(254 days)
For in vitro diagnostic use as a control for calibration verification of the IMMULITE® Unconjugated Estriol (uE3) assays on the IMMULITE 1000, 2000 systems.
The calibration verification material is assayed control with four levels. The analyte levels for IMMULITE 1000 are 0.00, 0.16, 2.70 and 11.2 ng/ml, and 0.00, 0.19, 2.90 and 12.00 for IMULITE 2000 systems. The matrix is estriol in processed horse serum.
One set of four vials, 2 mL each, containing low, intermediate and high levels of unconjugated estriol in processed horse serum, with preservative, and an unconjugated estriol free sample. The calibration verification material (CVM) is supplied in liquid form, ready to use. Store unopened materials refrigerated at 2-8°C until expiration date. CVM's are for single use only, immediately after opening, and discard after use.
Here's a breakdown of the acceptance criteria and the study details for the IMMULITE® Unconjugated Estriol (uE3) Calibration Verification Material, based on the provided 510(K) summary:
1. Table of Acceptance Criteria and Reported Device Performance:
The primary acceptance criteria for this device revolve around the precision and accuracy of the assigned values for the Calibration Verification Materials (CVMs) across different platforms.
| Parameter | Acceptance Criteria (IMMULITE/IMMULITE® 1000) | Reported Device Performance (IMMULITE/IMMULITE® 1000) | Acceptance Criteria (IMMULITE® 2000) | Reported Device Performance (IMMULITE® 2000) |
|---|---|---|---|---|
| CVM 1 (0 ng uE3/mL) | Range: 0 - 0.07 ng uE3/mL | CVM 1 = 0.0 ng/mL (Target) | Range: 0 - 0.07 ng uE3/mL | CVM 1 = 0.0 ng/mL (Target) |
| Precision: NA | Consistent with 0 target | Precision: NA | Consistent with 0 target | |
| CVM 2 (Target u: 0.16 ng uE3/mL) | Range: 0.07 - 0.25 ng uE3/mL | Within range (based on value assignment) | Range: 0.09 - 0.29 ng uE3/mL | Within range (based on value assignment) |
| Precision: <15% CV | Met (<15% CV) | Precision: <15% CV | Met (<15% CV) | |
| CVM 3 (Target u: 2.7 ng uE3/mL) | Range: 1.91 - 3.49 ng uE3/mL | Within range (based on value assignment) | Range: 2.03 - 3.77 ng uE3/mL | Within range (based on value assignment) |
| Precision: <10% CV | Met (<10% CV) | Precision: <10% CV | Met (<10% CV) | |
| CVM 4 (Target u: 11.2 ng uE3/mL) | Range: 8.73 - 13.7 ng uE3/mL | Within range (based on value assignment) | Range: 8.86 - 15.14 ng uE3/mL | Within range (based on value assignment) |
| Precision: <10% CV | Met (<10% CV) | Precision: <10% CV | Met (<10% CV) | |
| Overall Acceptance | CVM ranges fall within established ranges for all levels. | Achieved | CVM ranges fall within established ranges for all levels. | Achieved |
Study Proving Device Meets Acceptance Criteria:
The study proving the device meets the acceptance criteria is detailed under the "Performance" section, specifically regarding "The traceability, value assignment, and stability of the Calibration Verification Materials."
2. Sample Size Used for the Test Set and the Data Provenance:
- Test Set Sample Size: The document states that "Production lots of calibration verification materials are value assigned against the reference calibrators using multiple reagent lots which are run on at least three different instruments." While a specific number for "multiple reagent lots" and the number of individual CVM units tested from each lot isn't provided, it implies a systematic testing approach across different manufacturing batches and analytical systems.
- Data Provenance: The data is generated internally by Siemens Healthcare Diagnostics. The specific country of origin for the data generation is not explicitly stated, but the applicant's address is Tarrytown, NY, USA, suggesting the studies were conducted within the US or under the purview of their US operations. The study is prospective in the sense that new production lots of CVMs are being analyzed for value assignment and validation.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:
This information is not directly applicable in the context of this device. This is a calibration verification material, not a diagnostic device that interprets medical images or generates a diagnosis requiring expert review for ground truth.
Instead, the "ground truth" for the uE3 values in the reference calibrators is established by GC-MS (Gas Chromatography-Mass Spectrometry). This is a highly accurate and precise analytical method considered a gold standard for quantifying specific analytes. The document mentions that these reference calibrators are assigned values using a panel of real serum samples from pregnancy patients, which were value-assigned by GC-MS.
4. Adjudication Method for the Test Set:
Not applicable. This is not a study requiring adjudication of diagnostic interpretations. The value assignment involves running the CVMs on multiple instruments and averaging the observed values.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is a chemical/analytical device (calibration verification material), not a diagnostic imaging or AI-assisted diagnostic device that involves human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This is a chemical/analytical device. The performance refers to the accuracy and precision of the CVMs when analyzed on the IMMULITE systems, not an algorithm's performance.
7. The Type of Ground Truth Used:
The ground truth for the uE3 values in the reference calibrators (which are then used to assign values to the CVMs) is established by GC-MS (Gas Chromatography-Mass Spectrometry). This is considered an analytical gold standard.
8. The Sample Size for the Training Set:
This information is not explicitly provided as a classical "training set" for an algorithm. However, the development of the "uE3 reference calibrator values" involved using a "panel of real serum samples from pregnancy patients" that were value-assigned by GC-MS. The size of this panel (i.e., the number of patient samples) is not specified.
9. How the Ground Truth for the Training Set was Established:
The ground truth for the uE3 reference calibrator values was established by analyzing a panel of real serum samples from pregnancy patients using GC-MS (Gas Chromatography-Mass Spectrometry). These GC-MS derived values then served as the basis for assigning values to the internal reference calibrators, which in turn are used to value assign the CVMs.
{0}------------------------------------------------
510(K) Summary of Safety and Effectiveness
IMMULITE® Unconjugated Estriol (uE3) Calibration Verification Material
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990, and 21 CFR 807.92.
A. 510(K) Number: K110061
B. Date of Preparation: September 14, 2011 (Revised)
C. Proprietary and Established Names:
IMMULITE® Unconjugated Estriol (uE3) Calibration Verification Material
Applicant:
Siemens Healthcare Diagnostics Inc., 511 Benedict Ave, Tarrytown, NY 10591
Ernest Joseph, Regulatory Affairs Manager
Office: (914) 524-2431 Fax: (914) 524-2601
E. Regulatory Information:
IMMULITE® Calibration Verification Material, uE3
- Regulation section: 21 CFR § 862.1660
-
- Classification: Class I
-
- Product Code: JJX
-
- Panel: Clinical Chemistry
F. Predicate Device:
The ADVIA Centaur® Enhanced Estradiol Master Curve Material is for in vitro diagnostic use in the verification of calibration and reportable Range in the ADVIA Centaur® Enhanced Estradiol (eE2) assay cleared under 510k number K102904.
G. Device Description:
IMMULITE unconjugated Estriol assays are traceable to an internal standard and manufactured using qualified materials and measurement procedures. Calibration
{1}------------------------------------------------
Verification Material (CVM) is traceable to this standard. For additional information, consult the assay instruction for use.
One set of four vials, 2 mL each, containing low, intermediate and high levels of unconjugated estriol in processed horse serum, with preservative, and an unconjugated estriol free sample. The calibration verification material (CVM) is supplied in liquid form, ready to use. Store unopened materials refrigerated at 2-8°C until expiration date. CVM's are for single use only, immediately after opening, and discard after use.
Statement of Intended Use:
For in vitro diagnostic use as a control for calibration of the IMMULITE® Unconjugated Estriol (uE3) assays on the IMMULITE/IMMULITE® 1000, and IMMULITE® 2000 systems.
Comparison to the Predicate Device: Similarities and Differences between the devices and the predicate are shown below:
| IMMULITE® Device | Predicate | |
|---|---|---|
| Item | IMMULITE® CalibrationVerification Material(K110061) | ADVIA Centaur® MasterCalibration Material(K102904) |
| Intended use | For in vitro diagnostic use asa control for calibrationverification of the IMMULITEUnconjugated Estriol (UE3)assays on theIMMULITE/IMMULITE 1000,IMMULITE 2000 systems. | The ADVIA ® CentaurEnhanced Estradiol MasterCurve Material is for in vitrodiagnostic use in theverification of calibrationand reportable range in theADVIA Centaur® EnhancedEstradiol (eE2) assay. |
| Assay | Estriol | Estradiol |
Similarities:
{2}------------------------------------------------
Differences:
| Item | Device | Device |
|---|---|---|
| IMMULITE® Calibration Verification Material(K110061) | ADVIA Centaur® Master Calibration Material(K102904) | |
| Form | Liquid | Lyophilized |
| Matrix | Estriol in processed horse serum | Estradiol in processed human serum |
| Number of Levels | 4 | 6 |
| Stability | Unopened -until expiration date on the label at 2-8 °C | Unopened - until expiration date on the label at 2-8°Creconstituted - 14 dayson-board - 6 hours |
Performance:
The traceability, value assignment, and stability of the Calibration Verification Materials have been validated following procedures of Siemens Healthcare Diagnostics.
Analyte Levels/ Targets:
| Value Assignment Targeted Concentration of levels | |
|---|---|
| IMMULITE® 1000 | IMMULITE® 2000 |
| CVM 1 = 0.0 ng/mL | CVM 1 = 0.0 ng/mL |
| CVM 2 = 0.16 ng/mL | CVM 2 = 0.19 ng/mL |
| CVM 3 = 2.7 ng/mL | CVM 3 = 2.9 ng/mL |
{3}------------------------------------------------
CVM 4 = 11.2 ng/mL
CVM 4 = 12.0 ng/mL
Matrix:
Normal horse serum
Constituents:
Unconjugated Estriol spiked in normal horse serum, with100 ppm Proclin 950 as a preservative.
Preparation method:
The IMMULITE uE3 Calibration Verification Material levels contain Estriol spiked in normal horse serum and 100 ppm Proclin 950 as a preservative.
uE3 CVM Value Assignment:
The uE3 reference calibrator values are assigned by using a panel of real serum samples from pregnancy patients which were value assigned by GC-MS. This reference calibrator lot was then used to assign values of uE3 CVM. The reference calibrator and CVM are run on both IMMULITE® 1000 and IMMULITE® 2000 systems. The average observed value for each CVM is used as their assigned value. Quality control is then performed by monitoring the recovery of controls.
Protocol:
Production lots of calibration verification materials are value assigned against the reference calibrators using multiple reagent lots which are run on at least three different instruments.
Calculations:
Calibration verification material values are derived using four-parameter weighted data algorithm reduction of the reference calibrator for each run on each instrument. Calibration verification material values are then averaged across all instruments.
CVM Release Ranges by Platform:
| IMMULITE/IMMULITE® 1000 | ||||
|---|---|---|---|---|
| CVM(ng uE3/mL) | Target u(ng uE3/mL) | Range(ng uE3/mL) | Precision(% CV) | |
| 0 | 0 | 0 - 0.07 | NA | |
| 0.16 | 0.16 | 0.07 - 0.25 | <15 | |
| 2.7 | 2.7 | 1.91 - 3.49 | <10 | |
| 11.2 | 11.2 | 8.73 - 13.7 | <10 |
{4}------------------------------------------------
| IMMULITE® 2000 | |||
|---|---|---|---|
| CVM(ng uE3/mL) | Target μ(ng uE3/mL) | Range(ng uE3/mL) | Precision(% CV) |
| 0 | 0 | 0 - 0.07 | NA |
| 0.19 | 0.19 | 0.09 - 0.29 | <15 |
| 2.9 | 2.9 | 2.03 - 3.77 | <10 |
| 12 | 12.0 | 8.86 -15.14 | <10 |
Assay Range (Target u)
0.07-12 ng uE3/mL
Validation of Value Assignment:
Controls and Calibration Verification Material (CVM) levels are run to examine against the specification.
Acceptance Criteria:
Calibration Verification Material (CVM) ranges fall within established ranges for all levels.
Calibration Interval:
Not Applicable for Calibration Verification Material (CVM)
Traceability:
Calibration Verification Material CVM traceability is internal reference calibrators. Reference calibrators are traceable to individual human samples assigned with GC-MS values.
Controls:
Commercially available controls are recommended.
QC Acceptance Limit:
To monitor system performance and chart trends, as a minimum requirement, it is recommended to assay two levels of quality control material on each day that samples are analyzed. It is also recommended to assay quality control samples when performing a two point Adjustment. Use of commercially available control materials with at least three levels is recommended. Control results are considered acceptable if they are within the range published for the system or within the range established by the customer using an appropriate internal laboratory quality control scheme.
{5}------------------------------------------------
The recommended assay adjustment interval was based on the analyte recovery of the controls and samples from kit reagents stored at 2-8°C over time. The control testing frequency is done according to standard practices in clinical laboratory quality control schemes.
Warnings and Precautions:
Please refer to attached IMMULITE Unconjuqated Estriol (uE3) Calibration Verification Material Instruction for Use document for Warnings and Precautions.
"For In Vitro Diagnostic Use" Statement:
Please refer to attached IMMULITE Unconjugated Estriol (uE3) Calibration Verification Material Instruction for Use document for the In Vitro Diagnostic Use Statement. For in vitro diagnostic use as a control for calibration of the IMMULITE Unconjugated Estriol (UE3) assays on the IMMULITE/IMMULITE 1000, 2000 and 2000 XPi systems.
Procedure:
Please refer to IMMULITE Calibration Verification Material "Instruction for Use" document for "Directions for Use" including reconstitution and preparation.
Limitations:
Please refer to IMMULITE Instruction for Use document for Limitations.
References:
Please refer to IMMULITE Unconjugated Estriol Instruction for Use document for References.
Technical Assistance:
Please refer to the IMMULITE Unconjugated Estriol Calibration Verification Material (CVM) Instruction for Use document and CVM labels for contact information concerning the manufacturer.
Conclusion statement:
Per 21 CFR 807.92(b)(3) the nonclinical tests conducted for the uE3 Calibration Verification Materials (CVM) demonstrate that the device is as safe and effective, and performs as well as the predicate device.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services (HHS). The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The caduceus is composed of three curved lines that resemble a stylized human figure. The logo is surrounded by a circular border with the words "U.S. Department of Health & Human Services" written around it.
Food and Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993
Image /page/6/Picture/4 description: The image shows a date, "SEP 21 2011". The date is written in a simple, sans-serif font. The letters and numbers are all capitalized. The date is likely part of a document or label.
Siemens Healthcare Diagnostics c/o Ernest Joseph Senior Manager, Regulatory Affairs 511 Benedict Avenue Tarrytown, NY. 10591, USA
Re: k | | 0061 Trade/Device Name: IMMULITE Unconjugated Estriol (uE3) Calibration Verification Material Regulation Number: 21 CFR 862.1660 Regulation Name: Quality Control Material (Assayed and Unassayed) Regulatory Class: Class I, Reserved Product Code: JJX Dated: July 21, 2011 Received: July 22, 2011
Dear Mr. Joseph
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR), Parts 800 to 895. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{7}------------------------------------------------
Page 2 -
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Viro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcessforYou/Industry/default.htm.
Sincerely yours.
signature
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Indications for Use Form
510(k) Number (if known)
Device Name: IMMULITE® Unconjugated Estriol (uE3) Calibration Verification Material (CVM)
Indications for Use:
For in vitro diagnostic use as a control for calibration verification of the IMMULITE® Unconjugated Estriol (uE3) assays on the IMMULITE 1000, 2000 systems.
The calibration verification material is assayed control with four levels. The analyte levels for IMMULITE 1000 are 0.00, 0.16, 2.70 and 11.2 ng/ml, and 0.00, 0.19, 2.90 and 12.00 for IMULITE 2000 systems. The matrix is estriol in processed horse serum.
X___________________________________________________________________________________________________________________________________________________________________________ Prescription Use (21 CFR 801 Subpart D)
Over The Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
CONS
Division Sign-Off Office of In Vitro Diagnostic devices Evaluation and Safety
510 (k) K110061
Page 1 of (
§ 862.1660 Quality control material (assayed and unassayed).
(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.