K Number
K110712
Date Cleared
2011-06-28

(106 days)

Product Code
Regulation Number
874.1820
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Suction Stimulator Probe is a dedicated manual surgical instrument that allows the surgeon to clear secretions and test surgical tissue with nerve stimulation at the same time and with the same instrument. It is intended for use only by a licensed physician and in conjunction with the Neurovision SE (Nerveana) Nerve locator Monitor System.

Device Description

The Suction Stimulator Probe is a disposable, single use, sterile instrument that combines concurrent suction and stimulation into one device for use during tissue dissection and nerve stimulation during surgery.

The instrument consists of a stainless steel shaft, partially insulated with biocompatible PTFE and ending with a ball-tip suction. Proximal connectors attach the instrument to a nerve stimulator. A tapered type adapter attaches the device to a standard suction catheter and a finger-hole allows for adjusting the suction strength manually by the surgeon. The insulated shaft has a 2mm diameter and 2 length options of 13 or 26 cm. The ball tip feature reduces the current density of stimulation, allows stimulation along right angles, such as drill holes, and prevents bone chip clogging of the instrument.

AI/ML Overview

The Neurovision Medical Products Inc. Suction Stimulator Probe (DryTouch Suction Stimulator Probe) is a Class II medical device (Product Code ETN) that combines concurrent suction and nerve stimulation in a single, disposable, sterile instrument for use during tissue dissection and nerve stimulation in surgery. It is
intended for use by a licensed physician in conjunction with the Neurovision SE (Nerveäna) Nerve locator Monitor System.

The provided document describes both the acceptance criteria and the study conducted to demonstrate substantial equivalence for the device. However, it is important to note that this is a Traditional 510(k) submission, which primarily focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics and performance data, rather than a standalone clinical study with extensive acceptance criteria explicitly defined in terms of clinical outcomes or diagnostic performance metrics often seen with AI/ML devices.

Here's an analysis of the provided information:

1. Table of Acceptance Criteria and Reported Device Performance

Given the nature of this 510(k) submission for a surgical instrument, the "acceptance criteria" are related to mechanical, electrical, and sterility performance, as well as substantial equivalence to predicate devices, rather than AI/ML diagnostic or prognostic metrics.

Acceptance Criteria (Bench Testing)Device Performance
Ball tip diameter measurementConfirmed (implicitly, as it contributed to the "quality of manufacturing" and "reliability of the surface area of the ball as indicating the current density of delivered stimulation")
Electrical continuity testingConfirmed (implicitly, as it contributed to "reliability of current delivered to the tip")
Suction testing with applied airConfirmed (implicitly, as it contributed to "gross delivery of air/suction as reflecting the suction power of the device")
SterilizationEO validation compliant with ISO 11135-1:2007; EO residuals analysis compliant with ISO 10993-7:2008; Sterility Assurance Level 10-6
Shelf Life1 year, confirmed by package integrity testing compliant with ISO 11607

Note: The document does not provide specific numerical targets or thresholds for "acceptance criteria" for the bench tests (e.g., a specific range for ball tip diameter or a minimum suction flow rate). Instead, it states that these tests were conducted to "assess performance and safety" and to "test the quality of the manufacturing, the reliability of the surface area of the ball as indicating the current density of delivered stimulation, the continuity of the device as reflecting the reliability of current delivered to the tip, and finally, the gross delivery of air/suction as reflecting the suction power of the device." The FDA's substantial equivalence determination implies that these criteria were met adequately.

2. Sample Size Used for the Test Set and Data Provenance

The study described is bench testing, not a clinical study involving a "test set" in the context of AI/ML or clinical efficacy.

  • Sample Size for Test Set: Not applicable in the context of clinical data or AI/ML evaluations. The bench tests would have used an unspecified number of physical devices.
  • Data Provenance: Not applicable in the context of country of origin of clinical data. The testing was likely conducted in a lab setting by the manufacturer (Neurovision Medical Products, Inc., in Ventura, CA, USA).
  • Retrospective or Prospective: Not applicable as it refers to bench testing on physical devices, not data collection from subjects.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Number of Experts: Not applicable. Ground truth for bench testing involves engineering measurements and adherence to standards, not human expert consensus for clinical image interpretation or diagnosis.
  • Qualifications of Experts: Not applicable.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. This concept primarily applies to clinical studies where multiple human readers or ground truth establishers might disagree, requiring a method to reach a definitive truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC Study Done: No. This is not mentioned. The device is a surgical instrument, and the submission focuses on its physical and functional equivalence to existing predicate devices through bench testing. It does not involve human readers evaluating cases with or without AI assistance.
  • Effect Size of Human Readers Improvement with AI vs. without AI Assistance: Not applicable.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Standalone Study Done: No. The device is a physical surgical instrument. The concept of an "algorithm only" performance study is not relevant here.

7. Type of Ground Truth Used

  • Type of Ground Truth: For the bench testing, the ground truth was based on engineering measurements, established industry standards (e.g., ISO 11135-1:2007, ISO 10993-7:2008, ISO 11607), and the functional specifications of the device. This is analogous to "physical specifications" or "performance metrics" derived from controlled experimental setups.

8. Sample Size for the Training Set

  • Sample Size for Training Set: Not applicable. There is no AI/ML component described that would require a "training set."

9. How the Ground Truth for the Training Set Was Established

  • How Ground Truth for Training Set Was Established: Not applicable.

In summary: The provided document is a 510(k) summary for a physical surgical instrument. The "study" proving acceptance criteria involves bench testing to confirm physical characteristics, electrical continuity, suction capability, and adherence to sterilization and shelf-life standards. The concept of AI/ML-specific criteria, such as test sets, training sets, expert ground truth, or MRMC studies, is not applicable to this type of device submission. Substantial equivalence was primarily demonstrated by comparing its technological characteristics and performance to legally marketed predicate devices through engineering and laboratory evaluations.

{0}------------------------------------------------

K 110712
P. 1 of 2

Image /page/0/Picture/1 description: The image shows the logo for Neurovision Medical. The word "neurovision" is in large, bold, black letters. Underneath the word is a line with a heartbeat symbol in the middle, and the word "MEDICAL" is in smaller, bold, black letters to the right of the line.

Traditional 510k - Suction Stimulator Probe

JUN 2 8 2011

5. 510K Summary

In accordance with Title 21 of the Code of Federal Regulations, Part 807 and in particular 21 CFR §807.92, the following summary of information is provided:

Applicant Information

Christine Vergelv Regulatory Affairs Manager Neurovision Medical Products, Inc. 2225 Sperry Ave., Suite 1000 Ventura, CA 93003 PH 805-866-6999 Fax: 413-410-4579 christie@neurovisionmedical.com

Date of SummaryJune 28, 2011
--------------------------------
Device IdentificationTrade Name: DryTouch Suction Stimulator ProbeCommon Name: Suction Stimulator Probe
ClassificationClass IINumber 874.1820, Nerve Stimulator / LocatorProduct code: ETN
Predicate DevicesManufacturer: Elmed Inc.Device Name: Frazier Suction Coagulator (part #23-5001)510k Number: Pre-amendment device
Manufacturer: Neurovision Medical Products Inc.Device Name: Hemostat/Stimulator Probe (SHEM-s)510K Number: K895676
Manufacturer: Axon Systems, Inc.Device Name: XPAK II510K Number: K090838

Device Description

The Suction Stimulator Probe is a disposable, single use, sterile instrument that combines concurrent suction and stimulation into one device for use during tissue dissection and nerve stimulation during surgery.

The instrument consists of a stainless steel shaft, partially insulated with biocompatible PTFE and ending with a ball-tip suction. Proximal connectors attach the instrument to a nerve stimulator. A tapered type adapter attaches the device to a standard suction catheter and a finger-hole allows for adjusting the suction strength manually by the surgeon. The insulated shaft has a 2mm diameter and 2 length options of 13 or 26 cm. The ball tip feature reduces the current density of stimulation, allows stimulation along right angles, such as drill holes, and prevents bone chip clogging of the instrument.

{1}------------------------------------------------

K110712
P. 2 of 2

Image /page/1/Picture/1 description: The image shows the logo for Neurovision Medical. The word "neurovision" is in large, bold, black letters. Underneath the word "neurovision" are three horizontal lines, followed by a graphic of a heartbeat, and then the word "MEDICAL" in smaller, black letters.

Traditional 510k - Suction Stimulator Probe

Intended Use

The Suction Stimulator Probe is a dedicated manual surgical instrument that allows the surgeon to clear secretions and test surgical tissue with nerve stimulation at the same time and with the same instrument. It is intended for use only by a licensed physician and in conjunction with the Neurovision SE (Nerveäna) Nerve locator Monitor System.

Technological Characteristics of Device in Relation to Predicate Devices

The Suction Stimulator Probe combines technology from several predicate devices: the Frazier Suction Coagulator (part #23-5001), the Hemostat/Stimulator Probe (SHEM-s) K895676, and the XPAX II Nerve Stimulation Probe (K090838).

  • The Suction Stimulator Probe is substantially equivalent to the Frazier Suction Coagulator with an extended shaft length: the body of the Suction Stimulator Probe is a Frazier Suction Coagulator with its shaft extended from 17.7cm to 26 cm or shortened to 13 cm.

  • The Suction Stimulator Probe's stimulating capability is substantially equivalent to the stimulating capability of the Hemostat/Stimulator Probe (SHEM-s) K895676: the device is coated with PTFE insulating material and a dedicated wire is added for connection to an approved nerve stimulator.

  • The Suction Stimulator probe is substantially equivalent to the Axon Systems XPAK II (K090838) in that both are stimulating, dissecting instruments with biocompatible electrical insulation, which attach to a nerve stimulator for nerve stimulation during surgery.

Assessment of Performance Data Used to Justify Substantial Equivalence

Bench testing to assess performance and safety included micrometer measurement of the ball tip diameter, electrical continuity testing, and suction testing with applied air. These testing criteria were selected to test the quality of the manufacturing, the reliability of the surface area of the ball as indicating the current density of delivered stimulation, the continuity of the device as reflecting the reliability of current delivered to the tip, and finally, the gross delivery of air/suction as reflecting the suction power of the device.

Sterilization:

  • EO validation is compliant with ISO 11135-1.2007. ●
  • EO residuals analysis is compliant with ISO 10993-7:2008. .
  • The Sterility Assurance Level is 10-6. .
  • . Device is not labeled non-pyrogenic.

Shelf Life Testing:

  • Shelf life is 1 year and was confirmed by package integrity testing compliant with ISO 11607. .

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002

Neurovision Medical Products, Inc. c/o Ms. Christine Vergely Regulatory Manager 2225 Sperry Avenue, Suite 1000 Ventura, CA 93003

JUN 2 8 2011

Re: K110712

Trade/Device Name: DryTouch Suction Stimulator Probe Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: Class II Product Code: ETN Dated: May 22, 2011 Received: May 24, 2011

Dear Ms. Vergely:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 - Ms. Christine Vergely

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Melvin R. Fahlman, Ph.D.

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Neurovision Medical. The word "neurovision" is in bold, black letters on the top line. Underneath the word is a line with a graphic and the word "MEDICAL" in smaller, black letters. The text "Traditional 510k - Suction Stimulator Probe" is at the bottom of the image.

Indications for Use

510(k): K110712

Device Name: Suction Stimulator Probe

Statement of Indications for Use for the Suction Stimulator Probe:

The Suction Stimulator Probe is a dedicated manual surgical instrument that allows the surgeon to clear secretions and test surgical tissue with nerve stimulation at the same time and with the same instrument. It is intended for use only by a licensed physician and in conjunction with the Neurovision SE (Nerveana) Nerve locator Monitor System.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Aper Douch

(Division SignsOff) urrence of CDRH, Office of Device Evaluation (ODE) Division of Ophthalmic, Neurological Nose and Throat Devices

Prescription Use
(Per 21 CFR 801.109)

510(k) Number K110712

§ 874.1820 Surgical nerve stimulator/locator.

(a)
Identification. A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.(b)
Classification. Class II.