K Number
K111206
Manufacturer
Date Cleared
2011-10-04

(158 days)

Product Code
Regulation Number
862.3040
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Teco Saliva Alcohol Test is a semi-quantitative screening test for measuring alcohol in human saliva. This strip specifies the relative Blood Alcohol Concentration (BAC) at 0.0%, 0.02%, 0.04%, 0.08%, and 0.30% cut-off levels. Results can be used for the indication of alcohol intoxication. For in vitro diagnostic use only.
The Teco Saliva Alcohol Test is a screening test for the semi-quantitative detection of the presence of ethyl alcohol in human saliva. The test strip indicates relative Blood Alcohol Concentrations (BAC) at 0.0%, 0.02%, 0.04%, 0.08%, and 0.30% cut-off levels. Results are used for the diagnosis of alcohol intoxication. The test is intended for both prescription and over-the-counter in vitro diagnostic use. The assay is a disposable test for one-time use.

Device Description

The Teco Saliva Alcohol Test is used to detect the amount of alcohol in the saliva after consumption. This is a semi-quantitative test that is read visually. The test strip determines the relative Blood Alcohol Concentration at 5 cut-off levels (negative, 0.02%, 0.04%, 0.08% and 0.3%). The device consists of a test strip (1 test strip per foil pouch). The resultant color on the test strip is compared to the color blocks printed on the pouch.
The Teco Saliva Alcohol Test is a plastic strip tipped with a reactive pad containing tetramethylbenzidine, alcohol oxidase, and peroxidase. The test is utilizes two coupled enzymatic reactions to create the visual changes in the presence of varying amounts of saliva alcohol. The first reaction consists of an oxidation reaction where the ethanol in the saliva is converted to aldehyde by the alcohol oxidase. The second reaction is the oxidation of TMB, the color indicator substrate, and is catalyzed by the enzyme peroxidase. The color change of the pad corresponds to the amount of present alcohol in the sample. The results are read by comparing the colored pad to the color chart found on the strip package

AI/ML Overview

This document describes the Teco Saliva Alcohol Test, a semi-quantitative screening test for measuring alcohol in human saliva. The device aims to determine relative Blood Alcohol Concentration (BAC) at specific cut-off levels.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state numerical acceptance criteria in a table format for performance metrics like sensitivity, specificity, or accuracy. It primarily focuses on demonstrating substantial equivalence to a predicate device (ACON Mission Saliva Alcohol Test).

However, it implicitly indicates that the device's performance is acceptable if it can achieve "comparable testing data" to the predicate device and "perform satisfactorily" in conjunction with various studies.

Performance MetricAcceptance Criteria (Implied)Reported Device Performance
Overall PerformanceComparable testing data to the legally marketable ACON Mission Saliva Alcohol Test. Performance is satisfactory when used appropriately.Clinical study results indicate that intended users were able to obtain comparable testing data when using the Teco Saliva Alcohol Test and the ACON Mission Saliva Alcohol test. Performance characteristics were verified through method comparison, precision, linearity, shelf life, and stress studies.

2. Sample Size Used for the Test Set and Data Provenance

The document states "Clinical study results indicate that the intended users were able to obtain comparable testing data." However, it does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective nature) for the clinical study.

3. Number of Experts and Qualifications for Ground Truth Establishment

The document does not provide information on the number of experts used to establish ground truth for the test set or their qualifications. The primary ground truth for comparison appears to be the results obtained from the predicate device.

4. Adjudication Method for the Test Set

The document does not describe an adjudication method for the test set. It mentions comparison with a predicate device, suggesting a direct comparison rather than an expert consensus/adjudication process for the Teco Saliva Alcohol Test's results themselves.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study. The evaluation focuses on the device's performance compared to a predicate, not on how human readers' performance improves with or without the device's assistance.

6. Standalone (Algorithm Only) Performance Study

The Teco Saliva Alcohol Test is described as a "visually read color change" device with the resultant color compared to color blocks printed on the pouch. This intrinsically involves human interpretation. Therefore, a standalone (algorithm only without human-in-the-loop performance) study is not applicable and not mentioned.

7. Type of Ground Truth Used

The primary "ground truth" used for evaluating the Teco Saliva Alcohol Test is the performance of a legally marketed predicate device (ACON Mission Saliva Alcohol Test). The clinical study aimed to show "comparable testing data" between the two devices. The predicate device itself would have had its own established ground truth (likely laboratory confirmation of BAC levels).

8. Sample Size for the Training Set

The document does not mention a training set or its sample size. This device is a chromogenic assay; it is not a machine learning or AI-based algorithm that typically requires explicit training datasets.

9. How Ground Truth for the Training Set Was Established

As there is no mention of a training set in the context of this device, the question of how its ground truth was established is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/32 description: The image shows a black hexagon with the letters 'TC' in white inside. The letters are in a simple, sans-serif font, with the 'T' slightly larger than the 'C'. The hexagon is solid black, providing a strong contrast to the white letters.

005 - 510(k) Summary

  • 4 2011

A. 510(k) Number:

  • B. Purpose for Submission: New Device
  • C. Analyte:
    • Saliva Alcohol
  • D. Type of Test:

Semi-quantitative, chromogenic assay; visually read color change

  • E. Applicant:
    Teco Diagnostics

  • F. Trade Name: Teco Saliva Alcohol Test

G. Regulatory Information:

    1. Regulation section:
      21 CFR 862.3040, Alcohol Test System
    1. Classification: Class II
    1. Product code: DIC
    1. Panel:

Toxicology (91)

H. Intended Use:

2. Indication(s) for use:

  • The Teco Saliva Alcohol Test is a semi-quantitative screening test for measuring alcohol in human saliva. This strip specifies the relative Blood Alcohol Concentration (BAC) at 0.0%, 0.02%, 0.04%, 0.08%, and 0.30% cut-off levels. Results can be used for the indication of alcohol intoxication. For in vitro diagnostic use only.
    1. Special conditions for use statement(s): The assay is a disposable test for one-time use.

The strip is for over-the-counter use.

I. Device Description:

The Teco Saliva Alcohol Test is used to detect the amount of alcohol in the saliva after consumption. This is a semi-quantitative test that is read visually. The test strip determines the relative Blood Alcohol Concentration at 5 cut-off levels (negative, 0.02%, 0.04%, 0.08% and 0.3%). The device consists of a test strip (1 test strip per foil pouch). The resultant color on the test strip is compared to the color blocks printed on the pouch.

J. Substantial Equivalence Information:

    1. Predicate device name(s):
      Mission Saliva Alcohol Test; Acon Inc.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows a black hexagon with the letters 'TC' in white inside. The letters are in a sans-serif font, with the 'T' slightly larger than the 'C'. The hexagon is oriented with one point facing downwards.

1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 463-1111 Toll Free: 1-800-222-9880 Fax: (714) 463-1169 www.tecodiag.com

005 - 510(k) Summary

    1. Predicate 510(k) number(s): K093879
    1. Comparison with predicate:
Trade NameTeco Saliva Alcohol TestACON Mission Test
ManufacturerTeco DiagnosticsACON Labs
510(k) numberN/AK093879
Intended use/indicationsfor useThe Saliva Alcohol Test is a rapidscreening test used for measuringalcohol in human saliva. Results areused in the diagnosis of alcoholintoxication.Same
MethodologyChromogenicSame
EnzymeAlcohol oxidaseSame
Sample TypesSalivaSame
Measuring UnitsBlood Alcohol Concentration(BAC%)Same
Measuring RangeNegative0.02%0.04%0.08%0.30%Same
ResultSemi-quantitativeSame
Read Time Window2 minutes2 to 3 minutes
Pad MethodologyCoupled enzymatic reactions usingalcohol oxidase on ethyl alcohol in thepresence of peroxidase and enzymesubstrate such as tetramethylbenzidine(TMB).Same

K. Test Principle:

The Teco Saliva Alcohol Test is a plastic strip tipped with a reactive pad containing tetramethylbenzidine, alcohol oxidase, and peroxidase. The test is utilizes two coupled enzymatic reactions to create the visual changes in the presence of varying amounts of saliva alcohol. The first reaction consists of an oxidation reaction where the ethanol in the saliva is converted to aldehyde by the alcohol oxidase. The second reaction is the oxidation of TMB, the color indicator substrate, and is catalyzed by the enzyme peroxidase. The color change of the pad corresponds to the amount of present alcohol in the sample. The results are read by comparing the colored pad to the color chart found on the strip package

L. Clinical Performance Characteristics:

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows a black hexagon with the letters "TC" in white inside. The letters are in a bold, sans-serif font, with the "T" slightly smaller than the "C". The hexagon is solid black, providing a strong contrast to the white letters.

1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 463-1111 Toll Free: 1-800-222-9880 Fax: (714) 463-1169 www.tecodiag.com

005 - 510(k) Summary

Clinical study results indicate that the intended users were able to obtain comparable testing data when using the Teco Saliva Alcohol Test and the legally marketable ACON Mission Saliva Alcohol test.

M. Conclusion:

The performance characteristics of the Teco Saliva Alcohol Test were verified using method comparison, precision, linearity, shelf life, and stress studies. Testing results indicate that the Teco Saliva Alcohol Test compared with the ACON Mission Saliva Alcohol tests perform satisfactorily when used appropriately, as outlined in the package insert.

The clinical studies results demonstrate a substantial equivalency on performance between Teco Saliva Alcohol Test with predicate test, ACON Mission Alcohol Saliva test.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an eagle or bird-like figure with three parallel lines above it.

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring. MD 20993

BET 04 2011-

Teco Diagnostics c/o Mr. James Li 1268 N. Lakeview Avenue Anaheim. CA 92807

રિદ: K111206

Trade Name: Teco Saliva Alcohol Test Regulation Number: 21 CFR 862,3040 Regulation Name: Alcohol test system Regulatory Class: Class II Product Code: DIC Dated: August 12, 2011 Received: August 17, 2011

Dear Mr. Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

Page 2 -

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

signature

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K111206

Device Name: Teco Saliva Alcohol Test

Indications For Use:

The Teco Saliva Alcohol Test is a screening test for the semi-quantitative detection of the presence of ethyl alcohol in human saliva. The test strip indicates relative Blood Alcohol Concentrations (BAC) at 0.0%, 0.02%, 0.04%, 0.08%, and 0.30% cut-off levels. Results are used for the diagnosis of alcohol intoxication. The test is intended for both prescription and over-the-counter in vitro diagnostic use. The assay is a disposable test for one-time use.

Prescription Use X (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use X (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k)_K 111200

Image /page/5/Picture/13 description: The image shows the words "Page" and "of" with blanks next to them. The first blank is next to the word "Page", and the second blank is next to the word "of". The number 1 is written in the second blank.

§ 862.3040 Alcohol test system.

(a)
Identification. An alcohol test system is a device intented to measure alcohol (e.g., ethanol, methanol, isopropanol, etc.) in human body fluids (e.g., serum, whole blood, and urine). Measurements obtained by this device are used in the diagnosis and treatment of alcohol intoxication and poisoning.(b)
Classification. Class II.