K Number
K110560

Validate with FDA (Live)

Date Cleared
2011-11-18

(263 days)

Product Code
Regulation Number
882.5550
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medtronic ARES™ Antibiotic-Impregnated Catheters are intended for use in the treatment of hydrocephalus as a component of a shunt system when draining or shunting of cerebrospinal fluid (CSF) is indicated.

Device Description

The Medtronic ARES™ Antibiotic-Impregnated Catheters are manufactured using barium sulfate-filled silicone elastomer and are impregnated with clindamycin hydrochloride and rifampicin.

ARES Antibiotic-Impregnated Ventricular Catheter
The ARES ventricular catheter measures 23 cm in length, 0.13 cm in inner diameter, and 0.25 cm in outer diameter. Lengths are marked in 1 cm intervals starting from 3 cm to 15 cm from the catheter tip, thus enabling the surgeon to gauge the depth of penetration of the catheter into the lateral ventricle. The proximal end of the catheter has 32 flow holes—four lines of eight holes spaced at 90° intervals around the catheter circumference. Components supplied with the ARES Ventricular Catheter include a pre-loaded stainless steel stylet and a Right Angle Clip, which is included to facilitate placement and use of the ventricular catheter.

ARES Antibiotic-Impregnated Peritoneal Catheter
The ARES peritoneal catheter measures 120 cm in length, 0.13 cm in inner diameter, and 0.25 cm in outer diameter. There are no length markers or wall slits on the catheter, and the tip is open ended. The catheter may be trimmed to the proper length.

AI/ML Overview

This 510(k) summary describes the Medtronic ARES™ Antibiotic-Impregnated Catheters and their substantial equivalence to the BACTISEAL® Catheter. The submission primarily relies on non-clinical performance and biocompatibility data, rather than clinical studies involving human patients or complex AI algorithms. Therefore, many of the requested elements pertaining to clinical studies, AI performance, ground truth, and expert evaluation are not applicable or cannot be extracted from this document.

Here's the information that can be extracted from the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicAcceptance Criteria (Requirement)Reported Device Performance (ARES™)
Performance TestingMeet requirements outlined in ISO 7197: Neurosurgical implants - Sterile, single use hydrocephalus shunts and components and ASTM 647-94 Standard Practice for Evaluating and Specifying Implantable Shunt Assemblies for Neurosurgical Application.Successfully tested to meet the applicable requirements outlined in ISO 7197 and ASTM 647-94. Additionally, non-clinical tests were conducted for specific characteristics.
Sterilization TestingSterility assurance level (SAL) of 10-6 according to ISO 17665: Sterilization of health care products - Moist heat.Validated using the Half Cycle method as outlined in ISO 17665 to a sterility assurance level (SAL) of 10-6.
Pyrogen TestingConform to FDA standards concerning pyrogen levels for devices in contact with cerebrospinal fluid, meeting a 2.15 EU/device specification.Meets a 2.15 EU/device specification utilizing the Kinetic-Chromogenic LAL method.
Biocompatibility Testing (General)Considered an implant device with tissue/bone contact and permanent contact duration of greater than 30 days, tested in accordance with ISO 10993-1:2009 and FDA Blue Book Memo, G95-1.A series of tests were conducted to demonstrate biocompatibility with the following specific results:
CytotoxicityNo biological reactivity.No biological reactivity (Grade 0) observed at 48 hours post exposure to the test article extract. (Conclusion: Non-cytotoxic)
SensitizationNo reaction at challenge following induction phase.Extracts elicited no reaction at the challenge (0% sensitization), following an induction phase. (Conclusion: Non-sensitizing)
Intracutaneous Irritation/ReactivityTest article extract sites not significantly greater biological reaction than control.Test article extract sites did not show a significantly greater biological reaction than the sites injected with the control article extract. (Conclusion: Non-irritant)
Acute Systemic ToxicityTest article extracts not significantly greater biological reaction than control.Test article extracts did not induce a significantly greater biological reaction than the control extracts. (Conclusion: Non-toxic (acute))
Bacterial Reverse Mutation Study (Non-antibiotic impregnated catheter)Non-mutagenic.DMSO and saline test article extracts were considered to be non-mutagenic to Salmonella typhimurium tester strains TA98, TA100, TA1535, and TA1537 and to Escherichia coli tester strain WP2uvrA. (Conclusion: Non-genotoxic)
Bacterial Reverse Mutation Assay (Mock-impregnated catheter)Non-mutagenic.DMSO and saline test article extracts were considered to be non-mutagenic to Salmonella typhimurium tester strains TA98, TA100, TA1535, and TA1537 and to Escherichia coli tester strain WP2uvrA. (Conclusion: Non-genotoxic)
In Vitro Mouse Lymphoma AssayMutant frequencies and cloning efficiencies within limits for negative response.Mutant frequencies and cloning efficiencies of preparations treated with test article were within the limits defined for a negative response. (Conclusion: Non-genotoxic)
Mouse Peripheral Blood Micronucleus StudyNo induction of micronuclei.The test article extracts did not induce micronuclei in mice. (Conclusion: Non-genotoxic)
Implantation - 13 week brain and subcutaneous implant & Sub-Chronic / Chronic ToxicityNo systemic signs of toxicity and no local adverse event.The test article does not appear to demonstrate any systemic signs of toxicity when implanted in the brain and subcutaneous tissue of New Zealand White rabbits for a period of 13 weeks. (Conclusion: No local adverse event, Non-toxic (sub-chronic))
Comparative Performance (Predicate Device)
Drug release kineticsDemonstration that antibiotics release kinetics over a 38-day period is the same for ARES and BACTISEAL.Antibiotic release kinetics measured over a 38-day period demonstrated the same as BACTISEAL antibiotic release rates.
Zone of Inhibition TestingDemonstration of antimicrobial activity for at least 31 days with statistically equivalent zones of inhibition for ARES and BACTISEAL.Antimicrobial activity demonstrated for at least 31 days. Measured ZOIs statistically equivalent to those of BACTISEAL.
Catheter Crush ResistanceARES catheter mean value must not be significantly lower than control group (BACTISEAL) mean using a two-sample t-test at 95% confidence.Pass
Drug ContentClindamycin: 0.150 ± 45%; Rifampicin: 0.054 ± 60%.Pass

2. Sample Size Used for the Test Set and Data Provenance

The document primarily details non-clinical (bench and in-vitro/in-vivo animal) tests. Therefore, the concept of a "test set" in the context of clinical data or AI performance (which would typically involve patient cases) is not directly applicable.

  • Bench tests: No specific sample sizes are provided within the summary for tests like performance testing or sterilization testing, although these would involve multiple units for validation.
  • Biocompatibility tests: Sample sizes are implicit in the descriptions of animal studies (e.g., "All animals increased in weight" implies multiple animals were used, but specific numbers are not given for each test). The studies used Salmonella typhimurium and Escherichia coli tester strains, mice, and New Zealand White rabbits.
  • Data Provenance: The data originates from internal testing conducted by Medtronic as part of their 510(k) submission. It's prospective in the sense that these tests were conducted specifically to support this submission. The origin is implied to be within controlled laboratory settings, likely in the US, given the submission is to the FDA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

Not Applicable. This submission relies on objective, measurable physical, chemical, and biological endpoints from non-clinical tests (bench, in-vitro, and in-vivo animal studies) rather than human expert interpretation of clinical data or images. Therefore, no experts were used to establish "ground truth" in the clinical AI sense.

4. Adjudication Method for the Test Set

Not Applicable. Since no human expert evaluation or clinical data interpretation was involved for establishing a "ground truth" or test set, no adjudication method was used.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not Applicable. This submission is for a physical medical device (antibiotic-impregnated catheter) and does not involve AI algorithms or human reading of medical cases.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

Not Applicable. This submission is for a physical medical device and does not involve an AI algorithm.

7. The Type of Ground Truth Used

The "ground truth" or reference standards used are established scientific and regulatory standards and validated test methodologies. These include:

  • Engineering Standards: ISO 7197, ASTM 647-94 for performance.
  • Sterilization Standards: ISO 17665.
  • Pyrogen Standards: ANSI/AAMI/ST72.
  • Biocompatibility Standards: ISO 10993-1, ISO 10993-3, ISO 10993-5, ISO 10993-6, ISO 10993-10, ISO 10993-11, and FDA Blue Book Memo, G95-1.
  • Quantitative Measurements: Drug release kinetics (mg/day), Zone of Inhibition (mm), Drug content (weight %), mechanical strength (crush resistance).
  • Biological Endpoints: Cytotoxicity (cellular response), Sensitization (immune response), Irritation (tissue reaction), Systemic Toxicity (organism-wide effects), Genotoxicity (DNA damage), Implantation responses (local tissue reaction in vivo).

Comparisons were also made directly against the predicate device (BACTISEAL® Catheter) for key performance characteristics to demonstrate "sameness."

8. The Sample Size for the Training Set

Not Applicable. This submission is for a physical medical device and does not involve AI algorithms that require a "training set."

9. How the Ground Truth for the Training Set was Established

Not Applicable. As there is no AI algorithm, there is no training set or ground truth for one.

{0}------------------------------------------------

510(k) SUMMARY Medtronic ARES™ Antibiotic Impregnated Catheters K110560

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.

Submitter Information
NameMedtronic Neurosurgery
Address125 Cremona DriveGoleta, CA 93117-5503USA
Phone number805-571-8400
Fax number805-571-8480
EstablishmentRegistration Number2021898
Name of contactpersonJeffrey Henderson
Date preparedNovember 17, 2011
Name of device
Trade or proprietarynameMedtronic ARES™ Antibiotic-Impregnated Catheter
Common or usualnameHydrocephalus Catheter
Classification nameShunt, Central Nervous System and Components
ClassificationClass II
PanelNeurological
Regulation21 CFR §882.5550
Product Code(s)JXG
Legally marketed device(s)to which equivalence isclaimedJohnson & Johnson (Codman & Shurtleff)BACTISEAL® Catheter (K003322)
Reason for 510(k)submissionNew device
Device descriptionThe Medtronic ARES™ Antibiotic-Impregnated Catheters aremanufactured using barium sulfate-filled silicone elastomerand are impregnated with clindamycin hydrochloride andrifampicin.ARES Antibiotic-Impregnated Ventricular CatheterThe ARES ventricular catheter measures 23 cm in length,0.13 cm in inner diameter, and 0.25 cm in outer diameter.Lengths are marked in 1 cm intervals starting from 3 cm to 15cm from the catheter tip, thus enabling the surgeon to gaugethe depth of penetration of the catheter into the lateralventricle. The proximal end of the catheter has 32 flowholes—four lines of eight holes spaced at 90° intervals
around the catheter circumference. Components suppliedwith the ARES Ventricular Catheter include a pre-loadedstainless steel stylet and a Right Angle Clip, which is includedto facilitate placement and use of the ventricular catheter.ARES Antibiotic-Impregnated Peritoneal CatheterThe ARES peritoneal catheter measures 120 cm in length,0.13 cm in inner diameter, and 0.25 cm in outer diameter.There are no length markers or wall slits on the catheter, andthe tip is open ended. The catheter may be trimmed to theproper length.
Intended use of the deviceThe Medtronic ARES™ Antibiotic-Impregnated Catheters areintended for use in the treatment of hydrocephalus as acomponent of a shunt system when draining or shunting ofcerebrospinal fluid (CSF) is indicated.
Indications for useThe Medtronic ARES™ Antibiotic-Impregnated Catheters areintended for use in the treatment of hydrocephalus as acomponent of a shunt system when draining or shunting ofcerebrospinal fluid (CSF) is indicated.

p. 1/5

{1}------------------------------------------------

Summary of the technological characteristics of the device compared to the predicate device

The Medtronic ARES™ Antibiotic-Impregnated catheter is substantially equivalent in device description, function, principle of operation, and basic composition to the BACTISEAL® predicate device. The ARES and BACTISEAL devices characteristics are summarized below.

CharacteristicNew Device [MedtronicAREST™ Antibiotic-Impregnated Catheters]Predicate [BACTISEAL®Catheter, K003322]
Catheter Body MaterialBaSO4 filled silicone elastomerBaSO4 filled silicone elastomer
Drugs Formulation0.054 weight % rifampicin0.150 weight % clindamycinhydrochloride0.054 weight % rifampicin0.150 weight % clindamycinhydrochloride
Catheter Body DimensionsVentricular: 0.050" ID x 0.100"ODPeritoneal: 0.050" ID x 0.100"ODVentricular: 0.055" ID x 0.106"OD or 0.050" ID x 0.100" ODPeritoneal: 0.039" ID x 0.0866"OD or 0.050" ID x 0.100" OD
Catheter LengthVentricular: 23 cmPeritoneal: 120 cmVentricular: 14, 15, or 23 cmPeritoneal: 120 cm
Length MarkersVentricular: numerical lengthmarkersPeritoneal: NoneVentricular: 1 dot at 10 cm & 2dots at 15 cmPeritoneal: None
Tip ConfigurationVentricular: Bullet shape with 32inlet holes (4 rows of 8 holes)Peritoneal: Open end, no slitsVentricular: Bullet shape with 24inlet holes (3 rows of 8 holes)Peritoneal: Open end, no slits

p.2/5

{2}------------------------------------------------

PERFORMANCE DATA

SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE

Performance Test Summary

The ARES™ Antibiotic-Impregnated Catheters have been designed and tested to meet the requirements of voluntary standards and FDA guidance documents applicable to the subject and predicate devices. Results of the non-clinical testing support the conclusion of substantial equivalence to the ARES™ Antibiotic-Impregnated Catheters to the predicate devices.

CharacteristicStandard/Test/FDA GuidanceResults Summary
Performance TestingPerformance testing wasperformed according to ISO7197: Neurosurgical implants -Sterile, single use hydrocephalusshunts and components andASTM 647-94 Standard Practicefor Evaluating and SpecifyingImplantable Shunt Assemblies forNeurosurgical ApplicationThe ARES Antibiotic-Impregnated Catheters havebeen designed and successfullytested to meet the applicablerequirements outlined in ISO7197 and ASTM 647-94
Sterilization TestingSterilization testing wasperformed according to ISO17665: Sterilization of healthcare products - Moist heatSterilization of ARESTMAntibiotic-ImpregnatedCatheters is validated usingthe Half Cycle method asoutlined in ISO 17665 to asterility assurance level (SAL)of 10-6.
Pyrogen TestingPyrogen testing was performedaccording to ANSI/AAMI/ST72Bacterial endotoxins - Testmethodologies, routine monitoringand alternatives to batch testingARES Antibiotic-ImpregnatedCatheters conform to FDAstandards concerning pyrogenlevels for devices in contactwith cerebrospinal fluid.Utilizing the Kinetic-Chromogenic LAL method, theproduct meets a 2.15EU/device specification.

p. 3/5

{3}------------------------------------------------

Biocompatibility Test Summary

For purposes of biocompatibility testing, the Medtronic ARES™ Antibiotic-Impregnated Catheters are considered an implant device with tissue/bone contact and permanent contact duration of greater than 30 days. The selection of biocompatibility testing for Medtronic ARES™ Antibiotic-Impregnated Catheters was based on this categorization of the device, and in accordance with ISO 10993-1:2009-Biological evaluation of medical devices-Part 1: Evaluation and the FDA Blue Book Memo, G95-1, Use of the International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing.

Test SystemStudy ResultsConclusion
Cytotoxicity - MEM Elution(ISO 10993-5)No biological reactivity (Grade 0) was observed at 48hours post exposure to the test article extract.Non-cytotoxic
Sensitization - KligmanMaximization(ISO 10993-10)The extracts of the test article elicited no reaction at thechallenge (0% sensitization), following an inductionphase.Non-sensitizing
IntracutaneousIrritation/Reactivity -Intracutaneous Injection(ISO 10993-10)All animals increased in weight and none exhibitedover signs of toxicity at any observation point. The testarticle extract sites did not show a significantly greaterbiological reaction than the sites injected with thecontrol article extract.Non-irritant
Acute Systemic Toxicity -Systemic Injection(ISO 10993-11)All animals increased in weight and none exhibitedovert signs of toxicity at any observation point. Thetest article extracts did not induce a significantly greaterbiological reaction than the control extracts.Non-toxic(acute)
Bacterial ReverseMutation Study - Non-antibiotic impregnatedcatheter(ISO 10993-3)DMSO and saline test article extracts were consideredto be non-mutagenic to Salmonella typhimurium testerstrains TA98, TA100, TA1535, and TA1537 and toEscherichia coli tester strain WP2uvrA.Non-genotoxic
Bacterial ReverseMutation Assay - Extracts- Mock-impregnatedcatheter(ISO 10993-3)DMSO and saline test article extracts were consideredto be non-mutagenic to Salmonella typhimurium testerstrains TA98, TA100, TA1535, and TA1537 and toEscherichia coli tester strain WP2uvrA.Non-genotoxic
In Vitro Mouse LymphomaAssay with ExtendedTreatment(ISO 10993-3)The mutant frequencies and cloning efficiencies ofpreparations treated with test article were within thelimits defined for a negative response.Non-genotoxic
Mouse Peripheral BloodMicronucleus Study(ISO 10993-3)The test article extracts did not induce micronuclei inmice.Non-genotoxic
Implantation - 13 weekbrain and subcutaneousimplant(ISO 10993-6)&Sub-Chronic / ChronicToxicity(ISO 10993-11)The test article does not appear to demonstrate anysystemic signs of toxicity when implanted in the brainand subcutaneous tissue of New Zealand White rabbitsfor a period of 13 weeks.No localadverse eventNon-toxic (sub-chronic)

p.4/5

{4}------------------------------------------------

K110560

Comparative Performance Information Summary
CharacteristicRequirementARESBACTISEAL
Drug release kineticsDemonstration ofantibiotics releasekinetics over a 38day period is thesame forARES andBACTISEALAntibiotics releasekinetics measured overa 38 day periodAntibiotic release ratesdemonstrated the sameas BACTISEALantibiotic release ratesAntibiotics releasekinetics measuredover a 38 day periodAntibiotic release ratesdemonstrated thesame as ARESantibiotic release rates
Zone of Inhibition TestingDemonstration ofantimicrobial activityfor at least 31 dayshas statisticallyequivalent zones ofinhibition forARES andBACTISEALAntimicrobial activitydemonstrated for atleast 31 daysMeasured ZOIsstatistically equivalentto those of BACTISEALAntimicrobial activitydemonstrated for atleast 31 daysMeasured ZOIsstatically equivalent tothose of ARES
Catheter CrushResistanceARES cathetermean value mustnot be significantlylower than controlgroup (BACTISEAL)mean using a twosample t-test at 95%confidencePassPass
Drug ContentClindamycin: 0.150± 45%Rifampicin: 0.054 ±60%PassPass
SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OFSUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION
No prospective clinical trials were conducted in support of this Traditional 510(k)

CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA

Medtronic Neurosurgery considers the Medtronic ARES™ Antibiotic-Impregnated Catheters to be the "same" as the predicate BACTISEAL® catheters according to the definition of medical devices containing antimicrobial agents considered by FDA to be "the same" as described in FDA Draft Guidance for Industry and FDA Staff - Premarket Notification (510(k)) Submissions for Medical Devices that Include Antimicrobial Agents (issued July 19, 2007). The indications for use, function, implantation techniques, performance characteristics, and test standards for both devices are substantially equivalent. Medtronic Neurosurgery considers the Medtronic ARES™ Antibiotic-Impregnated Catheters to be substantially equivalent to the previously cleared Codman BACTISEAL® Catheters.

p.5/5

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three curved lines representing its wings. The eagle is positioned inside a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" surrounding it.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

NOV 1 8 2011

Medtronic, Inc. c/o Mr. Jeffrey Henderson Vice President, Quality and Regulations Affairs 125 Cremona Drive Goleta, CA 93117

Re: K110560

Trade/Device Name: Medtronic Ares™ Antibiotic-Impregnated Catheter Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt and Components Regulatory Class: Class II Product Code: JXG Dated: November 11, 2011 Received: November 14, 2011

Dear Mr. Henderson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Melvin R. Felder, MD

Malvina B. Eydelman, M)D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

K110560

  1. Indications for Use Statement

510(k) Number (if known):

To be determined

Device Name:

Indications for Use:

ARES™ Antibiotic-Impregnated The Medtronic Catheters are intended for use in the treatment of hydrocephalus as a component of a shunt system when draining or shunting of cerebrospinal fluid (CSF)

ARES™ Antibiotic-Impregnated Catheters

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Prescription Use _ (Part 21 CFR 801 Subpart D)

AND/OR

is indicated.

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number k//0560

§ 882.5550 Central nervous system fluid shunt and components.

(a)
Identification. A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.(b)
Classification. Class II (performance standards).