K Number
K111260
Manufacturer
Date Cleared
2012-07-26

(448 days)

Product Code
Regulation Number
862.1810
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The A/C Enzymatic Vitamin B6 Assay is intended for the quantitative in vitro diagnostic determination of pyridoxal 5'-phosphate (PLP, vitamin Bo) in EDTAhuman plasma. The device will be used to monitor PLP concentrations in plasma for aid in diagnosis of vitamin B6 deficiency. The A/C Enzymatic Vitamin B6 Assay is for IN VITRO DIAGNOSTIC USE ONLY.

Device Description

The A/C Enzymatic Vitamin B6 Assay is calibrated with external standardization and matrix-matched calibration solutions. Two sources of quality control material (A low and high level of PLP) are assaved in each run together with A/C Calibrators and samples for the verification of the accuracy and precision of the A/C Enzymatic Vitamin B6 Assay.

The A/C Enzymatic Vitamin B6 Assay uses the apo form of recombinant PLPdependent enzyme, homocysteine-a.y-lyase (rHCYase). The restoration of enzymatic activity by reconstitution of the holo-enzyme is linearly dependent on the amount of PLP bound to apo-enzyme. Nanomolar concentrations of PLP can then be measured by the conversion of millimolar concentrations of homocysteine to hydrogen sulfide, which is determined using DBPDA, the combination of which forms a chromophore, the absorbance is read with 96-well plate absorbance reader.

The A/C Enzymatic Vitamin B6 Assay is a three-step reaction with four reagents. which runs at 37°C or room temperature. The total assay takes 90 minutes. The assay can be performance on 96-well absorbance reader with 660-690m filter.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the A/C Enzymatic Vitamin B6 Assay, based on the provided 510(k) summary:

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly derived from the comparison to the predicate device. The study aims to demonstrate substantial equivalence, meaning the new device performs similarly to the already legally marketed predicate (ALPCO Vitamin B6 REA method). The performance is assessed through correlation and regression analysis, and bias.

Acceptance Criteria (Implicit)Reported Device Performance
Strong linear correlation (r) to predicate devicer = 0.909
Regression equation close to ideal y = xy = 0.969x + 7.6
Low bias/average difference compared to predicate device2.42 nmol/L (average difference between A/C and ALPCO assays)
Measurement range comparable to predicate deviceA/C: 16.3 - 189.3 nmol/L; ALPCO: 15.8 - 185.8 nmol/L
Ability to quantify PLP in EDTA-human plasmaQuantitative determination of pyridoxal 5'-phosphate (PLP) in EDTA-human plasma.
Aid in diagnosis of vitamin B6 deficiencyIntended to monitor PLP concentrations for aid in diagnosis.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Fifty-four (54) EDTA-plasma samples.
  • Data Provenance: Not explicitly stated regarding country of origin. The study appears to be retrospective, using existing EDTA-plasma samples. No information suggests prospective data collection.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not applicable and not provided in the summary. For this type of in vitro diagnostic device (IVD) assay, the "ground truth" for the test set is established by the results obtained from the predicate device itself, as the goal is to demonstrate equivalence, not diagnostic accuracy against a clinical gold standard (like pathology).

4. Adjudication Method for the Test Set

Not applicable. There was no expert adjudication involved in comparing the results between the new device and the predicate device. The comparison was a direct statistical analysis of the numerical results generated by both assays.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for diagnostic imaging devices or other assessments where human interpretation is a key component, comparing human performance with and without AI assistance. This 510(k) pertains to a laboratory assay (an IVD) where the device provides a direct quantitative result, not an interpretation requiring human readers.

6. Standalone (Algorithm-Only) Performance Study

Yes, implicit in the study. The comparison of the A/C Enzymatic Vitamin B6 Assay to the ALPCO Vitamin B6 REA assay is inherently a standalone performance study. The algorithm (assay) generates a quantitative value, which is then directly compared to the values generated by the predicate assay. There is no "human-in-the-loop" component for the measurement itself.

7. Type of Ground Truth Used

The "ground truth" for this comparative study was the results obtained from the predicate device (ALPCO Vitamin B6 REA assay). The new device's performance was evaluated by its agreement with these predicate device results. This is a common approach for IVD devices demonstrating substantial equivalence.

8. Sample Size for the Training Set

The 510(k) summary does not explicitly mention a "training set" in the context of machine learning, as this is a biochemical assay, not an AI/ML algorithm that requires training data in the typical sense. The assay is developed based on biochemical principles and optimized through research and development, but there's no data provided about a distinct "training set" for an algorithm.

9. How the Ground Truth for the Training Set Was Established

As noted above, the concept of a "training set" for establishing "ground truth" is not directly applicable in the summary provided for this type of enzymatic assay. The assay's analytical characteristics and performance are based on its underlying biochemical mechanism and empirical validation, rather than learning from a labeled training dataset.

{0}------------------------------------------------

K III 260

510 (K) Summaries

JUL 26 2012

A/C Enzymatic Vitamin B6 Assay

Summary of Safety and Effectiveness Information Supporting a Substantially Equivalent Determination

  1. Submitter's Name:

Submitter: A/C DIAGNOSTICS LLC 7917 Ostrow Street San Diego, CA 92111 Phone: (858) 654-2555 Fax: (858) 268-4175 Email: all@anticancer.com

Contact Person:

Qinghong Han M.D. Principal Investigator of the Device

Date of Summary Preparation: March 15, 2011

    1. Device Information
      Device Name: A/C Enzymatic Vitamin B6 Assay

Classification Panel: Clinical Chemistry

Device Classification: II

    1. Regulatory Information:
      Product Code: JIT, Regulation Section: 21 CFR 862.1150- Calibrator

JJX, Regulation Section: 21 CFR 862.1660-Quality Control Material (Assayed and Unassay)

{1}------------------------------------------------

    1. Predicate Device Information
    • (1) Predicate device name:
      • BUHLMAA Vitamin B6 REA American Laboratory Products CO., LTD PO BOX 451 Windham, NH 03087 Tel: (603) 893-8914
      • (2) Predicate 510(k) number K955561
  1. Information of Manufacturer Manufacturer: Bioserv Corporation 5340 Eastgate Mall San Diego, CA 92121 Tel: (858) 450-3123 Fax: (858) 450-0785 FDA establishment registration number: US FDA 2027352 Contact Person: Mary Richardson Quality Assurance Manager

  2. Statement of Intended Use

The A/C Enzymatic Vitamin B6 Assay is intended for the quantitative in vitro diagnostic determination of pyridoxal 5'-phosphate (PLP, vitamin Bo) in EDTAhuman plasma. The device will be used to monitor PLP concentrations in plasma for aid in diagnosis of vitamin B6 deficiency. The A/C Enzymatic Vitamin B6 Assay is for IN VITRO DIAGNOSTIC USE ONLY.

  1. Description of Device

The A/C Enzymatic Vitamin B6 Assay is calibrated with external standardization and matrix-matched calibration solutions. Two sources of quality control material (A low

{2}------------------------------------------------

and high level of PLP) are assaved in each run together with A/C Calibrators and samples for the verification of the accuracy and precision of the A/C Enzymatic Vitamin B6 Assay.

The A/C Enzymatic Vitamin B6 Assay uses the apo form of recombinant PLPdependent enzyme, homocysteine-a.y-lyase (rHCYase). The restoration of enzymatic activity by reconstitution of the holo-enzyme is linearly dependent on the amount of PLP bound to apo-enzyme. Nanomolar concentrations of PLP can then be measured by the conversion of millimolar concentrations of homocysteine to hydrogen sulfide, which is determined using DBPDA, the combination of which forms a chromophore, the absorbance is read with 96-well plate absorbance reader.

The A/C Enzymatic Vitamin B6 Assay is a three-step reaction with four reagents. which runs at 37°C or room temperature. The total assay takes 90 minutes. The assay can be performance on 96-well absorbance reader with 660-690m filter.

7. Method comparison

To establish equivalence to an existing device, and thus establish the safety and effectiveness, the A/C Enzymatic Vitamin Bo Assay was compared to ALPCO Vitamin B6 REA method (K# 955561).

The comparison of the A/C Enzymatic Vitamin B6 Assay to the ALPCO Vitamin B6 REA assay was carried for fifty four EDTA-plasma samples. Each sample was analyzed in duplicate with both methods. The correlation and regression analysis yielded y = 0.969x + 7.6 with a correlation coefficient of r = 0.909. The bias analysis of both methods by the difference (A/C Enzymatic Bg – ALPCO Bg) vs PLPCO Bg REA is shown in the Difference plot. The range of values was 16.3 - 189.3 nmol/L (mean = 6.3) for the A/C Enzymatic Vitamin B6 Assay and 15.8 -185.8 mmol/l (mean = 68.7) for ALPCO Vitamin B& REA Assay. The average difference exhibited by A/C Enzymatic Vitamin B6 Assay and ALPCO B6 REA in this study was 2.42 nmol/L.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three tail feathers, representing the three branches of government. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration

10903 New Hampshire Avenue Silver Spring, MD 20993

AntiCancer. Inc c.o Dr. Oinghong Han 7917 Ostrow St. San Diego, CA 92111

JUL 26 2012

k11260 Re:

Trade Name: A/C Enzymatic Vitamin B6 Assay Regulation Number: 21 CFR §862.1810 Regulation Name: Vitamin B12 test system Regulatory Class: Class II Product Codes: CDD Dated: July 2, 2012 Received: July 12, 2012

Dear Dr Han:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you dent operation an Noro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, oblease note the regulation entitled, "Misbranding by reference to premarket notification" (21 protess note are to garding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrio's (OSB's) Division of Postmarket Surveillance at (301) 01nee of Bur vinialio and 22dding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical

CITY rar 805), preaso go to implement htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...

You may obtain other general information on your responsibilities under the Act from the Tou may obtain other gefarers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm

Sincerely yours,

Countney H. Lias, Ph.D. Director

Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Form

510(k) Number (if known): K111260

Device Name: A/C ENZYMATIC VITAMIN B6 ASSAY

Indications for Use:

INDICATION FOR USE STATEMENT:

  • The A/C Enzymatic Vitamin B6 Assay is intended for the quantitative determination . of pyridoxal 5'-phosphate (PLP, vitamin Bo) in EDTA-plasma.
  • The device will monitor vitamin Bo (PLP) status in human plasma for aid in diagnosis . of vitamin B6 deficiency.
  • The A/C Enzymatic Vitamin B6 Assay is for IN VITRO DIAGNOSTIC USE . ONLY.

AND/OR Over-The-Counter Use Prescription Use

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER

PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Rute Chuler

Division Sign-Off

Office of In Vitro Diagnostic Device

Evaluation and Safety

510(k) 1112-60

Page 1 of

§ 862.1810 Vitamin B

12 test system.(a)
Identification. A vitamin B12 test system is a device intended to measure vitamin B12 in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of anemias of gastrointestinal malabsorption.(b)
Classification. Class II.