(365 days)
The Biolis 24i Clinical Chemistry Analyzer is a discrete photometric clinical chemistry analyzer. The device is intended to duplicate manual analytical procedures by automating various steps such as pipetting, heating, measuring color intensity, and reporting results. The device is intended to be used with certain materials to measure various analytes of diagnostic interest including glucose.
The Biolis 24i analyzer with glucose hexokinase assay is intended to measure glucose quantitatively in human serum. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic glycemia, and of the pancreatic islet cell carcinoma.
Using photometry, the Biolis 24i instrument measures the glucose concentration in serum by monitoring the change in absorbance at 340 nm. Additionally, the Biolis 24i with Ion-Selective Elective module additionally measures the concentration of the electrolytes, sodium, potassium and chloride in serum, using indirect potentiometry.
Here's an analysis of the provided text regarding the acceptance criteria and study for the Tokyo Boeki Medisys Inc. Biolis 24i Clinical Chemistry Analyzer, specifically for glucose measurement:
Acceptance Criteria and Device Performance for Biolis 24i Clinical Chemistry Analyzer (Glucose)
1. Table of Acceptance Criteria and Reported Device Performance
The provided 510(k) summary does not explicitly state pre-defined acceptance criteria values for each performance characteristic. Instead, it presents the results of the performance studies and implicitly suggests that these results demonstrated substantial equivalence to the predicate device. For the purpose of this response, I infer the reported performance as the outcome that met the (unspecified) acceptance criteria for substantial equivalence.
| Performance Characteristic | Acceptance Criteria (Inferred) | Reported Device Performance (Biolis 24i) |
|---|---|---|
| Method Comparison (vs. Predicate SYNCHRON CX 7) | Correlation Coefficient (R) close to 1 | GLU: 0.999 |
| Slope (Least-Squares) close to 1 | GLU: 0.974 | |
| Y-axis intercept close to 0 | GLU: 2.22 | |
| Precision (Repeatability) | Low Coefficient of Variation (CV%) | Serum 1: 1.3% Serum 2: 1.3% Serum 3: 1.2% Serum 4: 1.1% Serum 5: 1.5% |
| Precision (Total Imprecision) | Low Coefficient of Variation (CV%) | Analyzer 2229450610: Control 1: 2.4% Serum Pool 1: 2.5% Serum Pool 2: 2.1% Control 2: 2.3% Analyzer 2227671109: Control 1: 1.9% Serum Pool 1: 2.3% Serum Pool 2: 1.9% Control 2: 2.2% |
| Linearity | Correlation close to 1 | 0.9982 |
| Slope close to 1 | 0.948 | |
| Intercept close to 0 | 2.45 | |
| Range of linearity | 25 - 500 mg/dL | |
| Sensitivity | Low values for LoB, LoD, LoQ | LoB: 3.64 mg/dL LoD: 5.64 mg/dL LoQ: 10 mg/dL |
| Interferences | < 10% bias or clinically insignificant interference at stated concentrations | Ascorbic acid: none Bilirubin: -3.3 mg/dL, -3.6% Hemoglobin: -3.6 mg/dL, -3.8% Lipemia: +4 mg/dL, +3.9% Metronidazole: +2.4 mg/dL, +3.1% Tetracycline: none |
2. Sample Size Used for the Test Set and Data Provenance
- Method Comparison: The document does not explicitly state the total number of samples used for the method comparison (correlation analysis) with the Beckman CX-7.
- Precision (Repeatability): 60 replicates for each of 5 serum samples.
- Precision (Total Imprecision): 80 replicates for each of 4 samples (2 controls, 2 serum pools) on each of two different analyzers.
- Linearity: The sample size for the linearity study is not explicitly stated, but it covered a range of 25 - 500 mg/dL.
- Interferences: The sample sizes for the interference study are not explicitly stated, but tests were performed at two glucose concentrations (e.g., 75 mg/dL and 140 mg/dL for Ascorbic acid).
- Data Provenance: The document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective clinical studies. Given the manufacturer's location (Japan) and the context of typical lab instrument validation, these are likely prospective laboratory studies rather than large-scale clinical trials. The samples used (serum, controls, serum pools) suggest laboratory-generated or purchased materials.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
This device is a clinical chemistry analyzer, which measures quantitative values of analytes. The "ground truth" for such devices is established by reference methods or validated calibrated materials, not typically by human expert consensus or adjudication on individual cases.
- For Method Comparison: The "ground truth" was essentially the measurements obtained from the predicate device, the Beckman CX-7.
- For Calibration: The calibrator (Pointe Scientific Chemistry Calibrator) was verified against NIST SRM 965b (Glucose in Frozen Serum), which serves as a highly reliable reference.
- No human experts were used to establish the ground truth for the analytical performance of the test set in this context.
4. Adjudication Method for the Test Set
Not applicable. As noted above, the ground truth for this type of device is established by reference methods/calibrators, not by human expert adjudication of individual cases.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. An MRMC study is not relevant for a standalone clinical chemistry analyzer. MRMC studies are typically used for diagnostic imaging devices where human readers interpret images or data, and the effect of AI assistance on their performance is evaluated. This device provides a quantitative measurement directly.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, this entire submission focuses on the standalone performance of the Biolis 24i analyzer as a device for measuring glucose quantitatively. There is no human-in-the-loop component in its measurement principle; it is an automated instrument.
7. The Type of Ground Truth Used
The ground truth used was based on:
- Reference measurements from a predicate device (Beckman CX-7) for method comparison studies.
- NIST Standard Reference Material (SRM 965b, Glucose in Frozen Serum) for verifying the calibrator used to establish the measurement accuracy.
- Known concentrations in control materials and serum pools for precision studies.
8. The Sample Size for the Training Set
The document does not explicitly mention a "training set" in the context of device development. Clinical chemistry analyzers are typically developed and validated using well-established analytical principles and internal testing, calibration, and verification processes rather than machine learning "training sets" in the way an AI algorithm might have one. The "training" would refer to the calibration of the instrument using known calibrators. The text states:
- The calibrator (Pointe Scientific Chemistry Calibrator) was assayed eight times over each of four analytical runs against NIST standard levels 3 and 4, each assayed in duplicate. This process helps "train" or verify the calibration curve for the device.
9. How the Ground Truth for the Training Set Was Established
As mentioned above, the "training" in this context refers to the calibration and verification process. The ground truth for the calibrator was established by comparing it to NIST SRM 965b, Glucose in Frozen Serum, which has certified glucose values (118.5 mg/dL and 294.5 mg/dL). This comparison allowed for the verification of the 185 mg/dL glucose set point for the Pointe Calibrator.
{0}------------------------------------------------
Tokyo Boeki Medisys Inc. 1-14-21 Higashi-Toyoda, Hino, Tokyo 191-0051, Japan Phone: +81-42-587-2965 Fax: +81-42-587-7781
DEC - 1 2011
5. 510(k) SUMMARY
Submitter's Name/Address
| Submitter's Name: | Tokyo Boeki Medisys Inc. |
|---|---|
| Submitter's Address: | 1-14-21 Higashi-Toyota, HinoTokyo 191-0052 |
| Phone: | +81-42-587-2965 |
| Fax: | +81-42-587-7781 |
| Establishment Registration Number: | 3004378324 |
| Owner/Operator Number: | 9060135 |
Contact Person (United States Agent)
| Name of Agent: | James M. Clinton |
|---|---|
| Agent's Business Name: | Quality and Regulatory Consulting, LLC |
| Street Address: | 5105 Fairoaks Road |
| Durham, NC 27712-2078 | |
| Phone: | 919-247-0479 |
| Fax: | 919-287-2551 |
| E-mail address: | clintonjm@earthlink.net |
Date of Preparation of this Summary:
Device Trade or Proprietary Name:
Device Common Name:
October 25, 2011
Biolis 24i, MGC 240 and Prestige 24i, (These are the same models except the names)
Clinical Chemistry Analyzer (with optional ISE Module)
Classification Numbers/Class:
75JJE, Class I 75CFR Class II
{1}------------------------------------------------
vo Boeki Medisv
1-14-21 Higashi-Tovoda. Hino. Tokvo 191-0 Phone: +81-42-587-2965 Fax: +81-42-587-7
510(k) Summary:
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(K) number is : K103531
Description:
Using photometry, the Biolis 24i instrument measures the glucose concentration in serum by monitoring the change in absorbance at 340 nm. Additionally, the Biolis 24i with lon-Selective Elective module additionally measures the concentration of the electrolytes, sodium, potassium and chloride in serum, using indirect potentiometry.
Intended Use:
The Biolis 24i Clinical Chemistry Analyzer is a discrete photometric clinical chemistry analyzer. The device is intended to duplicate manual analytical procedures by automating various steps such as pipetting, heating, measuring color intensity, and reporting results. The device is intended to be used with certain materials to measure various analytes of diagnostic including glucose. An optional lon Selective Electrode Module is intended to measure sodium, potassium and chloride.
The Biolis 24i analyzer with glucose hexokinase assay is intended to measure glucose quantitatively in human serum. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic glycemia, and of the pancreatic islet cell carcinoma.
Substantial Equivalence:
Substantial equivalence has been demonstrated between the Beckman CX-7 (K904219) and the Biolis 24i. These analyzers are calibrated with known concentration calibrator material and both measure specific concentrations using photometry.
| Item | New DeviceBiolis 24i | PredicateSYNCHRON CX 7 (K904219) |
|---|---|---|
| General | ||
| System Principle | Discrete, single line random access, multi-test analysis | Discrete, random access, multi-test analysis |
| Throughput | 240 tests | 225 tests/hr (photometric only) |
| Configuration | Analytical unit, Control unit | Analytical unit, Control unit |
| Measurement modes | Absorbance | Absorbance |
Comparison with predicate:
{2}------------------------------------------------
Tokyo Boeki Medisys Inc.
1-14-21 Higashi-Toyoda, Hino, Tokyo 191-0051, Japan
Phone: +81-42-587-2965
| Detector | Photo-diode | Diffraction grating, photodiode array |
|---|---|---|
| Optical system | Wavelength range of 340 to 800nm | 340, 380, 410, 470, 520, 560, 600, 650, 670, 700 nm |
| Light source | Tungsten halogen lamp | xenon |
| Reaction cuvettes | Plastics, semi disposal | quartz |
| Path length | 8mm | 5 mm |
| Reaction time | Maximum 10min. | Maximum 12 min. |
| Incubation temperature | 37°C +/- 0.1°C | same |
| Glucose | Glucose hexokinase 510(k): K802810 | |
| Intended use | Quantitative determination of glucose in serum | Quantitative determination of glucose in serum plasma, urine, CSF |
| Method | Photometric endpoint using glucose hexokinase. | Same |
| Sample type | Serum | Serum, plasma, urine, CSF |
| Sample Volume | 3 uL | same |
| Reaction Time | 5 min` Analysis timeRead period: 52 - 54 points(15 seconds per point) | 5 min |
The validated system is described below.
.. .
.
| Analyzer: | Tokyo Boeki BioLis 24i Analyzer, software revision: 1.26Serial numbers are listed on individual data sheets. | ||
|---|---|---|---|
| Reagent: | Carolina Liquid Chemistries Glucose Reagent, Kit product no. BL-208 (alsopackaged as AU-208). Lot numbers are listed on individual data sheets. | ||
| Applicationparameters: | Sample volume: 3 µLReagent volume: 310 µLWavelengths: 340 / 405 nmReaction type: EndpointRead period: 52 - 54 | ||
| Calibrator: | Pointe Scientific Chemistry Calibratorproduct no. C7506-50, lot 11802, exp. March 2014Premarket clearance reference no.: K070207 | ||
| The 185 mg/dL glucose set point for the Pointe Calibrator was verified for theCarolina Liquid Chemistries Glucose Reagent by comparing the calibrator to NISTSRM 965b, Glucose in Frozen Serum. The Pointe Calibrator was assayed eighttimes over each of four analytical runs against NIST standard levels 3 and 4 whichwere each assayed in duplicate. The glucose concentration of the Pointe Calibratorwas calculated for each run by linear interpolation the NIST assay values and theirrespective certified values of 118.5 mg/dL and 294.5 mg/dL. The mean glucoseresult of the Pointe Calibrator over the four runs was 185.6 mg/dL. |
{3}------------------------------------------------
Tokyo Boeki Medisys Inc.
1-14-21 Higashi-Toyoda, Hino, Tokyo 191-0051, Japan Phone: +81-42-587-2965 Fax: +81-42-587-7781
Performance Characteristics
Method Comparison
A correlation analysis between the Beckman CX-7 and the Biolis 24i yielded the following results:
| RepresentativeMethod | CorrelationCoefficient | Slope(Least-Squares) | Y-axis intercept |
|---|---|---|---|
| GLU | 0.999 | 0.974 | 2.22 |
Precision
Repeatability
| Sample | N | Mean | SD | CV(%) |
|---|---|---|---|---|
| Serum 1 | રેત્વે છે. વિશ્વિત દિવસ (Career) તાલુકાઓ પૈકીના એક એવા ગામના લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલ | 60.3 | 0.8 | 1.3% |
| Serum 2 | 60 | 106.4 | 1.4 | 1.3% |
| Serum 3 | 60 | 116.8 | 1.4 | 1.2% |
| Serum 4 | 60 | 192.1 | 2.2 | 1.1% |
| Serum 5 | 60 | 446.2 | 6.8 | 1.5% |
Total Imprecision Study
| Analyzer s/n | Sample | n | mean | SD | %CV |
|---|---|---|---|---|---|
| 2229450610 | Control 1 | 80 | 63.2 | 1.5 | 2.4% |
| 2229450610 | Serum Pool 1 | 80 | 118.1 | 2.9 | 2.5% |
| 2229450610 | Serum Pool 2 | 80 | 188.8 | 4.1 | 2.1% |
| 2229450610 | Control 2 | 80 | 447.0 | 10.5 | 2.3% |
| 2227671109 | Control 1 | 80 | 63.2 | 1.2 | 1.9% |
| 2227671109 | Serum Pool 1 | 80 | 117.6 | 2.7 | 2.3% |
| 2227671109 | Serum Pool 2 | 80 | 186.9 | 3.6 | 1.9% |
| 2227671109 | Control 2 | 80 | 445.9 | 9.8 | 2.2% |
Linearity
The linearity test yielded the following results:
| Correlation | 0.9982 |
|---|---|
| Slope | 0.948 |
| Intercept | 2.45 |
| Range | 25 - 500 mg/dL |
{4}------------------------------------------------
Tokyo Boeki Medisys Inc.
1-14-21 Higashi-Toyoda, Hino, Tokyo 191-0051, Japan Phone: +81-42-587-2965 Fax: +81-42-587-7781
Sensitivity
| mg/dL | |
|---|---|
| LoB | 3.64 |
| LoD | 5.64 |
| LoQ | 10 |
Interferences
| Interferent | InterferentConcentration | GlucoseConcentration | ObservedInterference |
|---|---|---|---|
| Ascorbic acid | 30 mg/L | 75 mg/dL140 mg/dL | none |
| Bilirubin | 4.8 mg/dL6.4 mg/dL | 90 mg/dL146 mg/dL | -3.3 mg/dL-3.6% |
| Hemoglobin | 240 mg/dL160 mg/dL | 76 mg/dL139 mg/dL | -3.6 mg/dL-3.8%* |
| Lipemia(from Intralipid) | 600 mg/dL80 mg/dL | 72 mg/dL142 mg/dL | + 4 mg/dL+ 3.9% |
| Metronidazole | 24 mg/L48 mg/L | 76 mg/dL139 mg/dL | + 2.4 mg/dL+ 3.1% |
| Tetracycline | 15 mg/L | 76 mg/dL141 mg/dL | none |
Conclusion
The data demonstrates that Biolis 241 is substantially equivalent to Beckman SYNCHRON CX 7 (K904219) and that the CLC Glucose Reagent is substantially equivalent to Beckman Glucose Reagent (K802810).
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with healthcare, featuring a staff with a serpent entwined around it.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Tokyo Boeki Medisys Inc. c/o James M Clinton US Agent 5105 Fairoaks Road, Durham, NC 27712
DEC - 1 2011
Re: K103531
Trade/Device Name: Biolis 24i Clinical Chemistry Analyzer Regulation Number: 21 CFR 862.1345 Regulation Name:Glucose test system Regulatory Class: II Product Code: CFR, JJE Dated: November 4, 2011 Received: November 8, 2011
Dear Mr. Clinton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{6}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical
Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance...
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm
Sincerely yours,
N
Countney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
INDICATIONS FOR USE
510(k) Number (if known): K103531
Device Name: Biolis 24i Clinical Chemistry Analyzer
Indications for Use:
The Biolis 24i Clinical Chemistry Analyzer is a discrete photometric clinical chemistry analyzer. The device is intended to duplicate manual analytical procedures by automating various steps such as pipetting, heating, measuring color intensity, and reporting results. The device is intended to be used with certain materials to measure various andytes af diagnostic interest including glucose.
The Biolis 24i analyzer with glucose hexokinase assay is intended to measure glucose quantitatively in human serum. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic glycemia, and of the pancreatic isle cell carcinoma.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Ruth Clark
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) 103531
Page 1 of
§ 862.1345 Glucose test system.
(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.