(25 days)
No
The document describes a standard infusion set and its components, with no mention of AI or ML capabilities. The performance studies focus on bench testing related to fluid dynamics and backflow prevention, not algorithmic performance.
No
The device, an infusion set, is used for the delivery of fluids, solutions, and drugs. While these substances may have therapeutic effects, the device itself is a delivery mechanism, not a therapeutic agent that directly treats or cures a condition. Its function is to facilitate the administration of therapies, not to provide therapy itself.
No
Explanation: The provided text describes infusion sets designed for fluid delivery. There is no mention of these devices being used to identify, measure, or monitor a medical condition, disease, or other health parameters for diagnostic purposes.
No
The device description clearly states that the device is a disposable physical product ("Hospira infusion sets are disposable devices for single patient use") and includes physical components ("These components including the back check valve which is the subject of this submission"). There is no mention of software as the primary or sole component.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly describes the delivery of fluids, solutions, drugs, etc., into the body via various routes (parenteral, enteral, intravenous, etc.). This is a therapeutic or supportive function, not a diagnostic one.
- Device Description: The description focuses on the physical components and function of an infusion set for delivering substances.
- Lack of Diagnostic Language: There is no mention of analyzing samples from the body (blood, urine, tissue, etc.) to diagnose a condition, monitor a disease, or screen for health issues.
- Performance Studies: The performance studies described relate to the function of the infusion set itself (e.g., reduction of backflow), not to the accuracy or reliability of a diagnostic test.
IVD devices are used to examine specimens derived from the human body to provide information for diagnostic purposes. This infusion set is used to administer substances to the human body.
N/A
Intended Use / Indications for Use
Infusion sets are intended for the delivery of fluids, solutions, drugs, agents, nutritionals, electrolytes, blood and blood products via parenteral, intravenous, intra-arterial, subcutaneous, epidural or irrigation routes of administration. These sets are intended for use primarily in the hospital setting and can be used in other acute and non-acute care areas, such as, but not limited to Home Care, Nursing Homes, Mobile Intensive Care, Ambulatory Infusion Centers, Hospice, Subacute facilities, Outpatient / Surgical Centers, Long Term Care, Urgent Care, Transport and in Physician offices.
Product codes
80-FPA, FPA
Device Description
Hospira infusion sets are intended for use as gravity sets or with dedicated Hospira Infusion Pumps. Hospira infusion sets are disposable devices for single patient use, which incorporate various set configurations and components. These components including the back check valve which is the subject of this submission may be shared across Hospira set families and used with sets designed for gravity or other pump platforms.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
primarily in the hospital setting and can be used in other acute and non-acute care areas, such as, but not limited to Home Care, Nursing Homes, Mobile Intensive Care, Ambulatory Infusion Centers, Hospice, Subacute facilities, Outpatient / Surgical Centers, Long Term Care, Urgent Care, Transport and in Physician offices.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Bench Performance Data: The Hospira Infusion sets with an alternate Back Check Valve passed all specified test requirements. Testing The Prospira Infosion sels with respect to reduction of backflow. The validation and verification testing acmonotiate oxeculier. Street user needs and design inputs for an Infusion set.
Biocompatibility:
- ISO 10993-5: 2009 - Cytotoxicity Pass
- ISO 10993-10: 2002 - Sensitization Pass
- ISO 10993-10: 2002 - Irritation / Intracutaneous Reactivity Pass
- ISO 10993-11:2006 - Systemic Toxicity (Acute) Pass
- ISO 10993-4:2002 - Hemocompatibility Pass
Sterility:
- SAL 10-6 ISO 11137-2:2006 - Sterility Pass
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 880.5440 Intravascular administration set.
(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
0
November 12, 2010
·
Section 6:
な
510(k) SUMMARY
DEC 1 0 2010
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92. Hospira Infusion Sets with alternate fixed diaphragm Back Check Valve
Submitter Information | |
---|---|
Name | Rebecca Andersen |
Address | 275 North Field Drive |
Phone number | 224-212-5270 |
Fax number | 224-212-5401 |
Establishment | |
Registration Number | Owner/Operator #9063339 |
Name of contact person | Rebecca Andersen |
Date prepared | November 12, 2010 |
Name of device | |
Trade or proprietary | |
name | Hospira Infusion Sets |
Common or usual name | IV Set |
Classification name | Infusion Sets Class II |
Classification panel | 80-FPA - General Hospital |
Regulation | 21 CFR Part 880.5440 |
Product Code(s) | 80-FPA |
Legally marketed device(s) to | |
which equivalence is claimed | Symbiq sets as cleared in K041550 |
Reason for 510(k) submission | Back Check Valve component |
Device description | Hospira infusion sets are intended for use as gravity sets or with dedicated |
Hospira Infusion Pumps. Hospira infusion sets are disposable devices for | |
single patient use, which incorporate various set configurations and | |
components. These components including the back check valve which is the | |
subject of this submission may be shared across Hospira set families and | |
used with sets designed for gravity or other pump platforms. | |
Intended use of the device | Infusion sets are intended for the delivery of fluids, solutions, drugs, agents, |
nutritionals, electrolytes, blood and blood products via parenteral, enteral, | |
intravenous, intra-arterial, subcutaneous, epidural or irrigation routes of | |
administration. These sets are intended for use primarily in the hospital setting and | |
can be used in other acute and non-acute care areas, such as, but not limited to | |
Home Care, Nursing Homes, Mobile Intensive Care, Ambulatory Infusion Centers, | |
Hospice, Subacute facilities, Outpatient / Surgical Centers, Long Term Care, Urgent | |
Care, Transport and in Physician offices. | |
Indications for use | These are general use sets which are intended for use primarily in the hospital |
setting and can be used in other acute and non-acute care areas, such as, but not | |
limited to Home Care, Nursing Homes, Mobile Intensive Care, Ambulatory Infusion | |
Centers, Hospice, Subacute facilities, Outpatient / Surgical Centers, Long Term | |
Care, Urgent Care, Transport and in Physician offices. |
1
November 12, 2010
Infusion Set Modification
Special 510(k)
510(k) SUMMARY
Section 6: | 510(k) SUMMARY | |
---|---|---|
Summary of the technological characteristics of the device compared to the predicate device | ||
Characteristic | Proposed Device | |
Hospira Infusion Set | Predicate Symbiq Infusion Set | |
K041550 | ||
Intended Use | Same | These sets are intended for use primarily in |
the hospital setting and can be used in other | ||
acute and non-acute care areas, such as, but | ||
not limited to Home Care, Nursing Homes, | ||
Mobile Intensive Care, Ambulatory Infusion | ||
Centers, Hospice, Subacute facilities, | ||
Outpatient / Surgical Centers, Long Term | ||
Care, Urgent Care, Transport and in | ||
Physician offices. | ||
Set Functionality | Same | Meets all performance V&V testing |
Visual characteristics | Similar Back Check valve is of | |
similar size, but is blue and clear | Back check valve is clear / white | |
Back Check Valve | Normally closed - One way check | |
valve in two piece housing with | ||
fixed silicone diaphragm | Normally closed - One way check valve in two | |
piece housing with floating silicone | ||
diaphragm | ||
Bonding processes | Same | Solvent bond / over post or socket bond |
Biocompatibility | Same | External Communicating - Blood path indirect |
-Prolonged contact | ||
Principle of Operation | Same | One way pressure differential |
SUMMARY OF NON CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE | ||
Performance Test Summary-New Device | ||
Characteristic | Standard/Test Method | Standard / Test Title Device Performance |
Biocompatibility | ISO 10993-5: 2009 | Cytotoxicity Pass |
Biocompatibility | ISO 10993-10: 2002 | Sensitization Pass |
Biocompatibility | ISO 10993-10: 2002 | Irritation / Intracutaneous |
Reactivity Pass | ||
Biocompatibility | ISO 10993-11:2006 | Systemic Toxicity (Acute) Pass |
Biocompatibility | ISO 10993-4:2002 | Hemocompatibility Pass |
SAL 10-6 | ISO 11137-2:2006 | Sterility Pass |
Summary discussion of Bench Performance Data
The Hospira Infusion sets with an alternate Back Check Valve passed all specified test requirements. Testing The Prospira Infosion sels with respect to reduction of backflow. The validation and verification testing acmonotiate oxeculier. Street user needs and design inputs for an Infusion set.
mod these doness most as attributes and device performance meet requirements of fire standards istead inthe
2
Section 6:
510(k) SUMMARY
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Statement of Safety and Efficacy:
The Hospira Infusion Sets with the alternate back check valve meet the functional claims, and intended use as described in the product labeling. The safety and effectiveness, are substantially equivalent to the predicate Hospira Infusion Sets as cleared in K041550
The claim for substantial equivalence is supported by the information provided in this Special 510(k) submission.
3
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Rebecca Anderson Associate Director, Global Regulatory Affairs, Devices HOSPIRA, Incorporated 275 North Field Drive Lake Forest, Illinois 60045
DEC 1 0 2000
Re: K103344
Trade/Device Name: Hospira Infusion Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: November 12, 2010 Received: November 15, 2010
Dear Ms. Anderson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mendments, or to devices that have been reclassified in accordance with the provisions of Amendinente, or to arrivand Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, lecturents for allinar regainst misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your (1 w/1), it thay of surfet is alla of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
4
Please be advised that FDA's issuance of a substantial equivalence determination does not Fleast oc advisod that 1 Dr. 5 lookines on that your device complies with other requirements Incall that FDA has made a decommanent in administered by other Federal agencies. of the Act of ally I oderal bates as requirements, including, but not limited to: registration 1 ou must comply with an the Fiel singles and 801); medical device reporting and fishing (21 CFR Pat 607), lacenad (21 CFR 803); good manufacturing (reporting of medical device the quality systems (QS) regulation (21 CFR Part 820); practice requirements as sectionic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm113809.htm for inttp://www.ida.gov//About Health's (CDRH's) Office of Compliance. Also, the Somer the regulation entitled, "Misbranding by reference to premarket notification" prease note the regarding the reporting the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the I ou him of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S.,M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
5
Infusion Set Modification Special 510(k)
. !
November 12, 2010
Section 5: Indications for Use
DEC 1 0 2010
510(k) Number (unknown at this time)
ﺘ
Device Name: Hospira Infusion Set
Infusion sets are intended for the delivery of fluids, solutions, drugs, agents, nutritionals, electrolytes, blood and blood products via parenteral, intravenous, intra-arterial, subcutaneous, epidural or irrigation routes of administration. These sets are intended for use primarily in the hospital setting and can be used in other acute and non-acute care areas, such as, but not limited to Home Care, Nursing Homes, Mobile Intensive Care, Ambulatory Infusion Centers, Hospice, Subacute facilities, Outpatient / Surgical Centers, Long Term Care, Urgent Care, Transport and in Physician offices.
| Prescription Use
X | AND/OR | Over-The-Counter Use
______ |
|-----------------------------|--------|--------------------------------|
| (Part 21 CFR 801 Subpart D) | | (Part 21 CFR 807 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Rill C. Chapman 12/10/10
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K103344