(279 days)
The Tracer Implantation Kit is indicated for use to radiographically and radioactively mark soft tissue for future therapeutic procedures.
The Tracer is indicated for permanent implantation in the prostate.
The Tracer is intended to be used in conjunction with the RealEye system as an adjunct in treatment planning and radiation therapy, to align and monitor the patient's position relative to the isocenter of a linear accelerator.
The IndeX Implantation Kit is indicated for use to radiographically mark soft tissue for future therapeutic procedures.
The IndeX is intended to be used as a radiographic fiducial marker for visualization using x-ray, CT, or Cone Beam CT.
The Tracer is a Platinum/Iridium radioactive soft tissue fiducial for implantation in the prostate, in or near the treatment target for radiotherapy treatments. The Tracer emits photons that enable its localization by the Navotek RealEye system that is designed to track localized gamma-emitting radioactive sources.
The implantation of the Tracer is performed using an Implantation Device, which is supplied with the Tracer pre-loaded within it. Therefore, the Tracer is supplied as a 'Tracer Implantation Kit' which includes the Tracer and the Implantation Device. The Tracer lmplantation Kit is provided in a single package, sterile (by gamma radiation) and ready for use. It is intended for single use only.
The IndeX is identical to the Tracer in all respects except that it is not radioactive. It is intended for use to radiographically mark soft tissue and to be visible in x-ray, CT, and Cone Beam CT images.
{"1. A table of acceptance criteria and the reported device performance": "The document describes performance testing that demonstrated the Tracer Implantation Device performs "according to specifications and function as intended." It also states that "differences in technology compared to the predicate device do not impact safety or effectiveness." However, specific quantifiable acceptance criteria (e.g., minimum accuracy rates, success rates, or ranges of performance metrics) are not explicitly provided in the document. Therefore, a table comparing acceptance criteria to reported performance cannot be fully constructed.", "2. Sample sized used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)": "The document mentions "clinical testing" but does not specify the sample size used for this testing. It also does not provide information on the country of origin of the data or whether the study was retrospective or prospective.", "3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)": "The document does not provide information regarding the number of experts used to establish ground truth or their qualifications.", "4. Adjudication method (e.g., 2+1, 3+1, none) for the test set": "The document does not specify any adjudication method for the test set.", "5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance": "A Multi-Reader, Multi-Case (MRMC) comparative effectiveness study was not performed or mentioned. The device is a fiducial marker, not an AI-assisted diagnostic tool for human readers.", "6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done": "This is a physical medical device (fiducial marker and implantation kit), not a software algorithm. Therefore, a standalone algorithm performance study is not applicable.", "7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)": "The document mentions "bench, animal, and clinical testing" as part of the performance evaluation. For a fiducial marker, ground truth would likely involve direct physical measurements (for bench testing), imaging verification (for animal and clinical testing), and potentially long-term follow-up for outcomes related to marker stability and efficacy in therapy. However, the document does not explicitly state the type of "ground truth" used for each test.", "8. The sample size for the training set": "This is a physical medical device. The concept of a "training set" as used in AI/ML is not applicable here.", "9. How the ground truth for the training set was established": "This question is not applicable as the device is not an AI/ML algorithm that requires a training set."}
{0}------------------------------------------------
510(k) SUMMARY
JAN 1 0 2012
Navotek Medical Ltd. - Tracer Implantation Kit
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Navotek Medical Ltd.
POB 201, 5 HaCarmel St.
Yokneam 20692, Israel
Phone: +972 72 270 9020 Facsimile: +972 72 270 9021 Contact Person: Moshe Solomon Date Prepared: March 30, 2011
Name of Device and Name/Address of Sponsor
Tracer Implantation Kit and IndeX Implantation Kit Navotek Medical Ltd. POB 201, 5 HaCarmel St. Yokneam 20692, Israel
Common or Usual Name
Soft tissue fiducial marker
Classification Name
Medical linear accelerator
Regulation Number
{1}------------------------------------------------
CFR 892.5050 (Medical charged-particle radiation therapy system)
Product Code
IYE
Device Class
=
Predicate Devices
Navotek RealEye System and Calypso 4D Localization System (Tracer)
Visicoil soft tissue marker (IndeX)
Intended Use / Indications for Use
Tracer:
The Tracer Implantation Kit is indicated for use to radiographically and radioactively mark soft tissue for future therapeutic procedures.
The Tracer is indicated for permanent implantation in the prostate.
The Tracer is intended to be used in conjunction with the RealEye system as an adjunct in treatment planning and radiation therapy, to align and monitor the patient's position relative to the isocenter of a linear accelerator.
IndeX:
The IndeX Implantation Kit is indicated for use to radiographically mark soft tissue for future therapeutic procedures.
The IndeX is intended to be used as a radiographic fiducial marker for visualization using x-ray, CT, or Cone Beam CT.
Technological Characteristics
The Tracer is a Platinum/Iridium radioactive soft tissue fiducial for implantation in the prostate, in or near the treatment target for radiotherapy treatments. The Tracer emits
{2}------------------------------------------------
photons that enable its localization by the Navotek RealEye system that is designed to track localized gamma-emitting radioactive sources.
The implantation of the Tracer is performed using an Implantation Device, which is supplied with the Tracer pre-loaded within it. Therefore, the Tracer is supplied as a 'Tracer Implantation Kit' which includes the Tracer and the Implantation Device. The Tracer lmplantation Kit is provided in a single package, sterile (by gamma radiation) and ready for use. It is intended for single use only.
The IndeX is identical to the Tracer in all respects except that it is not radioactive. It is intended for use to radiographically mark soft tissue and to be visible in x-ray, CT, and Cone Beam CT images.
Performance Data
Performance testing demonstrates that the Tracer Implantation Device perform according to specifications and function as intended. Performance testing further demonstrates that differences in technology compared to the predicate device do not impact safety or effectiveness.
Biocompatibility testing was performed according to ISO 10993- Part 1 (2003) and the FDA Blue book memorandum G95-1. Testing included platinum leakage testing with respect to platinum toxicity and radioactive leakage testing to demonstrate that the Tracer meets the requirements for a sealed source according to ISO 2919 with classification ISO/99/C53211(X). Bench, animal, and clinical testing also was conducted and demonstrated substantial equivalence to the predicate device.
All testing that is not directly related to radioactivity applies to both the Tracer and the IndeX.
Substantial Equivalence
The Tracer and its predicates have the same intended use and similar indications, and similar technological characteristics and principles of operation. The identified technological differences do not raise any new types of safety or effectiveness questions. Performance testing demonstrates that the identified technological differences between the Tracer and the predicate device do not impact safety or effectiveness. In particular, bench, animal, and clinical testing results demonstrate that the safety and performance of the Tracer are similar
{3}------------------------------------------------
to that of its predicate device. Thus, the Tracer is substantially equivalent to its predicate device.
The IndeX and its predicate have the same intended use and indications, and similar technological characteristics and principles of operation. The identified technological differences do not raise any new types of safety or effectiveness questions. Performance testing demonstrates that the identified technological differences between the IndeX and the predicate device do not impact safety or effectiveness. In particular, bench, animal, and clinical testing results demonstrate that the safety and performance of the IndeX are similar to that of its predicate device. Thus, the IndeX is substantially equivalent to its predicate device.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing or supporting another, with flowing lines suggesting movement or care.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Navotek Medical Ltd. % Mr. Steven B. Datlof, M.D., J.D. Regulatory Counsel Hogan Lovells US LLP 1835 Market Street, 29th Floor PHILADELPHIA PA 19103
JAN 1 0 2012 ·
Re: K110971
Trade/Device Name: Tracer Implantation Kit and IndeX Implantation Kit Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE, NEU Dated: December 2, 2011 Received: December 2, 2011
Dear Mr. Datlof:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice medical device related adverse oviality systems (QS) regulation (21 CFR Part 820). This letter requirements as be forth in keting your device as described in your Section 510(k) premarket with anow you to begin finding of substantial equivalence of your device to a legally marketed nonicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 11 you desire specific ad reserved in Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket 5 150. Thise, product 1000 and 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary Pastel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use Statement
510(k) Number (if known): K110971
Device Name: Tracer Implantation Kit and IndeX Implantation Kit
Indications for Use:
The Tracer Implantation Kit is indicated for use to radiographically and radioactively mark soft tissue for future therapeutic procedures.
The Tracer is indicated for permanent implantation in the prostate.
The Tracer is intended to be used in conjunction with the RealEye system as an adjunct in treatment planning and radiation therapy, to align and monitor the patient's position relative to the isocenter of a linear accelerator.
The IndeX Implantation Kit is indicated for use to radiographically mark soft tissue for future therapeutic procedures.
The IndeX is intended to be used as a radiographic fiducial marker for visualization using x-ray, CT, or Cone Beam CT.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary Patel
(Division Sign-Off)
Division of Radiological Devices
Diagnostic Device Evaluation and Safety
ffice of In Vitro Diagno
K110971
Page 1 of 1
§ 892.5050 Medical charged-particle radiation therapy system.
(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.