K Number
K111346
Device Name
RESOLUTIONMD MOBILE MODEL RMB-MOB-2X
Date Cleared
2011-09-09

(119 days)

Product Code
Regulation Number
892.2050
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The ResolutionMD™ Mobile software is a software-based Picture Archiving and Communication System (PACS) used with general purpose computing servers and specific mobile devices. It provides for communication, storage, reformatting, rendering on the server component and communication and display of DICOM 3.0-compliant CT and MR medical images as well as reports on the mobile device. The ResolutionMD Mobile provides wireless and portable access to medical images. The device is intended for use as a diagnostic, review, and analysis tool by trained professionals such as radiologists, physicians and technologists. This device is not intended to replace full workstations and should be used only when there is no access to a workstation. The ResolutionMD Mobile is not to be used for mammography.
Device Description
The ResolutionMD™ Mobile software is a software-based Picture Archiving and Communication System (PACS) used with general purpose computing servers and specific mobile devices. It provides for communication, storage, reformatting, rendering on the server component and communication and display of DICOM 3.0-compliant CT and MR medical images as well as reports on the mobile device.
More Information

Not Found

Not Found

No
The document describes a standard PACS system for viewing medical images and reports, with no mention of AI or ML capabilities.

No
The device is described as a Picture Archiving and Communication System (PACS) for viewing and analyzing medical images (CT and MR), intended for use as a "diagnostic, review, and analysis tool." It does not provide treatment or modify physiological functions.

Yes

The text explicitly states: "The device is intended for use as a diagnostic, review, and analysis tool".

Yes

The device is explicitly described as "software-based" and its function is centered around communication, storage, reformatting, rendering, and display of medical images and reports. While it interacts with servers and mobile devices, the device itself is the software component.

Based on the provided information, the ResolutionMD™ Mobile software is not an In Vitro Diagnostic (IVD) device.

Here's why:

  • IVD devices are used to examine specimens taken from the human body. This software deals with medical images (CT and MR) and reports, which are not specimens in the context of IVD testing.
  • The intended use is for diagnostic review and analysis of medical images. This is a different function than performing tests on biological samples.
  • The device description focuses on image handling and display. It doesn't mention any processes related to analyzing biological fluids, tissues, or other specimens.

Therefore, the ResolutionMD™ Mobile software falls under the category of medical imaging software, not an IVD.

N/A

Intended Use / Indications for Use

The ResolutionMD™ Mobile software is a software-based Picture Archiving and Communication System (PACS) used with general purpose computing servers and specific mobile devices. It provides for communication, storage, reformatting, rendering on the server component and communication and display of DICOM 3.0-compliant CT and MR medical images as well as reports on the mobile device.

The ResolutionMD Mobile provides wireless and portable access to medical images. The device is intended for use as a diagnostic, review, and analysis tool by trained professionals such as radiologists, physicians and technologists. This device is not intended to replace full workstations and should be used only when there is no access to a workstation.

The ResolutionMD Mobile is not to be used for mammography.

Product codes

LLZ

Device Description

The ResolutionMD™ Mobile software is a software-based Picture Archiving and Communication System (PACS) used with general purpose computing servers and specific mobile devices. It provides for communication, storage, reformatting, rendering on the server component and communication and display of DICOM 3.0-compliant CT and MR medical images as well as reports on the mobile device.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

DICOM 3.0-compliant CT and MR medical images

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

trained professionals such as radiologists, physicians and technologists.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Not Found

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

Not Found

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

0

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with three lines representing its wings and legs. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter of the circle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Mr. Kyle Peterson Director, Regulatory & Corporate Affairs Calgary Scientific. Inc. 1210 2014 Avenue SE. Suite 208 Calgery, AB T2G 1MB CANADA

Re: K111346

ે ગામ

Trade/Device Name: ResolutionMDTM Mobile Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 29, 2011 Received: August 1. 2011

Dear Mr. Peterson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to 'devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of

1

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Mary Pastel

Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

2

SCIENTIF

Indications for Use Statement

Calgary Scientific, Inc., Suite 208 - 1210 20th Ave. SE, Calgary, Alberta, CANADA Applicant: T2G 1M8

510(k) Number:4000602- K | | | | 346

Device Name: ResolutionMD™ Mobile

Indications for Use:

The ResolutionMD™ Mobile software is a software-based Picture Archiving and Communication System (PACS) used with general purpose computing servers and specific mobile devices. It provides for communication, storage, reformatting, rendering on the server component and communication and display of DICOM 3.0-compliant CT and MR medical images as well as reports on the mobile device.

The ResolutionMD Mobile provides wireless and portable access to medical images. The device is intended for use as a diagnostic, review, and analysis tool by trained professionals such as radiologists, physicians and technologists. This device is not intended to replace full workstations and should be used only when there is no access to a workstation.

The ResolutionMD Mobile is not to be used for mammography.

Prescription Use _ × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mary S Patil

Office of In Vitro Diagnostic Device Evaluation and Safety

510K K11346