Intended Use
For use in calibration of the Albumin, Calcium, Cholesterol, Creatinine, Glucose, Iron, Lactate, Magnesium, Phosphorus, Total Protein, Triglyceride, Urea and Uric Acid assays. This is an in vitro diagnostic product intended for use as a calibration serum in clinical chemistry assays. The MCC contains 13 analytes in a liquid human serum based matrix. The concentrations and activities are suitable for calibration of the Abbott ARCHITECT c8000 System. Constituent concentrations are available at 2 levels.
Device Description
A multi-Analyte Calibrator - human serum based, liquid ready-to-use calibrator containing Albumin, Calcium, Cholesterol, Creatinine, Glucose, Iron, Lactate, Magnesium, Phosphorus, Total Protein, Triglyceride, Urea and Uric Acid.
More Information

Not Found

No
The document describes a multi-analyte calibrator for in vitro diagnostic assays and does not mention any AI or ML components.

No.
This device is an in vitro diagnostic product used for calibrating clinical chemistry assays, not for diagnosing, treating, or preventing disease in a patient.

No

This device is a calibrator, used to ensure the accuracy of diagnostic assays, rather than performing a diagnosis itself.

No

The device is a multi-analyte calibrator, which is a physical substance (liquid human serum based matrix) used for calibrating in vitro diagnostic assays. It is not software.

Yes, this device is an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use/Indications for Use: Explicitly states "This is an in vitro diagnostic product intended for use as a calibration serum in clinical chemistry assays."
  • Device Description: Describes a "multi-Analyte Calibrator - human serum based, liquid ready-to-use calibrator" used in clinical chemistry.
  • Performance Studies: Mentions studies conducted on "Architect c8000 Systems," which are clinical chemistry analyzers.
  • Predicate Device: Lists a predicate device with a K number (K981706) and name ("Abbott Clinical Chemistry MCC"), indicating a previous regulatory submission for a similar IVD device.

All of these points strongly indicate that this device is intended for use in testing biological samples outside of the body to provide information for medical purposes, which is the definition of an In Vitro Diagnostic device.

N/A

Intended Use / Indications for Use

For use in calibration of the Albumin, Calcium, Cholesterol, Creatinine, Glucose, Iron, Lactate, Magnesium, Phosphorus, Total Protein, Triglyceride, Urea and Uric Acid assays.

The Abbott Clinical Chemistry Multiconstituent Calibrator is an in vitro diagnostic product intended for use as a calibration serum in clinical chemistry assays. The MCC contains 13 analytes in a liquid human serum based matrix. The concentrations and activities are suitable for calibration of the Abbott ARCHITECT c8000 System. Constituent concentrations are available at 2 levels.

Product codes (comma separated list FDA assigned to the subject device)

JIX

Device Description

Original: A multi-Analyte Calibrator - human serum based, liquid ready-to-use calibrator containing Albumin, Calcium, Cholesterol, Creatinine, Glucose, Phosphorus, Total Protein, Triglyceride, Urea and Uric Acid.

Modified: A multi-Analyte Calibrator - human serum based, liquid ready-to-use calibrator containing Albumin, Calcium, Cholesterol, Creatinine, Glucose, Iron, Lactate, Magnesium, Phosphorus, Total Protein, Triglyceride, Urea and Uric Acid.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

The modified Abbott Clinical Chemistry Multiconstituent Calibrator is The modified Abbott Chineal Chemistry Mialiconstituent Calibrator Calibrator as demonstrated by results obtained in the studies.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K981706

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 862.1150 Calibrator.

(a)
Identification. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

0

510(k) Summary

Submitter's Name/Address

Abbott Laboratories 1921 Hurd Drive Irving, TX 75038

Contact Person Linda Morris Senior Regulatory Specialist, MS 2-11 Regulatory Affairs (972) 518-6711 Fax (972) 518-7479 Email: Linda.Morris@Abbott.com

Date of Preparation of this Summary:

Device Trade or Proprietary Name:

January 14, 2011

Abbott Clinical Chemistry Multiconstituent Calibrator

Device Common/Usual Name or Classification Name:

Classification Number/Class:

21 CFR 862.1150/ Class II

Multiconstituent Calibrator (MCC)

(Calibrator, Multi-Analyte Mixture) JIX

Product Code:

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: (k)103403______________________________________________________________________________________________________________________________________

Device Description:

Original { Abbott Clinical Chemistry MCC (K981706)

A multi-Analyte Calibrator - human serum based, liquid ready-to-use calibrator containing Albumin, Calcium, Cholesterol, Creatinine, Glucose, Phosphorus, Total Protein, Triglyceride, Urea and Uric Acid.

Modified Abbott Clinical Chemistry MCC (K103403)

A multi-Analyte Calibrator - human serum based, liquid ready-to-use calibrator containing Albumin, Calcium, Cholesterol, Creatinine, Glucose, Iron, Lactate, Magnesium, Phosphorus, Total Protein, Triglyceride, Urea and Uric Acid.

3

FEB - 4 2011

1

Description of modifications:

The modified Abbott Clinical Chemistry Multiconstituent Calibrator is substantially equivalent to the original Multiconstituent Calibrator (K981706). The modifications consisted of certification of the following new analytes:

  • l. Iron
    1. Lactate
  • · 3. Magnesium

These modifications did not significantly change the safety and effectiveness of the device as demonstrated in the Performance Characteristics Summary.

Intended Use:

For use in calibration of the Albumin, Calcium, Cholesterol, Creatinine, Glucose, Iron, Lactate, Magnesium, Phosphorus, Total Protein, Triglyceride, Urea and Uric Acid assays.

Indications for use:

This is an in vitro diagnostic product intended for use as a calibration serum in clinical chemistry assays. The MCC contains 13 analytes in a liquid human serum based matrix. The concentrations and activities are suitable for calibration of the Abbott ARCHITECT c8000 System. Constituent concentrations are available at 2 levels.

Predicate Device Description: MCC (K981706)

  • Human serum based, liquid ready-to-use .
  • Containing Albumin, Calcium, Cholesterol, Creatinine, Glucose, Phosphorus, . Total Protein, Triglyceride, Urea and Uric Acid
  • Shelf-life of 20 months demonstrated by real time testing .

Modified MCC Device:

  • Human serum based, liquid ready-to-use. .
  • Containing Albumin, Calcium, Cholesterol, Creatinine, Glucose, Iron, Lactate, . Magnesium, Phosphorus, Total Protein, Triglyceride, Urea and Uric Acid
  • Shelf-life of 20 months was demonstrated by comparison with predicate using . accelerated studies.
  • The modifications are the addition of Iron, Lactate and Magnesium (at . physiological levels) to the formulation.
  • These modifications did not significantly change the safety and effectiveness of . The device. The stability data confirms that the stability of the new formulation the device. The stability claims of the previous formulation and traccability to connorms to the blackly the same reference methods has been demonstrated to be substantially equivalent to the predicate device.

2

In determining substantial equivalence, a direct comparison of new and predicate device was performed, refer to table below.

AttributesPredicate MCC (k)981706Modified MCC (k)103403
Intended UseFor use in the calibration of
Albumin, Calcium,
Cholesterol, Creatinine,
Glucose, Phosphorus, Total
Protein, Triglyceride Urea
and Uric Acid assays.Same plus Iron, Lactate,
and Magnesium
ComponentHuman serum matrix
containing Albumin,
Calcium, Cholesterol,
Creatinine, Glucose,
Phosphorus, Total Protein,
Triglyceride Urea and Uric
Acid.Same plus Iron, Lactate,
and Magnesium
Storage2 to 8° CSame
Shelf-life Stability20 months from date of
manufactureSame
Open-vial Stability7 days when stored at 2 to 8
°C or 24 hours when stored
at 15 to 30 °CSame

Value assignment and Traceability:

Value assignment is performed as follows:

  • The testing is performed independently for each of the analytes, using one Abbott . Architect c8000 System.
  • For each analyte, system suitability (validity) is verified by calibrating and testing . controls with pre-defined ranges.
  • NIST standards are used for applicable analytes (refer to table on the next page . for a list of all standards used).
  • The standards may need to be weighed, dissolved and/or diluted to prepare a . solution of appropriate concentration for the value assignment.
  • 20 replicates of each tested calibrator level and of each respective standard . solution are tested.
  • The mean analyte values are calculated using the concentration of the standard . The mount and the absorbance values of the standard and the tested calibrator.
  • Solution analyte values are verified by calibrating the system with the newly . I he mean andryte varant, and measuring the reference standard solutions.
  • Acceptance criterion: the mean value obtained for the reference standard solution . must be within 5% from the known standard concentration.

3

AnalyteReference MaterialReference Method
AlbuminERM-DA470Gravimetric
CalciumNIST SRM 956Coulometric Titration
CholesterolHuman Cholesterol
(Abell-Kendall
verification)Volumetric
CreatinineNIST SRM 967 and
914IDMS
GlucoseNIST SRM 965IDMS
IronNIST SRM 3126Gravimetric
Lactic AcidReagent grade lactateGravimetric
MagnesiumNIST SRM 956IDMS
PhosphorusNIST 186-1/2186-IGravimetric
Total ProteinNIST SRM 927Gravimetric
TriglycerideACS Grade GlycerolGravimetric
Urea NitrogenNIST SRM 909IDMS
Uric AcidNIST SRM 909IDMS

Standardization/ Traceability information:

Stability:

The expiration date claim is 20 months from product dispensing. Open vial stability claim is 7 days at 2-8 °C and 24 hours at room temperature (15-30 °C).

Data to support shelf-life stability of 20 months was established by accelerated stability studies on Architect c8000 Systems.

Open vial stability was established by incubating opened calibrator vials at 30°C for 28 Open vial stability was colloried i with 3 calibrator lots. Results of the open vials means of at least three replicates) were compared to those of freshly opened calibrator vials.

Conclusion:

The modified Abbott Clinical Chemistry Multiconstituent Calibrator (MCC) is The modified Abbott Chineal Chemistry Mialiconstituent Calibrator Calibrator as demonstrated by results obtained in the studies.

4

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002

FEB 0 4 2011

Abbott Laboratories c/o Linda Morris Senior Regulatory Specialist 1921 Hurd Drive Irving, TX. 75038, USA

Re: K103403 Trade/Device Name: Abbott Clinical Chemistry Multiconstituent Calibrator Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator, Multi-Analyte Mixture Regulatory Class: Class II Product Code: JIX Dated: January 21, 2011 Received: January 21, 2011

Dear Ms. Morris

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce. prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFF) Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

5

Page 2 -

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation (21 CFR Parts 801 and 5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

G.C.

Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

6

Indications for Use Form

510(k) Number (if known): (k)103403

Device Name: Abbott Clinical Chemistry Multiconstituent Calibrator

Indications for Use:

The Abbott Clinical Chemistry Multiconstituent Calibrator is an in vitro diagnostic product intended for use as a calibration serum in clinical chemistry assays. The MCC contains 13 analytes in a liquid human serum based matrix. The concentrations and activities are suitable for calibration of the Abbott ARCHITECT c8000 System. Constituent concentrations are available at 2 levels.

Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

· (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign Off

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K103403