K Number
K103760

Validate with FDA (Live)

Device Name
TITAN
Manufacturer
Date Cleared
2011-05-05

(133 days)

Product Code
Regulation Number
882.1900
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Titan with DPOAE440 is intended for use in the audiologic evaluation and documentation of ear disorders using Distortion Product Otoacoustic Emissions. The target population for Titan with DPOAE440 includes all ages.

The Titan with ABRIS440 is intended for use in the audiologic evaluation and documentation of ear and nerve disorders using auditory evoked potentials from the inner ear, the auditory nerve and the brainstem. The target population for Titan with ABRIS440 is newborns.

Device Description

The instrument is audiometric equipment used for testing middle ear, inner ear and brainstem abnormalities.

The Titan is a combination of:

  • Audiometric device for measurement of aural acoustic impedance (IMP) (Cleared . under K083861)
  • Audiometric device for measurement of otoacoustic emissions (OAE) .
  • Audiometric device for measurement of auditory brainstem responses (ABR) ●

The Titan can have one of three functions depending on license:

The Titan IMP test is intended to change the air pressure in the external auditory canal and measure and graph the mobility characteristics of the tympanic membrane to evaluate the functional condition of the middle ear.

The Titan OAE test generates an acoustic stimulus in the ear canal at a moderate intensity level. The stimulus vibrates a thin membrane in the cochlear, and activates the outer hair cells located on the membrane. This activation is propagated back through the middle ear as sound into the ear canal. Microphone(s) detects OAE-related sounds.

The Titan ABR test produces a short acoustic stimulus and measures via cutaneous electrodes the auditory evoked potentials from the inner ear, the auditory nerve and the brainstem. The test uses an automatic detection algorithm to determine if an ABR is evoked in response to the short acoustic stimulus. The output of the device is either PASS (ABR detected) or REFER (ABR is not detected).

The Titan with DPOAE440 is intended as clinical audiometer. The Titan with ABRIS440 is intended as a screener audiometer.

AI/ML Overview

The Interacoustics Titan with ABRIS440 and Titan with DPOAE440 devices are audiometric equipment used for testing middle ear, inner ear, and brainstem abnormalities. The acceptance criteria and the study proving the device meets these criteria are detailed below.

1. Table of Acceptance Criteria and Reported Device Performance

The provided document evaluates the Titan with ABRIS440 against the Maico MB 11 (predicate device) and the Titan with DPOAE440 against the DPOAE20 (predicate device). The "Equivalence" column in the tables below serves as the acceptance criteria and indicates whether the new device met the criteria (Same, Similar).

Titan with ABRIS440 vs. Maico MB 11 (Predicate Device)

DescriptionMaico, MB 11 (Predicate)Titan with ABRIS 440 (New)Equivalence
Indications for useThe Maico MB 11 with BERAphone and the MB 11 Standard are intended to generate evoked auditory brainstem response (ABR) -based measurements for newborn hearing screening, providing a PASS/REFER result.The Titan with ABRIS440 is intended for use in the audiologic evaluation and documentation of ear and nerve disorders using auditory evoked potentials from the inner ear, the auditory nerve and the brainstem.Similar (Wording just a little different)
Target populationNewbornsNewbornsSame
Intended userThe User does not require special technical skills or interpretation of the results.The User does not require special technical skills or interpretation of the results.Same
Anatomical sitesExamination of Ear and hearing nervesExamination of Ear and hearing nervesSame
Safety standardsIEC 60601-1IEC 60601-1Same
Performance standardIEC 60645-7 (Type 2)IEC 60645-7 (Type 2)Same
Device TypeScreening device (PASS/REFER) resultScreening device (PASS/REFER) resultSame
System Configuration1-channel ABR system operated through a base unit connected via a USB port to any standard computer (Laptop Computer recommended)1-channel ABR system operated through a handheld base unit. The base unit can be operated stand alone or PC controlled through USB or Bluetooth.Similar (Titan can work as stand alone)
Display informationTesting sequence along with current EEG parameters (number of sweeps, number of artifacts, SNR, and artifact rejection system setting)PASS/REFER status, indicated with value between 0 and 100% where 100% indicates a pass. EEG peak or RMS value, rejection status, residual noise and what transducer(s) are detected.Similar (Minor differences in displayed information)
Electrode quality checkYESYESSame
Impedance Test (Quality control)Indicator in three levels. Red: above 10kOhm Yellow: between 10 & 8kOhm Green: below 8kOhmBefore recording: Electrode impedance is measured if they are above 10kOhm, below 10 kOhm or below 3 kOhm.Similar (Difference in mid-range due to measurement frequencies, Titan may require more careful electrode placement)
Binaural screeningYESYESSame
Pre-amplifier channels11Same
StimulusChirpsChirpsSame
Pre-amplifier Gain86 dB64 dBSimilar (Depends on amplifier characteristics, no influence on measuring results)
FilteringHW HP 30Hz filters of 1st orderTwo HW HP filters of 1st order: Cut-off frequencies 7Hz and 45HzSimilar (Both look at frequencies above 80Hz; response identical at that frequency. Basic purpose is to eliminate DC components)
MaskingNoneNoneSame
TransducersBERAphone TDH39 EAR 3A Insert phonesTitan Probe TDH39 EAR 3A Insert phonesSimilar (Titan probe is dedicated for Titan as BERAphone for MB 11)
Stimulus rate92.5/s90/sSimilar
Screening level35dBHL35dBHLSame
Screening measurement180 seconds180 secondsSame

Titan with DPOAE440 vs. DPOAE20 (Predicate Device)

DescriptionDPOAE20 (Predicate)DPOAE440 (New)Equivalence
Indications for useThe Interacoustics DPOAE20 system is for use in the audiologic evaluation and documentation of ear disorders using Distortion Product Otoacoustic Emission tone stimuli. The presence of Otoacoustic emissions suggests normal outer hair cell function within cochlea, which in turn suggests normal hearing. OAEs are recorded using an OAE probe which is placed in the ear canal. The OAE response from the ear is recorded and processed by the Eclipse and the DPOAE2O software and then displayed on the computer screen for evaluation.The Titan with DPOAE440 is intended for use in the diagnostic audiologic evaluation and documentation of ear disorders using Distortion Product Otoacoustic Emissions.Similar (Wording just a little different)
Target populationThe patient group includes all agesThe patient group includes all agesSame
Intended userEar, Nose, and Throat doctors, Neurology specialists, Audiologist and other health- professionals concerned with measuring auditory functions.The Titan is designed to be used only by skilled personnel such as audiologists, ENT surgeons, and doctors, hearing healthcare professionals or personnel with similar level of education. The device should not be used without the necessary knowledge and training to understand its use and how results should be interpreted.Same
Evaluation methodDPOAEDPOAESame
Safety standardIEC 60601-1IEC 60601-1Same
EMC standardIEC 60601-1-2IEC 60601-1-2Same
Performance standardIEC 60645-6 (Type 2)IEC 60645-6 (Type 2)Same
Test signalsIEC 60645-1/ANSI S3.6, IEC 60645-3IEC 60645-1/ANSI S3.6, IEC 60645-3Same
Stimulus: Frequency Range500-8000 Hz500 to 10000 HzSimilar (Extended range is a technical improvement due to better technology and probes)
Stimulus: Frequency Step50 Hz25 HzSimilar (Better resolution possible due to better technology)
Stimulus: Level30-75 dB SPL (70 dB above 6kHz)30 to 80 dB SPL (75 dB SPL for 6kHz and 65 dB SPL for 8kHz and 10kHz)Similar (Extended range is a technical improvement due to better technique and probes)
Stimulus: Level Step1 dB SPL1 dB SPLSame
TransducerDedicated DPOAE20/TEOAE25 probeDedicated IOW ProbeSimilar (New transducer due to hardware design, but same characteristics as predicate)
Probe detectionNoneAuto detectionDifferent (Probe unit has built-in memory for calibration values)
Recording Analysis timeMinimum 2 sec to unlimited test time.Minimum 2 seconds to unlimited timeSame
Recording A/D Resolution16 bit, 7.3 Hz resolution24 bit, 5.38 Hz resolutionSimilar (Better resolution due to better technology)
Artifact Reject System-30 -> +30 dB SPL or off.-30 to +30 dB SPL or offSame
Stimulus ToleranceAdjustable between 1 and 10 dBAdjustable between 1 and 10 dBSame
SNR CriteriaAdjustable between 1 and 20 dBAdjustable between 3 and 25 dBSimilar (Limits adjusted to better fit real-world practice, giving higher evidence for response)
Probe check window256 points frequency response of the ear canal due to a click stimulus presented with a rate of 100 Hz at 80 dB SPL1024 points frequency response of the ear canal due to a click stimulusSimilar (Better resolution due to better technology used)
DP-Response window4096 points frequency response4096 points frequency responseSame
Display General Display gainApplicable during testingApplicable during testingSame
DisplayStimulus level and typeStimulus level and type, Basic and advanced viewSimilar (A simpler (basic) view has been introduced)
DatabaseSupported by Interacoustics OtoAccess™ database. Operating under NOAH is possible, using IA NOAHLink.Supported by Interacoustics OtoAccess™ database. Operating under NOAH is possible, using IA NOAHLink.Same

2. Sample size used for the test set and the data provenance

The provided document states: "No clinical tests were performed". Instead, the device's substantial equivalence was determined based on its conformity to international standards and comparison to predicate devices, which implies that no specific test set of patient data was used for a clinical performance study as part of this submission. Therefore, data provenance is not applicable.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable, as no clinical tests were performed and no ground truth was established by experts for a test set in this submission. The comparison relies on technical specifications and international standards.

4. Adjudication method for the test set

Not applicable, as no clinical tests were performed and no ground truth was established by experts for a test set in this submission.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The document does not describe an MRMC study or AI assistance. The device is a diagnostic tool, not an AI-powered diagnostic aide.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

For the Titan with ABRIS440, the device includes an "automatic detection algorithm to determine if an ABR is evoked in response to the short acoustic stimulus. The output of the device is either PASS (ABR detected) or REFER (ABR is not detected)." This indicates a standalone algorithmic performance component. However, no specific standalone performance study results, such as sensitivity or specificity, are presented in this submission. The study relies on comparison to predicate devices and adherence to standards.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

Given the statement "No clinical tests were performed," there was no direct patient-level ground truth established for the purpose of this 510(k) submission. The ground truth for the equivalence claims is implicitly based on the established safety and performance characteristics of the predicate devices and adherence to relevant international standards (IEC 60601-1, IEC 60601-1-2, IEC 60645-7, IEC 60645-6).

8. The sample size for the training set

Not applicable, as no clinical tests were performed and the device is not described as utilizing a machine learning algorithm that would require a training set in the context of this 510(k) submission.

9. How the ground truth for the training set was established

Not applicable, as no training set was used in the context of this 510(k) submission.

{0}------------------------------------------------

Premarket Notification Submission 510(k)

MAY - 5 2011

510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

Titan Titan with DPOAE440 Titan with ABRIS440

SUBMITTER INFORMATION Company Name Company Address

Company Phone Company Fax Company e-mail Contact Person

Date Summary Prepared

DEVICE IDENTIFICATION Trade Name Common Name Classification Name

Product Code Panel Device Class

Interacoustics A/S Drejervaenget 8 Assens, DK-5610, Denmark +45 6371 3555 +45 6371 3522 info@interacoustics.com Erik Nielsen, Director, Quality and Regulatory Affairs Apr. 28 2011

Titan Audiometric equipment Otoacoustic emission device, Evoked response auditory stimulator and measurements EWO / GWJ Ear Nose & Throat / Neurology Class II

SUBSTANTIAL EQUIVALENCE (TITAN WITH DPOAE440) Predicate Device DPOAE20 (Eclipse) Manufacturer Interacoustics 510(k) No. KOROS39 Date Cleared 05/30/2006

SUBSTANTIAL EQUIVALENCE (TITAN WITH ABRIS440) Predicate Device Maico MB 11 Manufacturer MAICO 510(k) No. K082035 Date Cleared 01/29/2009

Section 05 p. 2 - 10

, •

Image /page/0/Picture/15 description: The image shows the Interacoustics logo. The logo consists of a globe-like icon on the left, followed by the word "Interacoustics" in bold, white letters. A registered trademark symbol is present to the right of the word. Below the logo, there is a tagline that is difficult to read due to its small size and resolution.

{1}------------------------------------------------

Device Description

The instrument is audiometric equipment used for testing middle ear, inner ear and brainstem abnormalities.

The Titan is a combination of:

  • Audiometric device for measurement of aural acoustic impedance (IMP) (Cleared . under K083861)
  • Audiometric device for measurement of otoacoustic emissions (OAE) .
  • Audiometric device for measurement of auditory brainstem responses (ABR) ●

The Titan can have one of three functions depending on license:

The Titan IMP test is intended to change the air pressure in the external auditory canal and measure and graph the mobility characteristics of the tympanic membrane to evaluate the functional condition of the middle ear.

The Titan OAE test generates an acoustic stimulus in the ear canal at a moderate intensity level. The stimulus vibrates a thin membrane in the cochlear, and activates the outer hair cells located on the membrane. This activation is propagated back through the middle ear as sound into the ear canal. Microphone(s) detects OAE-related sounds.

The Titan ABR test produces a short acoustic stimulus and measures via cutaneous electrodes the auditory evoked potentials from the inner ear, the auditory nerve and the brainstem. The test uses an automatic detection algorithm to determine if an ABR is evoked in response to the short acoustic stimulus. The output of the device is either PASS (ABR detected) or REFER (ABR is not detected).

The Titan with DPOAE440 is intended as clinical audiometer. The Titan with ABRIS440 is intended as a screener audiometer.

Indications for Use

The Titan with DPOAE440 is intended for use in the audiologic evaluation and documentation of ear disorders using Distortion Product Otoacoustic Emissions. The target population for Titan with DPOAE440 includes all ages.

The Titan with ABRIS440 is intended for use in the audiologic evaluation and documentation of ear and nerve disorders using auditory evoked potentials from the inner ear, the auditory nerve and the brainstem. The target population for Titan with ABRIS440 is newborns.

K103760

Section 05 p. 3 - 10

Image /page/1/Picture/18 description: The image shows the logo for Interacoustics. The logo features a stylized globe on the left, followed by the word "Interacoustics" in a bold, sans-serif font. The logo is set against a black background.

leading diagnostic solutions

{2}------------------------------------------------

Technological Characteristics

Titan consists of a handheld unit named Titan, Titan Cradle and PC software (Titan Suite, Titan modules/ Titan Applications). The measurements are controlled by the handheld unit. A license system makes it possible within each configuration to select which functionality the user wants to be incorporated in the system.

A connection box (shoulder box) enables different types of accessories to be connected to the platform (depending on the module)

The Titan platform can connect to PC software via a Bluetooth connection.

Image /page/2/Picture/6 description: The image shows a sequence of handwritten alphanumeric characters. The characters are "K103760" and are written in a simple, slightly slanted font. The characters are all connected, and the overall impression is that of a quickly written note.

Section 05 p. 4 - 10

Image /page/2/Picture/8 description: The image shows the Interacoustics logo. The logo consists of a globe-like symbol on the left and the word "Interacoustics" in bold, white letters on a black background. Below the logo, there is a tagline that reads "leading diagnostic solutions".

{3}------------------------------------------------

DescriptionMaico, MB 11Titan with ABRIS 440Equivalence
Indications for useThe Maico MB 11 withBERAphone and the MB 11Standard are intended togenerate evoked auditorybrainstem response (ABR) -based measurements fornewborn hearing screening,providing a PASS/REFER result.The Titan with ABRIS440 isintended for use in the audiologicevaluation and documentation ofear and nerve disorders usingauditory evoked potentials fromthe inner ear, the auditory nerveand the brainstem.SimilarThe wordingis just a littledifferent
Target populationNewbornsNewbornsSame
Intended userThe User does not require specialtechnical skills or interpretationof the results.The User does not require specialtechnical skills or interpretation ofthe results.Same
Anatomical sitesExamination of Ear and hearingnervesExamination of Ear and hearingnervesSame
Safety standardsIEC 60601-1IEC 60601-1Same
PerformancestandardIEC 60645-7 (Type 2)IEC 60645-7 (Type 2)Same
Device TypeScreening device(PASS/REFER) resultScreening device(PASS/REFER) resultSame
SystemConfiguration1 -channel ABR system operatedthrough a base unit connectedvia a USB port to any standardcomputer (Laptop Computerrecommended)1 -channel ABR system operatedthrough a handheld base unit. Thebase unit can be operated standalone or PC controlled through USBor Bluetooth.Similar Theonly differentis that Titancan work asstand alone
Display information,Testing sequence along withcurrent EEG parameters (numberof sweeps, number of artifacts,SNR, and artifact rejectionsystem setting)PASS/REFER status, indicated withvalue between 0 and 100% where100% indicates a pass. EEG peak orRMS value, rejection status,residual noise and whattransducer(s) are detected.SimilarSome minordifferences indisplayedinformation
Electrode qualitycheckYESYESSame
Impedance Test(Quality control)Indicator in three levels.Red: above 10kOhmYellow: between 10 & 8kOhmGreen: below 8kOhmBefore recording:Electrode impedance is measured ifthey are above 10kOhm, below 10kOhm or below 3 kOhm.SimilarThedifference inthe mid rangeis due tomeasurementfrequencies
DescriptionMaico, MB 11Titan with ABRIS 440Equivalence
and hasinfluence onthe results.The Titan mayrequire a littlemorecarefulnesswhile placingelectrodes.
Binaural screeningYESYESSame
Pre-amplifierchannels11Same
StimulusChirpsChirpsSamePlease refertocomparisonbetween MB-11 andABRIS440 inattachment12-E
Pre-amplifier Gain86 dB64 dBSimilarDepends onamplifiercharacteristicsas CMMR andSN ratio. Haveno influenceon measuringresults.
FilteringHW HP 30Hz filters of 1st orderTwo HW HP filters of 1st order:Cut-off frequencies 7Hz and 45HzSimilarBoth look atfrequenciesabove 80Hzand at thatfrequency theresponsefrom the twofilters areidentical. Thebasic purposeof the filters
DescriptionMaico, MB 11Titan with ABRIS 440Equivalence
are toeliminate DCcomponents
MaskingNoneNoneSame
TransducersBERAphoneTDH39EAR 3A Insert phonesTitan ProbeTDH39EAR 3A Insert phonesSimilarThe Titanprobe is adedicatedprobe forTitan as theBERAphone isa dedicatedprobe for MB11
Stimulus rate92.5/s90/sSimilar
Screening level35dBHL35dBHLSame
Screeningmeasurement180 seconds180 secondsSame

.

Comparison Titan with ABRIS440 and predicate device

K103760

Section 05 p. 5 - 10

.

(1) Interacoustics® toading dlagnostic solullans

{4}------------------------------------------------

Section 05 p. 6 - 10

Image /page/4/Picture/4 description: The image shows the Interacoustics logo. The logo consists of a globe-like icon on the left, followed by the word "Interacoustics" in a bold, sans-serif font. Below the word, there is a tagline that reads "leading diagnostic solutions" in a smaller font. The logo is set against a black background.

{5}------------------------------------------------

Premarket Notification Submission 510(k)

K103760

Section 05 p. 7 - 10

Image /page/5/Picture/5 description: The image shows the Interacoustics logo. The logo consists of a globe-like icon on the left, followed by the word "Interacoustics" in bold, white letters on a black background. Below the logo, the words "leading diagnostic solutions" are written in a smaller font.

{6}------------------------------------------------

DPOAE20 (Predicate)DPOAE440 (New)Comments
Indications for useThe Interacoustics DPOAE20 system is for use in the audiologic evaluation and documentation of ear disorders using Distortion Product Otoacoustic Emission tone stimuli. The presence of Otoacoustic emissions suggests normal outer hair cell function within cochlea, which in turn suggests normal hearing. OAEs are recorded using an OAE probe which is placed in the ear canal. The OAE response from the ear is recorded and processed by the Eclipse and the DPOAE2O software and then displayed on the computer screen for evaluation.The Titan with DPOAE440 is intended for use in the diagnostic audiologic evaluation and documentation of ear disorders using Distortion Product Otoacoustic Emissions.SimilarThe wording is just a little different
Target populationThe patient group includes all agesThe patient group includes all agesSame
Intended userEar, Nose, and Throat doctors, Neurology specialists, Audiologist and other health- professionals concerned with measuring auditory functions.The Titan is designed to be used only by skilled personnel such as audiologists, ENT surgeons, and doctors, hearing healthcare professionals or personnel with similar level of education. The device should not be used without the necessary knowledge and training to understand its use and how results should be interpreted.Same
Evaluation methodDPOAEDPOAESame
Safety standardIEC 60601-1IEC 60601-1Same
EMC standardIEC 60601-1-2IEC 60601-1-2Same
Performance standardIEC 60645-6 (Type 2)IEC 60645-6 (Type 2)Same
DPOAE20 (Predicate)DPOAE440 (New)Comments
Test signalsIEC 60645-1/ANSI S3.6 , IEC60645-3IEC 60645-1/ANSI S3.6 , IEC60645-3Same
Stimulus:
Frequency Range500-8000 Hz500 to 10000 HzSimilarThe DPOAE has anextended range. This isa technicalimprovement due tobetter technology andprobes.
Frequency Step50 Hz25 HzSimilarA technical betterresolution is possibledue to bettertechnology.
Level30-75 dB SPL ( 70 dB above6kHz)30 to 80 dB SPL (75 dB SPLfor 6kHz and 65 dB SPL for8kHz and 10kHz)SimilarThe DPOAE has anextended range. This isa technicalimprovement due tobetter technique andprobes.
Level Step1 dB SPL1 dB SPLSame
TransducerDedicatedDPOAE20/TEOAE25 probeDedicatedIOW ProbeSimilarNew transducer due tothe hardware design ofTitan is different fromDPOAE20 hardware.The transducer has thesame characteristics asthe predicate. Thetransducer is not newcompared to theprevious version ofTitan cleared underK083861
Probe detectionNoneAuto detectionDifferentThe probe unit hasbuild-in memory for allprobe relevantcalibration values.
Recording
Analysis timeMinimum 2 sec to unlimitedtest time.Minimum 2 seconds tounlimited time
A/D Resolution16 bit, 7.3 Hz resolution24 bit, 5.38 Hz resolutionSimilarBetter resolution dueto better technology
DPOAE20 (Predicate)DPOAE440 (New)Comments
Artifact Reject System-30 -> +30 dB SPL or off.-30 to +30 dB SPL or offSame
Stimulus ToleranceAdjustable between 1 and 10 dBAdjustable between 1 and 10 dBSame
SNR CriteriaAdjustable between 1 and 20 dBAdjustable between 3 and 25 dBSimilarThe limits has beadjusted to better fit toreal world practice.This givers higherevidence for response.(1 dB is in fact toosmall to use as criteria)
Probe check window256 points frequencyresponse of the ear canaldue to a click stimuluspresented with a rate of 100Hz at 80 dB SPL1024 points frequencyresponse of the ear canaldue to a click stimulusSimilarBetter resolution dueto better technologyused
DP-Responsewindow4096 points frequencyresponse4096 points frequencyresponseSame
Display
General Display gainApplicable during testingApplicable during testingSame
DisplayStimulus level and typeStimulus level and type,Basic and advanced viewSimilarA simpler (basic) viewhas been introduces
Database
Supported by InteracousticsOtoAccess™ database.Supported byInteracoustics OtoAccess™database.Same
Operating under NOAH ispossible, using IANOAHLink.Operating under NOAH ispossible, using IANOAHLink.Same

- : -

ı

Image /page/6/Picture/4 description: The image shows a sequence of handwritten characters, which appear to be alphanumeric. The sequence reads 'K103760'. The characters are written in a simple, clear style, with each digit and letter easily distinguishable. The writing is in black against a white background.

Section 05 p. 8 - 10

:・

Image /page/6/Picture/6 description: The image shows the logo for Interacoustics, a company that provides leading diagnostic solutions. The logo consists of a stylized globe on the left, followed by the company name in bold, white letters on a black background. Below the company name is the tagline "leading diagnostic solutions" in a smaller font.

{7}------------------------------------------------

Section 05 p. 9 - 10

Image /page/7/Picture/4 description: The image shows the logo for Interacoustics, a company that provides diagnostic solutions. The logo consists of a globe-like icon and the company name in a bold, sans-serif font. Below the company name is the tagline "leading diagnostic solutions" in a smaller font.

{8}------------------------------------------------

Premarket Notification Submission 510(k)

Test summary

The Titan was tested according to current standards for OAE and ABR and was found to conform to the standards. No clinical tests were performed, but based on the fulfillment of the international standards for OAE and ABR, and the comparison to predicate devices we believe the device is safe and effective.

Conclusion

We have compared the intended use and performance characteristics with the predicate devices. The Titan was tested according to current standards and the differences found between the devices were related to functionality, not in relation to safety and efficiency. The Titan conforms to the current standards.

The Titan with DPOAE440 and Titan with ABRIS440 were found to be substantially equivalent to the predicate devices in technological characteristics and indications for use.

Section 05 p. 10 - 10

Image /page/8/Picture/9 description: The image shows the logo for Interacoustics. The logo consists of a globe-like icon on the left and the word "Interacoustics" in a bold, sans-serif font on the right. The background is black, and the text and icon are white.

Image /page/8/Picture/10 description: The image shows a sequence of characters that appear to be handwritten. The characters are 'K103760', with each digit and letter clearly distinguishable. The handwriting style is casual, with some characters slightly slanted and connected.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines forming its body and wings. The eagle is facing left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room ~WO66-G609 Silver Spring, MD 20993-0002

Interacoustics A/S c/o Erik Nielsen Director, Quality and Regulatory Affairs Drejervaenget 8 Assens, DK-5610 Denmark

MAY - 5 2011

Re: K103760

Trade/Device Name: · Titan with ABRIS440 and Titan with DPOAE440 Regulation Number: 21 CFR 882.1900 Regulation Name: Evoked Response Auditory Stimulator Regulatory Class: Class II Product Code: GWJ, EWO Dated: April 7, 2011 Received: April 12, 2011

Dear Mr. Nielsen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class-II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{10}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Rithm, h

Malvina B. Eydelman, M.D Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

103760

Premarket Notification Submission 510(k)

Indications for Use

Applicant: Interacoustics A/S

510(k) Number (if known): K103760

Device Name: Titan with DPOAE440, Titan with ABRIS440

Indications for Use:

The Titan with DPOAE440 is intended for use in the audiologic evaluation and documentation of ear disorders using Distortion Product Otoacoustic Emissions. The target population for Titan with DPOAE440 includes all ages.

The Titan with ABRIS440 is intended for use in the audiologic evaluation and documentation of ear and nerve disorders using auditory evoked potentials from the inner ear, the auditory nerve and the brainstem. The target population for Titan with ABRIS440 is newborns.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR

(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear,
Nose and Throat Devices

K103760

510(k) Number

Image /page/11/Picture/17 description: The image shows the Interacoustics logo. The logo consists of a globe-like icon on the left and the word "Interacoustics" in bold, white letters on a black background. Below the logo, there is a tagline that reads "leading diagnostic solutions" in a smaller font size.

§ 882.1900 Evoked response auditory stimulator.

(a)
Identification. An evoked response auditory stimulator is a device that produces a sound stimulus for use in evoked response measurements or electroencephalogram activation.(b)
Classification. Class II (performance standards).