K Number
K111300
Date Cleared
2011-08-12

(95 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Green, Orange, Violet and White Nitrile examination gloves, Powder Free, Non - Sterile are disposable device and made of Synthetic Polymer that exhibits rubber like characteristics intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Powder Free Green, Orange, Violet and White Nitrile Examination Gloves, Non Sterile is a disposable device intended for medical purposes that is worn on the examiner's to prevent contamination between patient and examiner.

AI/ML Overview

The provided document describes the performance testing and acceptance criteria for "Senstouch Nitrile Examination Gloves, Powder Free, Non Sterile". This device is a Class I medical device and the submission focuses on its physical and chemical properties rather than an AI/ML algorithm. Therefore, many of the requested categories (e.g., number of experts, adjudication methods, MRMC studies, standalone algorithm performance, training set details) are not applicable.

Here's a breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

RequirementStandardAcceptance CriteriaReported Device Performance
Physical DimensionD 6319-00aE3Length ≥ 230 mmLength = 246.77 mm
Width = 95 ± 10 mmWidth = 94.46 mm
Thickness: ≥ 0.05 mm (Finger, Palm, Cuff)Thickness: Finger = 0.150 mm, Palm = 0.095 mm, Cuff = 0.082 mm
Physical Properties (Unaged)D 6319-00aE3TS = 14 MPaTS = 15.9 MPa
UE = 500 %UE = 583.4 %
Physical Properties (Aged)D 6319-00aE3TS = 14 MPaTS = 22.3 MPa
UE = 400 %UE = 569.2 %
Freedom from PinholesD 6319-00aE3, FDA 21 CFR 800.20Acc / Rej = 3 / 40 piece found
Moisture ContentIn-house

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.