K Number
K110410
Manufacturer
Date Cleared
2011-08-04

(171 days)

Product Code
Regulation Number
882.1310
Panel
NE
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

When the Low output is selected the MS-120BK is used as a nerve stimulator for surgical procedures and brain mapping during treatment of patients with seizure disorders.

When High output is selected the MS-120BK is used for the intraoperative diagnosis of acute dysfunction in corticospinal axonal conduction brought about by mechanical trauma (traction, shearing, laceration, or compression) or vascular insufficiency.

The system is intended for use by qualified medical personnel within a hospital or clinical environment. The stimulator is available for use on any patients as determined by the qualified medical personnel.

Device Description

The Nihon Kohden MS-120BK is a dual (high & low) output constant current stimulator. The high and low setting, stimulation current and frequency of stimulation is selected by the user. The Nihon Kohden MS-120BK is connected to the MEE 1000A through the JB-116BK or JB-132BK amplifier.

In the Low output setting the Nihon Kohden MS-120BK applies cortical stimulation energy through the Nihon Kohden stimulation pod (JS-102B) which is connected to commercially available cortical electrodes (strip and grid electrodes).

In the High output setting, the MS-120BK outputs electrostimulation pulse through the Nihon Kohden extension cord (BM-121B) which is connected to commercially available stimulation electrode(s).

AI/ML Overview

This document describes a 510(k) submission for the Nihon Kohden MS-120BK Electric Stimulator (Extension Unit), primarily focusing on demonstrating its substantial equivalence to previously cleared predicate devices. Therefore, the study doesn't present typical "acceptance criteria" and "reported device performance" in the way one might expect for a novel AI/software device. Instead, the focus is on showing technical characteristics are equivalent and that the device complies with recognized safety standards.

Here's an analysis based on the provided text, addressing your points where information is available:

1. Table of Acceptance Criteria and Reported Device Performance

The submission demonstrates equivalence by comparing the technical characteristics of the new device to predicate devices. The "acceptance criteria" here are effectively that the new device's specifications fall within acceptable ranges similar to or better than the predicate devices, and that it complies with relevant safety and performance standards. The "reported device performance" is a direct comparison of these specifications.

CharacteristicPredicate Device (Nihon Kohden SEN-4100A)Predicate Device (Digitimer D185)New Device (Nihon Kohden MS-120BK)Acceptance Criteria (Implied)Reported Device Performance
High Output Setting
Voltage1000V1000V350VSimilar or acceptable range350V
Current1A (1k Ω load)1A (1k Ω load)200mA (1k Ω load)Similar or acceptable range200mA (1k Ω load)
Maximum Stimulation Energy50mJ50mJ40mJSimilar or acceptable range40mJ
Output Impedanceless than 120 Ωless than 120 Ω50 ΩSimilar or acceptable range50 Ω
Stimulation ModeConstant VoltageConstant VoltageConstant CurrentDemonstrates different, but acceptable and defined modeConstant Current
Pulse Duration50 μs50 μs50-1000 μsSimilar or acceptable range50-1000 μs
Output Frequency Range1 Hz1 Hz0.1-1HzSimilar or acceptable range0.1-1Hz
Number of Pulse Trains1-91-91-9Matches1-9
Interpulse Interval1.0-9.9ms1.0-9.9ms1.0-1000msSimilar or acceptable range1.0-1000ms
Limitation of Voltage/Current and PulsesVaries with predicatesVaries with predicatesVaries (e.g., 200mA: 3 Pulses)Similar or acceptable rangeVaries
Polarity ChangeOKOKOKMatchesOK
MonophasicOKOKOKMatchesOK
BiphasicNONOOKDemonstrates additional capabilityOK
AlternateNONOOKDemonstrates additional capabilityOK
Maximum Charge50 micro-Coulomb50 micro-Coulomb200 micro-CoulombSimilar or acceptable range200 micro-Coulomb
Pulse ShapeRectangularRectangularRectangularMatchesRectangular
Maximum Energy Per Pulse50mJ50mJ40mJSimilar or acceptable range40mJ
Maximum Energy Per Second150mJ150mJ40mJSimilar or acceptable range40mJ
Electrode Minimum Size0.5cm20.5cm20.5cm2Matches0.5cm2
Maximum Charge Density100 Micro-C/cm2100 Micro-C/cm2400 Micro-C/cm2Similar or acceptable range400 Micro-C/cm2
Interface ControlMEE-1000ANOOnly MEE-1000ASpecifies compatibilityOnly MEE-1000A
Trigger InputOKOKOK*2MatchesOK*2
Trigger OutputOKOKOK*2MatchesOK*2
Foot Switch ControlOKOKOK*2MatchesOK*2
Low Output Setting
Constant Current StimulatorYes (Nicolet)Yes (Ojemann)YesMatchesYes
Maximum Stimulation Charge15 micro-Coulomb (Nicolet)20 micro-Coulomb (Ojemann)4.5 micro-CoulombSimilar or acceptable range4.5 micro-Coulomb
Current Stimulation Range0.1 to 15mA (peak) (Nicolet)0 to 10mA (peak) (Ojemann)0 to 15mA (peak)Similar or acceptable range0 to 15mA (peak)
Stimulation Frequency1 to 100Hz (Nicolet)5, 10, 20, 50, 75, 100Hz (Ojemann)0.1 to 50HzSimilar or acceptable range0.1 to 50Hz
Stimulation Pulse Width Duration0.1 to 1.0 msec Per Phase (Nicolet)0.1 to 2.0 msec Per Phase (Ojemann)0.05 to 0.3 msec Per PhaseSimilar or acceptable range0.05 to 0.3 msec Per Phase
Pulse ShapeRectangular (Nicolet)Rectangular (Ojemann)RectangularMatchesRectangular
Electrode Minimum Size0.04cm2 (Nicolet)0.5cm2 (Ojemann)0.04cm2Similar or acceptable range0.04cm2
Maximum Charge Density375 micro-C/cm2 (Nicolet)40 micro-C/cm2 (Ojemann)113 micro-C/cm2Similar or acceptable range113 micro-C/cm2
Maximum RMS Current Density Per Pulse119 mA RMS/cm2 (Nicolet)8.9 mA RMS/cm2 (Ojemann)46 mA RMS/cm2Similar or acceptable range46 mA RMS/cm2

2. Sample Size Used for the Test Set and Data Provenance

This is a physical medical device (electrical stimulator), not an AI/software device that processes data. Therefore, the concept of a "test set" with "data provenance" (country of origin, retrospective/prospective) in the context of image analysis or diagnostic algorithms does not apply here.

The "testing" mentioned is focused on verifying the device's hardware and software performance against its specifications and compliance with safety standards. The document states:

  • "Testing of the Nihon Kohden System with the MS-120BK was performed in compliance with the Nihon Kohden Corporation design control process."
  • "Testing included: Software and hardware verification and validation, and the device is in compliance with the following voluntary industrial standards..." (followed by a list of IEC, EN, and CAN/CSA standards).

This implies standard engineering and regulatory compliance testing rather than a clinical trial with patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. As noted above, this device does not function by interpreting patient data for diagnosis, so there is no "ground truth" established by experts in this context. The "truth" is whether the device meets its technical specifications and safety standards.

4. Adjudication Method for the Test Set

Not applicable for the same reasons as point 3.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

No. This type of study is typically performed for diagnostic or AI-assisted devices to assess the impact on human reader performance. The MS-120BK is an electrical stimulator, and its effectiveness is determined by its physical output and adherence to specifications, not by how it affects human interpretation of data.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No. This is a physical stimulus device, not an algorithm. Its operation is inherently with a "human-in-the-loop" (qualified medical personnel).

7. The Type of Ground Truth Used

The "ground truth" for this type of device is its technical specifications and compliance with international safety and performance standards. Testing involved verifying that the device's electrical outputs, timing, and other physical parameters matched the design specifications and met the requirements of standards like IEC 60601-1, IEC 60601-1-2, IEC 60601-2-40, etc.

8. The Sample Size for the Training Set

Not applicable. This device does not use machine learning algorithms that require a "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable for the same reasons as point 8.

In summary, the provided document is a 510(k) summary for a traditional medical device (an electric stimulator). The "study" referenced is the internal design control process and associated testing to ensure the device meets its technical specifications safely and effectively, and is substantially equivalent to predicate devices, rather than a clinical study evaluating an AI/software's diagnostic or interpretive performance.

§ 882.1310 Cortical electrode.

(a)
Identification. A cortical electrode is an electrode which is temporarily placed on the surface of the brain for stimulating the brain or recording the brain's electrical activity.(b)
Classification. Class II (performance standards).