K Number
K103560
Manufacturer
Date Cleared
2011-05-18

(166 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SHASER IPL Hair Removal System is an over the counter device intended for removal of unwanted hair.

Device Description

The SHASER IPL Hair Removal System is a personal, light-based, hair reduction system intended to be sold over-the-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology. The SHASER IPL Hair Removal System consists of a powered base unit and a tethered handpiece from which the IPL flashes are delivered to the patient's skin. The system will be sold with light cartridges, cleaning cloth, and user's manual.

AI/ML Overview

Here's an analysis of the provided text regarding the Shaser IPL Hair Removal System's acceptance criteria and study details:

Acceptance Criteria and Device Performance

Acceptance Criterion (Implicit)Reported Device Performance
Hair reduction at 3 months post-final treatmentAverage reduction of 37%
Hair reduction at 6 months post-final treatmentAverage reduction of 48%
Safety and effectiveness for over-the-counter useDemonstrated through clinical trial
Compliance with 21CFR § 1040.10 & 1040.11 (Performance Standards)Device complies with these standards

Study Details

  1. Sample size used for the test set and data provenance:

    • Sample Size: 111 test sites from an unspecified number of subjects (approximately 50% female and 50% male). The exact number of individual subjects is not explicitly stated, but it implies more than one since "approximately 50% female and 50% male subjects" is mentioned.
    • Data Provenance: The document does not specify the country of origin. The study was a "clinical trial," which implies a prospective design. Diverse demographical sub-groups were included.
  2. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

    • This information is not provided in the document. The method for assessing hair count reduction (ground truth) is stated as "Hair counts were taken before treatment and 3 and 6 months following the final treatment," but it does not specify who performed these counts or their qualifications.
  3. Adjudication method for the test set:

    • This information is not provided in the document.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    • No, an MRMC comparative effectiveness study was not done. The study described is a clinical trial to demonstrate the device's standalone effectiveness, not a comparison involving human readers with and without AI assistance.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, the described clinical trial evaluated the standalone performance of the Shaser IPL Hair Removal System. There is no mention of an algorithm or AI assistance; it's a device performance study.
  6. The type of ground truth used:

    • The ground truth was based on measured hair counts ("Hair counts were taken before treatment and 3 and 6 months following the final treatment"). This is objective, quantitative data.
  7. The sample size for the training set:

    • This device is an IPL hair removal system, not an AI or algorithm-based device that would typically have a "training set" in the machine learning sense. Therefore, this information is not applicable/provided. The clinical trial is the performance validation of the physical device.
  8. How the ground truth for the training set was established:

    • As in point 7, this question is not applicable as there is no mention of a training set for an AI model.

{0}------------------------------------------------

3560

e ① of ①

MAY 18 2011

510(k) Summary of Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Submitter Name:SHASER Inc.
Submitter Address :130 New Boston St., Ste 201, Woburn, MA 01801
Contact Person:Daniel L. Roth
Phone Number:(781)995-2246
Fax Number:(781)933-4640
Date Prepared:November, 2010
Device Trade Name:Shaser IPL Hair Removal System
Device Common Name:Light-based Hair Removal System
Classification Name:Laser surgical instrument for use in general and plasticsurgery and in dermatology, product code GEX
Predicate device:Home Skinovations Flash N' Go, K082298Tria Beauty Tria Laser Hair Removal System, K090820
Reason for submission:Not previously marketed in the USA

Device Description:

The SHASER IPL Hair Removal System is a personal, light-based, hair reduction system intended to be sold over-the-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology.

The SHASER IPL Hair Removal System consists of a powered base unit and a tethered handpiece from which the IPL flashes are delivered to the patient's skin. The system will be sold with light cartridges, cleaning cloth, and user's manual.

Intended Use:

The SHASER IPL Hair Removal System is an over the counter device intended for removal of unwanted hair.

Performance Data

The device complies with the following U.S. Food and Drug Administration performance standards: 21CFR & 1040.10 & 1040.11. A clinical trial was conducted to demonstrate the safety and effectiveness of the SHASER IPL Hair Removal System for over-the-counter use for hair removal. In a clinical trial, 111 test sites including arm, leg, chest, abdomen, bikini line and underarms were treated 3 times at 2 week intervals on a population of approximately 50% female and 50% male subjects from diverse demographical sub-groups. Hair counts were taken before treatment and 3 and 6 months following the final treatment. The results showed an average reduction of 37% and 48% at each respective end point.

Substantial Equivalence

Technological Characteristics and Comparison to Predicate Device(s):

The SHASER IPL Hair Removal System is substantially equivalent to the Home Skinovations Flash N' Go, K082298, and the Tria Beauty Tria Laser Hair Removal System, K090820.

The SHASER IPL Hair Removal System has the same intended uses and similar indications, technological characteristics, and principles of operation as the predicate devices. Any minor differences between the SHASER IPL Hair Removal System and its predicate devices raise no new questions of safety or effectiveness nor change the device's intended therapeutic effect in comparison to its predicates. Performance data demonstrate that SHASER IPL Hair Removal System is as safe and effective as its predicate devices for the stated indications. Thus, SHASER IPL Hair Removal System is substantially equivalent.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its wings or body.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Shaser, Inc. % The CRO Group, Inc. Mr. Glen Emelock 32 Harrison Street Melrose, Massachusetts 02176

MAY 1 8 2511

Re: K103560

Trade/Device Name: Shaser IPL Hair Removal System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: ONF Dated: May 11, 2011

Received: May 16, 2011

Dear Mr. Emelock:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Mr. Glen Emelock

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

A.J. B. Ri. Rete

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

ن 2 2 5 6 ت 510(k) Number:

Device Name: SHASER IPL Hair Removal System

Indications For Use:

The SHASER IPL Hair Removal System is an over the counter device intended for removal of unwanted hair.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use X (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Nel RPOgden for mxn

(Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K103560

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.