K Number
K110650

Validate with FDA (Live)

Date Cleared
2011-06-30

(118 days)

Product Code
Regulation Number
888.3080
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Zyston Curve Interbody Spacer System is indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

The Zyston Curve Interbody Spacer System is designed for use with autograft to facilitate fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine. The Zyston Curve Interbody Spacer System may also be implanted using the Accuvision System to provide the surgeon with a minimally invasive approach for posterior or posterolateral spinal surgery.

Device Description

The purpose of this submission is the introduction of a new intervertebral body fusion device. The Zyston Curve Interbody Spacer System is a spacer constructed of medical grade Polyetheretherketone (PEEK) with a titanium alloy threaded insert, and tantalum radiographic markers for spinal applications.

AI/ML Overview

Here's an analysis of the provided text regarding the Zyston Curve Interbody Spacer System, focusing on acceptance criteria and performance studies:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Mechanical Testing Standards)Reported Device Performance
Static Axial Compression (ASTM F-2077)Met or exceeded predicate
Dynamic Axial Compression (ASTM F-2077)Met or exceeded predicate
Static Compression-Shear (ASTM F-2077)Met or exceeded predicate
Dynamic Compression-Shear (ASTM F-2077)Met or exceeded predicate
Subsidence (ASTM F-2267 and ASTM F-2077)Met or exceeded predicate

2. Sample Size Used for the Test Set and Data Provenance

The provided document describes mechanical performance testing for a medical device (Zyston Curve Interbody Spacer System). These tests are typically conducted on device samples, not on patient data.

  • Sample Size for Test Set: The document does not specify the exact number of devices tested for each mechanical test. This information is often found in detailed test reports, not summarized in regulatory submissions like this 510(k).
  • Data Provenance: Not applicable in the context of mechanical testing. The data is generated in a laboratory setting. If it were a clinical study, provenance might refer to the country of origin of patients or the type of clinical trial (retrospective/prospective); however, there is no clinical study described.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable to this type of study. Mechanical testing involves objective measurements based on standardized protocols (ASTM standards) and engineering principles, not expert consensus or interpretation of clinical data.

4. Adjudication Method for the Test Set

This information is not applicable. Adjudication methods (like 2+1, 3+1) are used in clinical studies where expert disagreement on ground truth is possible. Mechanical tests rely on direct measurements and adherence to testing standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The document explicitly describes mechanical testing in a lab setting, not a clinical study involving human readers or patient cases.

  • Effect size of human readers with AI vs. without AI assistance: Not applicable, as this was not an AI-assisted diagnostic study.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No, a standalone algorithm performance study was not done. This document concerns a physical medical implant device and its mechanical properties, not a software algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the mechanical performance tests is established by the ASTM standards themselves (F-2077, F-2267) which define the methods, parameters, and acceptable ranges or comparison criteria for evaluating the mechanical properties of intervertebral body fusion devices. The performance of the Zyston Curve was then measured against these standards and compared to legally marketed predicate devices.

8. The Sample Size for the Training Set

This information is not applicable. There is no "training set" as this is not a machine learning or AI algorithm. The device's design is based on engineering principles and existing predicate devices, not trained on a dataset.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable, as there is no training set involved in the mechanical testing of this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for Biomet Spine. The word "BIOMET" is in a bold, sans-serif font, with each letter connected to the next. Below the word "BIOMET" is the word "SPINE" in a smaller, sans-serif font. The logo is black and white.

JUN 3 0 2011

510(k) Summary

Preparation Date:May 9, 2011
Applicant/Sponsor:Biomet Spine
100 Interpace Parkway
Parsippany, NJ 07054
Contact Person:Margaret F. Crowe
Phone: 973-299-9300, ext. 2260
Fax: 973-257-0232
Trade name:Zyston Curve Interbody Spacer System
Common Name:Non-cervical (lumbar) spinal spacer
Classification Name:Intervertebral body fusion device, 21 CFR §888.3080
Device Panel/ProductCode:Orthopedic MAX

Device Description:

The purpose of this submission is the introduction of a new intervertebral body fusion device. The Zyston Curve Interbody Spacer System is a spacer constructed of medical grade Polyetheretherketone (PEEK) with a titanium alloy threaded insert, and tantalum radiographic markers for spinal applications.

Indications for Use:

The Zyston Curve Interbody Spacer System is indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

The Zyston Curve Interbody Spacer System is designed for use with autograft to facilitate fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine. The Zyston Curve Interbody Spacer System may also be implanted using the Accuvision System to provide the surgeon with a minimally invasive approach for posterior or posterolateral spinal surgery.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for Biomet Spine. The logo is in black and white and features the word "BIOMET" in large, bold letters. Below the word "BIOMET" is the word "SPINE" in smaller letters. The logo is simple and modern.

Summary of Technologies:

The technological characteristics (material, design and sizing) of the Zyston Curve Spacer is the same as, or similar to, the predicate devices. Examples of predicate devices include:

  • Ibex Spinal Spacer System (K091381). .
  • . Expandable PEEK Implant (K082406)
  • ESL Spacer (K092574) .
  • Signature™ TLIF Spacer (K072970) .
  • Loop™ Interbody Fusion Device (K080401) .

Performance Data

Mechanical testing recommended in the special controls guidance document entitled "Class II Special Controls Guidance Document: Intervertebral Body Fusion Device" was conducted. The testing conducted, along with the ASTM standard, are listed below:

    1. Static Axial Compression (ASTM F-2077)
    1. Dynamic Axial Compression (ASTM F-2077)
    1. Static Compression-Shear (ASTM F-2077)
    1. Dynamic Compression-Shear (ASTM F-2077)
    1. Subsidence (ASTM F-2267 and ASTM F-2077)

The Zyston Curve's mechanical performance met or exceeded that of the predicate device(s).

Substantial Equivalence:

The Zyston Curve Interbody Spacer System is substantially equivalent to its predicate devices with respect to intended use and indications, technological characteristics, and principles of operation and do not present any new issues of safety or effectiveness. The predicates listed above are distributed for similar indications, and have similar design features.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUN 3 0 2011

Biomet Spinc % Ms. Margaret F. Crowe 100 Interpace Parkway Parsippany, New Jersey 07054

Re: K110650

Trade/Device Name: Zyston Curve Interbody Spacer System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: May 12, 2011 Received: May 13, 2011

Dear Ms. Crowe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 - Ms. Margaret F. Crowe

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

E.L. Keith

-C3/ Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Zyston Curve Interbody Spacer System

Indications for Use:

The Zyston Curve Interbody Spacer System is indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

The Zyston Curve Interbody Spacer System is designed for use with autograft to facilitate fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine. The Zyston Curve Interbody Spacer System may also be implanted using the Accuvision System to provide the surgeon with a minimally invasive approach for posterior or posterolateral spinal surgery.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Diysion of Surgical, Orthopedic, and Restorative Devices

KIJO650 510(k) Number_

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.