K Number
K110520

Validate with FDA (Live)

Device Name
BIOLIS 12I
Date Cleared
2012-03-23

(394 days)

Product Code
Regulation Number
862.1600
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Biolis 12i is a discrete photometric chemistry analyzer with ion-selective electrode (ISE), with direct quantitative measurement of sodium, potassium, chloride, and glucose in serum. It is a device intended for the in-vitro, spectrophotometric determination of general chemistry assays for clinical use. The Biolis 12i includes an optional lon Selective Electrode (ISE) module for the measurement of sodium, potassium and chloride in serum. The Biolis 12i is not for Point-Of-Care testing. It is for vitro diagnostic use only.

Sodium measurements are used in the diagnosis and treatment diseases involving electrolyte imbalance.

Potassium measurements monitor electrolyte balance and in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.

Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders.

The Biolis 12i analyzer with glucose hexokinase assay is intended to measure glucose quantitatively in human serum. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic glycemia, and of the pancreatic islet cell carcinoma.

Device Description

Using photometry, the Biolis 12i instrument measures the glucose concentration in serum by monitoring the change in absorbance at 340 nm. Additionally, the Biolis 12i with lon-Selective Elective module additionally measures the concentration of the electrolytes, sodium, potassium and chloride in serum, using indirect potentiometry.

AI/ML Overview

The provided 510(k) summary describes the Biolis 12i, a clinical chemistry analyzer with an optional Ion-Selective Electrode (ISE) module, and its performance for measuring glucose, sodium, potassium, and chloride in serum. The study focuses on demonstrating substantial equivalence to predicate devices (Sirrus for Glucose and Prestige 24i for ISE).

Here's an analysis of the acceptance criteria and study as requested, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state numerical "acceptance criteria" in the format of a predefined pass/fail threshold. Instead, it presents performance characteristics (correlation, linearity, precision, and interference) and asserts "substantial equivalence" to the predicate devices. The implicit acceptance criterion is that the performance of the Biolis 12i must be comparable or equivalent to the predicate devices.

Performance MetricAnalyteAcceptance Criteria (Implicit: Comparable to Predicate)Reported Device Performance (Biolis 12i)
Correlation
(vs. Predicate)SodiumComparable to Prestige 24i0.9872 (Correlation Coefficient)
PotassiumComparable to Prestige 24i0.9992 (Correlation Coefficient)
ChlorideComparable to Prestige 24i0.9922 (Correlation Coefficient)
GlucoseComparable to Sirrus0.9975 (Correlation Coefficient)
Linearity
Range (Serum)SodiumReportable range100 - 200 mmol/L
PotassiumReportable range1 - 10 mmol/L
ChlorideReportable range70 - 200 mmol/L
GlucoseReportable range25 - 540 mg/dL
Precision
Within Run %CVSodiumLow variability0.85, 0.61, 0.93
PotassiumLow variability0.90, 0.91, 0.94
ChlorideLow variability0.79, 0.83, 0.69
GlucoseLow variability1.31, 1.19, 1.04
Day-by-Day %CVSodiumLow variability0.4, 0.4, 0.3
PotassiumLow variability0.8, 0.9, 0.4
ChlorideLow variability1.0, 0.7, 0.4
Between-run %CVGlucoseLow variability0.87, 0.78, 0.70
InterferencesAll AnalytesNo significant interference at specified concentrationsNo significant interference observed at specified concentrations (Bilirubin, Hemoglobin, Lipemia, Lithium Chloride, Sodium Bromide, Sodium Salicylate, Sodium Thiocyanate)
Sensitivity
Minimum Detectable ValueGlucoseQuantifiable low limit7.83 mg/dL
Stability
Calibration Stability CV (%)GlucoseLow variability2.1, 2.5, 2.1

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Sample Size: The document does not explicitly state the total number of patient samples (serum) used for the correlation, linearity, and interference studies.
    • For precision studies:
      • ISE (Sodium, Potassium, Chloride): N=20 for Within Run (repeated measurements of 3 samples), N=15 for Day-by-Day (measurements of 3 samples over 15 days).
      • Glucose: N=20 for Within-run (repeated measurements of 3 samples), N=20 for Between-run (measurements of 3 samples over 20 runs, likely implying 20 days if one run/day).
  • Data Provenance: The document does not specify the country of origin of the data or explicitly state whether the study was retrospective or prospective. Given that the manufacturer is in Japan, it's possible the studies took place there, but this is not confirmed in the text.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

  • Experts: Not applicable. For this type of in vitro diagnostic device, "ground truth" is established by reference methods or predicate devices, not human expert consensus on interpretations. The ground truth for the correlation studies was the measurements obtained from the predicate devices (Prestige 24i for ISE and Sirrus for glucose) and for linearity studies, it would be prepared known concentrations. For calibrator verification, the ground truth for glucose was NIST SRM 965b.
  • Qualifications: Not applicable in the context of expert interpretation for ground truth.

4. Adjudication Method for the Test Set:

  • Adjudication Method: Not applicable. This is an IVD device measuring analytes, not making diagnostic interpretations that require human adjudication. The performance is assessed by direct comparison to established methods or known concentrations.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance:

  • This is not applicable. The Biolis 12i is an automated clinical chemistry analyzer, not an AI-assisted diagnostic tool that involves human readers interpreting images or data outputs and making diagnoses. Therefore, an MRMC study and effects on human reader improvement are not relevant to this device.

6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done:

  • Yes, this study essentially represents standalone performance. The Biolis 12i is an automated analyzer. The performance characteristics (correlation, linearity, precision, interference) are measured for the device itself using predefined protocols, without a "human-in-the-loop" interaction for interpretation that would alter its output. The results are quantitative measurements directly from the algorithm/instrument.

7. The Type of Ground Truth Used:

  • For Correlation Studies: The ground truth was the measurements obtained from the predicate devices (Sirrus for Glucose, Prestige 24i for Na, K, Cl).
  • For Linearity Studies: The ground truth was based on a range of known concentrations (likely serially diluted or prepared samples).
  • For Calibrator Verification (Glucose): The ground truth was NIST SRM 965b, Glucose in Frozen Serum, which is a certified reference material.

8. The Sample Size for the Training Set:

  • Not applicable / Not disclosed. Automated clinical chemistry analyzers typically do not have a "training set" in the machine learning sense. Their operational parameters are often determined by engineering design, chemical principles, and calibration curves established with known standards. The document clarifies how calibrators were verified but does not mention a distinct "training set" for an algorithm.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable. As a traditional in vitro diagnostic device, it does not involve machine learning "training sets" and associated ground truth establishment in the way AI/ML devices do. Its "calibration" is performed using known concentration calibrator materials and validated against reference standards. For instance, the glucose calibrator was verified against NIST SRM 965b.

{0}------------------------------------------------

K 10520

1-14-21 Higashi-Toyoda, Hino, Tokyo 191-0052, Japan Phone: +81-42-587-2965 Fax: ++81-42-587-7781 MAR 2 3 2012

510(k) SUMMARY

Submitter's Name/Address

Submitter's Name:

Submitter's Address:

Phone:

Fax:

Establishment Registration Number: Owner/Operator Number:

Tokyo Boeki Medisys Inc. 1-14-21 Higashi-Toyoda, Hino Tokyo 197-0823 +81-42-532-2771 +81-42-532-2772 3004378324 9060135

Contact Person (United States Agent)

Name of Agent:James M. Clinton
Agent's Business Name:Quality and Regulatory Consulting, LLC
Street Address:5105 Fairoaks RoadDurham, NC 27712-2078
Phone:919-247-0479
Fax:919-287-2551
E-mail address:clintonim@earthlink.net

Date of Preparation of this Summary:

Device Trade or Proprietary Name:

Device Common Name:

Classification Numbers/Class:

اللهرجما (اللانيل) الكان

January 6, 2012

Biolis 12i

Clinical Chemistry Analyzer (with optional ISE Module)

75JJE,Class I
75JGS,Class II
75CEM,Class II
75CGZ,Class II
75CFR,Class II

-1-

{1}------------------------------------------------

disys System I

1-14-21 Higashi-Toyoda, Hino, Tokyo 191-0052, Ja Phone: +81-42-587-2965 Fax: ++81-42-587-7781

510(k) Summary:

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(K) number is : K110520

Description:

Using photometry, the Biolis 12i instrument measures the glucose concentration in serum by monitoring the change in absorbance at 340 nm. Additionally, the Biolis 12i with lon-Selective Elective module additionally measures the concentration of the electrolytes, sodium, potassium and chloride in serum, using indirect potentiometry.

Intended Use:

The Biolis 12i is a discrete chemistry analyzer with ion-selective electrode (ISE), with direct quantitative measurement of sodium, potassium, chloride, and glucose in serum. It is a device intended for the in-vitro, spectrophotometric determination of general chemistry assays for clinical use. The Biolis 12i includes an optional lon Selective Electrode (ISE) module for the measurement of sodium, potassium and chloride in serum. The Biolis 12i is not for Point-Of-Care testing. It is for vitro diagnostic use only.

Sodium measurements are used in the diagnosis and treatment diseases involving electrolyte imbalance.

Potassium measurements monitor electrolyte balance and in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.

Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders.

The Biolis 12i analyzer with glucose hexokinase assay is intended to measure glucose quantitatively in human serum. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic glycemia, and of the pancreatic isle cell carcinoma

Substantial Equivalence:

Substantial equivalence has been demonstrated between Sirrus (K0421169) running Glucose reagents (K971467) and the Biolis 12i for measuring glucose in serum. These analyzers are calibrated with known concentration calibrator material and both measure specific concentrations using photometry and electrolytes using identical ion selective electrode modules In addition, substantial equivalence has been demonstrated between the Prestige 24i (K040958) and

-2-

{2}------------------------------------------------

1-14-21 Higashi-Toyoda, Hino, Tokyo 191-0052, Japan Phone: +81-42-587-2965 Fax: ++81-42-587-7781

the Biolis 12i with an optional lon Selective Electrode Module for measuring sodium, potassium and chloride in serum. These two analyzers are used to analyze for electrolytes. These analyzers are calibrated with known concentration calibrator material and both utilize lon-Selective Electrodes.

Comparison table with Sirrus (Glucose)

ItemNew DeviceBiolis 12iPredicateSirrus(K042169)
General
System PrincipleDiscrete, single line random access, multi-test analysisDiscrete, random access, multi-test analysis
Throughput90 tests240 tests
ConfigurationAnalytical unit, Control unitAnalytical unit, Control unit
Measurement modesAbsorbanceAbsorbance
DetectorPhoto-diodePhoto-diode
Optical systemWavelength range of 340 to 800nmWavelength range of 340 to 800nm
Light sourceTungsten halogen lampTungsten halogen lamp
Reaction cuvettesPlastics, semi disposalPlastics, semi disposal
Path length8mm8mm
Reaction timeMaximum 10min.Maximum 10min.
Incubation temperature37°C +/- 0.1°C37°C +/- 0.1°C
Glucose
Intended useQuantitative determination ofglucose in serumQuantitative determination ofglucose in serum
MethodPhotometric endpoint usingglucose hexokinase.Photometric endpoint usingglucose
Sample typeSerumSerum,
Sample Volume2 uL3 uL
Wavelength340 / 405 nm340 / 405 nm
Reaction typeEndpointEndpoint
Read PointRead period: 19 - 20 cycles(30 seconds per cycle)Read period: 50 - 52 cycles(15 seconds per cycle)

Comparison Table with Prestige 24i (ISE)

ItemNew DeviceBiolis 12iPredicatePrestige 24i (K040958)
General
System PrincipleDiscrete, single line randomaccess, multi-test analysisDiscrete, random access,multi-test analysis
Throughput100 tests including ISE tests400 tests including ISE tests
ConfigurationAnalytical unit, Control unitAnalytical unit, Control unit
Measurement modesAbsorbanceAbsorbance
DetectorPhoto-diodePhoto-diode
Optical systemWavelength range of 340 toWavelength range of 340 to

. ·

{3}------------------------------------------------

Tokyo Boeki Medisys System Ltd.
1-14-21 Higashi-Toyoda, Hino, Tokyo 191-0052, Japan
Phone: +81-42-587-2965

800nm800nm
Light sourceTungsten halogen lampTungsten halogen lamp
Reaction cuvettesPlastics, semi disposalPlastics, semi disposal
Path length8mm8mm
Reaction timeMaximum 10min.Maximum 10min.
Incubation temperature37°C +/- 0.1°C37°C +/- 0.1°C
ISEThe ISE module is operated onthe Biolis 12i integrated system.The ISE module is operated onthe Biolis 24i integrated system
Intended useQuantitative determination ofNa, K and ClQuantitative determination ofNa, K and Cl
MethodIon selective electrodeIon selective electrode
Sample typeSerumSerum
Sample Volume60 uL60 uL
Analysis time100 seconds100 seconds

The validated system is described below.

. :

Analyzer:Tokyo Boeki BiOLiS 12i Analyzer with Ion -selective electrode module usingdirect potentiometry.
Serial numbers are listed on individual study reports
SoftwareInterface software version: 1.70ISE ROM software version: 3.05
Reagent:Carolina Liquid Chemistries Glucose Reagent, Kit product no. BL-208 (alsopackaged as AU-208). Lot numbers are listed on individual study reports
Calibratorfor glucosePointe Scientific Chemistry Calibratorproduct no. C7506-50, lot 11802, exp. March 2014
Premarket clearance reference no.: K070207
The 185 mg/dL glucose set point for the Pointe Calibrator was verified for theCarolina Liquid Chemistries Glucose Reagent by comparing the calibrator toNIST SRM 965b, Glucose in Frozen Serum. The Pointe Calibrator wasassayed eight times over each of four analytical runs against NIST standardlevels 3 and 4 which were each assayed in duplicate. The glucoseconcentration of the Pointe Calibrator was calculated for each run by linearinterpolation the NIST assay values and their respective certified values of 118.5mg/dL and 294.5 mg/dL. The mean glucose result of the Pointe Calibrator overthe four runs was 185.6 mg/dL.
Calibratorfor ISECalibrator 1 and Calibrator 2Manufacturer: Tokyo Boeki Medisys
Lot numbers are listed on individual study reports
Premarket clearance reference no: K040958

{4}------------------------------------------------

1-14-21 Higashi-Toyoda, Hino, Tokyo 191-0052, Japan Phone: +81-42-587-2965 Fax: ++81-42-587-7781

Performance Characteristics:

ISE

A correlation analysis between the Prestige 24i and the Biolis 12i yielded the following results:

RepresentativeMethodCorrelationCoefficientSlope(Least-Squares)Y-axis intercept
1Sodium0.98721.0204-3.1903
2Potassium0.99921.0255-0.1457
3Chloride0.99220.97361.8580

The linearity test yielded the following results:

Linearity
Sodium100 - 200 mmol/L (Serum)
Potassium1 - 10 mmol/L (Serum)
Chloride70 - 200 mmol/L (Serum)

The precision test results:

ItemSample 1%CVSample 2%CVSample 3%CV
4Within RunN=20Sodium0.850.610.93
Potassium0.900.910.94
Chloride0.790.830.69
5Day byDay-RunN=15Sodium0.40.40.3
Potassium0.80.90.4
Chloride1.00.70.4

Interferences

The Interference test yielded the following results:

No significant interference were observed for the substances at the concentration levels as follows

SubstanceNormalAbnormal
Bilirubin F19.7 mg/dL19.7 mg/dL
Bilirubin C21 mg/dL21 mg/dL
Hemoglobin488 mg/dL488 mg/dL
Lipemia1000mg/dL500 mg/dL

{5}------------------------------------------------

1-14-21 Higashi-Toyoda, Hino, Tokyo 191-0052, Japan Phone: +81-42-587-2965 Fax: ++81-42-587-7781

NormalAbnormal
Lithium Chloride3.20 mmol/L3.20 mmol/L
Sodium Bromide37.60 mmol/L37.60 mmol/L
Sodium Salicylate4.34 mmol/L4.34 mmol/L
Sodium Thiocyanate6.88 mmol/L6.88 mmol/L

:

1

:

:

:

Conclusion:

:

:

.

:

: :

:

.

: 上一篇:

. . . . .

:

:

.

. '

:

.

The lon Selective Electrode performance data for Sodium, Potassium and Chloride demonstrates that Biolis 12i is substantially equivalent to the Prestige 24i (K040958).

:

    1. 1

.

.

.

:

.

. : :

.

· . '

.

{6}------------------------------------------------

1-14-21 Higashi-Toyoda, Hino, Tokyo 191-0052, Japan Phone: +81-42-587-2965 Fax: ++81-42-587-7781

Glucose Reagent

Method Comparison

A correlation analysis between the Sirrus and the Biolis 12i yielded the following results:

RepresentativeMethodCorrelationCoefficientSlope(Least-Squares)Y-axis intercept
GLU0.99751.0447-4.845

Precision

The precision test results:

ItemSample 1CV (%)Sample 2CV (%)Sample 3CV (%)
1Within-runN=20GLU1.311.191.04
2Between-runN=20GLU0.870.780.70

Linearity

The linearity test yielded the following results:

Correlation0.9997
Slope0.96
Intercept-2.66
Range25 - 540mg/dL

Sensitivity

The minimum detection limit test yielded the following results:

Minimum Detectable value7.83 mg/dL
--------------------------------------

{7}------------------------------------------------

1-14-21 Higashi-Toyoda, Hino, Tokyo 191-0052, Japan Phone: +81-42-587-2965 Fax: ++81-42-587-7781

Interferences

  • The Interference test yielded the following results: .
    No significant interference were observed at the following concentration levels
NormalAbnormal
Hemoglobin500mg/dL500mg/dL
Bilirubin20 mg/dL20 mg/dL
Lipemia1000 mg/dL1000 mg/dL

Stability Summary

The calibration stability test yielded the following results:

Sample 1Sample 2Sample 3
CV (%)2.12.52.1

Conclusion:

The glucose performance data demonstrates that Biolis 12i is substantially equivalent to Sirrus (K 042169)

{8}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around a central emblem. The emblem features a stylized depiction of an eagle or bird-like figure, with flowing lines suggesting wings or feathers.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Tokyo Boeki Medisys Inc. c/o James Clinton c/o Quality & Regulatory Consulting, LLC. 5105 Fairoaks Rd. Durham, NC 27712

MAR 2 3 2012

K110520 Re:

Trade Name: BIOLIS 12i Regulation Number: 21 CFR §862.1665 Regulation Name: Sodium test system Regulatory Class: Class II Product Codes: JGS, CEM, CGZ, CFR, JJE Dated: February 29, 2012 Received: March 6, 2012

Dear Mr. Clinton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{9}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 796 tt the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical

Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm

Sincerely yours,

N

Counney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

INDICATIONS FOR USE

510(k) Number (if known): K110520

Device Name: Biolis 12i

Indications for Use:

The Biolis 12i is a discrete photometric chemistry analyzer with ion-selective electrode (ISE), with direct quantitative measurement of sodium, potassium, chloride, and glucose in serum. It is a device intended for the in-vitro, spectrophotometric determination of general chemistry assays for clinical use. The Biolis 12i includes an optional lon Selective Electrode (ISE) module for the measurement of sodium, potassium and chloride in serum. The Biolis 12i is not for Point-Of-Care testing. It is for vitro diagnostic use only.

Sodium measurements are used in the diagnosis and treatment diseases involving electrolyte imbalance.

Potassium measurements monitor electrolyte balance and in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.

Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders.

The Biolis 12i analyzer with glucose hexokinase assay is intended to measure glucose quantitatively in human serum. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic glycemia, and of the pancreatic islet cell carcinoma.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Rute Chube

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

110520 510(k)________________________________________________________________________________________________________________________________________________________________________

Page 1 of |

§ 862.1600 Potassium test system.

(a)
Identification. A potassium test system is a device intended to measure potassium in serum, plasma, and urine. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.(b)
Classification. Class II.