K Number
K103377
Manufacturer
Date Cleared
2011-05-18

(182 days)

Product Code
Regulation Number
870.1330
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Charter™ Guidewires are intended for use in the coronary and peripheral vasculature.

Device Description

The Navilyst Medical Guidewire is a disposable medical device designed for single use only. It consists of a PTFE coated 140cm or 180cm, 0.014" or 0.018" diameter stainless steel core wire, one end of which is reduced in diameter over a 43cm approx. segment in a progressive fashion through a centreless grinding operation. The profile of this reduced section affords the product a reduced area of stiffness and can be varied to produce various levels of support.

The distal part of the reduced section is covered with a 3cm, 0.010" platinum tungsten spring coil. This provides greater visibility on x-ray equipment. A 39cm approx. length of black / grey Estane 88A radiopaque heat shrink polymer tubing is applied over the tapered distal end of the wire and ground to form a constant outer diameter, OD, equivalent in diameter to the main core body. Coatings are placed on the device to improve the lubricity and ease in its advancement through the guide catheter and the blood vessel

AI/ML Overview

The provided text describes the 510(k) application for the Navilyst Charter Guidewire, focusing on its substantial equivalence to predicate devices based on in vitro bench testing and biological safety evaluations. There is no mention of a diagnostic device, AI/ML algorithm, or clinical study involving human readers or ground truth established by experts.

Therefore, most of the requested information cannot be extracted from the given text.

Here is what can be inferred:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Defined in Product Specification)Reported Device Performance
Tensile StrengthMet acceptance criteria
Torque StrengthMet acceptance criteria
Outer Diameter measurementMet acceptance criteria
Torque ResponseMet acceptance criteria
Catheter CompatibilityMet acceptance criteria
Coating Adherence/Coating IntegrityMet acceptance criteria
Tip FlexibilityMet acceptance criteria

2. Sample size used for the test set and the data provenance

  • Sample Size: Not specified for each test. The document states "Charter™ Guidewires (in its various configurations)" were tested, implying multiple units were used.
  • Data Provenance: In vitro bench testing. The country of origin for the data is not specified, but the applicant is Brivant Ltd, Ireland. The testing was prospective for the Charter Guidewire.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. No experts were used for establishing ground truth as this was an in vitro bench testing study for a medical device (guidewire), not a diagnostic algorithm.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. No adjudication method was used as this was an in vitro bench testing study.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was conducted. This document describes the approval of a medical device (guidewire), not a diagnostic AI.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not for an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable. The "ground truth" for the bench tests would be the established engineering specifications and performance standards for guidewires, as referenced by the "FDAs guidance document "Coronary and Cerebrovascular Guidewire Guidance, Jan 1995" and the product specification.

8. The sample size for the training set

Not applicable. There is no training set mentioned, as this is not an AI/ML algorithm.

9. How the ground truth for the training set was established

Not applicable. There is no training set mentioned.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Lake Region Medical. The logo consists of a stylized graphic above the words "Lake Region" in a bold, sans-serif font. Below "Region" is the word "Medical" in a smaller font size with a small solid circle to the right of the word.

K103377

Navilyst Charter 510(k) Application

510(k) Summary

as required by 21 CFR 807.92(c)

Device NameCharter™ Guidewire
Submittersname/contactdetailsBrivant Ltd,Parkmore West Business Park,Galway,IrelandContact Details:Tomas FureyOperations Manager,Tel: +353 91 385037Fax: +353 91 766598
SummaryPreparation Date10th November 2010
Device Name &ClassificationTrade Name: Charter™ GuidewireCommon Name: GuidewireClassification Name: Catheter, GuidewireDevice Classification: Class II, 21 CFR §870.1330Product Code: DQX
Intended UseIntended Use:The Charter™ Guidewires are intended for use in the coronary and peripheralvasculature.Contraindications:The Charter™ Guidewire is not intended for use in the cerebral vasculature.Patients judged not acceptable for percutaneous intervention (PCI)
Device DescriptionThe Navilyst Medical Guidewire is a disposable medical device designed forsingle use only. It consists of a PTFE coated 140cm or 180cm, 0.014" or 0.018"diameter stainless steel core wire, one end of which is reduced in diameter overa 43cm approx. segment in a progressive fashion through a centreless grindingoperation. The profile of this reduced section affords the product a reduced areaof stiffness and can be varied to produce various levels of support.The distal part of the reduced section is covered with a 3cm, 0.010" platinumtungsten spring coil. This provides greater visibility on x-ray equipment. A 39cmapprox. length of black / grey Estane 88A radiopaque heat shrink polymer tubingis applied over the tapered distal end of the wire and ground to form a constantouter diameter, OD, equivalent in diameter to the main core body.Coatings are placed on the device to improve the lubricity and ease in itsadvancement through the guide catheter and the blood vessel
Predicate DevicesManufacturer510kDate
Brivant Ltd, Brivant GuidewireK06055107 Jun 2006
Brivant Ltd, Streamer GuidewireK08309407 July 2009
Brivant Ltd, Courier GuidewireK07308209 Jan 2008
Principle ofOperationThe Charter™ Guidewire is operated manually by a manual process
Comparison ofTechnologicalCharacteristicsThe Charter™ wire has minor differences in construction to the approved BrivantGuidewire. These include Same diameter range as predicate devices Same basic construction technology as the predicate devices. Different polymer and hydrophilic coating materials to the predicate devices providing improved radiopacity. Slightly shorter overall length than predicate devices. Minor differences in profiles of the distal tip area of the guidewire changes to meet customer performance requirements. These changes are within ranges of the predicate devices.
In vitro bench testing was performed to support a determination of substantial equivalence (refer to performance testing below) between the Charter™ Guidewires (in its various configurations) and the predicate devices.The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use and performs comparably to the predicate devices. The differences in construction between the Charter™ wire and the predicate devices raise no new issues of safety and effectiveness such that the Charter™ Guidewire is considered substantially equivalent to the predicate devices.
PerformanceTesting (non-clinical)In vitro bench tests were carried out to demonstrate equivalence with reference to the FDAs guidance document "Coronary and Cerebrovascular Guidewire Guidance, Jan 1995".The following bench tests were performed: Tensile Strength Torque Strength Outer Diameter measurement Torque Response Catheter Compatibility Coating Adherence/Coating Integrity Tip Flexibility The results from these performance evaluations demonstrated that the Charter™ Guidewire met the acceptance criteria defined in the product specification and performed comparably to the predicate device.
Biological Safety of the device has been established through biocompatibility testing carried out in compliance with ISO 10993-1 under VP-0427The following biocompatibility tests were performed
Test DescriptionTest Method
Cytotoxicity StudyQualitative Evaluation – Dye Exclusion &Microscopial Evaluation
Cytotoxicity StudyQuantitative Evaluation - MTT or XTT Assay
Irritation TestIntracutaneous Injection (ISO10993-10)
Acute Systemic Toxicity TestSystemic InjectionISO10993-11
Acute Systemic Toxicity TestISO-Rabbit-Pyrogen
Haemocompatibility TestHaematology: Haemolysis - rabbit blood- Direct(Complete ASTM Method)
Haemocompatibility TestHaematology:In-vitro Haemocompatiblity Assay
Haemocompatibility TestCoagulation: The Prothrombin Time Assay (PT)
Haemocompatibility TestCoagulation: The Unactivated PartialThromboplastin Time Assay (UPTT)
Haemocompatibility TestThrombosis: In Vivo Thrombogenicity in Dogs
Haemocompatibility TestComplement Activation
Haemocompatibility TestLee& White Coagulation Assay
ConclusionsBased on safety and performance testing, technological characteristics and the
indications for use for the device, the Charter™ Guidewire has been
demonstrated to be appropriate for its intended use and is considered to besubstantially equivalent to the predicate devices.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a logo for Lake Region Medical. The logo consists of a stylized graphic above the words "Lake Region" in a bold, serif typeface. Below "Lake Region" is the word "Medical" in a smaller, less bold font.

Navilyst Charter 510(k) Application

·

:

{2}------------------------------------------------

:

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features an abstract image of an eagle with its wings spread, symbolizing protection and strength. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Brivant, LTD. c/o Mr. Tomas Furey Operations Manager Parkmore West Business Park Galway Ireland

MAY 18 2011

Re: K103377

Trade/Device Name: Charter Guidewire Model 45-281, 45-282, 45-283 Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter, Guidewire Regulatory Class: Class II Product Code: DQX Dated: April 14, 2011 Received: April 18, 2011

Dear Mr. Furey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{4}------------------------------------------------

Page 2 - Mr. Tomas Furey

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

C. Tom Zuckerman, M.D.

Bram Zuckerman, M.D Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

K103377 510(k) Number (if known):

Charter Guidewire Device Name:

Indications For Use: Charter Guidewires are intended for use in the coronary and peripheral vasculature.

Contraindications: The Charter Guidewire is not intended for use in the cerebral vasculature. Patients judged not acceptable for percutaneous intervention (PCI).

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular Devices

510(K) number K103377

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.