
JAN 1 8 2012
Re: K111494
Trade/Device Name: Handheld Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: December 16, 2011 Received: December 19, 2011
Dear Ms. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
5
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm 1 1 5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Anthony D. Watson, B.S., M.S., M.B.A.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
6
Indications for Use
510(k) Number (if known): pending
Device Name: Handheld Pulse Oximeter_
Indications for Use:
The MD300M122/MD300M222/MD300M322 handheld pulse oximeter is intended The MD300M122/MD300MZ22/MD300M222 fitunctional pulse oxygen saturation
for continuous monitoring, spot-checking of functional pulse oxygens in hospitals and for continuous monitoring, spot-checking of functions. Partin and provins and home care.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(Please do not write below this line-continue on another page IF NEEDED)
Moll J. Ward br Lester Schultheis
(Division Sign-Off) Division of Anesthesiology, General Hospital nfection Control, Dental Devices
510(k) Number: