(196 days)
The neonatal bilirubin test used on RAPIDPoint® 405 systems is an in vitro diagnostic test for the determination of total neonatal bilirubin (nBili) concentration in the whole blood of newborn infants. Measurement of nBili aids in assessing the risk of kernicterus. This test is intended for use in point of care or laboratory settings.
Neonatal Bilirubin (nBili) is a new parameter offered on the RAPIDPoint 405 (RP405) blood gas system. The RP405 system is a point of care and laboratory testing blood gas analyzer and currently measures a variety of parameters that have been previously cleared. Enabling the nBili measurement is accomplished through software design changes introduced in Software Version 3.7. No hardware or mechanical changes were needed.
Here's a breakdown of the acceptance criteria and study information for the Siemens RAPIDPoint® 405 System Neonatal Bilirubin (nBili) Test, based on the provided 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (for substantial equivalence to predicate) | Reported Device Performance (nBili internal & external evaluation) |
|---|---|
| Correlation coefficient (r) value > 0.90 | Correlation coefficient (r) value was within the acceptance criteria (>0.90) |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Approximately 200 samples (combining internal and external evaluations).
- Data Provenance:
- Internal Evaluation: Unconjugated bilirubin in oxygenated cord whole blood (mimics neonatal samples). This suggests laboratory-controlled samples, likely prepared in a laboratory setting.
- External Evaluation: Intended use neonatal whole blood samples from multiple point-of-care sites. This indicates prospective, real-world clinical samples collected from neonates within various healthcare facilities.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not specify the number of experts used or their qualifications for establishing ground truth. The ground truth was established by comparison to the predicate devices.
4. Adjudication Method for the Test Set
The document does not describe an adjudication method for the test set. Performance was assessed by comparing the device's measurements directly to those of the predicate devices.
5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study
No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This device is an automated in vitro diagnostic system for measuring bilirubin, not an imaging device requiring interpretation by human readers. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply here.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Yes, a standalone study was done. The "Assessment of Performance" section describes the device's performance in measuring nBili directly, without human intervention in the measurement process itself. The system uses "iterative least squares analysis" and "correction for hematocrit to produce nBili results," indicating an automated algorithm.
7. Type of Ground Truth Used
The ground truth used was comparison to a predicate device. Specifically, "Specimens were evaluated against the RAPIDLab 1245 and 1265 predicate devices." This means the measurements from the RAPIDPoint® 405 System were compared against the established measurements from the legally marketed and previously cleared predicate devices to demonstrate substantial equivalence.
8. Sample Size for the Training Set
The document does not provide information on a separate training set or its sample size. This type of 510(k) submission for an in vitro diagnostic device, especially one involving a software update to an existing platform for a new parameter, often focuses on validation against a predicate rather than detailed machine learning model training data.
9. How the Ground Truth for the Training Set Was Established
Since no independent training set or machine learning model training details are described, the document does not explain how ground truth for a training set was established. The focus is on demonstrating clinical performance and substantial equivalence.
{0}------------------------------------------------
510(k) Summary - Revised
t
AUG 15 2011
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
| Submitter informationContact person: | Steven GoldbergSpecialist, Point of Care Regulatory Affairs |
|---|---|
| Address: | Siemens Healthcare DiagnosticsPoint of Care Products2 Edgewater DriveNorwood, MA 02062 |
| Phone: | 781-269-3941 |
| Date summary prepared: | July 11, 2011 |
| Device InformationProprietary Name: | RAPIDPoint® 405 System Neonatal Bilirubin (nBili) Tes |
| Common name: | Neonate bilirubin on blood gas system |
| Classification name: | Bilirubin in the neonate test system |
| Classification number: | 21 CFR 862.1113, Class I, reserved, MQM |
| Classification panel: | Clinical Chemistry and Clinical Toxicology Devices |
Predicate Device
| Element | Predicate Device |
|---|---|
| Device Name | Neonatal Bilirubin on RAPIDLab® 1200 Blood GasSystem |
| Common Name | Neonate bilirubin on blood gas system |
| 510(k) Number | K073537 |
| Manufacturer | Siemens Healthcare Diagnostics Limited |
Device Description
Neonatal Bilirubin (nBili) is a new parameter offered on the RAPIDPoint 405 (RP405) blood gas system. The RP405 system is a point of care and laboratory testing blood gas analyzer and currently measures a variety of parameters that have been previously cleared. Enabling the nBili measurement is accomplished through software design changes introduced in Software Version 3.7. No hardware or mechanical changes were needed.
{1}------------------------------------------------
510(k) Summary - Revised
The nBili test is intended for in vitro diagnostic use by healthcare professionals for the determination of total neonatal bilirubin concentration in the whole blood samples of newborn infants. Measurement of nBili aids in assessing the risk of kernicterus.
Statement of Intended Use
The neonatal bilirubin test used on RAPIDPoint® 405 systems is an in vitro diagnostic test for the determination of total neonatal bilirubin (nBili) concentration in the whole blood of newborn infants. Measurement of nBili aids in assessing the risk of kernicterus. This test is intended for use in point of care or laboratory settings.
Summary of Technological Characteristics
The RAPIDPoint 405 system uses multiple wavelength spectrophotometry (CO-oximetry) to measure the transmission of light through a sample of neonate whole blood to determine concentrations of hemoglobin derivatives and bilirubin.
The RAPIDPoint 405 system aspirates the whole blood sample at the sample port and then transfers the sample to the CO-ox sample chamber. As the sample flows through an optical chamber, the CO-ox optics head directs light through the sample and to a polychromator that measures the intensity of transmitted light at different wavelengths.
Iterative least squares analysis is used to determine raw bilirubin values. Raw values are then corrected for hematocrit to produce nBill results.
Assessment of Performance
Studies were conducted to demonstrate the performance of the RAPIDPoint 405 with nBill parameter and assess substantial equivalence against the Siemens Healthcare Diagnostics RAPIDLab 1245 and 1265 (predicate device).
The nBili internal evaluation study entailed testing concentrations of unconjugated bilirubin in oxygenated cord whole blood (which mimics neonatal samples) across the reporting ranges. The nBili external evaluation study included testing at multiple point of care sites with intended use neonatal whole blood samples. Combining these internal and external evaluations for a total of approximately 200 samples, the correlation coefficient (r) value was within the acceptance criteria (>0.90). Specimens were evaluated against the RAPIDLab 1245 and 1265 predicate devices and based on performance data analyzed, it was concluded that the predetermined acceptance criteria were met.
In addition, information on Software Development Life Cycle including software requirements specifications, risk management report, and overall verification and validation results were included to provide additional assurance of device performance.
{2}------------------------------------------------
510(k) Summary - Revised
Conclusion
In conclusion, the studies completed demonstrate that the RAPIDPoint® 405 System Neonatal Bilirubin (nBili) Test is similar to the RAPIDLab 1245 and 1265 predicate device in both Technological Characteristics and Intended Use. The data presented in the submission is a summary of internal evaluation, external clinical evaluation, and software development information. This information provides assurance that the RAPIDPoint 405 system measuring neonate bilirubin has demonstrated substantial equivalence to the currently marketed Siemens Healthcare Diagnostics RAPIDLab 1245 and 1265 predicate device across the reporting range.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. U.S.A." arranged around the perimeter. Inside the circle is a stylized graphic of a human figure with outstretched arms, representing care and protection.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Siemens Healthcare Diagnostics c/o Steven Goldberg 2 Edgewater Drive Norwoord, MA 02062
K110277 Re:
Trade Name: Siemens Healthcare Diagnostics, RAPIDPoint® 405 System Neonatal Bilirubin (nBili) Test Regulation Number: 21 CFR §862.1113 Regulation Name: Bilirubin in the Neonate Test System Regulatory Class: Class I, reserved Product Codes: MQM Dated: July 11, 2011 Received: July 12, 2011
Dear Mr. Goldberg,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2 -
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
CJC.
Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
: ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' ' '
Enclosure
{5}------------------------------------------------
Indications for Use Form
510(k) Number (if known):
Device Name:
RAPIDPoint® 405 System
Intended Use:
The RAPIDPoint 405 system is intended for in vitro diagnostic use and is designed to provide the determination in whole blood for the following parameters:
- Partial pressure of carbon dioxide .
- Partial pressure of oxygen .
- pH .
- Sodium .
- . Potassium
- Ionized calcium .
- Chloride .
- Glucose .
- Total hemoglobin and fractions: fO2Hb, fCOHb, fMetHb, fHHb .
- Neonatal bilirubin .
This test system is intended for use in point of care or laboratory settings.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign Off
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
K110277 510(k)
{6}------------------------------------------------
510(k) Number: K110277 RAPIDPoint® 405 System Device Name:
The following list includes the Indications for Use information for each analyte measured on the Rapidpoint 405 System:
pCO2, p02, pH. Measurements of blood gases (pCO2, p02) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.
Neonate Bilirubin. A bilirubin (total and unbound) in the neonate test system is a device intended to measure the levels of bilirubin (total and unbound) in the blood (serum) of newborn infants to aid in indicating the risk of bilirubin encephalopathy (kernicterus).
Sodium. Sodium measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
Potassium. Potassium measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
Chloride. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
Ionized calcium. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
CORS
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) | K110277 | | | | | | | | |
{7}------------------------------------------------
Indications for Use Form
510(k) Number: K110277 RAPIDPoint® 405 System Device Name:
Glucose. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Total hemoglobin. Total hemoglobin measurements are used to determine the hemoglobin content of human blood.
Oxyhemoglobin. Oxyhemoglobin measurements are used to measure the hemoglobin content of whole blood for the detection of anemia.
Carboxyhemoglobin. Carboxyhemoglobin measurements are used to determine the carboxyhemoglobin (the compound formed when hemoglobin is exposed to carbon monoxide) content of human blood as an aid in the diagnosis of carbon monoxide poisoning.
Sulfhemoglobin, Sulfhemoglobin measurements are used to determine the sulfhemoglobin (a compound of sulfur and hemoglobin) content of human blood as an aid in the diagnosis of sulfhemoglobinemia (presence of sulfhemoglobin in the blood due to drug administration or exposure to a poison).
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign Off
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K110277 -
§ 862.1113 Bilirubin (total and unbound) in the neonate test system.
(a)
Identification. A bilirubin (total and unbound) in the neonate test system is a device intended to measure the levels of bilirubin (total and unbound) in the blood (serum) of newborn infants to aid in indicating the risk of bilirubin encephalopathy (kernicterus).(b)
Classification. Class I.