Intended Use

The angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.

The Integrated Innova - s5i system option is indicated for use in conjunction with single plane and biplane angiographic X-ray systems.

The Integrated Innova - s5i system option simplifies the clinical workflow associated with the use of Volcano s5i systems by:
(1) automatically synchronizing the patient demographic and medical exam information (patient name, date of birth, etc.) from angiographic X-ray systems with Volcano s5i systems,
(2) providing a remote access to commonly used Volcano s5i system functions from the angiographic X-ray systems user interface,
(3) displaying the Volcano s5i systems output on the monitor display solutions of the angiographic X-ray systems.

Device Description

The integrated INNOVA - s5i system option provides enhanced connectivity with Volcano's intravascular imaging and pressure system.

AI/ML Overview

The provided 510(k) summary for the "Integrated Innova - s5i system option" indicates that no clinical studies were performed to support substantial equivalence. Therefore, there is no detailed information available regarding acceptance criteria based on clinical performance, a study proving device performance against such criteria, sample sizes, expert qualifications, ground truth establishment, or multi-reader multi-case studies related to AI performance.

The submission primarily focuses on demonstrating substantial equivalence based on non-clinical tests and technological similarity to its predicate device.

Here's a breakdown of the information that can be extracted from the provided text based on your request, with an emphasis on what is not available from this specific document:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (from a clinical performance standpoint)Reported Device Performance (from a clinical performance standpoint)
Not explicitly defined in terms of clinical outcomes or metrics within this 510(k) summary.Not applicable, as no clinical studies were performed.
  • Note: The acceptance criteria for this device, as implied by the submission, revolve around its functional performance in integrating with the Volcano s5i system and maintaining the safety and effectiveness profile of the predicate device. This is assessed through non-clinical means like software validation, risk analysis, and verification/validation testing.

2. Sample size used for the test set and the data provenance:

  • Not applicable: No clinical "test set" in the context of diagnostic or AI performance was used, as no clinical studies were conducted. The device's integration and functional aspects were likely tested using internal data or simulated environments, but details are not provided in this summary.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable: As no clinical studies or test sets requiring expert ground truth were submitted, this information is not available.

4. Adjudication method for the test set:

  • Not applicable: No clinical test set to adjudicate.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No: No MRMC study was conducted. This device is not an AI algorithm for diagnostic interpretation; rather, it's an integration option for existing medical equipment to simplify workflow. Therefore, "human readers improve with AI vs without AI assistance" is not a relevant metric for this specific device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Not applicable: This device is not a standalone algorithm. It's an integration option for an existing X-ray system with an IVUS system, aiming to streamline workflow and display information.

7. The type of ground truth used:

  • Not applicable: No clinical ground truth was used for performance evaluation in this 510(k) submission. For the non-clinical tests, the "ground truth" would be against design specifications and established functional requirements.

8. The sample size for the training set:

  • Not applicable: This device is not described as utilizing machine learning or AI that would require a "training set" in the conventional sense for diagnostic algorithm development. Its function is connectivity and display.

9. How the ground truth for the training set was established:

  • Not applicable: As no training set was used in the context of AI/ML, this information is irrelevant for this specific device.

Summary from the document regarding meeting acceptance criteria:

The document explicitly states: "The subject of this premarket submission, integrated Innova - s5i system option, did not require clinical studies to support substantial equivalence."

Instead, the device's substantial equivalence was determined based on:

  • Compliance with voluntary standards: (Detailed in Section 9 of the premarket submission, though not included in this excerpt).
  • Quality assurance measures applied during development:
    • Risk Analysis
    • Software Requirements Specifications
    • Requirements Reviews
    • Software design specification document
    • Design Reviews
    • Traceability between requirements & hazards towards their associated mitigations
    • Testing on unit level (Module verification)
    • Integration testing (System verification)
    • Performance testing (Verification)
    • Safety testing (Verification)
    • Simulated use testing (Validation)

The implied acceptance criteria are that these non-clinical tests demonstrate the integrated system performs its intended functions (patient data synchronization, remote access to IVUS functions, display of IVUS output) reliably, safely, and effectively without introducing new safety concerns or compromising the performance of the predicate device. The "study" proving this would be the compilation of results from the various verification and validation activities listed above.

{0}------------------------------------------------

K111209
P. 1 of 2

JUL 26 2011

Image /page/0/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined and enclosed within a circular shape. The letters and the circle are black, creating a strong contrast against the white background.

GE Healthcare

Integrated Innova - s5i system option - 510(k) Premarket Notification

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

Date: April 27th, 2011
Submitter:GE HEALTHCAREGE Medical Systems SCS283, rue de la minière78530 BUCFRANCE
Primary Contact Person:Fayçal KHERRARegulatory Affairs LeaderGE Medical Systems SCS283, rue de la minière78530 BUCFRANCET: +33 1 30 70 40 82Email: Faycal.kherra@ge.com
Secondary Contact Person:Carol AlloianRegulatory Affairs LeaderGE Healthcare, QARA Regions - Americas9900 W innovation driveWauwatosa, WI, USA, 53226-4856T: (847) 244-8327F: (847) 589-8524Email: carol.alloian@ge.com
Device/Trade Name:Integrated Innova - s5i system option (Formerly known as InnovaSystem with IVUS Option)
Common/Usual Name:Integrated Innova - s5i system option (Formerly known as InnovaSystem with IVUS Option)
Classification Names:Product Code:IZI : SYSTEM, X-RAY, ANGIOGRAPHICMQB : SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
Predicate Device(s):K061163 : INNOVA x1x1¹⁰ & X100¹⁰ systems with IVUS option
Device Description:The integrated INNOVA - s5i system option provides enhancedconnectivity with Volcano's intravascular imaging and pressure system.
Intended Use:The angiographic X-ray systems are indicated for use in generatingfluoroscopic and rotational images of human anatomy forcardiovascular, vascular and non-vascular, diagnostic and interventionalprocedures.
The Integrated Innova - s5i system option is indicated for use inconjunction with single plane and biplane angiographic X-ray systems.
The Integrated Innova - s5i system option simplifies the clinical workflowassociated with the use of Volcano s5i systems by:(1) automatically synchronizing the patient demographic andmedical exam information (patient name, date of birth, etc.) fromangiographic X-ray systems with Volcano s5i systems,(2) providing a remote access to commonly used Volcano s5i systemfunctions from the angiographic X-ray systems user interface,(3) displaying the Volcano s5i systems output on the monitor displaysolutions of the angiographic X-ray systems.
Technology:The integrated Innova - s5i system option employs the samefundamental scientific technology as its predicate devices.
Determination of SubstantialEquivalence:Summary of Non-Clinical Tests:The integrated Innova - s5i system option complies with voluntarystandards as detailed in Section 9 of this premarket submission. Thefollowing quality assurance measures were applied to the developmentof the system:• Risk Analysis
• Software Requirements Specifications• Requirements Reviews• Software design specification document.
• Design Reviews• Traceability between requirements & hazards towards their associatedmitigations as appropriate and V&V
• Testing on unit level (Module verification)• Integration testing (System verification)• Performance testing (Verification)• Safety testing (Verification) Simulated use testing (Validation)
Summary of Clinical Tests:The subject of this premarket submission, integrated Innova - s5i systemoption, did not require clinical studies to support substantial equivalence.
Conclusion:GE Healthcare considers the integrated Innova - s5i system option to beas safe and as effective as the predicate devices, and its performance issubstantially equivalent to the predicate devices.

{1}------------------------------------------------

KIII 209
P. 2 of Z

GE Healthcare

Integrated Innova - s5i system option - 510(k) Premarket Notification

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular pattern around the eagle. The eagle is facing to the right. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Mr. Faycal Kherra Regulatory Affairs Leader GE Healthcare - GE Medical Systems SCS 283, rue de la minière BUC, 78530 FRANCE

JUL 2 6 2311

Re: K111209

Trade/Device Name: Integrated Innova-s5i system option Regulation Number: 21 CFR 892.1600 Regulation Name: Angiographic x-ray system Regulatory Class: II Product Code: IZI Dated: April 27, 2011 Received: April 29, 2011

Dear Mr. Kherra:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becales on and the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to conimores that have been reclassified in accordance with the provisions of the Federal Food, Drug, devices that have boon roomstiled in assee approval of a premarket approval application (PMA). and Cosmetic (110.) that to hovice, subject to the general controls provisions of the Act. The I ou mayy are sprovisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such 11 your device to to to stime major regulations affecting your device can be found in Title 21, accitional corners Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I least be advised hit i Dri i issualite our device complies with other requirements of the Act that I Dri has intate a and regulations administered by other Federal agencies. You must or any I caeral statutes and regencients, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{3}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours.

Mary Patel

Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

KI11209

  • 510(k) Number (if known): K || | 20 9
  • · Device Name: Integrated Innova s5i system option

• Indications for Use:

The angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for cardiovascular and non-vascular, diagnostic and interventional procedures.

The Integrated Innova - s5i system option is indicated for use in conjunction with single plane and biplane angiographic X-ray systems.

The Integroted Innova - 55i system option simplifies the clinical workflow associated with the use of Volcano s5i systems by:

(1) automatically synchronizing the patient demographic and medical exam information lpatient name, date of birth, etc.) from angiographic X-ray systems with Volcano s5i systems,

(2) providing a remote access to commonly used Volcano s5i system functions from the angiographic X-ray systems user interface,

(3) displaying the Volcano s5i systems output on the monitor display solutions of the angiographic X-ray systems.

Prescription Use X ___________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mancy S. Patil

Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety

510K K11209

Prescription Use (Per 21 CFR 801.109)

§ 892.1600 Angiographic x-ray system.

(a)
Identification. An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a contrast medium. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.