K Number
K111342
Date Cleared
2011-11-25

(196 days)

Product Code
Regulation Number
878.4580
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Schott Light Guide Cable is indicated for use in transmitting light for illumination purposes from an illuminator to various instruments such as headlights, microscopes and endoscopes, providing illumination in body cavities during examinations or surgical procedures.

Device Description

The Light guide cable consists of glass fibers which are glued or fused into ferrules on both ends. The ends are polished to effectively transmit light and are equipped with standard endoscope connectors to transmit visible light from the light source to the intended instrument. The fibers are covered by Silicone sheathing for ease of handling and protecting the fibers from damage. The device is intended to be Ethylene Oxide (EtO) sterilized by the user. Sterilization information is contained in section 14.

AI/ML Overview

This looks like a 510(k) premarket notification for a Light Guide Cable, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with detailed acceptance criteria for a new, innovative device. Therefore, much of the requested information (like efficacy studies, sample sizes for test and training sets, number of experts for ground truth, MRMC studies, etc.) is not typically included in such a submission.

Here's a breakdown of what can be extracted and what cannot:

1. Table of acceptance criteria and reported device performance:

This document does not provide a table with specific quantitative acceptance criteria (e.g., minimum light transmission efficiency, specific durability metrics) and corresponding reported performance values. The performance testing is described at a high level.

Acceptance Criteria (Implied)Reported Device Performance
Electrical safetyConforms to appropriate FDA Recognized Consensus Standards
Sterilization validationConforms to appropriate FDA Recognized Consensus Standards
Substantial Equivalence to Predicate Device (Isolux America K991208) in:Demonstrated based on intended use, typical clinical use, and operational and fundamental technological characteristics. Any differences do not affect safety or effectiveness.
Function: Transmit light for illuminationSuccessfully transmits visible light from light source to intended instrument.
Design: Glass fibers, ferrules, polished ends, silicone sheathingMeets design specifications.

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not specified. The document states "performance testing includes electrical safety testing and sterilization validation." This typically involves a defined number of units for testing, but the specific sample size, especially for comparing performance metrics against a predicate, is not detailed.
  • Data Provenance: Not applicable in the context of device performance data for a 510(k) of this type. Performance testing is done in a lab setting by the manufacturer, not from clinical patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

Not applicable. This device is a light guide cable, not a diagnostic or AI-driven device requiring expert-established ground truth for its performance assessment in a clinical context. Its performance is evaluated through engineering and sterilization tests.

4. Adjudication method for the test set:

Not applicable. There's no adjudication needed as it's not a diagnostic or classification task.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This is not an AI-assisted diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This is a physical medical device (a cable), not an algorithm.

7. The type of ground truth used:

  • Technical Specifications/Standards: The "ground truth" for electrical safety and sterilization validation would be the relevant FDA Recognized Consensus Standards (e.g., IEC standards for electrical safety, ISO standards for sterilization).
  • Predicate Device Characteristics: For substantial equivalence, the characteristics of the Isolux America K991208 Endoscopic Fiber Optic Cable serve as the de facto "ground truth" for comparison.

8. The sample size for the training set:

Not applicable. This is not a machine learning or AI device that requires a training set.

9. How the ground truth for the training set was established:

Not applicable.

{0}------------------------------------------------

510(K) SUMMARY

NOV 2 5 2011

Schott North America Light Guide Cable (per 21CFR 807.92)

K111342
1/2

1. SUBMITTER/510(K) HOLDER

Schott North America 122 Charlton Street Southbridge, MA 01550

Contact Person:Nilesh Samant
Telephone:1-508-765-9744

Date Prepared: May 11, 2011

2. DEVICE NAME

Proprietary Name: Common/Usual Name: Classification Name:

Light Guide Cable Fiberoptic Cable Surgical Lamp

3. PREDICATE DEVICE

Endoscopic Fiber Optic Cable marketed by Isolux America, K991208

4. DEVICE DESCRIPTION

The Light guide cable consists of glass fibers which are glued or fused into ferrules on both ends. The ends are polished to effectively transmit light and are equipped with standard endoscope connectors to transmit visible light from the light source to the intended instrument. The fibers are covered by Silicone sheathing for ease of handling and protecting the fibers from damage. The device is intended to be Ethylene Oxide (EtO) sterilized by the user. Sterilization information is contained in section 14.

5. INTENDED USE

The Light Guide Cable is indicated for use in transmitting light for illumination purposes from an illuminator to various instruments such as headlights, microscopes and endoscopes, providing illumination in body cavities during examinations or surgical procedures.

{1}------------------------------------------------

K 111342 2/2

6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

The characteristics of the Light Guide Cable are substantially equivalent to the following current legally marketed predicate devices based on intended use, typical o linical use, and operational and fundamental technological characteristics.

Endoscopic Fiber Optic Cable marketed by Isolux America, K991208

A detailed side-by-side comparison of the Light Guide Cable with the identified predicate device is provided in the substantial equivalence discussion in this premarket notficulon.

7. PERFORMANCE TESTING

The Light Guide Cable performance testing includes electrical safety testing and sterilization validation to the appropriate FDA Recognized Consensus Stardands The results of this performance testing conclude that the technological characteristics have not diminished the safety and effectiveness of the Light Guide Cable when compared to the predicate device.

8. CONCLUSION

Based on the similarities in indication for use, design, functional, and operational features the Light Guide Cable has demonstrated substantial equivalence to the listed legally marketed predicate device and any differences do not affect the product's safety or effectiveness.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three parallel lines forming the wings and a snake-like form at the bottom. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

NOV 25 2011

Schott North America % Medical Device Consultants, Inc. Mr. Jeffrey Roberts 40 Plain Street North Attleboro, Massachusetts 02760

Re: K111342

Trade/Device Name: Light Guide Cable Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulatory Class: Class II Product Code: HBI Dated: November 09, 2011 Received: November 10, 2011

Dear Mr. Roberts:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

Page 2 – Mr. Jeffrey Roberts

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Erich Keith

So Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE 4.

K111342

510(k) Number (if known): Not yet assigned

Device Name:

Light Guide Cable

Indications for Use:

The Schott Light Guide Cable is indicated for use in transmitting light for illumination purposes from an illuminator to various instruments such as headlights, microscopes and endoscopes, providing illumination in body cavities during examinations or surgical procedures.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Keith

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

K111342 510(k) Number

May 11, 2011

§ 878.4580 Surgical lamp.

(a)
Identification. A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.(b)
Classification. Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.