K Number
K110915
Device Name
CERVICAL CAGE
Manufacturer
Date Cleared
2011-07-22

(112 days)

Product Code
Regulation Number
888.3080
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cervical Cage is intended for spinal fusion procedures at one level (C2 to T1) in skeletally mature patients with degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implants are intended to be packed with autogenous bone graft. The Cervical Cage is intended to be used with a supplemental fixation system.

Device Description

The Cervical Caqe is an intervertebral body fusion orthosis intended to be used in cervical spinal fusion surgery. It provides mechanical support to the spine and protects the bone graft from excessive loads so that bone healing can occur.

AI/ML Overview

Cervical Cage Acceptance Criteria and Study Details

The provided document describes the Eisertech, LLC Cervical Cage, an intervertebral fusion device. The acceptance criteria and supporting study are focused on the device's mechanical performance and substantial equivalence to legally marketed predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Implied from Design Requirements)Reported Device Performance
Maintain spacing between two vertebral bones following cervical discectomy until fusion occurs.The device output data showed performance meeting or exceeding the design input requirements for all conditions. (Based on testing per ASTM F2077 - static axial compression, dynamic axial compression, static torsion, dynamic torsion, and ASTM F2267 - resistance to subsidence, and expulsion testing). The Cervical Cage performs at least as safely and effectively as a legally marketed predicate device.
Withstand biomechanical loads expected in the spine."Test results relative to those loading conditions (e.g. design output data) were compared to the design input criteria. The device output data showed performance meeting or exceeding the design input requirements for all conditions." (Specific load values or thresholds are not detailed in the provided text, but the general claim of meeting or exceeding is made).
Provide mechanical support and protect bone graft from excessive loads."The Cervical Cage is an intervertebral body fusion orthosis intended to be used in cervical spinal fusion surgery. It provides mechanical support to the spine and protects the bone graft from excessive loads so that bone healing can occur." (This is a statement of intended function based on the design, with the performance confirmation being the overall mechanical testing).

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: The document does not explicitly state the sample size used for the mechanical testing of the Cervical Cage. The phrase "the Cervical Cage was tested" implies a singular or small number of devices were tested for each scenario.
  • Data Provenance: The data is from prospective bench testing of the device prototypes, conducted by Eisertech, LLC. The "country of origin of the data" is not explicitly stated but can be inferred as the USA, where Eisertech, LLC is located.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • This information is not applicable to this type of device and study. The "ground truth" for the mechanical performance of an implant like the Cervical Cage is established through objective, standardized biomechanical testing against defined loads and established material properties, not through expert human interpretation of images or clinical outcomes.

4. Adjudication Method for the Test Set:

  • This information is not applicable. Adjudication methods are typically relevant for studies involving human interpretation or clinical endpoints (e.g., radiology reads). For biomechanical testing, the results are typically quantitative and objective measurements.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation of medical images. The Cervical Cage is a surgical implant, and its evaluation focuses on its mechanical integrity and biocompatibility.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

  • Yes, the testing described is a standalone (algorithm only without human-in-the-loop performance). The "algorithm" in this context refers to the device itself and its inherent mechanical properties, which were tested objectively without human involvement in interpreting the results beyond reading the quantitative data.

7. The Type of Ground Truth Used:

  • The ground truth used for this mechanical performance study is based on engineering design requirements and established biomechanical standards (ASTM F2077 and ASTM F2267). These standards define the parameters and methodologies for testing intervertebral body fusion devices, and the device's ability to "meet or exceed" these accepted engineering benchmarks serves as the "ground truth" for its mechanical fitness.

8. The Sample Size for the Training Set:

  • This information is not applicable. The Cervical Cage is a physical medical device, not an AI or software algorithm that undergoes machine learning training. There is no concept of a "training set" in this context. The design of the device would have involved engineering principles and potentially iterative prototyping and testing, but not in the sense of a machine learning training set.

9. How the Ground Truth for the Training Set Was Established:

  • This information is not applicable, as there is no "training set" for this type of device. The "ground truth" for the device's design and mechanical properties would be established through engineering calculations, material science, and the aforementioned ASTM standards.

{0}------------------------------------------------

K110915 Page 1/2

JUL 222 2011

Company:Eisertech, LLCSan Diego, California 92103
Contact:Lukas Eisermannlukas@eisertech.com888-262-2817x101
Trade Name:Cervical Cage
Common Name:CervicalIntervertebral FusionDevice with Bone Graft
Classification Name:Orthosis, spinal intervertebral fusion
Regulation Number:888.3080
Product Code:ODP

Substantial Equivalence

510(k) Summary:

Eisertech, LLC believes that the Eisertech, LLC Cervical Cage is substantially equivalent to the Aesculap CeSpace PEEK Spinal Implant System (K083311), the Spinal Elements Crystal Cervical Interbody System (K073351), the Signus Medical, LLC Rabea implant (K082848), the Nuvasive CoRoent-S (K081611), and the Alphatec Spine, Inc. Novel Cervical Spine System (K081730).

Description of device and its intended use

The Cervical Cage is intended for spinal fusion procedures at one level (C2 to T1) in skeletally mature patients with degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implants are intended to be packed with autogenous bone graft. The Cervical Cage is intended to be used with a supplemental fixation system.

Description of device design requirements

The Cervical Cage design must maintain the spacing between two vertebral bones following cervical discectomy until fusion occurs.

ldentification of the risk analysis method

Risks were qualitatively summarized and addressed by quantitatively analyzing specific in-vivo device performance requirements. The biomechanical loads that the device is expected to be subjected to were described and used as design input criteria. Test results relative to those loading conditions (e.g. design output data) were compared to the design input criteria. The device output data showed

{1}------------------------------------------------

performance meeting or exceeding the design input requirements for all conditions.

Discussion of the device characteristics

The Cervical Caqe is an intervertebral body fusion orthosis intended to be used in cervical spinal fusion surgery. It provides mechanical support to the spine and protects the bone graft from excessive loads so that bone healing can occur.

Description of the performance aspects

The Cervical Cage was tested by the methods described in ASTM F2077, including static axial compression, dynamic axial compression, static torsion, and dynamic torsion. Testing per ASTM F2267 to quantify the potential for device subsidence was also conducted. The resistance to expulsion was evaluated by performing expulsion testing against grade 15 polyurethane foam with 100 N axial preload.

The Cervical Cage performs at least as safely and effectively as a legally marketed predicate device.

Reliance on standards

Standards relevant to the methods in which the testing was conducted were relied upon. These include ASTM F2077 and ASTM F2267. However, no performance standard exists for intervertebral body fusion orthoses.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Eisertech, LLC % Mr. Lukas Eisermann 2555 Front Street San Diego, California 92103

ງປະ 2-2-2011

Re: K110915

Trade/Device Name: Cervical Cage Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: ODP Dated: June 29, 2011 Received: June 30, 2011

Dear Mr. Eisermann:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 - Mr. Lukas Eisermann

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm

Sincerely yours,

yours,

Melkerson

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known):

Cervical Cage Device Name:

Indications for Use:

The Cervical Cage is intended for spinal fusion procedures at one level (C2 to T1) in skeletally mature patients with degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implants are intended to be packed with autogenous bone graft. The Cervical Cage is intended to be used with a supplemental fixation system.

Prescription Use
(Part 21 CFR 801 Subpart D)

AND/OR

Over-the-counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page _________________________________________________________________________________________________________________________________________________________________________


(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

KI10915 510(k) Number_

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.