(42 days)
The Tända Max System is intended for the treatment of wrinkles, rhytides and fine lines in the periorbital region.
Tända Max is a modular system that offers simplicity in use and convenience. The system can operate while connected directly to an electrical outlet or can be used in cordless mode drawing upon its rechargeable batteries to deliver the treatment. The system components include the control unit, the treatment head, recharging stand, Pre-treatment gel, AC adapter and goggles.
Tända Max utilizes red light at 660 nm. The unit is applied directly to the skin to ensure consistent administration of light during each treatment. The surface that comes into contact with the skin has been selected to ensure the light is administered to the skin while providing a smooth surface for cleaning.
The provided text describes the Tända Max System, an LED light system intended for the treatment of wrinkles, rhytides, and fine lines in the periorbital region. However, the document explicitly states that "additional clinical information is not necessary" and therefore no clinical study was performed to demonstrate device effectiveness.
The submission focuses primarily on comparing the Tanda Max System to predicate devices based on technological characteristics and non-clinical testing.
Given this, I cannot provide detailed information for many of your requested points as a clinical study proving the device meets acceptance criteria was not conducted or reported in this document.
Acceptance Criteria and Reported Device Performance:
Since no clinical study was performed, there are no specific clinical acceptance criteria or reported device performance metrics from such a study. The "acceptance criteria" appear to be met by demonstrating substantial equivalence to predicate devices and adherence to non-clinical safety standards.
| Acceptance Criteria Category | Reported Device Performance (or how it was met) |
|---|---|
| Intended Use | Same as predicate: Treatment of wrinkles, rhytides and fine lines in the periorbital region. |
| Output | Same as predicate: 50 mW/cm² |
| Wavelength | Same as predicate: 660 nm (red light) |
| Treatment Duration | Same as predicate: 160 seconds |
| Treatment Regimen | Same as predicate: Two treatments per week for six weeks |
| Electrical Safety | Meets IEC 60601-1 (electrical safety) |
| EMC and EMI | Meets IEC 60601-1-2 (EMC and EMI) |
| Laser Safety | Meets IEC 60825 (Laser Safety) |
| Biocompatibility | Meets ISO 10993 (biocompatibility) |
Study Details (Not Applicable/Information Not Provided):
- Sample size used for the test set and the data provenance: Not applicable, no clinical test set mentioned.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, no clinical test set mentioned.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable, no clinical test set mentioned.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, no clinical study was done, and this device is not an AI-assisted diagnostic tool for human readers.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, no standalone performance study was done as it's not an algorithm-only device.
- The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable, no clinical ground truth established or used.
- The sample size for the training set: Not applicable, no training set for a clinical study or AI algorithm mentioned.
- How the ground truth for the training set was established: Not applicable, no training set or ground truth established.
Summary of Determination:
The FDA's review concluded that the Tända Max System is substantially equivalent to legally marketed predicate devices based on its intended use, technological characteristics (output, wavelength, treatment duration, regimen), and adherence to safety and biocompatibility standards. This substantial equivalence determination allowed the device to be marketed without requiring new clinical data, as explicitly stated in section 9: "Based on the above information, additional clinical information is not necessary."
{0}------------------------------------------------
510(k) Summary of Safety and Effectiveness
Submitted by: Pharos Life Corporation 11-380 Jamieson Parkway Cambridge, Ontario N3C 4N4
JAN - 3 2011
The assigned 510(k) number is: K103415
-
- Date Prepared: Dec 8 2010
-
- Contact Person: Gordon Wehner Tel: 519-651-1177 Fax: 519-651-2277
3. Device Name and Classification:
| Device Name | Tanda Max |
|---|---|
| Classification Name | Laser Surgical Instrument for use in general andplastic surgery and in dermatology |
| Common Name | LED Light System |
| Device Classification | 2 |
| Review Panel | General and Plastic Surgery |
| Product Code | GEX |
| Regulation Number | 878.4810 |
4. Predicate Devices:
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------TICALL---------------- | and and the first from the consistence and and and and and and and and the comments ofBUNDER LEASE CARL & L E C.------------------------------------------------------------------------------------------------------------------------------------------------------------------------------SIDERESand a can annual a comments on control and control and control and concession for the first and a comments of the comments of the comments of the comments of the comments of | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------NAME AND ANDREWrogium |
|---|---|---|---|
| A RELIE MALE IN CHARRENE------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | AMAR RANCE LEBELLERA . A SENS ANDREW AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND | -------------------------------------------------------------------------.------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | .A LINE MILLE LE LEGION LE LEAR LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE |
{1}------------------------------------------------
5. Device Description
ਾਂ
Tända Max is a modular system that offers simplicity in use and convenience. The system can operate while connected directly to an electrical outlet or can be used in cordless mode drawing upon its rechargeable batteries to deliver the treatment. The system components include the control unit, the treatment head, recharging stand, Pre-treatment gel, AC adapter and goggles.
Tanda Max utilizes red light at 660 nm. The unit is applied directly to the skin to ensure consistent administration of light during each treatment. The surface that comes into contact with the skin has been selected to ensure the light is administered to the skin while providing a smooth surface for cleaning.
6. Intended Use:
The Tanda Max System is intended for the treatment of wrinkles, rhytides and fine lines in the periorbital region.
7. Comparison of Technological Differences
The technological characteristics of Tanda Max are equivalent to the listed predicate devices. Specifically, Tanda Max:
-
- Has the same intended use as the predicate (i.e., Treatment of wrinkles, rhytides and fine lines;
-
- Has the same output (i.e., 50 mW/cm2) as the predicate;
-
- Utilizes the same wavelength (i.e., 660 nm) as the predicate device;
-
- Utilizes the same treatment duration (i.e., 160 seconds) as the predicate device;
-
- Utilizes the same treatment regimen (i.e., two treatments per week for six weeks) as the predicate device.
Any differences between the Tanda Max System and the predicate device are not significant to its safety or effectiveness for its intended use.
8. Summary of non-clinical testing
Tanda Max meets the requirements of IEC 60601-1(electrical safety), IEC 60601-1-2 (EMC and EMI), IEC 60825 (Laser Safety), and ISO 10993 (biocompatibility).
9. Clinical Information
Based on the above information, additional clinical information is not necessary.
10. Conclusion
Tanda Max meets the safety and efficacy requirements necessary for its intended use.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, a symbol often associated with medicine and healthcare, with a triple helix design representing DNA. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol. The logo is simple and monochromatic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WQ66-G609 Silver Spring. MI) 20993-0002
Pharos Life Corporation % Regulatory Technology Services, LLC Mr. Mark Job 1394 25" Street, NW Buffalo, Minnesota 55313
JAN - 3 - 11
Re: K103415
Trade/Device Name: Tända Max System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: ONE Dated: December 21, 2010 Received: December 22, 2010
Dear Mr. Job:
:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
AZ. B. 2
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number: K | O 3 Y | را
Device Name: Tända Max System
Indications For Use:
JAN - 3 2011.
· The Tända Max System is intended for the treatment of wrinkles, rhytides and fine lines in the periorbital region.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil RP Ryden for mkn
(Division Sign-Off Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K103415
§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.
(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.