(100 days)
The Royal Imperial Nitrile Examination Gloves Textured Powder-Free Blue is a single-use disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
The Royal Imperial Nitrile Examination Gloves Textured Powder-Free Blue is a class I medical device classified under Product Code LZA. It is a single-use disposable device that meets all requirements of ASTM D 6319-10.
The Royal Imperial Nitrile Examination Gloves Textured Powder-Free Blue were evaluated for substantial equivalence based on a comparison to a predicate device and adherence to recognized consensus standards. The study demonstrates the device meets the acceptance criteria through non-clinical testing.
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria | Device Performance |
|---|---|
| Compliance with ASTM D6319-10 (Standard Specification for Nitrile Examination Gloves for Medical Application) | The device "[meets] all requirements of ASTM D 6319-10." |
| Compliance with ASTM D3578-05 (Standard Specification for Rubber Examination Gloves) | The device "was tested in accordance, and found to be in compliance, with ASTM D3578-05." |
| Compliance with ASTM D6124-06 (Standard Test Method for Residual Powder on Medical Gloves) | The device "was tested in accordance, and found to be in compliance, with ASTM D6124-06." |
| Compliance with ISO 10993-10 (Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization) | The device "was tested in accordance, and found to be in compliance, with [...] ISO 10993-10." |
| Substantial Equivalence to predicate device (Ultrawin SDN BHD Non-Sterile, Powder Free Nitrile Examination Gloves - K090828) | The device "are substantially equivalent to the Ultrawin SDN BHD Non-Sterile, Powder Free Nitrile Examination Gloves (K090828)" and "passed all the tests associated with the FDA-recognized consensus standards listed above". |
2. Sample size used for the test set and the data provenance:
The document does not specify the exact sample sizes used for each test within the ASTM and ISO standards. However, the tests were "non-clinical testing" as indicated by "Summary of Non-clinical Testing Data." The provenance of the data is from the manufacturer's testing of their product, Koon Seng SDN. BHD., located in Tangkak, Johor, Malaysia. The testing is reported retrospectively as part of the 510(k) submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not explicitly provided in the document. For these types of glove tests, the "ground truth" is typically established by the standardized methods and criteria defined within the ASTM and ISO standards themselves, rather than through expert consensus in the same way clinical diagnostic studies would. The testing would be performed by qualified laboratory personnel following these established protocols.
4. Adjudication method for the test set:
Adjudication methods like 2+1 or 3+1 are not applicable here. The evaluation involves standardized laboratory tests against predefined physical and chemical specifications in the ASTM and ISO standards. The results of these tests are objectively measured and compared directly to the passing criteria within the standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This type of study is relevant for diagnostic imaging or AI-assisted clinical decision-making devices, not for patient examination gloves. The evaluation of these gloves focuses on physical and biological material properties and performance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This is not applicable as the device is a medical device (examination gloves), not an algorithm or AI system. The performance evaluated is the physical properties and biocompatibility of the gloves themselves, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
The "ground truth" for the device's performance is established by the specified requirements and testing methodologies outlined in the referenced FDA-recognized consensus standards: ASTM D3578-05, ASTM D6124-06, ASTM D6319-10, and ISO 10993-10. These standards define the acceptable performance characteristics (e.g., tensile strength, elongation, barrier integrity, residual powder, biocompatibility) that the gloves must meet.
8. The sample size for the training set:
This is not applicable, as there is no training set for a physical medical device like examination gloves. Training sets are used in the context of machine learning or AI models.
9. How the ground truth for the training set was established:
This is not applicable, as there is no training set.
{0}------------------------------------------------
K110960
JUL 1 4 2011
Royal Imperial Nitrile Examination Gloves -Textured. Powder-Free Blue. 510(k). Summary__________________________________________________________________________________________________________________________________
Date prepared: June 3, 2011
Name and Address of Sponsor
Koon Seng SDN. BHD. 16058, Kaw. Perindustrian Tangkak, 84900 Tangkak, Johor, Malaysia Ph: 606-979 1778 Fax: 606-979 1779 Email: koonsen@po.jaring.my
Name and Address of Official Correspondent
Regulatory Insight, Inc. 5401 S. Cottonwood Ct. Greenwood Village, Colorado 80121 Contact: Mr. Kevin Walls, RAC Tel: 720-962-5412 Fax: 720-962-5413 Email: kevin@reginsight.com
Device Name
Trade Name: Royal Imperial Nitrile Examination Gloves Textured Powder-Free Blue Common Name: Patient examination glove Classification Name: Polymer patient examination glove
Classification, Panel and Product Code
Class 1. General Hospital, LZA
Substantial Equivalence
The Royal Imperial Nitrile Examination Gloves Textured Powder-Free Blue are substantially equivalent to the Ultrawin SDN BHD Non-Sterile, Powder Free Nitrile Examination Gloves (K090828).
Device Description
The Royal Imperial Nitrile Examination Gloves Textured Powder-Free Blue is a class I medical device classified under Product Code LZA. It is a single-use disposable device that meets all requirements of ASTM D 6319-10.
{1}------------------------------------------------
KII0960
Indications for Use
The Royal Imperial Nitrile Examination Gloves Textured Powder-Free Blue is a singleuse disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Intended Use
The Royal Imperial Nitrile Examination Gloves Textured Powder-Free Blue is intended to be worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Summary of Non-clinical Testing Data
The Roval Imperial Nitrile Examination Gloves Textured Powder-Free Blue was tested in accordance, and found to be in compliance, with ASTM D3578-05, ASTM D6124-06, ASTM D6319-10 and ISO 10993-10. The test reports and the Forms FDA 3654 may be found in the body of the 510(k) premarket notification.
Conclusion Statement
Based on the fact that the Royal Imperial Nitrile Examination Gloves Textured Powder-Free Blue passed all the tests associated with the FDA-recognized consensus standards listed above, the Royal Imperial Nitrile Examination Gloves Textured Powder-Free Blue are substantially equivalent to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three curved lines representing its wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The logo is black and white and appears to be a scanned image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Koon Seng SDN. BHD. C/O Mr. Kevin Walls, RAC Principal Consultant Regulatory Insight, Incorporated 5401 South Cottonwood Court Greenwood Village, Colorado 80121
JUL 1 4 2011
Re: K110960
Trade/Device Name: Royal Imperial Nitrile Examination Gloves Textured Powder-Free Blue Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: June 3, 2011 Received: June 7, 2011
Dear Mr. Walls:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Walls
:
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/Repo rtaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Anthony D, Owen
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): K110960
Device Name: Royal Imperial Nitrile Examination Gloves Textured Powder-Free Blue
Indications for Use: The Royal Imperial Nitrile Examination Gloves Textured Powder-Free Blue is a single-use disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use × (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ronie Rangsuluri
·Division Sign-Off) Division of Anesthesiology, General Hospital infection Control, Dental Devices
510(k) Number: K110960
Page 1 of 1
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.