K Number
K111474
Device Name
BINGO PRO
Date Cleared
2011-11-02

(159 days)

Product Code
Regulation Number
N/A
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

BINGO PRO is an electronic device used for precise apex localization and working length determination during root canal treatment. The device enables to obtain correct results in canals with different conditions - dry or wet.

Device Description

BINGO PRO is a modern apex locator intended for precise localization of root canal apex. The measurements in BINGO PRO are performed utilizing AC signals at two frequencies -500 Hz and 8 kHz. The frequencies are alternated and not mixed, eliminating the need for signal mixing and frequency discrimination electronic circuits. The patented signal measuring method utilized in BINGO PRO is based on measurements of RMS (Root Mean Square) level of the signal. Advanced 3D user interface implemented in BINGO PRO is based on high resolution TFT color graphic display. "Live" real time presentation of endodontic file movement inside the canal is designed to make dentist's work easier and to increase his confidence. Display indicators are carefully designed to be intuitively understood and to serve for instant troubleshooting during device usage. BINGO PRO shows the movement of the file inside the canal from the beginning of the measurements to the end, providing uninterrupted feedback to the dentist. ROOT WIZARD file tracking algorithm enables full-scale display of the file movement during the treatment while APICAL ZOOM feature enables high-resolution display of the file advance in preapical and apical zones. Large, clearly recognizable graphical and numerical readings in the APICAL ZOOM are designed to enable precise control over the file advance matching the individual technique of the dentist. Visual information is accompanied by optional audio signals. Numerical values and the numerical scale shown in the APICAL ZOOM do not represent actual distance from the apex in mm; they serve as a convenient reference to judge the file tip position in relation to the apex. Operation of BINGO PRO is fully automatic, no manual calibrations or adjustments are required. The measured signal is analyzed and automatic adjustments are made if required. The device may operate within different conditions in the root canal: dry or wet. BINGO PRO may only be used with stainless steel or nickel titanium endodontic files. According to the modern approach, the final distance of the probe tip from the apex should vary in different cases: for example, in non-contaminated root canal the tip should be about 0.5 mm before the minor apical foramen; in case of bacterial contamination the canal should be cleaned to the apex. Such cases may occur in different teeth of the same patient or even in the different canals of the same root. VIRTUAL APEX feature implemented in BINGO PRO enables to mark a predetermined indication position on the display in the APICAL ZOOM area. When VIRTUAL APEX feature is enabled, the dentist gets clear visual and audio feedback that the color bar indicating file tip position has reached the pre-selected point in the apical zone. If the virtual apex is set at the fifth bar, for example, the visual mark appears at 0.5 position. Additionally, special easily recognizable "VIRTUAL APEX" audio signal is activated when the color bar indicator reaches the virtual apex mark. Utilizing the VIRTUAL APEX feature, the dentist may choose to proceed either to the VIRTUAL APEX mark in non-contaminated canals or to the APEX indication in contaminated canals using visual or audio feedback. Numerical value of virtual apex mark does not represent actual distance from the apex in mm. Built-in Demo mode of BINGO PRO enables easy simulation of all stages of the treatment and is designed to simplify familiarization of the user with the device.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the BINGO PRO electronic apex locator, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The 510(k) summary does not explicitly define acceptance criteria in terms of specific performance metrics or thresholds (e.g., "accuracy must be X% within Y mm"). Instead, the acceptance is based on demonstrating substantial equivalence to a predicate device (Bingo-1020) and showing that the BINGO PRO provides "clinically acceptable results" through comparison with another established device, the Dentaport ZX.

Therefore, the "acceptance criteria" can be inferred as:

  • Equivalence to Predicate Device: The BINGO PRO should have the same intended use and fundamental scientific technology as the Bingo-1020.
  • Clinically Acceptable Performance: The BINGO PRO's apex localization results should be equivalent to those of a recognized and cleared reference device (Dentaport ZX) in ex-vivo testing.

Here's a table based on this interpretation:

Acceptance Criteria (Inferred)Reported Device Performance
Substantial Equivalence to Predicate Device (Bingo-1020): Same intended use and fundamental technology.- Same intended use (precise apex localization and working length determination during root canal treatment). - Same fundamental scientific technology (electronic apex locator, RMS signal measurement at two frequencies: 500 Hz and 8 kHz). - Differences in design features (e.g., display type, battery, weight) were deemed not to raise new questions of safety or effectiveness.
Clinically Acceptable Apex Localization: Performance equivalent to a recognized reference device.- Ex-vivo test compared BINGO PRO results to Dentaport ZX apex locator results. - Conclusion: "the apex localization obtained with both devices is essentially equivalent and that BINGO PRO provides clinically acceptable results."

Study Details:

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document only states "ex-vivo test was performed on extracted teeth." It does not specify the number of extracted teeth used in the test set.
  • Data Provenance:
    • Country of Origin: Not explicitly stated for the extracted teeth. The manufacturer is Forum Engineering Technologies (96) Ltd., located in Rishon Lezion, Israel. The predicate device's original manufacturer is also listed as Forum Engineering Technologies (96) Ltd., (Israel).
    • Retrospective or Prospective: Not specified, but "ex-vivo test" implies a prospective collection of extracted teeth for the specific purpose of this evaluation.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • The document does not specify the number of experts or their qualifications for establishing ground truth in the ex-vivo test. It only mentions that the results from BINGO PRO were compared to a "reference device - Dentaport ZX apex locator." For ex-vivo studies involving extracted teeth, ground truth is typically established by direct visual measurement or microscopy after sectioning the tooth, rather than by human experts interpreting images from the device. However, the exact method for ground truth establishment for this study is not detailed.

4. Adjudication Method for the Test Set

  • The document does not describe an adjudication method. This type of study (comparing a new device's measurements to a reference device or objective physical measurements) typically doesn't involve expert adjudication in the same way an image interpretation study would.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

  • No, an MRMC comparative effectiveness study was not done. The study described is a direct comparison of a new device (BINGO PRO) against a cleared reference device (Dentaport ZX) in an ex-vivo setting to assess measurement equivalence, not a study evaluating human reader performance with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Yes, a standalone performance study was done. The BINGO PRO is an electronic apex locator that provides a measurement directly. The ex-vivo test compares the measurements produced by the device itself (BINGO PRO) to measurements generated by another device (Dentaport ZX). There is no mention of a human-in-the-loop component for the device's operation or measurement interpretation in this performance evaluation.

7. The Type of Ground Truth Used

  • The document states that the results from BINGO PRO were compared to the results of the "reference device - Dentaport ZX apex locator." This implies that the ground truth for the performance claim was established by the measurements from the predicate/reference device (Dentaport ZX). For apex locators, the ultimate ground truth typically involves physical measurement of the tooth, often by sectioning, but here the equivalence is drawn to another device.

8. The Sample Size for the Training Set

  • Not applicable. The BINGO PRO is an electronic apex locator that uses AC signals and RMS level measurements. Its operation is "fully automatic" and does not appear to involve machine learning or AI models that require a separate training set. The device operates based on fixed algorithms and electrical principles, not pattern recognition learned from data.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As the device does not appear to utilize a training set in the machine learning sense, there is no ground truth establishment for a training set.

{0}------------------------------------------------

Forum Engineering Technologies (96) Ltd.
BINGO PRO 510(k) File510(k) SummaryRev. 04
Document number:5-B31-003.FDA
Effective Date:Oct 23, 2011
Page 1 of 6

KII474

"510(k) Summary"

NOV - 2 2011

BINGO PRO, Electronic Apex Locator

The following 510(k) Summary of Safety and Effectiveness has been prepared pursuant to requirements for 510(k) summaries specified in 21 CFR § 807.92(a).

Section's content

  • રે ! . 807.92(a)(1)-Owner & Submitter's Details
  • 807.92(a)(2) Candidate Device Details. 5.2.
  • રે રે જેવી સાં 807.92(a)(3) - Predicate Device Identification
  • 807.92(a)(4) Device Description ર્સ વર્ષ
  • 807.92(a)(5) Intended Use 5.5.
  • 807.92(a)(6) Substantial Equivalence Comparison Table ર્ડ.
  • 807.92(b)(1) Brief discussion of the nonclinical tests 5.7.
  • 807.92(b)(3) Conclusions ર જ

5.1. Owner & Submitter Details: [ 807.92(a)(1) }

Owner & Submitter Name:

Address:

  • 1 Platin St., New Industrial Zone, Rishon Lezion 75653, Israel. Phone: +972-3-9625517 Fax number: +972-3-9613355 info@forumtec.net E-mail Name of Contact Person: Ms. Yuliya Yutkevich
    Forum Engineering Technologies (96) Ltd.

US Agent:

NORMAN F.ESTRIN, PH.D., ESTRIN CONSULTING GROUP, INC. (ECG) 9109 Copenhaver Drive, Potomac, MD 20854 Phone: +001-301-279-2899 +001-301-294-0126 Fax: estrin@yourfdaconsultant.com Email:

{1}------------------------------------------------

Image: Forum Technologies LogoForum Engineering Technologies (96) Ltd.
BINGO PRO 510(k) File510(k) SummaryRev. 04
Document number:5-B31-003.FDAEffective Date: Oct 23, 2011Page 2 of 6
5.2. Candidate Device Details:[ 807.92(a)(2) ]
Trade Name:BINGO PRO – Electronic Apex Locator
Trade Name:BINGOS PRO - Electronic Apex Locator
Common Name:Apex Locator
Classification Name:Locator, Root Apex
Product Code:LQY -- Locator, ROOT APEX
Review Panel:Dental
Device Class:Unclassified

5.3. Predicate Device Identification: [ 807.92(a)(3) ]

Devices to which substantial equivalence is claimed:

Table 5.3: Predicate Device Identification
-----------------------------------------------------
PredicateDevice NameName ofManufacturerName of Applicant510(k)Number
Bingo-1020Forum EngineeringTechnologies (96) Ltd.,(Israel)DENT CORP, Research& Development (USA)K992233

5.4. Device Description: [ 807.92(a)(4) ]

BINGO PRO is a modern apex locator intended for precise localization of root canal apex.

The measurements in BINGO PRO are performed utilizing AC signals at two frequencies -500 Hz and 8 kHz. The frequencies are alternated and not mixed, eliminating the need for signal mixing and frequency discrimination electronic circuits. The patented signal measuring method utilized in BINGO PRO is based on measurements of RMS (Root Mean Square) level of the signal.

Advanced 3D user interface implemented in BINGO PRO is based on high resolution TFT color graphic display. "Live" real time presentation of endodontic file movement inside the canal is designed to make dentist's work easier and to increase his confidence. Display indicators are carefully designed to be intuitively understood and to serve for instant troubleshooting during device usage.

BINGO PRO shows the movement of the file inside the canal from the beginning of the measurements to the end, providing uninterrupted feedback to the dentist. ROOT WIZARD file tracking algorithm enables full-scale display of the file movement during the treatment while APICAL ZOOM feature enables high-resolution display of the file advance in preapical and apical zones. Large, clearly recognizable graphical and numerical readings in the APICAL ZOOM are designed to enable precise control over the file advance matching the individual technique of the dentist. Visual information is accompanied by optional audio

{2}------------------------------------------------

Image: Forum Technologies LogoForum Engineering Technologies (96) Ltd.
BINGO PRO 510(k) File510(k) SummaryRev. 04
Document number:5-B31-003.FDAEffective Date: Oct 23, 2011Page 3 of 6

signals. Numerical values and the numerical scale shown in the APICAL ZOOM do not represent actual distance from the apex in mm; they serve as a convenient reference to judge the file tip position in relation to the apex.

Operation of BINGO PRO is fully automatic, no manual calibrations or adjustments are required. The measured signal is analyzed and automatic adjustments are made if required. The device may operate within different conditions in the root canal: dry or wet. BINGO PRO may only be used with stainless steel or nickel titanium endodontic files.

According to the modern approach, the final distance of the probe tip from the apex should vary in different cases: for example, in non-contaminated root canal the tip should be about 0.5 mm before the minor apical foramen; in case of bacterial contamination the canal should be cleaned to the apex. Such cases may occur in different teeth of the same patient or even in the different canals of the same root. VIRTUAL APEX feature implemented in BINGO PRO enables to mark a predetermined indication position on the display in the APICAL ZOOM area. When VIRTUAL APEX feature is enabled, the dentist gets clear visual and audio feedback that the color bar indicating file tip position has reached the pre-selected point in the apical zone. If the virtual apex is set at the fifth bar, for example, the visual mark appears at 0.5 position. Additionally, special easily recognizable "VIRTUAL APEX" audio signal is activated when the color bar indicator reaches the virtual apex mark. Utilizing the VIRTUAL APEX feature, the dentist may choose to proceed either to the VIRTUAL APEX mark in non-contaminated canals or to the APEX indication in contaminated canals using visual or audio feedback. Numerical value of virtual apex mark does not represent actual distance from the apex in mm.

Built-in Demo mode of BINGO PRO enables easy simulation of all stages of the treatment and is designed to simplify familiarization of the user with the device.

5.5. Intended Use: [807.92(a)(5)]

BINGO PRO is an electronic device used for precise apex localization and working length determination during root canal treatment. The device enables to obtain correct results in canals with different conditions - dry or wet.

5.6. Substantial Equivalence Comparison Table: [ 807.92(a)(6) ]

LineNo.DeviceCharacteristicsPredicate DeviceBingo-1020Candidate DeviceBINGO PRO
1.Device definitionElectronic apex locatorThe same as in Bingo-1020.
2.Intended UsePrecise apex localizationduring root canal treatment.The same as in Bingo-1020.

Table 5.6: Substantial Equivalence Comparison

{3}------------------------------------------------

Image: Forum Technologies LogoForum Engineering Technologies (96) Ltd.
BINGO PRO 510(k) File510(k) SummaryRev. 04

Document number:

510(k) Summary

5-B31-003.FDA

Effective Date: Oct 23, 2011

LineDevicePredicate DeviceCandidate Device
No.CharacteristicsBingo-1020BINGO PRO
3.Indications for useBingo-1020 is a moderndevice for precise apexlocalization during rootchannel treatment.Bingo-1020 is distinguishedby increasing precision at0.1mm on wet/dry, largegraphic display wherecurrent position of endo fileis reflected and otheressential information isdisplayed.BINGO PRO is an electronicdevice used for precise apexlocalization and workinglength determination duringroot canal treatment. Thedevice enables to obtaincorrect results in canals withdifferent conditions - dry orwet.
4.Where to be used(clinics, home etc.)This product must only beused in hospitalenvironments, clinics ordental offices by qualifieddental personnel.The same as in Bingo-1020.
5.Device categoryActive, invasiveThe same as in Bingo-1020.
6.Power SourceLow voltage NiMHrechargeable batteries (3.6V)Low voltage NiMHrechargeable batteries (2.4V)
7.External chargerInput: 115V/50-60HzOutput: 6V DC @ 200mAInput: 120V/50-60HzOutput: 6V DC @ 500mA.
8.CurrentConsumptionMaximum -- 50 mA DCMaximum - 250 mA DC.
9.Method ofcalculating locationof root canal apexRMS functions of themeasured signals at twofrequencies are used tocalculate the test scores,which are compared tostatistically predefinedthresholds.The same as in Bingo-1020.
10.DisplayCustom monochrome LCD3.5" Color TFT Display
11.ButtonsThree pushbuttons:1. On / Off2. Sound control3. MODEThe same as in Bingo-1020.
12.Sound indicationPiezzo transducer withsound level control (high,medium, low, mute)The same as in Bingo-1020.

{4}------------------------------------------------

Forum Engineering Technologies (96) Ltd.
BINGO PRO 510(k) FileDocument number:5-B31-003.FDA510(k) SummaryEffective Date: Oct 23, 2011Rev. 04Page 5 of 6
LineNo.DeviceCharacteristicsPredicate DeviceBingo-1020Candidate DeviceBINGO PRO
13.Adjustment beforemeasurementNot requiredThe same as in Bingo-1020
14.CalibrationNot requiredThe same as in Bingo-1020
15.Measuring signalamplitudeNominal - doesn't exceed25 VACThe same as in Bingo-1020
14.CalibrationNot requiredThe same as in Bingo-1020.
15.Measuring signal amplitudeNominal - doesn't exceed 25 mV AC.The same as in Bingo-1020.
16.Frequencies used for measurements500 Hz and 8 kHzThe same as in Bingo-1020.
17.Weight430 Gr300 Gr
18.Dimensions160 x 95 x 35 mm74 x 120 x 70 mm
19.Endodontic Files to be used with the deviceBingo-1020 may only be used with stainless steel or nickel titanium endodontic files.The same as in Bingo-1020.
20.Type of ConnectorThe type of connector used – Stereo plug.The type of connector used – Micro-USB plug.
21.Automatic Turn-off functionThe device turns off automatically after 5 min. of idle state.The same as in Bingo-1020.
22.Display Zoom AreaZoom area with the following graphical indications of file tip position:1.0, 0.9, 0.8, 0.7, 0.6, 0.5, 0.4, 0.3, 0.2, 0.1, and 0.0.Graphical indication of over-instrumentation.Zoom area with the following graphical indications of file tip position:2.0, 1.9, 1.8, 1.7, 1.6, 1.5, 1.4, 1.3, 1.2, 1.1, 1.0, 0.9, 0.8, 0.7, 0.6, 0.5, 0.4, 0.3, 0.2, 0.1, and 0.0.Additionally file tip position is indicated in numerical form.Over-instrumentation: graphical indication and additional alphanumerical presentation.
23.Virtual Apex featureBasic Virtual Apex with audio feedback.Advanced Virtual Apex with visual and audio feedback.
24.Training modeTutor Mode is implemented to demonstrate device operation and to shorten learning curve of the user.DEMO mode is used for the same purposes.

.

·

·

... ,

{5}------------------------------------------------

Forum Engineering Technologies (96) Ltd.
BINGO PRO 510(k) File510(k) SummaryRev. 04
Document number:5-B31-003.FDAEffective Date: Oct 23, 2011Page 6 of 6

5.7. Brief discussion of the nonclinical tests [807.92(b)(1)]

To evaluate the performance of BINGO PRO apex locator, ex-vivo test was performed on extracted teeth. The results obtained with BINGO PRO were compared to the results of reference device - Dentaport ZX apex locator, which is used in many scientific studies and has FDA clearance. The conclusion of the test was that the apex localization obtained with both devices is essentially equivalent and that BINGO PRO provides clinically acceptable results.

5.8. Conclusions: [ 807.92(b)(3) ]

  • BINGO PRO has the same intended use and fundamental scientific technology as . its predicate device - Bingo-1020 (K992233).
  • BINGO PRO was evaluated against its predicate device, and was found to be ● Substantially Equivalent.

{6}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Yuliya Yutkevich Quality Assurance & Regulatory Affairs Manager Forum Engineering Technologies (96) Limited 1 Platin Street New Industrial Zone Rishon Lezion Israel 75653

NOV - 2 2011

Re: K111474

Trade/Device Name: BINGO PRO Regulation Number: Unclassified Regulation Name: None Regulatory Class: None Product Code: LQY Dated: October 25, 2011 Received: October 28, 2011

Dear Ms. Yutkevich:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however. that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

Page 2- Ms. Yutkevich

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Nh for

Anthony Watson, MS, MBA Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

ForumForum Engineering Technologies (96) Ltd.
BINGO PRO 510(k) FileIndication for Use Statement
Document number: 5-B31-002.FDAEffective Date: Aug 18, 2011
Rev. 02
Page 1 of 1

Indication for Use Statement

510(k) Number (if known): K111474

Device Name:

BINGO PRO

Indications for Use:

BINGO PRO is an electronic device used for precise apex localization and working length determination during root canal treatment. The device enables to obtain correct results in canals with different conditions - dry or wet.

Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Rumm

(Division Sign-Off) יוvision of Anesthesiology, General Hospital fection Control, Dental Devices

510(k) Number: K111474

N/A