K Number
K110508
Device Name
SUGLUE-10
Manufacturer
Date Cleared
2011-06-03

(101 days)

Product Code
Regulation Number
872.3275
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
  • Final cementing of all-ceramic, composite, or metal inlays, onlays, crowns and bridges; 2-3-unit Maryland bridges and 3-unit inlay/onlay bridges (excluded for patients with bruxism or periodontitis)
  • Final cementing of posts and screws .
  • Final cementation of all-ceramic, or composite veneers ●
  • Final cementation of all-ceramic, composite, or metal restorations on implant . abutments
Device Description

Suglue-10, manufactured by 3M ESPE, is a dual-curing resin cement supplied in an automix syringe. It is intended to be used for the adhesive cementation of indirect restorations and will be available in various shades. Suglue-10 is intended to be used in combination with an adhesive, e.g. Adhesive EXL 759 (by 3M ESPE). The Adhesive EXL 759 can be used either as a "Total Etch" or "Self-Etch" procedure. The additional etching of the tooth structure increases the adhesive strength of the adhesive even further. Suglue-10 contains bi-functional (meth)acrylate. The proportion of inorganic fillers is about 43% by volume; the grain size (D 90%) is about 13 µm. The mixing ratio, based on volume, is 1 part base paste : 1 part catalyst.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Suglue-10 device:

The provided document is a 510(k) Summary of Safety and Effectiveness, which focuses on demonstrating substantial equivalence to predicate devices rather than proving specific clinical efficacy of the device as a standalone product through extensive clinical trials. Therefore, much of the information typically found in a study demonstrating a device meets acceptance criteria (especially for AI/ML devices) will not be present.

Based on the provided information, the "acceptance criteria" are the performance limits defined by the ISO standards and the "device performance" refers to the results obtained for Suglue-10 in comparison to these limits and its predicate device.


1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricMethodAcceptance Criteria (Limit)Suglue-10 ResultsPredicate (Malta) Results
Film thicknessISO 4049< 50 μm10 ± 2 μm13 ± 1 μm
Working timeISO 4049> 60 secpasspass
Setting timeISO 4049< 10 min03:30 min03:30 min
RadiopacityISO 4049> 1.0 mm2.3 mm1.8 mm
Flexural Strength
   dark curedISO 4049> 50 MPa77 ± 5 MPa84 ± 12 MPa
   light curedISO 4049> 50 MPa111 ± 7 MPa111 ± 16 MPa
Compressive Strength
   dark curedISO 9917na [MPa]278 ± 9 MPa263 ± 9 MPa
   light curedISO 9917na [MPa]286 ± 36 MPa256 ± 36 MPa
Surface Hardness
   dark curedISO 2039-1na [MPa]187 ± 12 MPa190 ± 20 MPa
   light curedISO 2039-1na [MPa]217 ± 83 MPa212 ± 30 MPa

Note: "na" for compressive strength and surface hardness limits indicates "not applicable" or that specific numerical limits were not defined in this summary, but rather the comparison was made to the predicate device.


2. Sample Size Used for the Test Set and Data Provenance

The document provides performance results but does not specify the sample sizes used for the various tests (e.g., how many samples were tested for film thickness, flexural strength, etc.).

There is also no explicit mention of data provenance in terms of country of origin or whether the data was retrospective or prospective. The manufacturer is 3M ESPE AG in Germany, implying the testing likely took place in Germany, but this isn't stated directly as "data provenance."


3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This type of information is not applicable and not provided in this document. The "tests" performed are objective, physical/mechanical property measurements based on ISO standards for dental cements, not subjective evaluations requiring expert interpretation or ground truth establishment in the context of clinical images or diagnostic decisions.


4. Adjudication Method for the Test Set

This is not applicable and not provided. The testing performed involves standardized physical and chemical measurements governed by ISO methods, not subjective assessments requiring adjudication.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

This is not applicable. Suglue-10 is a dental cement, not an AI/ML device or a diagnostic imaging tool that would involve human readers or AI assistance in interpretation. Therefore, an MRMC study and related effect sizes are irrelevant to this submission.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This is not applicable. Suglue-10 is a physical dental material, not an algorithm or software device.


7. The Type of Ground Truth Used

The "ground truth" for the performance data is based on standardized physical and chemical measurements conducted according to established International Organization for Standardization (ISO) methods (ISO 4049, ISO 9917, ISO 2039-1) for dental materials. These methods define how the properties are measured objectively, not through expert consensus, pathology, or outcomes data in a clinical sense.


8. The Sample Size for the Training Set

This is not applicable and not provided. As Suglue-10 is a physical product and not an AI/ML device, there is no "training set" in the context of machine learning.


9. How the Ground Truth for the Training Set Was Established

This is not applicable. As Suglue-10 is a physical product and not an AI/ML device, there is no "training set" or "ground truth for the training set."


Summary of the Study Proving Device Meets Acceptance Criteria:

The "study" presented here is a premarket notification (510(k)) submission for a dental cement. The purpose of this submission is to demonstrate substantial equivalence to existing legally marketed predicate devices, not to conduct a comprehensive clinical trial or establish de novo safety and effectiveness.

The study proves the device meets "acceptance criteria" by showing that its physical and chemical properties (film thickness, working time, setting time, radiopacity, flexural strength, compressive strength, surface hardness) are within acceptable limits defined by international ISO standards and are comparable to or better than a legally marketed predicate device (Malta).

The document states: "In summary, it can be concluded that Suglue-10 is as safe and effective as the predicate devices: Malta by 3M ESPE, Germany (K100756), Panavia F 2.0 by Kuraray Medical Inc., Japan (K032455), and Nexus 3 by Kerr Corporation, U.S.A. (K062519)." This conclusion is based on the comparative performance data, chemical composition analysis, and comparison of indications for use. Biocompatibility testing was also carried out, though the specific results are not detailed in this summary.

{0}------------------------------------------------

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS: K110508

Submitter

JUN - 3 2011

Company:3M ESPE AG
Street:ESPE Platz
ZIP-Code, City:D-82229 Seefeld
Federal State:Bavaria
Country:Germany
Establishment Registration Number9611385
Official Correspondent:Dr. Desi W. Soegiarto,Regulatory Affairs Specialist
Phone:+49-8152-700 1169
Fax:+49-8152-700 1869
E-mail:desi.soegiarto@mmm.com
Date:May 04, 2011

Name of Device

………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Proprietary Name:Suglue-10
Classification Name:Dental cement other than zinc oxide-
eugenol
Common Name:Adhesive resin cement

Predicate Devices

Malta by 3M ESPE, GermanyK100756
Panavia F 2.0 by Kuraray Medical Inc., JapanK032455
Nexus 3 by Kerr Corporation, U.S.A.,K062519

{1}------------------------------------------------

K110508

Description for the Premarket Notification

Suglue-10 is classified as a Dental cement other than zinc oxide-eugenol (21 C.F.R. §872.3275 [b]) because it is a device composed of various materials other than zinc oxide-eugenol.

Suglue-10, manufactured by 3M ESPE, is a dual-curing resin cement supplied in an automix syringe. It is intended to be used for the adhesive cementation of indirect restorations and will be available in various shades.

Suglue-10 is intended to be used in combination with an adhesive, e.g. Adhesive EXL 759 (by 3M ESPE). The Adhesive EXL 759 can be used either as a "Total Etch" or "Self-Etch" procedure. The additional etching of the tooth structure increases the adhesive strength of the adhesive even further.

Suglue-10 contains bi-functional (meth)acrylate. The proportion of inorganic fillers is about 43% by volume; the grain size (D 90%) is about 13 µm. The mixing ratio, based on volume, is 1 part base paste : 1 part catalyst.

Predicate devices to which Suglue-10 has been compared are Malta by 3M ESPE, Germany (K100756), Panavia F 2.0 by Kuraray Medical Inc., Japan (K032455) and Nexus 3 by Kerr Corporation, U.S.A. (K062519).

Technological Characteristics

As its predicate devices, Suglue-10 is a dual-curing, resin-based cement. Suglue-10 cement is based on the Malta material (K100756, by 3M ESPE, Germany). As Suglue-10 is designed to compete with the latest state of the art adhesive resin cements, its in vitro performance was compared to the newer system Nexus 3 and the still widely used Panavia F2.0.

The intended use of Suglue-10 is comparable to the area of the intended use of the predicate devices of Suglue-10.

In this 510(k) premarket notification Suglue-10 has been compared to its predicate devices with regard to chemical composition, performance data and indications for use.

Biocompatibility testing was carried out.

{2}------------------------------------------------

K110508

Performance Data

The following table shows the performance data of Suglue-10 and its predicate device Malta (K100756):

Suglue-10Malta
MethodLimitResults
Film thicknessISO 4049< 50 μm$10 \pm 2$$13 \pm 1$
Working timeISO 4049> 60 secpasspass
Setting timeISO 4049< 10 min03:3003:30
RadiopacityISO 4049> 1.0 mm2.31.8
Flexuralstrengthdark curedISO 4049> 50 MPa$77 \pm 5$$84 \pm 12$
light cured> 50 MPa$111 \pm 7$$111 \pm 16$
Compressivestrengthdark curedISO 9917na [MPa]$278 \pm 9$$263 \pm 9$
light curedna [MPa]$286 \pm 36$$256 \pm 36$
Surfacehardnessdark curedISO 2039-1na [MPa]$187 \pm 12$$190 \pm 20$
light curedna [MPa]$217 \pm 83$$212 \pm 30$

In summary, it can be concluded that Suglue-10 is as safe and effective as the predicate devices: Malta by 3M ESPE, Germany (K100756), Panavia F 2.0 by Kuraray Medical Inc., Japan (K032455), and Nexus 3 by Kerr Corporation, U.S.A. (K062519).

Indications for Use:

  • · Final cementing of all-ceramic, composite, or metal inlays, onlays, crowns and bridges; 2-3-unit Maryland bridges and 3-unit inlay/onlay bridges (excluded for patients with bruxism or periodontitis)
  • Final cementing of posts and screws .
  • Final cementation of all-ceramic, or composite veneers ●
  • Final cementation of all-ceramic, composite, or metal restorations on implant . abutments

{3}------------------------------------------------

K110508

Conclusion of Substantial Equivalence

The comparison for chemistry, performance data and indications for use shows that Suglue-10 is substantially equivalent to the predicate devices: Malta by 3M ESPE, Germany (K100756), Panavia F 2.0 by Kuraray Medical Inc., Japan (K032455) and Nexus 3 by Kerr Corporation, U.S.A. (K062519).

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle or bird symbol, with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Desi W. Soegiarto 3M ESPE AG ESPE Platz Seefeld, Bavaria Germany D-82229

Re: K110508

Trade/Device Name: Suglue-10 Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: II Product Code: EMA Dated: May 16, 2011 Received: May 23, 2011

JUN - 3 2011

Dear Ms. Soegiarto:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2- Ms. Soegiarto

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Wh for

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known):

Suglue-10 Device Name:

Indications For Use: • • Final cementing of all-ceramic, composite, or metal inlays, onlays, crowns and bridges; 2-3-unit Maryland bridges and 3-unit inlay/onlay bridges (excluded for patients with bruxism or periodontitis)

  • Final cementing of posts and screws .
  • Final cementation of all-ceramic, or composite veneers .
  • Final cementation of all-ceramic, composite, or metal restorations on . implant abutments

× Prescription Use . {Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Dunne

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page 1 of __1

510(k) Number: K110508

§ 872.3275 Dental cement.

(a)
Zinc oxide-eugenol —(1)Identification. Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.(2)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 872.9.(b)
Dental cement other than zinc oxide-eugenol —(1)Identification. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.(2)
Classification. Class II.